Surgilon recallZ-0844-2024 · Covidien LP · initiated Dec 21, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Covidien LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Specific lots of sutures were sterilized with gamma doses that exceeded the range approved. The extra gamma levels may decrease the strength of the sutures over time, which could result in harms such as wound dehiscence and hemorrhage/blood loss/ bleeding at a critical level.
FDA root cause category: Process control
What the company is doing
On December 21, the firm notified affected consignees via URGENT MEDICAL DEVICE RECALL letters. Customers were instructed to identify, quarantine, and return all unused, non-expired affected sutures to Medtronic. Please contact [email protected] for the Return Good Authorization (RGA). Customers will receive credit for returned affected product. If purchased from a distributor, please contact your distributor directly to arrange for the return of product back to your distributor. Pass on the recall notice to all those who need to be aware within your organization, or to any organization where the affected sutures have been transferred or distributed. If you have questions regarding this recall, please call Medtronic Quality Assurance at 800-255-6774 option 1, then option 1 ***Updated 4/1/2024*** Firm identified additional models and lots affected by the gamma radiation issue and sent additional notification to customers on January 25, 2024.
Recall details
- Recall number
- Z-0844-2024
- Status
- Open, Classified
- Initiated by the company
- Dec 21, 2023
- Posted by the FDA
- Jan 29, 2024
- Recalling firm
- Covidien LP
- Quantity
- 46,812 US; 68,976 OUS
- Product code
- GAR
- FDA submissions
- K981582
- Distribution
- Worldwide US Nationwide distribution including in the states of AL, AZ, CA, FL, GA, HI, IA, IL, IN, MD, MI, MN, MO, MS, NC, NE, NJ, NV, NY, OH, OK, OR, PA, PR, TN, TX, VA, WA, WV and the countries of Australia, Bahrain, Belgium, Bulgaria, Canary Islands, China, Cyprus, Czech Republic, Finland, France, Germany, Greece, Hong Kong, Hungary, Italy, Japan, Korea, Republic Of Kosovo, Mongolia, Netherlands, New Zealand, Panama, Poland, Portugal, Puerto Rico, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, United Kingdom.
Product description
Surgilon Braided Nylon sutures: 88861919-31 SURGILON* 4-0 BLK 7X75CMPCT 88861919-41 SURGILON* 3-0 BLK 7X75CMPCT 88861919-51 SURGILON* 2-0 BLK 7X75CMPCT 88861919-71 SURGILON* 1 BLK 7X75CM PCT
Code information (models, lots, serial numbers)
Model 88861919-31 SURGILON* 4-0 BLK 7X75CMPCT UDI-DI 10884521071728, 20884521071725 Lots D1L1929RY D2B2243RY Model 88861919-41 SURGILON* 3-0 BLK 7X75CMPCT UDI-DI 10884521071742, 20884521071749 Lot D2B0176RY Model 88861919-51 SURGILON* 2-0 BLK 7X75CMPCT UDI-DI 10884521071766 Lots D2D2437RY UDI-DI 20884521071763 Lots D2D2437RY D2E0905RY D2E0916RY Model 88861919-71 SURGILON* 1 BLK 7X75CM PCT UDI-DI 20884521071794 Lots D1M0640RY D1M1494RY
Devices named in this recall
4 devices in the FDA device registry whose identifier appears in the recall's code information.
- Surgilon 88861919-41Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE2 recalls2 reports
- Surgilon 88861919-31Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE2 recalls0 reports
- Surgilon 88861919-51Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE2 recalls0 reports
- Surgilon 88861919-71Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE2 recalls0 reports
Product lines
Other devices under the same FDA submission
45 devices cleared under K981582 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Dermalon 88861749-24Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE1 recall4 reports
- Dermalon 88861741-11Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE1 recall2 reports
- Dermalon 88861757-31Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE1 recall1 report
- Dermalon 88861799-31Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE1 recall1 report
- Dermalon 88861747-11Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE1 recall0 reports
- Dermalon 88861757-41Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE1 recall0 reports
- Surgilon 88861883-51Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE1 recall0 reports
- Surgilon 88861915-71Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE1 recall0 reports
- Surgilon 88861919-61Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE1 recall0 reports
- Surgilon 88861932-51Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE1 recall0 reports
- Surgilon 88861971-71Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE1 recall0 reports
- Surgilon 88861985-71Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 93768).
- Z-0845-2024 · Open, ClassifiedSofsilk Braided Silk sutures: CS-211 SOFSILK* 4-0 BLK 45CM CVF21 X36 GS-831 SOFSILK* 4-0 BLK 75CM V20 X36 S-176 SOFSILK* 0 BLK 12X45CM PCT X24 S-182 SOFSILK* 5-0 BLK 12X45CM PCT X…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page