Skip to content

AirLife Adult Manual Resuscitator, Oxygen Reservoir Bag… recallZ-1060-2024 · Vyaire Medical · initiated Jan 10, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Vyaire Medical or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.

FDA root cause category: Process control

What the company is doing

Vyaire issued an URGENT: Customer notification Letter to its consignees on 01/10/2024 via letter delivered by UPS Next Day Air. The notice explained the problem, complaints received, and requested the consignee discard or destroy all affected devices. If the consignee further distributed the product, they were directed to forward a copy of the notice to those parties.

Recall details

Recall number
Z-1060-2024
Status
Open, Classified
Initiated by the company
Jan 10, 2024
Posted by the FDA
Feb 25, 2024
Recalling firm
Vyaire Medical
Quantity
3,636 units
Product code
BTM
Distribution
Worldwide distribution. US nationwide, Brazil, Canada, Switzerland, France, UK, Mexico, Netherlands, Saudi Arabia, Sweden, and South Africa

Product description

AirLife Adult Manual Resuscitator, Oxygen Reservoir Bag, Adult Mask, CO2 Detector, REF 2K8004C2

Code information (models, lots, serial numbers)

UDI/DI Case: 50190752114119, Each: 10190752114111; All manufacturing dates of 2017 and prior, and any products without a manufacturing date.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 93821).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page