IdentiTi recallZ-1182-2024 · Alphatec Spine, Inc. · initiated Jan 26, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Alphatec Spine, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to reports of intraoperative graft bolt implantation breakages.
FDA root cause category: Device Design
What the company is doing
On 01/26/2024, the firm emailed an "URGENT-MEDICAL DEVICE RECALL NOTIFICATION" Letter to customer informing them that, The Surgical Technique Guide, LIT-85141, and Instructions for Use, INS-130, were revised to require hole preparation prior to insertion of an IdentiTi SA Graft Bolt, as well as to add a new caution statement, to confirm with fluoroscopy that the graft bolt insertion angle is as close as possible to a 40-degree trajectory. The revised labeling mitigates potential fracturing of the Graft Bolt intra-operatively or post-operatively. Customer are instructed to: Please share this notice with all of those who need to be aware within your organization or any organization where the affected product has been transferred. - Inform doctors and surgical staff of the labeling (surgical technique and instructions for use) - Please fill out the last page of this letter to confirm that you have read this notification and will complete the necessary actions as described in this notification. - If you have any questions/concerns or require assistance with the correction, please contact by e-mailing [email protected] or calling 1-800-922-1356.
Recall details
- Recall number
- Z-1182-2024
- Status
- Open, Classified
- Initiated by the company
- Jan 26, 2024
- Posted by the FDA
- Feb 23, 2024
- Recalling firm
- Alphatec Spine, Inc.
- Quantity
- 12,777 devices
- Product code
- OVD
- FDA submissions
- K232097
- Distribution
- Worldwide - U.S. Nationwide distribution including in the states of AL, AZ, CA, CO, CT, FL, GA, IA, ID, IN, KS, KY, MD, MI, MN, MS, NC, NH, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, SD, TN, TX, UT, WA and WI. The country of New Zealand.
Product description
IdentiTi ALIF Standalone Interbody System: Part Number/Description 132-10-85-250 / IdentiTi ALIF SA Graft Bolt, 08.5 x 25 mm 132-10-85-300 / IdentiTi ALIF SA Graft Bolt, 08.5 x 30 mm 132-10-85-350 / IdentiTi ALIF SA Graft Bolt, 08.5 x 35 mm 132-10-90-250 / IdentiTi ALIF SA Graft Bolt, 09 x 25 mm 132-10-90-300 / IdentiTi ALIF SA Graft Bolt, 09 x 30 mm 132-10-90-350 / IdentiTi ALIF SA Graft Bolt, 09 x 35 mm 232-10-85-250 / IdentiTi ALIF SA Graft Bolt, 08.5 x 25 mm 232-10-85-300 / IdentiTi ALIF SA Graft Bolt, 08.5 x 30 mm 232-10-85-350 / IdentiTi ALIF SA Graft Bolt, 08.5 x 35 mm 232-11-90-250 / IdentiTi ALIF SA Cannulated Graft Bolt, 09 x 25 mm 232-11-90-300 / IdentiTi ALIF SA Cannulated Graft Bolt, 09 x 30 mm 232-11-90-350 / IdentiTi ALIF SA Cannulated Graft Bolt, 09 x 35 mm The IdentiTi ALIF Standalone Interbody System is an integrated intervertebral body fusion device for use in anterior lumbar interbody fusion (ALIF) procedures.
