Custom procedural convenience kits and trays, general &… recallZ-0825-2024 · American Contract Systems, Inc. · initiated Nov 29, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from American Contract Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.

FDA root cause category: Process control

What the company is doing

Owens & Minor issued an URGENT: MEDICAL DEVICE CORRECTION notice to it consignees on 11/29/2023 via email. Then notice explained the problem, risk to health, and requested the following actions be taken: If you received any of the batches/lots identified, please segregate and quarantine all affected product. Add warning labels to all kits/cases impacted: A labeling template is enclosed for printing warning labels which should be affixed to all affected product in inventory. This label should be affixed to each affected kit and case in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing product labeling. " Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have. " Distributors: If any of the affected kits identified above were further distributed, please ensure all end-users are appropriately notified of this field correction and maintain records of effectiveness. Report any quality problems or adverse events associated with the products listed in this notification to [email protected].

Recall details

Recall number
Z-0825-2024
Status
Open, Classified
Initiated by the company
Nov 29, 2023
Posted by the FDA
Jan 26, 2024
Recalling firm
American Contract Systems, Inc.
Quantity
7881 units
Product code
LRO
Distribution
US Nationwide distribution.

Product description

Custom procedural convenience kits and trays, general & plastic surgery, labeled as: a) BRACHY THERAPY PROCEDURE PAC, kit number AGBR90E; b) BRACHY THERAPY PROCEDURE PAC, kit number AGBR90E; c) MINI MINOR AGH ASC PACK, kit number AGMM60M; d) OUTPATIENT PORT ACCESS KIT 307907, kit number AHPO55J; e) General Pack, kit number BBGP31C; f) FRACTURE TABLE PACK, kit number FDFT75N; g) Arthroscopy Pack, kit number HCAR04K; h) BARIATRIC PACK, kit number HGBA92F; i) LAPAROTOMY PACK, kit number HNLP12J; j) MIDLINE CHANGE KIT, kit number HSMD20E; k) Minor General, kit number JSMI10I; l) ENDO VASCULAR PACK, kit number LVEV61; m) MAJOR PACK, kit number LVMJ21; n) HEAD AND NECK PACK, kit number MMHN37F; o) KIT, MINOR U BAR, kit number MMNU34E; p) KIT, MAJOR UNIVERSAL, kit number MMUN28E; q) VAG PLASTY PACK, kit number OSVP16; r) DR L MAJOR, kit number SALM40H; s) DR L MAJOR, kit number SALM40I; t) MINOR BASIN SAH, kit number SAMB25E; u) MAJOR SAH, kit number SAMJ36G; v) MINOR PACK, kit number SMMN65; w) General Pack, kit number THGN59D; x) MINOR PACK, kit number TPMN15; y) MAJOR UNIVERSAL PACK, kit number TPUN48

Code information (models, lots, serial numbers)

Show the full code information (3k characters)

