Scorpio recallZ-0888-2024 · Howmedica Osteonics Corp. · initiated Dec 1, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Howmedica Osteonics Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
FDA root cause category: Nonconforming Material/Component
What the company is doing
Stryker issued Urgent Medical Device Recall Letter (Product Field Action #: 3306872)through third party Sedgwick via UPS on 12/1/23. Letter states reason for recall, health risk and action to take: Please inform users of this Urgent Medical Device Recall and forward this notice to all individuals who need to be made aware. 2. Immediately check all stock areas and/or operating room storage to determine if any devices from the affected product list are at your facility. 3. Quarantine and discontinue use of the affected product in Part/Lot Number Attachment: PFA 3306872 starting on page 3. 4. Complete and sign the enclosed Urgent Medical Device Recall Business Reply Form and email to [email protected] 5. Please contact your Local Sales Office or your Stryker Sales Representative directly for product replacement and inventory questions. 6. Please return ALL affected product to: Stryker Orthopaedics/PFA Product Returns Attn: Distribution Inventory Team 325 Corporate Drive Dock M-East Mahwah, NJ 07430 Ref. PFA 3306872; Please include Tracking information on the package. If you have any questions or concerns after reviewing this letter, please contact Customer Service at (201) 831-5000. For questions pertaining to the recall, email [email protected]
Recall details
- Recall number
- Z-0888-2024
- Status
- Open, Classified
- Initiated by the company
- Dec 1, 2023
- Posted by the FDA
- Jan 31, 2024
- Recalling firm
- Howmedica Osteonics Corp.
- Quantity
- 269 units US; 307 units OUS
- Product code
- JWH
- FDA submissions
- K962152
- Distribution
- Nationwide Foreign: Argentina, Australia, Brazil, Canada, Chile, China, Columbia, India, Japan, Korea, Latin America, Mexico, Poland, Sweden, Switzerland, Taiwan, UK, & Venlo (Netherlands). ***Updated 1/25/24*** Foreign distribution added: Austria, Finland, Germany, Ireland, Hong Kong, New Zealand, Norway, and Spain.
Product description
SCORPIO TOTAL KNEE MEDIALIZED DOME PATELLA-SZ #7 -Intended for knee replacement Part Number: 73-0710
Code information (models, lots, serial numbers)
GTIN: 07613327050295 Lot Numbers: RLAR, 2P84, EW6V, 90AV, 4JHN, 65A3, 22MP, NNE7, Y373, 58LA, VXAL, 4201, A2AN, 5YP8, X0P9, 9A8T,6LN7, WEL3, NDK3, 25WJ, T2JW, YVWT, 7DR0, DMVL.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Scorpio 73-0710Howmedica Osteonics Corp. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER2 recalls0 reports
Product lines
Other devices under the same FDA submission
7 devices cleared under K962152 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Scorpio 73-2710Howmedica Osteonics Corp. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER3 recalls0 reports
- Scorpio 73-2910Howmedica Osteonics Corp. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER2 recalls1 report
- Scorpio 73-0910Howmedica Osteonics Corp. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER2 recalls1 report
- Scorpio 73-0110Howmedica Osteonics Corp. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER2 recalls0 reports
- Scorpio 73-2110Howmedica Osteonics Corp. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall0 reports
- Scorpio 73-0510Howmedica Osteonics Corp. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall0 reports
- Scorpio 71-4509LHowmedica Osteonics Corp. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 93687).
