LUCEA LED recallZ-0759-2024 · Getinge Usa Sales Inc · initiated Nov 8, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Getinge Usa Sales Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
A potential for a light system to fall in the operating room.
FDA root cause category: Equipment maintenance
What the company is doing
An URGENT MEDICAL DEVICE CORRECTION letter dated 11/8/23 was sent to customers. Actions to be taken by the customer: Our records indicate that you have been identified as having purchased at least one model of one of these product ranges. Please identify your affected units using the referenced devices on page 1 and follow the instructions below. Please distribute this notice and forward the link below to access manuals to anyone within your organization who may need to use the manuals, and especially to your service or service provider who performs your maintenance. Light system fixing and other replacements: Regarding: - the fixing of the light systems: suspensions fixing screws, adapter fixing screws, bushing fixing screws - the brake screws - the safety segments - the batteries, a dedicated instruction to illustrate and facilitate understanding of what to replace and what to order for replacement has been issued. The instruction is available on a dedicated space on the web portal https://www.getinge.com/int/campaigns/maintenance-and-service-on-or-light-system/, and provided upon request by emailing [email protected] Should you have questions or require additional information, please do not hesitate to contact Maquet SAS/Getinge or your local representative. Please complete and sign the attached MEDICAL DEVICE CORRECTION - RESPONSE FORM (page 7) to acknowledge that you have received this notification. Return the completed form to Maquet/Getinge by e-mailing a scanned copy to [email protected] or by faxing the form to (866) 350-0897. If you are a distributor who has shipped any affected products to customers, please forward this document to their attention for appropriate action. Actions to be taken by Maquet SAS / Getinge: As the majority of customers may not have regular contact with Getinge services or authorized Getinge representatives and therefore may not be aware of current maintenance recommendations, Maquet SAS/Getinge is providin
Recall details
- Recall number
- Z-0759-2024
- Status
- Open, Classified
- Initiated by the company
- Nov 8, 2023
- Posted by the FDA
- Jan 19, 2024
- Recalling firm
- Getinge Usa Sales Inc
- Quantity
- 236793 units in total
- Product code
- FTD
- FDA submissions
- K113679
- Distribution
- US Nationwide. Global Distribution.
Product description
Maquet Lucea - Lucea10/40, Lucea50/100 OR Light Systems, Model Numbers and Catalog Numbers LCA 10 - ARD2LCA00040C, LCA 10 - ARDLCA109000C, LCA 10 - ARDLCA109002C, LCA 10 - ARDLCA109006C, LCA 10 - ARDLCA209008A, LCA 10 - ARDLCA209008C, LCA 10 - ARDLCA209009A, LCA 10 - ARDLCA209009C, LCA 10 - ARDLCA209010A, LCA 10 - ARDLCA209010C, LCA 10 - ARDLCA409000C, LCA 10 - ARDLCA409001C, LCA 10 - ARDLCA409002A, LCA 10 - ARDLCA409002C, LCA 10 - ARDLCA409003C, LCA 10 - ARDLCA409004A, LCA 10 - ARDLCA409004C, LCA 10 - ARDLCA409005C, LCA 10 DES - ARDLCA409001A, LCA 10 MOBILE - ARDLCA309008A, LCA 10 RAIL - ARDLCA409005A, LCA 10 WALL - ARDLCA109006A, LCA 40 - ARDLCA109001C, LCA 40 - ARDLCA208000A, LCA 40 - ARDLCA208000C, LCA 40 - ARDLCA208001A, LCA 40 - ARDLCA208001C, LCA 40 - ARDLCA209000A, LCA 40 - ARDLCA209000C, LCA 40 - ARDLCA209001A, LCA 40 - ARDLCA209001C, LCA 40 - ARDLCA209002A, LCA 40 - ARDLCA209002C, LCA 40 - ARDLCA209003A, LCA 40 - ARDLCA209003C, LCA 40 - ARDLCA209014A, LCA 40 - ARDLCA209014C, LCA 40 - ARDLCA209017A, LCA 40 - ARDLCA209017C, LCA 40 - ARDLCA209018A, LCA 40 - ARDLCA209018C, LCA 40 - ARDLCA209019A, LCA 40 - ARDLCA209019C, LCA 40 - ARDLCA209020A, LCA 40 - ARDLCA209020C, LCA 40 - ARDLCA209021A, LCA 40 - ARDLCA209021C, LCA 40 MOBILE - ARDLCA309009A, LCA 40 WALL - ARDLCA109001A, LCA 40GS - ARDLCA109009A, LCA 40GS - ARDLCA109009C, LCA 100 - ARDLCA109005A, LCA 100 - ARDLCA109005C, LCA 100 - ARDLCA219000A, LCA 100 MOBILE - ARDLCA309006A, LCA 100 MOBILE B - ARDLCA309007A, LCA 100 SAT - ARDSAT209004A, LCA 100 SAT - ARDSAT209005A, LCA 100 V - ARDLCA219000C, LCA 100DF - ARDLCA109008A, LCA 100DF - ARDLCA109008C, LCA 100DF - ARDLCA219003A, LCA 100DF - ARDSAT209008A, LCA 100DF - ARDSAT209009A, LCA 100DF V - ARDLCA219003C, LCA 50 - ARDLCA109004A, LCA 50 - ARDLCA109004C, LCA 50 - ARDLCA209012A, LCA 50 - ARDLCA209012C, LCA 50 - ARDSAT209002A, LCA 50 MOBILE - ARDLCA309004A, LCA 50 MOBILE B - ARDLCA309005A, LCA 50 SAT - ARDSAT209003A, LCA 50 WALL 11 - ARD2LCA00140A, LCA 50DF - ARDLCA109007A, LCA 50DF - ARDLCA109007C, LCA 50DF - ARDLCA209015A, LCA 50DF - ARDLCA209015C, LCA 50DF - ARDSAT209006A, LCA 50DF - ARDSAT209007A, LCA DUO 100 - ARDLCA219001A, LCA DUO 100 V - ARDLCA219001C, LCA DUO 100+50 - ARDLCA219002A, LCA DUO 100+50 V - ARDLCA219002C, LCA DUO 100+50DF - ARDLCA219005A, LCA DUO 100+50DF V - ARDLCA219005C, LCA DUO 100DF - ARDLCA219004A, LCA DUO 100DF V - ARDLCA219004C, LCA DUO 50 - ARDLCA209013A, LCA DUO 50 - ARDLCA209013C, LCA DUO 50DF - ARDLCA209016A, LCA DUO 50DF - ARDLCA209016C, LCA100 MOBILE UPGRADE - ARDLCA506006A, LCA100 UPGRADE - ARDLCA506004A, LCA100 UPGRADE 24V - ARDLCA506005A, LCA100 WALL UPGRADE - ARDLCA506007A, LCA50 UPGRADE - ARDLCA506000A, LCA50 UPGRADE 24V - ARDLCA506001A, LCA50 UPGRADE WALL MOUNTED - ARDLCA506003A, SAT LCA 100 - ARDSAT209004C, SAT LCA 100 - ARDSAT209005C, SAT LCA 100DF - ARDSAT209008C, SAT LCA 100DF - ARDSAT209009C, SAT LCA 50 - ARDSAT209002C, SAT LCA 50 - ARDSAT209003C, SAT LCA 50DF - ARDSAT209006C, SAT LCA 50DF - ARDSAT209007C, LCA 10 MOB - ARDLCA309000A, LCA 10 MOB - ARDLCA309000C, LCA 10 MOB - ARDLCA309002A, LCA 10 MOB - ARDLCA309002C, LCA 10 MOB - ARDLCA309008C, LCA 100 MOBILE - ARDLCA309006C, LCA 100 MOBILE B - ARDLCA309007C, LCA 40 MOB - ARDLCA309001A, LCA 40 MOB - ARDLCA309001C, LCA 40 MOB - ARDLCA309003A, LCA 40 MOB - ARDLCA309003C, LCA 40 MOB - ARDLCA309009C, LCA 40GS MOBILE - ARDLCA309010A, LCA 40GS MOBILE - ARDLCA309010C, LCA 50 MOBILE - ARDLCA309004C, LCA 50 MOBILE B - ARDLCA309005C,
Code information (models, lots, serial numbers)
All systems affected UDI-DI: 3700712415921 3700712415853 3700712415846 3700712415877 3700712415938 3700712415860 3700712400750 3700712400613 3700712400712 3700712400729 3700712401115 3700712401122 03700712400767 3700712400620 3700712401139 3700712401146 3700712400774 3700712400637 3700712401153 3700712400736 3700712400743 3700712401160 03700712400781 3700712400644 3700712401177 3700712401184 3700712400651 3700712400668 3700712400675 3700712400682 3700712400699 03700712400705 03700712404710 03700712404697 03700712404703 03700712404727 03700712404666 03700712404673 03700712404680
Devices named in this recall
28 devices in the FDA device registry whose identifier appears in the recall's code information.
