Vitality® 07.02000.118Recalls & reported problems
Biomet Spine LLC · Class 2
All models of Vitality (38)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 17
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 17 adverse event reports that list this exact device (DI 00889024002159), received between Oct 6, 2023 and Oct 16, 2023 — classified as 17 malfunctions.
The most frequently reported specific device problem is “Inaccurate Information” (17 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Oct 16, 2023Class IIOpen, Classified
Spinal fixation system 7.5mmx 50mm screws are packaged with the correct labeling, but the screws themselves were laser marked with an incorrect size description of 7.5 x 40; use of incorrect screws could lead to screw perforation and indention of vascular or other structures.
Z-0639-2024 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
Most reported problems
Device problems
- Inaccurate Information17
Patient problems
- No Clinical Signs, Symptoms or Conditions17
Recent adverse event reports
The 17 most recent of 17 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 16, 2023Inaccurate Information; Inaccurate Information
IT WAS REPORTED THE LENGTH OF THE SIZE LASER MARKED ON SEVEN VITALITY SCREWS DOES NOT MATCH THE SIZE THAT CORRELATES WITH THE LASER MARKED PART NUMBER. THE LASER MARKED PART NUMBER IS 07.02000.118, WHICH SHOULD BE A 7.5 X 50 SCREW. HOWEVER, THE SIZE IS LASER MARKED 7.5 X 40MM. THIS WAS FOUND IN A WAREHOUSE AND DID NOT HAVE ANY PATIENT INVOLVEMENT. THIS IS REPORT SEVEN OF SEVEN FOR THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17942738 ↗MalfunctionOct 16, 2023Inaccurate Information; Inaccurate Information
IT WAS REPORTED THE LENGTH OF THE SIZE LASER MARKED ON SEVEN VITALITY SCREWS DOES NOT MATCH THE SIZE THAT CORRELATES WITH THE LASER MARKED PART NUMBER. THE LASER MARKED PART NUMBER IS 07.02000.118, WHICH SHOULD BE A 7.5 X 50 SCREW. HOWEVER, THE SIZE IS LASER MARKED 7.5 X 40MM. THIS WAS FOUND IN A WAREHOUSE AND DID NOT HAVE ANY PATIENT INVOLVEMENT. THIS IS REPORT SIX OF SEVEN FOR THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17942688 ↗MalfunctionOct 16, 2023Inaccurate Information; Inaccurate Information
IT WAS REPORTED THE LENGTH OF THE SIZE LASER MARKED ON SEVEN VITALITY SCREWS DOES NOT MATCH THE SIZE THAT CORRELATES WITH THE LASER MARKED PART NUMBER. THE LASER MARKED PART NUMBER IS 07.02000.118, WHICH SHOULD BE A 7.5 X 50 SCREW. HOWEVER, THE SIZE IS LASER MARKED 7.5 X 40MM. THIS WAS FOUND IN A WAREHOUSE AND DID NOT HAVE ANY PATIENT INVOLVEMENT. THIS IS REPORT FIVE OF SEVEN FOR THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17942537 ↗MalfunctionOct 16, 2023Inaccurate Information; Inaccurate Information
IT WAS REPORTED THE LENGTH OF THE SIZE LASER MARKED ON SEVEN VITALITY SCREWS DOES NOT MATCH THE SIZE THAT CORRELATES WITH THE LASER MARKED PART NUMBER. THE LASER MARKED PART NUMBER IS 07.02000.118, WHICH SHOULD BE A 7.5 X 50 SCREW. HOWEVER, THE SIZE IS LASER MARKED 7.5 X 40MM. THIS WAS FOUND IN A WAREHOUSE AND DID NOT HAVE ANY PATIENT INVOLVEMENT. THIS IS REPORT FOUR OF SEVEN FOR THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17942418 ↗MalfunctionOct 16, 2023Inaccurate Information; Inaccurate Information
IT WAS REPORTED THE LENGTH OF THE SIZE LASER MARKED ON SEVEN VITALITY SCREWS DOES NOT MATCH THE SIZE THAT CORRELATES WITH THE LASER MARKED PART NUMBER. THE LASER MARKED PART NUMBER IS 07.02000.118, WHICH SHOULD BE A 7.5 X 50 SCREW. HOWEVER, THE SIZE IS LASER MARKED 7.5 X 40MM. THIS WAS FOUND IN A WAREHOUSE AND DID NOT HAVE ANY PATIENT INVOLVEMENT. THIS IS REPORT THREE OF SEVEN FOR THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17942350 ↗MalfunctionOct 16, 2023Inaccurate Information; Inaccurate Information