a) AGBR90E, UDI/DI 00191072036865, Lot Numbers: 959231, exp. 02/10/2024; 681221, exp. 11/15/2023; 721221, exp. 09/30/2023; 771221, exp. 08/147/2023; 827221, exp. 06/22/2023; 897221, exp. 04/13/2023; b) AGMM60M, UDI/DI 00191072063786, Lot Numbers: 939231, exp. 03/02/2024; 638221, exp. 4528812/28/2023; 694221, exp. 11/02/2023; 734221, exp. 09/23/2023; 769221, exp. 08/19/2023; c) AHPO55J, UDI/DI 00191072171566, Lot Numbers: 682221, exp. 11/14/2023; 661221, exp. 12/5/2023; 977231, exp. 1/23/2024; 945231, exp. 2/24/2024; 891231, exp. 4/19/2024; d) BBGP31C, UDI/DI none, Lot Numbers: 687221, exp. 11/9/2023; 738221, exp. 9/19/2023; 805221, exp. 7/14/2023; 870221, exp. 5/10/2023; e) FDFT75N, UDI/DI none, Lot Numbers: 773221, exp. 8/15/2023; 714221, exp. 10/13/2023; 990231, exp. 1/10/2024; 935231, exp. 3/6/2024; 890231, exp. 4/20/2024; 870231, exp. 5/10/2024; f) HCAR04K, UDI/DI 00191072114174, Lot Numbers: 757221, exp. 08/31/2023; g) HGBA92F, UDI/DI none, Lot Numbers: 757221, exp. 8/31/2023; 695221, exp. 11/1/2023; 635221, exp. 12/31/2023; h) HNLP12J, UDI/DI 00191072146274, Lot Numbers: 805221, exp. 7/14/2023; 742221, exp. 9/15/2023; 673221, exp. 11/23/2023; 924231, exp. 3/17/2024; i) HSMD20E, UDI/DI 00191072130068, Lot Numbers: 898221, exp. 4/12/2023; 772221, exp. 8/16/2023; 984231, exp. 1/16/2024; j) JSMI10I, UDI/DI none, Lot Numbers: 766221, exp. 08/22/2023; k) LVEV61, UDI/DI 00191072033475, Lot Numbers: 678221, exp. 11/18/2023; 637221, exp. 12/29/2023; l) LVMJ21, UDI/DI 00191072139023, Lot Numbers: 833221, exp. 6/16/2023; 764221, exp. 8/24/2023; 717221, exp. 10/10/2023; 988231, exp. 1/12/2024; 954231, exp. 2/15/2024; m) MMHN37F, UDI/DI none, Lot Numbers: 639221, exp. 12/27/2023; n) MMNU34E, UDI/DI none, Lot Numbers: 928231, exp. 3/13/2024; 889231, exp. 4/21/2024; o) MMUN28E, UDI/DI none, Lot Numbers: 889231, exp. 04/21/2024; p) OSVP16, UDI/DI 00191072156570, Lot Numbers: 827221, exp. 6/17/2023; q) OSVP16, UDI/DI 00191072170835, Lot Numbers: 700221, exp. 10/23/2023; 644221, exp. 12/4/2023; 987231, exp. 11/20/2023; r) SALM40H, UDI/DI none, Lot Numbers: 826221, exp. 6/23/2023; 888221, exp. 4/22/2023; s) SALM40I, UDI/DI none, Lot Numbers: 687221, exp. 11/9/2023; 756221, exp. 9/1/2023; 984231, exp. 1/16/2024; t) SAMB25E, UDI/DI none, Lot Numbers: 843221, exp. 6/6/2023; 894221, exp. 4/16/2023; 871221, exp. 5/9/2023; 686221, exp. 11/10/2023; 758221, exp. 8/30/2023; 793221, exp. 7/26/2023; 974231, exp. 1/26/2024; u) SAMJ36G, UDI/DI none, Lot Numbers: 660221, exp. 12/6/2023; 704221, exp. 10/23/2023; 752221, exp. 9/5/2023; 800221, exp. 7/19/2023; v) SMMN65, UDI/DI 00191072188472, Lot Numbers: 969231, exp. 1/31/2024; 926231, exp. 3/15/2024; w) SMMN65, UDI/DI none, Lot Numbers: 919231, exp. 3/22/2024; x) THGN59D, UDI/DI 00191072191069, Lot Numbers: 805221, exp. 7/14/2023; 721221, exp. 10/6/2023; y) TPMN15, UDI/DI 00191072171771, Lot Numbers: 737221, exp. 9/20/2023; 702221, exp. 10/25/2023; 664221, exp. 12/2/2023; 990231, exp. 1/10/2024; 960231, exp. 2/9/2024; 939231, exp. 3/2/2024; 903231, exp. 4/7/2024; z) TPUN48, UDI/DI 00191072171832, Lot Numbers: 738221, exp. 9/19/2023; 688221, exp. 11/8/2023; 637221, exp. 12/29/2023; 967231, exp. 2/2/2024; 924231, exp. 3/17/2024

Devices named in this recall

12 devices in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 93656).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page