Scorpio recallZ-0899-2024 · Howmedica Osteonics Corp. · initiated Dec 1, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Howmedica Osteonics Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
FDA root cause category: Nonconforming Material/Component
What the company is doing
Stryker issued Urgent Medical Device Recall Letter (Product Field Action #: 3306872)through third party Sedgwick via UPS on 12/1/23. Letter states reason for recall, health risk and action to take: Please inform users of this Urgent Medical Device Recall and forward this notice to all individuals who need to be made aware. 2. Immediately check all stock areas and/or operating room storage to determine if any devices from the affected product list are at your facility. 3. Quarantine and discontinue use of the affected product in Part/Lot Number Attachment: PFA 3306872 starting on page 3. 4. Complete and sign the enclosed Urgent Medical Device Recall Business Reply Form and email to [email protected] 5. Please contact your Local Sales Office or your Stryker Sales Representative directly for product replacement and inventory questions. 6. Please return ALL affected product to: Stryker Orthopaedics/PFA Product Returns Attn: Distribution Inventory Team 325 Corporate Drive Dock M-East Mahwah, NJ 07430 Ref. PFA 3306872; Please include Tracking information on the package. If you have any questions or concerns after reviewing this letter, please contact Customer Service at (201) 831-5000. For questions pertaining to the recall, email [email protected]
Recall details
- Recall number
- Z-0899-2024
- Status
- Open, Classified
- Initiated by the company
- Dec 1, 2023
- Posted by the FDA
- Jan 31, 2024
- Recalling firm
- Howmedica Osteonics Corp.
- Product code
- JWH
- FDA submissions
- K030978
- Distribution
- Nationwide Foreign: Argentina, Australia, Brazil, Canada, Chile, China, Columbia, India, Japan, Korea, Latin America, Mexico, Poland, Sweden, Switzerland, Taiwan, UK, & Venlo (Netherlands). ***Updated 1/25/24*** Foreign distribution added: Austria, Finland, Germany, Ireland, Hong Kong, New Zealand, Norway, and Spain.
Product description
Scorpio NRG Tibial Brg Insert Assy -Intended for knee replacement Part Number: 82-2-0908
Code information (models, lots, serial numbers)
GTIN: 04546540400734 Lot Numbers: 42874701, 42912101, 42995601
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Scorpio 82-2-0908Howmedica Osteonics Corp. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall1 report
Product lines
Other devices under the same FDA submission
11 devices cleared under K030978 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Scorpio 82-3-0318Howmedica Osteonics Corp. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall0 reports
- Scorpio 82-3-0918Howmedica Osteonics Corp. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER4 reports
- Scorpio 82-3-0310Howmedica Osteonics Corp. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER2 reports
- Scorpio 82-3-0512Howmedica Osteonics Corp. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER2 reports
- Scorpio 82-3-0910Howmedica Osteonics Corp. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER2 reports
- Scorpio 82-3-0312Howmedica Osteonics Corp. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 report
- Scorpio 82-3-0508Howmedica Osteonics Corp. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 report
- Scorpio 82-3-0510Howmedica Osteonics Corp. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 report
- Scorpio 82-3-0515Howmedica Osteonics Corp. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 report
- Scorpio 82-3-0715Howmedica Osteonics Corp. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 report
- Scorpio 81-4405RHowmedica Osteonics Corp. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 93687).
