Skip to content

ECHELON ENDOPATH recallZ-0834-2024 · Ethicon Endo-Surgery Inc · initiated Dec 11, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Ethicon Endo-Surgery Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

There is the potential for incomplete staple line formation during the firing sequence, without most or all staples, which may lead to insufficient tissue apposition that could require surgical intervention. This may result in bleeding, prolonged surgery, or surgical delay.

FDA root cause category: Under Investigation by firm

What the company is doing

On December 11, 2023, the firm notified affected consignees through URGENT VOLUNTARY MEDICAL DEVICE RECALL (REMOVAL) letters. Customers were instructed to cease use and distribution of affected product lots. Customers should examine their inventory immediately to determine if they have affected product. Affected product should be quarantined. Unused affected product should be returned. If you require any assistance with returning product, please contact Sedgwick at 888-591-5087 and reference Event # 6925. If you have additional questions regarding this voluntary medical device recall, please contact the Ethicon Resource Department at 1-877-ETHICON (1-877-384-4266). The Ethicon Resource Department is open Monday through Friday, 8:00 AM to 5:00 PM ET.

Recall details

Recall number
Z-0834-2024
Status
Open, Classified
Initiated by the company
Dec 11, 2023
Posted by the FDA
Jan 26, 2024
Recalling firm
Ethicon Endo-Surgery Inc
Quantity
22,616 (US), 4,476 (OUS)
Product code
GDW
FDA submissions
K163454
Distribution
Worldwide distribution - US Nationwide and the countries of Canada and Brazil.

Product description

ENDOPATH ECHELON 45mm RELOADS WITH GRIPPING SURFACE TECHNOLOGY (GST), ECHELON Endopath Stapler, Endoscopic Linear Cutter Reload (+Gripping Surface technology), 45MM WHITE VASCULAR/THIN, Product No. GST45W

Code information (models, lots, serial numbers)

UDI-DI 10705036014928 Lots 442C02 and 442C03

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page