Skip to content

Senographe Pristina recallZ-0862-2024 · GE Medical Systems, SCS · initiated Dec 26, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from GE Medical Systems, SCS or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

X-ray exposure termination audible signal on Senographe Pristina with Efficiency Suite 2 intermittently not emitted per 21CFR 1020.31(j) requirements.

FDA root cause category: Radiation Control for Health and Safety Act

What the company is doing

GE HealthCare will, without charge, remedy the defect or bring the product into compliance. If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.

Recall details

Recall number
Z-0862-2024
Status
Open, Classified
Initiated by the company
Dec 26, 2023
Posted by the FDA
Feb 8, 2024
Recalling firm
GE Medical Systems, SCS
Quantity
180 units (60 US; 120 OUS)
Product code
MUE
FDA submissions
K162268, K182951, K193334, K211215, K173576
Distribution
Us Nationwide-Worldwide Distribution

Product description

Senographe Pristina

Code information (models, lots, serial numbers)

GTIN(s): 195278022745, 195278276971, 195278628916, 195278661999, 195278665379, 840682118460, 840682142052, 840682145879

Devices named in this recall

7 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

1 device cleared under K162268, K182951, K193334, K211215, K173576 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page