Custom procedural convenience kits and trays, general… recallZ-0827-2024 · American Contract Systems, Inc. · initiated Nov 29, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from American Contract Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.
FDA root cause category: Process control
What the company is doing
Owens & Minor issued an URGENT: MEDICAL DEVICE CORRECTION notice to it consignees on 11/29/2023 via email. Then notice explained the problem, risk to health, and requested the following actions be taken: If you received any of the batches/lots identified, please segregate and quarantine all affected product. Add warning labels to all kits/cases impacted: A labeling template is enclosed for printing warning labels which should be affixed to all affected product in inventory. This label should be affixed to each affected kit and case in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing product labeling. " Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have. " Distributors: If any of the affected kits identified above were further distributed, please ensure all end-users are appropriately notified of this field correction and maintain records of effectiveness. Report any quality problems or adverse events associated with the products listed in this notification to [email protected].
Recall details
- Recall number
- Z-0827-2024
- Status
- Open, Classified
- Initiated by the company
- Nov 29, 2023
- Posted by the FDA
- Jan 26, 2024
- Recalling firm
- American Contract Systems, Inc.
- Quantity
- 2717 units
- Product code
- OJG
- Distribution
- US Nationwide distribution.
Product description
Custom procedural convenience kits and trays, general hospital use, labeled as: a) IR NEURO PACK, kit number AGNE20J; b) ADULT CRANIOTOMY PACK, kit number LLCR99; c) LAMINECTOMY PACK, kit number LLLC26; d) KIT, CRANIOTOMY, kit number MMCR11J; e) KIT, NEURO SPINE, kit number MMNS20H; f) IR NEURO PACK, kit number OSIN77; g) NEURO SAH, kit number SANE32H; h) NEURO SAH, kit number SANE32I; i) S78FCH CRANI PACK (PS 132432), kit number UICR10U; j) CRANI PACK (PS039498), kit number UICR33AW; k) Laminectomy, kit number UILM77AR; l) DIAGNOSTIC NEURO IR TRAY, kit number WENE21D; m) DIAGNOSTIC NEURO IR TRAY, kit number WENE21F; n) DIAGNOSTIC NEURO IR TRAY, kit number WENE21H; o) DIAGNOSTIC NEURO IR TRAY, kit number WENE21H-01; p) NEURO IR TRAY, kit number WENI56F; q) NEURO IR TRAY, kit number WENI56H; r) NEURO IR TRAY, kit number WENI56H-01
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
a) AGNE20J, UDI/DI 00191072183187, Lot Numbers: 934231, exp. 10/30/2023; 900231, exp. 4/10/2024; 990231, exp. 1/10/2024; 687221, exp. 11/9/2023; 720221, exp. 10/7/2023; 757221, exp. 8/31/2023; 870221, exp. 5/10/2023; 702221, exp. 10/25/2023; 795221, exp. 7/24/2023; b) LLCR99, UDI/DI 00191072162755, Lot Numbers: 756221, exp. 9/1/2023; 720221, exp. 10/7/2023; 654221, exp. 12/12/2023; 949231, exp. 2/20/2024; c) LLLC26, UDI/DI 00191072171900, Lot Numbers: 749221, exp. 9/8/2023; 695221, exp. 11/1/2023; 997231, exp. 1/3/2024; 920231, exp. 3/21/2024; d) MMCR11J, UDI/DI none, Lot Numbers: 744221, exp. 9/13/2023; 939231, exp. 3/2/2024; 876231, exp. 5/4/2024; e) MMNS20H, UDI/DI none, Lot Numbers: 963231, exp. 2/6/2024; 959231, exp. 2/10/2024; f) OSIN77, UDI/DI 00191072170705, Lot Numbers: 780221, exp. 8/8/2023; 728221, exp. 9/29/2023; 668221, exp. 11/28/2023; 991231, exp. 1/9/2024, 960231, exp. 2/9/2024; 939231, exp. 3/2/2024; UDI/DI 00191072190031, Lot Numbers: 912231, exp. 3/29/2024; 903231, exp. 4/7/2024; g) SANE32H, UDI/DI none, Lot Numbers: 838221, exp. 6/11/2023; 889221, exp. 4/21/2023; h) SANE32I, UDI/DI none, Lot Numbers: 673221, exp. 11/23/2023; 730221, exp. 9/27/2023; 784221, exp. 8/4/2023; 911231, exp. 3/30/2024; 969231, exp. 1/31/2024; i) UICR10U, UDI/DI none, Lot Numbers: 990231, exp. 1/10/2024; 921231, exp. 3/20/2024; 885231, exp. 4/25/2024; j) UICR33AW, UDI/DI none, Lot Numbers: 658221, exp. 12/8/2023; 931231, exp. 3/10/2024; 876231, exp. 5/4/2024; k) UILM77AR, UDI/DI 00191072139184, Lot Numbers: 953231, exp. 2/16/2024; 905231, exp. 4/5/2024; l) WENE21D, UDI/DI 00191072167958, Lot Numbers: 839221, exp. 6/10/2023; m) WENE21F, UDI/DI 00191072167958, Lot Numbers: 819221, exp. 6/30/2023; n) WENE21H, UDI/DI 00191072179326, Lot Numbers: 729221, exp. 9/28/2023; 686221, exp. 11/10/2023; o) WENE21H-01, UDI/DI 00191072179326, Lot Numbers: 980231, exp. 1/20/2024; p) WENI56F, UDI/DI 00191072169341, Lot Numbers: 799221, exp. 7/20/2023; q) WENI56H, UDI/DI 00191072179333, Lot Numbers: 700221, exp. 10/27/2023; r) WENI56H-01, UDI/DI 00191072179333, Lot Numbers: 652221, exp. 12/14/2023; 996231, exp. 1/4/2024
Devices named in this recall
7 devices in the FDA device registry whose identifier appears in the recall's code information.