- LUCEA LED® ARDLCA209012AMaquet Sas · Light, Surgical, Ceiling Mounted1 recall20 reports
- LUCEA LED® ARDLCA219001AMaquet Sas · Light, Surgical, Ceiling Mounted1 recall7 reports
- LUCEA LED® ARDLCA219003AMaquet Sas · Light, Surgical, Ceiling Mounted1 recall5 reports
- LUCEA LED® ARDLCA219000AMaquet Sas · Light, Surgical, Ceiling Mounted1 recall3 reports
- LUCEA LED® ARDLCA219002AMaquet Sas · Light, Surgical, Ceiling Mounted1 recall2 reports
- LUCEA LED® ARDLCA309005AMaquet Sas · Light, surgical, floor standing1 recall2 reports
- LUCEA LED® ARDLCA219005AMaquet Sas · Light, Surgical, Ceiling Mounted1 recall1 report
- LUCEA LED® ARDLCA219004AMaquet Sas · Light, Surgical, Ceiling Mounted1 recall1 report
- LUCEA LED® ARDLCA309007AMaquet Sas · Light, surgical, floor standing1 recall1 report
- LUCEA LED® ARDLCA109005AMaquet Sas · Lamp, surgical1 recall1 report
- LUCEA LED® ARDLCA209015AMaquet Sas · Light, Surgical, Ceiling Mounted1 recall0 reports
- LUCEA LED® ARDLCA209013AMaquet Sas · Light, Surgical, Ceiling Mounted1 recall0 reports
- LUCEA LED® ARDLCA209016AMaquet Sas · Led light source1 recall0 reports
- LUCEA LED® ARDLCA109008AMaquet Sas · Led light source1 recall0 reports
- LUCEA LED® ARDLCA109004AMaquet Sas · Lamp, surgical1 recall0 reports
- LUCEA LED® ARDLCA109007AMaquet Sas · Lamp, surgical1 recall0 reports
- LUCEA LED® ARDSAT209005AMaquet Sas · Light, Surgical, Ceiling Mounted1 recall0 reports
- LUCEA LED® ARDSAT209009AMaquet Sas · Light, Surgical, Ceiling Mounted1 recall0 reports
- LUCEA LED® ARDSAT209002AMaquet Sas · Light, Surgical, Ceiling Mounted1 recall0 reports
- LUCEA LED® ARDSAT209006AMaquet Sas · Light, Surgical, Ceiling Mounted1 recall0 reports
- LUCEA LED® ARDLCA506000AMaquet Sas · Led light source1 recall0 reports
- LUCEA LED® ARDLCA506003AMaquet Sas · Lamp, surgical1 recall0 reports
- LUCEA LED® ARDLCA506005AMaquet Sas · Lamp, surgical1 recall0 reports
- LUCEA LED® ARDLCA506006AMaquet Sas · Led light source1 recall0 reports
- LUCEA LED® ARDLCA109006AMaquet Sas · Device, Medical Examination, Ac Powered1 recall0 reports
- LUCEA LED® ARDLCA209021AMaquet Sas · Device, Medical Examination, Ac Powered1 recall0 reports
- LUCEA LED® ARDLCA309008AMaquet Sas · Device, Medical Examination, Ac Powered1 recall0 reports
- LUCEA LED® ARDLCA309009AMaquet Sas · Device, Medical Examination, Ac Powered1 recall0 reports
Product lines
Other devices under the same FDA submission
2 devices cleared under K113679 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- LUCEA LED® ARDLCA309004AMaquet Sas · Light, surgical, floor standing4 reports
- LUCEA LED® ARDLCA309006AMaquet Sas · Light, surgical, floor standing1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 93567).
- Z-0757-2024 · Open, ClassifiedMaquet PowerLEDII OR Light System, Model Names and Catalog Numbers PWDII77SF - ARDPWT629100A, PWDII555DF - ARDPWT629101A, PWDII555SF - ARDPWT629102A, PWDII755DF - ARDPWT629103A, PW…
- Z-0758-2024 · Open, ClassifiedMaquet Volista, Volista Access, Volista Access II, Volista Standop OR Light System, Model Numbers and Catalog Numbers ROL B FOR VLT400SF STP - ARDROL309001A, ROL B FOR VLT400SF STP…
- Z-0760-2024 · Open, ClassifiedMaquet Rolite OR Light Systems, Model Numbers and Catalog Numbers ROL PWDII50SF - ARDPWT309001A, ROL PWDII50SF - ARDPWT309002A, ROL PWDII50SF - ARDPWT309009A, ROL PWDII50SF - ARDPW…
- Z-0761-2024 · Open, ClassifiedMaquet Equipment OR Light Systems, Model Numbers and Catalog Numbers BRAINLAB CAMERA - ARDEQT269000A, BRAINLAB CAMERA - ARDEQT269000C, BRAINLAB CAMERA - ARDEQT269001A, BRAINLAB CAM…
- Z-0762-2024 · Open, ClassifiedMaquet Orchide OR Light Systems, Model Numbers and Catalog Numbers OHDII FHD QL VP01 - ARD568803935, OHDII FHD QL AIR03 U - ARD568803952, OHDII FHD QL AIR03 E - ARD568803967, OHDII…
- Z-0763-2024 · Open, ClassifiedMaquet PowerLED/HLED and PowerLED300 OR Light Systems, Model Numbers and Catalog Numbers EPS MB1 - ARDINT059002A, EPS MB1 - ARDINT059002C, EPS MB1 TR - ARDINT059000A, EPS MB1 TR - …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page