Skip to content

IdentiTi 232-10-85-350Recalls & reported problems

Alphatec Spine, Inc. · IdentiTi ALIF SA Graft Bolt, Ø8.5 x 35 mm · Class 2 · Implant

All models of IdentiTi (31)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00190376370019), received between Aug 7, 2022 and Sep 29, 2022 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Jan 26, 2024Class IIOpen, Classified

    Due to reports of intraoperative graft bolt implantation breakages.

    Z-1182-2024 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break1
  • Fracture1

Patient problems

  • Failure of Implant1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 29, 2022Fracture; Fracture

    IT WAS REPORTED THE SURGEON APPLIED TOO MUCH TORQUE ON THE DRIVER CAUSING THE GRAFT BOLT TO FRACTURE DURING INSERTION. A PORTION OF THE GRAFT BOLT REMAINS IN THE PATIENT.

    Patient problems: Failure of Implant

    MAUDE report 15508547 ↗
  • MalfunctionAug 7, 2022Break; Break

    AFTER INSERTING A SCREW, THE SURGEON DECIDED TO USE A LONGER ONE. THE SHORT SCREW WAS REMOVED ALONG WITH SOME BONE; THE LONGER SCREW SHEARED ONCE THE DISTAL PORTION ENGAGED THE BONE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15180711 ↗

Other Intervertebral fusion device with integrated fixation, lumbar devices

All Intervertebral fusion device with integrated fixation, lumbar devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page