- Z-0869-2024 · Open, ClassifiedSCORPIO RECESSED PATELLA- Intended for Knee Replacement Part Number: 3044-0026
- Z-0870-2024 · Open, ClassifiedSCORPIO RECESSED PATELLA- Intended for knee replacement Part Number: 3044-0028
- Z-0871-2024 · Open, ClassifiedSCORPIO RECESSED PATELLA- Intended for knee replacement Part Number: 3044-0030
- Z-0872-2024 · Open, ClassifiedSCORPIO RECESSED PATELLA-Intended for knee repalcment Part Number: 3044-0032
- Z-0873-2024 · Open, ClassifiedSERIES II TIBIAL BEAR INSERT-Intended for knee replacement Part Number: 3052-0515
- Z-0874-2024 · Open, ClassifiedSERIES II TIBIAL BEAR INSERT-Intended for knee replacement Part Number: 3052-0524
- Z-0875-2024 · Open, ClassifiedSCORPIO TS TIB INSERT -Intended for knee replacement Part Number: 72-4-0321
- Z-0876-2024 · Open, ClassifiedSCORPIO TS TIB INSERT -Intended for knee replacement Part Number: 72-4-0510
- Z-0877-2024 · Open, ClassifiedSCORPIO TS TIB INSERT-Intended for knee replacement Part Number:72-4-7512
- Z-0878-2024 · Open, ClassifiedSCORPIO TS TIB INSERT -Intended for knee replacement Part Number: 72-4-7514
- Z-0879-2024 · Open, ClassifiedSCORPIO TS TIB INSERT -Intended for knee replacment Part Number: 72-4-7516
- Z-0880-2024 · Open, ClassifiedSCORPIO TS TIB INSERT. Intended for knee replacement Part Number: 72-4-7518
- Z-0881-2024 · Open, ClassifiedSCORPIO TS TIB INSERT -Intended for knee replacement Part Number: 72-4-7521
- Z-0882-2024 · Open, ClassifiedSCORPIO TS TIB INSERT -Intended for knee replacement Part Number: 72-4-7524
- Z-0883-2024 · Open, ClassifiedScorpio-Flex Ttl Kn P-S Tib Brg Insrt Asy- Intended for knee replacement Part Number: 72-15-0324
- Z-0884-2024 · Open, ClassifiedScorpio-Flex Ttl Kn P-S Tib Brg Insrt Asy- -Intended for knee replacement Part Number: 72-16-0908
- Z-0885-2024 · Open, ClassifiedScorpio-Flex Ttl Kn P-S Tib Brg Insrt Asy- -Intended for knee replacement Part Number: 72-16-0910
- Z-0886-2024 · Open, ClassifiedSCORPIO TOTAL KNEE MEDIALIZED DOME PATELLA-SZ #11 -Intended for knee replacement Part Number: 73-0110
- Z-0887-2024 · Open, ClassifiedSCORPIO TOTAL KNEE MEDIALIZED DOME PATELLA-SZ #5 -Intended for knee replacement Part Number: 73-0510
- Z-0889-2024 · Open, ClassifiedSCORPIO TOTAL KNEE MEDIALIZED DOME PATELLA-SZ #9 -Intended for knee replacement Part Number: 73-0910
- Z-0890-2024 · Open, ClassifiedSCORPIO TOTAL KNEE CONCENTRIC DOME PATELLA-SZ #11 -Intended for knee replacement Part Number: 73-2110
- Z-0891-2024 · Open, ClassifiedSCORPIO TOTAL KNEE CONCENTRIC DOME PATELLA-SZ #7 -Intended for knee replacement Part Number: 73-2710
- Z-0892-2024 · Open, ClassifiedSCORPIO TOTAL KNEE CONCENTRIC DOME PATELLA-SZ #9 -Intended for knee replacement Part Number: 73-2910
- Z-0893-2024 · Open, ClassifiedSCORPIO U-DOME PATELLA -Intended for knee replacement Part Number: 73-3110
- Z-0894-2024 · Open, ClassifiedSCORPIO U-DOME PATELLA -Intended for knee replacement Part Number: 73-3308
- Z-0895-2024 · Open, ClassifiedSCORPIO U-DOME PATELLA -Intended for knee replacement Part Number: 73-3508
- Z-0896-2024 · Open, ClassifiedSCORPIO U-DOME PATELLA -Intended for knee replacement Part Number: 73-3708
- Z-0897-2024 · Open, ClassifiedSCORPIO U-DOME PATELLA -Intended for knee replacement Part Number: 73-3710
- Z-0898-2024 · Open, ClassifiedSCORPIO U-DOME PATELLA -Intended for knee replacement Part Number: 73-3910
- Z-0899-2024 · Open, ClassifiedScorpio NRG Tibial Brg Insert Assy -Intended for knee replacement Part Number: 82-2-0908
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page