IT WAS REPORTED THE LENGTH OF THE SIZE LASER MARKED ON SEVEN VITALITY SCREWS DOES NOT MATCH THE SIZE THAT CORRELATES WITH THE LASER MARKED PART NUMBER. THE LASER MARKED PART NUMBER IS 07.02000.118, WHICH SHOULD BE A 7.5 X 50 SCREW. HOWEVER, THE SIZE IS LASER MARKED 7.5 X 40MM. THIS WAS FOUND IN A WAREHOUSE AND DID NOT HAVE ANY PATIENT INVOLVEMENT. THIS IS REPORT TWO OF SEVEN FOR THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17942327 ↗MalfunctionOct 16, 2023Inaccurate Information; Inaccurate Information
IT WAS REPORTED THE LENGTH OF THE SIZE LASER MARKED ON SEVEN VITALITY SCREWS DOES NOT MATCH THE SIZE THAT CORRELATES WITH THE LASER MARKED PART NUMBER. THE LASER MARKED PART NUMBER IS 07.02000.118, WHICH SHOULD BE A 7.5 X 50 SCREW. HOWEVER, THE SIZE IS LASER MARKED 7.5 X 40MM. THIS WAS FOUND IN A WAREHOUSE AND DID NOT HAVE ANY PATIENT INVOLVEMENT. THIS IS REPORT ONE OF SEVEN FOR THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17942104 ↗MalfunctionOct 16, 2023Inaccurate Information; Inaccurate Information
IT WAS REPORTED THAT THE LENGTH OF THE SIZE LASER MARKED ON TWO VITALITY SCREWS DOES NOT MATCH THE SIZE THAT CORRELATES WITH THE LASER MARKED PART NUMBER. THE LASER MARKED PART NUMBER IS 07.02000.118, WHICH SHOULD BE A 7.5 X 50 SCREW. HOWEVER, THE SIZE IS LASER MARKED 7.5 X 40MM. THIS WAS FOUND IN A WAREHOUSE AND DID NOT HAVE ANY PATIENT INVOLVEMENT. THIS IS REPORT TWO OF TWO FOR THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17940798 ↗MalfunctionOct 16, 2023Inaccurate Information; Inaccurate Information
IT WAS REPORTED THAT THE LENGTH OF THE SIZE LASER MARKED ON TWO VITALITY SCREWS DOES NOT MATCH THE SIZE THAT CORRELATES WITH THE LASER MARKED PART NUMBER. THE LASER MARKED PART NUMBER IS 07.02000.118, WHICH SHOULD BE A 7.5 X 50 SCREW. HOWEVER, THE SIZE IS LASER MARKED 7.5 X 40MM. THIS WAS FOUND IN A WAREHOUSE AND DID NOT HAVE ANY PATIENT INVOLVEMENT. THIS IS REPORT ONE OF TWO FOR THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17940796 ↗MalfunctionOct 6, 2023Inaccurate Information; Inaccurate Information
IT WAS REPORTED THAT THE LENGTH OF THE SIZE LASER MARKED ON A VITALITY SCREW DOES NOT MATCH THE SIZE THAT CORRELATES WITH THE LASER MARKED PART NUMBER. THE LASER MARKED PART NUMBER IS 07.02000.118, WHICH SHOULD BE A 7.5 X 50 SCREW. HOWEVER, THE SIZE IS LASER MARKED 7.5 X 40MM. THIS WAS FOUND DISCOVERED IN THE STERILE DEPARTMENT WHEN PLACING THE SCREW INTO THE STERILIZATION CADDY. THERE WAS NO PATIENT INVOLVEMENT.THIS IS REPORT EIGHT OF EIGHT FOR THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17890657 ↗MalfunctionOct 6, 2023Inaccurate Information; Inaccurate Information
IT WAS REPORTED THAT THE LENGTH OF THE SIZE LASER MARKED ON A VITALITY SCREW DOES NOT MATCH THE SIZE THAT CORRELATES WITH THE LASER MARKED PART NUMBER. THE LASER MARKED PART NUMBER IS 07.02000.118, WHICH SHOULD BE A 7.5 X 50 SCREW. HOWEVER, THE SIZE IS LASER MARKED 7.5 X 40MM. THIS WAS FOUND DISCOVERED IN THE STERILE DEPARTMENT WHEN PLACING THE SCREW INTO THE STERILIZATION CADDY. THERE WAS NO PATIENT INVOLVEMENT. THIS IS REPORT SEVEN OF EIGHT FOR THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17890650 ↗MalfunctionOct 6, 2023Inaccurate Information; Inaccurate Information
IT WAS REPORTED THAT THE LENGTH OF THE SIZE LASER MARKED ON A VITALITY SCREW DOES NOT MATCH THE SIZE THAT CORRELATES WITH THE LASER MARKED PART NUMBER. THE LASER MARKED PART NUMBER IS 07.02000.118, WHICH SHOULD BE A 7.5 X 50 SCREW. HOWEVER, THE SIZE IS LASER MARKED 7.5 X 40MM. THIS WAS FOUND DISCOVERED IN THE STERILE DEPARTMENT WHEN PLACING THE SCREW INTO THE STERILIZATION CADDY. THERE WAS NO PATIENT INVOLVEMENT. THIS IS REPORT SIX OF EIGHT FOR THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17890617 ↗MalfunctionOct 6, 2023Inaccurate Information; Inaccurate Information
