Skip to content

Vitality recallZ-0639-2024 · Zimmer Biomet Spine Inc. · initiated Oct 16, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Zimmer Biomet Spine Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Spinal fixation system 7.5mmx 50mm screws are packaged with the correct labeling, but the screws themselves were laser marked with an incorrect size description of 7.5 x 40; use of incorrect screws could lead to screw perforation and indention of vascular or other structures.

FDA root cause category: Nonconforming Material/Component

What the company is doing

On 10/16/23, recall notices were sent to customers asking them to do the following: 1) Locate, quarantine, and return affected screws to the recalling firm. 2) Complete and return the inventory return certificate form via email to [email protected] Recall-related questions or concerns can be directed to please email [email protected].

Recall details

Recall number
Z-0639-2024
Status
Open, Classified
Initiated by the company
Oct 16, 2023
Posted by the FDA
Jan 4, 2024
Recalling firm
Zimmer Biomet Spine Inc.
Quantity
3
Product code
NKB
FDA submissions
K150896
Distribution
US Nationwide distribution in the states of MN, NE.

Product description

Vitality Polyaxial Screw, 7.5mmx 50mm, REF: 07.02000.118, used with Vitality and Vital Spinal Fixation System

Code information (models, lots, serial numbers)

UDI-DI: 00889024002159, Lot: W834061

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page