- ACS UILM77ARAmerican Contract Systems, Inc. · Neurological tray1 recall0 reports
- ACS LLCR99HAmerican Contract Systems, Inc. · Neurological tray1 recall0 reports
- ACS WENE21DAmerican Contract Systems, Inc. · Neurological tray1 recall0 reports
- ACS WENI56FAmerican Contract Systems, Inc. · Neurological tray1 recall0 reports
- ACS LLLC26IAmerican Contract Systems, Inc. · Neurological tray1 recall0 reports
- ACS AGNE20JAmerican Contract Systems, Inc. · Neurological tray1 recall0 reports
- ACS OSIN77FAmerican Contract Systems, Inc. · Neurological tray1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 93656).
- Z-0817-2024 · Open, ClassifiedCustom procedural convenience kits and trays, cardiac, labeled as: a) HEART CATHETERIZATION TRAY, kit number JRHC45G; b) TAVR PACK, kit number LLTV10; c) CATH PED LF SJH, kit n…
- Z-0818-2024 · Open, ClassifiedProcedural convenience kits and trays, cardiac, labeled as: a) EP LAB BI V IMPLANT PACK, kit number AGBI64J; b) LAB DEVICE IMPLANT, kit number AGDI40J; c) LAB DEVICE IMPLANT, k…
- Z-0819-2024 · Open, ClassifiedCustom procedural convenience kits and trays, cardiac, labeled as: a) CENTRAL LINE INSERTION PACK, kit number LLCE52; b) Adult Central Line Insertion Supply Kit, kit number UICL…
- Z-0820-2024 · Open, ClassifiedCustom procedural convenience kits and trays, ear, nose & throat, labeled as: a) T AND A PACK AGH ASC, kit number AGTA48J; b) OPSC EAR PACK, kit number LLEO66; c) T AND A PACK,…
- Z-0821-2024 · Open, ClassifiedCustom procedural convenience kits and trays, gastroenterological & urological, labeled as: AHN BIOPSY STERILE, kit number AHBP12A
- Z-0822-2024 · Open, ClassifiedCustom procedural convenience kits and trays, gastroenterological & urological, labeled as: a) LAP KIDNEY MODULE, kit number AGKD50L; b) ROBOTIC PROSTATE PACK - 264852, kit numb…
- Z-0823-2024 · Open, ClassifiedCustom procedural convenience kits and trays, urological, labeled as: a) UROLOGY DAVINCI PACK, kit number HGUR87G; b) CYSTO PACK, kit number LVCY20; c) URO ROBOTIC PACK, kit nu…
- Z-0825-2024 · Open, ClassifiedCustom procedural convenience kits and trays, general & plastic surgery, labeled as: a) BRACHY THERAPY PROCEDURE PAC, kit number AGBR90E; b) BRACHY THERAPY PROCEDURE PAC, kit nu…
- Z-0826-2024 · Open, ClassifiedCustom procedural convenience kits and trays, general hospital use, labeled as: a) Baby PICC Kit, kit number UIBP90J; b) Baby PICC Kit, kit number UIBP90K; c) PORT PACK (PS 058…
- Z-0828-2024 · Open, ClassifiedCustom procedural convenience kits and trays, Obstetrical & Gynecological use, labeled as: a) LOWER PELVIC PACK, kit number AGLP52L; b) OB/GYN DAVINCI PACK A RG BASIN, kit numb…
- Z-0830-2024 · Open, ClassifiedCustom procedural convenience kits and trays, Obstetrical & Gynecological use, labeled as: a) C-SECTION PACK, kit number FDCS01AA; b) C-SECTION PACK, kit number FDCS01AB; c) C-…
- Z-0832-2024 · Open, ClassifiedCustom procedural convenience kits and trays, Ophthalmic use, labeled as: a) EYE PACK, kit number LVEY27; b) Laminectomy, kit number UIOL34AG
- Z-0833-2024 · Open, ClassifiedCustom procedural convenience kits and trays, Othopedic use, labeled as: a) HIP FRACTURE, kit number AHHF15C; b) HIP FRACTURE, kit number AHHF15D; c) Hand Pack, kit number BBHP…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page