AirLife Adult Manual Resuscitator, Variable Volume Oxygen… recallZ-1064-2024 · Vyaire Medical · initiated Jan 10, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Vyaire Medical or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.
FDA root cause category: Process control
What the company is doing
Vyaire issued an URGENT: Customer notification Letter to its consignees on 01/10/2024 via letter delivered by UPS Next Day Air. The notice explained the problem, complaints received, and requested the consignee discard or destroy all affected devices. If the consignee further distributed the product, they were directed to forward a copy of the notice to those parties.
Recall details
- Recall number
- Z-1064-2024
- Status
- Open, Classified
- Initiated by the company
- Jan 10, 2024
- Posted by the FDA
- Feb 25, 2024
- Recalling firm
- Vyaire Medical
- Quantity
- 1,020,330 units
- Product code
- BTM
- Distribution
- Worldwide distribution. US nationwide, Brazil, Canada, Switzerland, France, UK, Mexico, Netherlands, Saudi Arabia, Sweden, and South Africa
Product description
AirLife Adult Manual Resuscitator, Variable Volume Oxygen Reservoir Tubing, Adult Mask, REF 2K8017
Code information (models, lots, serial numbers)
UDI/DI Case: 50190752114294, Each: 10190752114296; All manufacturing dates of 2017 and prior, and any products without a manufacturing date.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- AirLife™ 2K8017Airlife · Cardiopulmonary resuscitation aid kit1 recall1 death2 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 93821).
- Z-1057-2024 · Open, ClassifiedAirLife Adult Manual Resuscitator, Oxygen Reservoir Bag, without Mask, REF 2K8000,
- Z-1058-2024 · Open, ClassifiedAirLife Adult Manual Resuscitator, 40" (1.0m) Oxygen Reservoir Tubing, without Mask, REF 2K8001
- Z-1059-2024 · Open, ClassifiedAirLife Adult Manual Resuscitator, Oxygen Reservoir Bag, Expiratory Filter, REF 2K8004F
- Z-1060-2024 · Open, ClassifiedAirLife Adult Manual Resuscitator, Oxygen Reservoir Bag, Adult Mask, CO2 Detector, REF 2K8004C2
- Z-1061-2024 · Open, ClassifiedAirLife Adult Manual Resuscitator, 40" (1.0m) Oxygen Reservoir Tubing, Adult Mask, REF 2K8005
- Z-1062-2024 · Open, ClassifiedAirLife Adult Manual Resuscitator, 40" (1.0m) Oxygen Reservoir Tubing, Adult Mask, Expiratory Filter, REF 2K8005F
- Z-1063-2024 · Open, ClassifiedAirLife Adult Manual Resuscitator, 40" (1.0 m) Oxygen Reservoir Tubing, Adult Mask, CO2 Detector, REF 2K8005C2
- Z-1065-2024 · Open, ClassifiedAirLife Adult Manual Resuscitator, 40" (1.0m) Oxygen Reservoir Tubing, Adult Mask, REF 2K8004
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page