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
Part Number / Product Description / UDI-DI Code / Lot Number 132-10-85-250 / IdentiTi ALIF SA Graft Bolt, 08.5 x 25 mm UDI-DI code: 00190376299716 Lot Numbers: 8746709, 8746724, TO00212 132-10-85-300 / IdentiTi ALIF SA Graft Bolt, 08.5 x 30 mm UDI-DI code: 00190376299730 Lot Numbers: 8746710, 8746725, TO00213 132-10-85-350 / IdentiTi ALIF SA Graft Bolt, 08.5 x 35 mm UDI-DI code: 00190376299754 Lot Numbers: 8746711, TO00218 132-10-90-250 / IdentiTi ALIF SA Graft Bolt, 09 x 25 mm UDI-DI code: 00190376299792 Lot Numbers: 8746712 132-10-90-300 / IdentiTi ALIF SA Graft Bolt, 09 x 30 mm UDI-DI code: 00190376299815 Lot Numbers: 6746713 132-10-90-350 / IdentiTi ALIF SA Graft Bolt, 09 x 35 mm UDI-DI code: 00190376299839 Lot Numbers: 8746714 232-10-85-250 / IdentiTi ALIF SA Graft Bolt, 08.5 x 25 mm UDI-DI code: 00190376369976 Lot Numbers: TO00373, TO00670, TO00694, TO00834, TO01038, TO01404, TO01514, TO01572, TO01573, TO01696, TO01974, TO01979, TO01982, TO02297, TO02393, TO03218, TO03465 232-10-85-300 / IdentiTi ALIF SA Graft Bolt, 08.5 x 30 mm UDI-DI code: 00190376369990 Lot Numbers: TO00376, TO00671, TO00695, TO00710, TO00835, TO01039, TO01405, TO01515, TO01574, TO01575, TO01697, TO01975, TO01980, TO01983, TO02298, TO02394, TO03219, TO03466 232-10-85-350 / IdentiTi ALIF SA Graft Bolt, 08.5 x 35 mm UDI-DI code: 00190376370019 Lot Numbers: TO00380, TO00672, TO00696, TO00836, TO01576, TO01577, TO02163, TO02395 232-11-90-250 / IdentiTi ALIF SA Cannulated Graft Bolt, 09 x 25 mm UDI-DI code: 00190376370293 Lot Numbers: TO00381, TO00673, TO00697, TO01277, TO01578, TO02396 232-11-90-300 / IdentiTi ALIF SA Cannulated Graft Bolt, 09 x 30 mm UDI-DI code: 00190376370316 Lot Numbers: TO00382, TO00674, TO00698, TO01040, TO01406, TO02164, TO02397, TO03220 232-11-90-350 / IdentiTi ALIF SA Cannulated Graft Bolt, 09 x 35 mm UDI-DI code: 00190376370330 Lot Numbers: TO00383, TO00699, TO01407, TO02398
Devices named in this recall
12 devices in the FDA device registry whose identifier appears in the recall's code information.
- IdentiTi 232-10-85-300Alphatec Spine, Inc. · Intervertebral fusion device with integrated fixation, lumbar1 recall8 reports
- IdentiTi 232-10-85-250Alphatec Spine, Inc. · Intervertebral fusion device with integrated fixation, lumbar1 recall5 reports
- IdentiTi 132-10-85-300Alphatec Spine, Inc. · Intervertebral fusion device with integrated fixation, lumbar1 recall3 reports
- IdentiTi 232-10-85-350Alphatec Spine, Inc. · Intervertebral fusion device with integrated fixation, lumbar1 recall2 reports
- IdentiTi 232-11-90-350Alphatec Spine, Inc. · Intervertebral fusion device with integrated fixation, lumbar1 recall1 report
- IdentiTi 132-10-85-250Alphatec Spine, Inc. · Intervertebral fusion device with integrated fixation, lumbar1 recall0 reports
- IdentiTi 132-10-85-350Alphatec Spine, Inc. · Intervertebral fusion device with integrated fixation, lumbar1 recall0 reports
- IdentiTi 132-10-90-250Alphatec Spine, Inc. · Intervertebral fusion device with integrated fixation, lumbar1 recall0 reports
- IdentiTi 132-10-90-300Alphatec Spine, Inc. · Intervertebral fusion device with integrated fixation, lumbar1 recall0 reports
- IdentiTi 132-10-90-350Alphatec Spine, Inc. · Intervertebral fusion device with integrated fixation, lumbar1 recall0 reports
- IdentiTi 232-11-90-250Alphatec Spine, Inc. · Intervertebral fusion device with integrated fixation, lumbar1 recall0 reports
- IdentiTi 232-11-90-300Alphatec Spine, Inc. · Intervertebral fusion device with integrated fixation, lumbar1 recall0 reports
Product lines
Other devices under the same FDA submission
2 devices cleared under K232097 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- IdentiTi 232-13-85-300Alphatec Spine, Inc. · Intervertebral fusion device with integrated fixation, lumbar1 report
- IdentiTi 232-13-90-300Alphatec Spine, Inc. · Intervertebral fusion device with integrated fixation, lumbar1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page