IT WAS REPORTED THAT THE LENGTH OF THE SIZE LASER MARKED ON A VITALITY SCREW DOES NOT MATCH THE SIZE THAT CORRELATES WITH THE LASER MARKED PART NUMBER. THE LASER MARKED PART NUMBER IS 07.02000.118, WHICH SHOULD BE A 7.5 X 50 SCREW. HOWEVER, THE SIZE IS LASER MARKED 7.5 X 40MM. THIS WAS FOUND DISCOVERED IN THE STERILE DEPARTMENT WHEN PLACING THE SCREW INTO THE STERILIZATION CADDY. THERE WAS NO PATIENT INVOLVEMENT.THIS IS REPORT FIVE OF EIGHT FOR THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17890614 ↗MalfunctionOct 6, 2023Inaccurate Information; Inaccurate Information
IT WAS REPORTED THAT THE LENGTH OF THE SIZE LASER MARKED ON A VITALITY SCREW DOES NOT MATCH THE SIZE THAT CORRELATES WITH THE LASER MARKED PART NUMBER. THE LASER MARKED PART NUMBER IS 07.02000.118, WHICH SHOULD BE A 7.5 X 50 SCREW. HOWEVER, THE SIZE IS LASER MARKED 7.5 X 40MM. THIS WAS FOUND DISCOVERED IN THE STERILE DEPARTMENT WHEN PLACING THE SCREW INTO THE STERILIZATION CADDY. THERE WAS NO PATIENT INVOLVEMENT.THIS IS REPORT FOUR OF EIGHT FOR THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17890586 ↗MalfunctionOct 6, 2023Inaccurate Information; Inaccurate Information
IT WAS REPORTED THAT THE LENGTH OF THE SIZE LASER MARKED ON A VITALITY SCREW DOES NOT MATCH THE SIZE THAT CORRELATES WITH THE LASER MARKED PART NUMBER. THE LASER MARKED PART NUMBER IS 07.02000.118, WHICH SHOULD BE A 7.5 X 50 SCREW. HOWEVER, THE SIZE IS LASER MARKED 7.5 X 40MM. THIS WAS FOUND DISCOVERED IN THE STERILE DEPARTMENT WHEN PLACING THE SCREW INTO THE STERILIZATION CADDY. THERE WAS NO PATIENT INVOLVEMENT. THIS IS REPORT THREE OF EIGHT FOR THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17890544 ↗MalfunctionOct 6, 2023Inaccurate Information; Inaccurate Information
IT WAS REPORTED THAT THE LENGTH OF THE SIZE LASER MARKED ON A VITALITY SCREW DOES NOT MATCH THE SIZE THAT CORRELATES WITH THE LASER MARKED PART NUMBER. THE LASER MARKED PART NUMBER IS 07.02000.118, WHICH SHOULD BE A 7.5 X 50 SCREW. HOWEVER, THE SIZE IS LASER MARKED 7.5 X 40MM. THIS WAS FOUND DISCOVERED IN THE STERILE DEPARTMENT WHEN PLACING THE SCREW INTO THE STERILIZATION CADDY. THERE WAS NO PATIENT INVOLVEMENT. THIS IS REPORT TWO OF EIGHT FOR THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17890520 ↗MalfunctionOct 6, 2023Inaccurate Information; Inaccurate Information
IT WAS REPORTED THAT THE LENGTH OF THE SIZE LASER MARKED ON A VITALITY SCREW DOES NOT MATCH THE SIZE THAT CORRELATES WITH THE LASER MARKED PART NUMBER. THE LASER MARKED PART NUMBER IS 07.02000.118, WHICH SHOULD BE A 7.5 X 50 SCREW. HOWEVER, THE SIZE IS LASER MARKED 7.5 X 40MM. THIS WAS FOUND DISCOVERED IN THE STERILE DEPARTMENT WHEN PLACING THE SCREW INTO THE STERILIZATION CADDY. THERE WAS NO PATIENT INVOLVEMENT. THIS IS REPORT ONE OF EIGHT FOR THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17890490 ↗
Other Orthosis, spinal pedicle fixation, for degenerative disc disease devices
- EXPEDIUM VERSE 199721000Medos International Sàrl · Orthosis, spinal pedicle fixation, for degenerative disc disease387 reports
- Sequoia® 3301-1Zimmer Spine, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease184 reports
- CD HORIZON® Spinal System 8590855Medtronic Sofamor Danek, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease179 reports
- Vitality® 07.02011.001Biomet Spine LLC · Orthosis, spinal pedicle fixation, for degenerative disc disease3 recalls0 reports
- EXPEDIUM VERSE 199721001SMedos International Sàrl · Thoracolumbosacral pedicle screw system124 reports
- LineSider LS-N6SA7550Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls1 report
- LineSider LS-N6SA6545Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls0 reports
- EXPEDIUM VERSE 199721000SMedos International Sàrl · Thoracolumbosacral pedicle screw system97 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page