HLS Set recallZ-1076-2024 · Maquet Medical Systems USA · initiated Jan 16, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Maquet Medical Systems USA or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The Emergency Priming Line, a component of the HLS Set, may have a compromised sterile barrier due to nonconformities at the bag seal. This issue may result in inflammation, sepsis, or infection.
FDA root cause category: Nonconforming Material/Component
What the company is doing
On January 16, 2024, the recalling firm began notifying affected customers via URGENT MEDICAL DEVICE CORRECTION letters. Customers were instructed to dispose of any Emergency Priming Line from a previously opened, affected HLS set. Customers should only use the Emergency Priming Line from a newly opened HLS Set. Device users should always have an HLS Set ready as a backup during therapy. The remaining components of the backup HLS Set must be disposed of after extraction of the Emergency Priming Line. Please reach out to Getinge Customer Support for the issuance of a replacement HLS Set at no cost. Customers may also return unopened/unexpired affected HLS Sets by contacting Customer Support. If an affected Emergency Priming Line is already in use, the firm advises it should remain in use due to increased potential risk of stopping this emergency procedure. Users should monitor the patient closely for signs of infection. Contact Getinge Customer Support: (888) 943-8872 (press option 2), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone). ***Updated 3/21/24*** Additional lots identified. Firm notified affected customers on March 21, 2024.
Recall details
- Recall number
- Z-1076-2024
- Status
- Open, Classified
- Initiated by the company
- Jan 16, 2024
- Posted by the FDA
- Feb 13, 2024
- Recalling firm
- Maquet Medical Systems USA
- Quantity
- 16520 worldwide; 6657 US
- Product code
- DTZ
- FDA submissions
- K112360
- Distribution
- Worldwide - US Nationwide distribution.
Product description
Model No. BEQ-HLS 5050 USA, HLS Set Advanced 5.0, Product Code/Part Number 70106.9077. Used with the CARDIOHELP heart-lung machine.
Code information (models, lots, serial numbers)
UDI-DI 04058863076355 Lots 3000301222, 3000319782, 3000327602 ***Added 3/21/24*** 3000333098, 3000352360, 3000357174
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- HLS Set BEQ-HLS 5050 USAMaquet Cardiopulmonary AG · Oxygenator, Cardiopulmonary Bypass3 recalls4 reports
Product lines
Other devices under the same FDA submission
3 devices cleared under K112360 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- HLS Set BEQ-HLS 7050 USAMaquet Cardiopulmonary AG · Oxygenator, Cardiopulmonary Bypass3 recalls13 deaths90 reports
- NA HLS Set Advanced 5.0, Small Adult, BIOLINE CoatingMaquet Cardiopulmonary AG · OXYGENATOR, CARDIOPULMONARY BYPASS3 recalls0 reports
- NA HLS Set Advanced 7.0, Adult, BIOLINE CoatingMaquet Cardiopulmonary AG · OXYGENATOR, CARDIOPULMONARY BYPASS2 recalls0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 93849).
- Z-1077-2024 · Open, ClassifiedBEQ-HLS 7050 USA, HLS Set Advanced 7.0, Product Code/Part Number 70106.9078. Used with the CARDIOHELP heart-lung machine.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page