- Z-0869-2024 · Open, ClassifiedSCORPIO RECESSED PATELLA- Intended for Knee Replacement Part Number: 3044-0026
- Z-0870-2024 · Open, ClassifiedSCORPIO RECESSED PATELLA- Intended for knee replacement Part Number: 3044-0028
- Z-0871-2024 · Open, ClassifiedSCORPIO RECESSED PATELLA- Intended for knee replacement Part Number: 3044-0030
- Z-0872-2024 · Open, ClassifiedSCORPIO RECESSED PATELLA-Intended for knee repalcment Part Number: 3044-0032
- Z-0873-2024 · Open, ClassifiedSERIES II TIBIAL BEAR INSERT-Intended for knee replacement Part Number: 3052-0515
- Z-0874-2024 · Open, ClassifiedSERIES II TIBIAL BEAR INSERT-Intended for knee replacement Part Number: 3052-0524
- Z-0875-2024 · Open, ClassifiedSCORPIO TS TIB INSERT -Intended for knee replacement Part Number: 72-4-0321
- Z-0876-2024 · Open, ClassifiedSCORPIO TS TIB INSERT -Intended for knee replacement Part Number: 72-4-0510
- Z-0877-2024 · Open, ClassifiedSCORPIO TS TIB INSERT-Intended for knee replacement Part Number:72-4-7512
- Z-0878-2024 · Open, ClassifiedSCORPIO TS TIB INSERT -Intended for knee replacement Part Number: 72-4-7514
- Z-0879-2024 · Open, ClassifiedSCORPIO TS TIB INSERT -Intended for knee replacment Part Number: 72-4-7516
- Z-0880-2024 · Open, ClassifiedSCORPIO TS TIB INSERT. Intended for knee replacement Part Number: 72-4-7518
- Z-0881-2024 · Open, ClassifiedSCORPIO TS TIB INSERT -Intended for knee replacement Part Number: 72-4-7521
- Z-0882-2024 · Open, ClassifiedSCORPIO TS TIB INSERT -Intended for knee replacement Part Number: 72-4-7524
- Z-0883-2024 · Open, ClassifiedScorpio-Flex Ttl Kn P-S Tib Brg Insrt Asy- Intended for knee replacement Part Number: 72-15-0324
- Z-0884-2024 · Open, ClassifiedScorpio-Flex Ttl Kn P-S Tib Brg Insrt Asy- -Intended for knee replacement Part Number: 72-16-0908
- Z-0885-2024 · Open, ClassifiedScorpio-Flex Ttl Kn P-S Tib Brg Insrt Asy- -Intended for knee replacement Part Number: 72-16-0910
- Z-0886-2024 · Open, ClassifiedSCORPIO TOTAL KNEE MEDIALIZED DOME PATELLA-SZ #11 -Intended for knee replacement Part Number: 73-0110
- Z-0887-2024 · Open, ClassifiedSCORPIO TOTAL KNEE MEDIALIZED DOME PATELLA-SZ #5 -Intended for knee replacement Part Number: 73-0510
- Z-0888-2024 · Open, ClassifiedSCORPIO TOTAL KNEE MEDIALIZED DOME PATELLA-SZ #7 -Intended for knee replacement Part Number: 73-0710
- Z-0889-2024 · Open, ClassifiedSCORPIO TOTAL KNEE MEDIALIZED DOME PATELLA-SZ #9 -Intended for knee replacement Part Number: 73-0910
- Z-0890-2024 · Open, ClassifiedSCORPIO TOTAL KNEE CONCENTRIC DOME PATELLA-SZ #11 -Intended for knee replacement Part Number: 73-2110
- Z-0891-2024 · Open, ClassifiedSCORPIO TOTAL KNEE CONCENTRIC DOME PATELLA-SZ #7 -Intended for knee replacement Part Number: 73-2710
- Z-0892-2024 · Open, ClassifiedSCORPIO TOTAL KNEE CONCENTRIC DOME PATELLA-SZ #9 -Intended for knee replacement Part Number: 73-2910
- Z-0893-2024 · Open, ClassifiedSCORPIO U-DOME PATELLA -Intended for knee replacement Part Number: 73-3110
- Z-0894-2024 · Open, ClassifiedSCORPIO U-DOME PATELLA -Intended for knee replacement Part Number: 73-3308
- Z-0895-2024 · Open, ClassifiedSCORPIO U-DOME PATELLA -Intended for knee replacement Part Number: 73-3508
- Z-0896-2024 · Open, ClassifiedSCORPIO U-DOME PATELLA -Intended for knee replacement Part Number: 73-3708
- Z-0897-2024 · Open, ClassifiedSCORPIO U-DOME PATELLA -Intended for knee replacement Part Number: 73-3710
- Z-0898-2024 · Open, ClassifiedSCORPIO U-DOME PATELLA -Intended for knee replacement Part Number: 73-3910
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page