GammaPod - Treatment Planning System-a component of the… recallZ-0750-2024 · Xcision Medical Systems, LLC · initiated Nov 30, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Xcision Medical Systems, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Wrong structures imported into the GammaPod Treatment Planning System.-system allowed an RTSS generated from a previous CT scan to be selected

FDA root cause category: Software design

What the company is doing

Xcision issued Urgent Field Correction Letter via email on 11/30/23. Letter states reason for recall, health risk and action to take: Please post this notification internally for awareness of GammaPod BY CUSTOMER/USER Treatment Planning System users. Xcision is working on an update to the TPS software to implement controls to identify not only the patient identification but also the series of which the structures were delineated. This change is expected for implementation in early 2024. If you need any further information or support concerning this issue, please contact Xcision Customer Support at 855-924-7466 (1-855- SUPPORT XCISION).

Recall details

Recall number
Z-0750-2024
Status
Open, Classified
Initiated by the company
Nov 30, 2023
Posted by the FDA
Jan 18, 2024
Recalling firm
Xcision Medical Systems, LLC
Quantity
4 units
Product code
IWB
FDA submissions
K172706, K180571
Distribution
Worldwide - US Nationwide distribution in the states of MD, PA, TX and the country of Italy.

Product description

GammaPod - Treatment Planning System-a component of the GammaPod Irradiation System. Ref: XMSGP030A09.07-0.01 Intended for use in the noninvasive stereotactic delivery of a radiation dose to a partial volume of the breast in conjunction with breast conserving treatment.

Code information (models, lots, serial numbers)

00850002615029(11)230321(21)TPS0018 (Distributed in the US) 00850002615029(11)230629(21)TPS0013 (Distributed in the US) 00850002615005(11)21H24(21)TPS0016 (Distributed in the US) 00850002615029(11)210629(21 )TPS0014 (Italy)

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page