Skip to content

VITROS recallZ-1225-2024 · Ortho-Clinical Diagnostics, Inc. · initiated Jan 26, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Ortho-Clinical Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A software defect affecting VITROS Systems running VITROS Software Versions 3.8.1 causes QC baseline statistics to not update as expected when changed by the user. This may cause erroneous patient results to be reported.

FDA root cause category: Software Design Change

What the company is doing

On January 26, 2024, the firm sent a customer letter to all customers who have V3.8.1 installed on their VITROS Analyzer. Customers were informed of this issue and instructed to follow the instructions when using the VITROS System onboard QC package to update the QC baseline. The software will be updated to mitigate both issues within the VITROS Software QC feature. The estimated availability is 2Q2024. If you have further questions, please contact Global Services Organization at 1-800-421-3311.

Recall details

Recall number
Z-1225-2024
Status
Open, Classified
Initiated by the company
Jan 26, 2024
Posted by the FDA
Mar 1, 2024
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Quantity
4 US; 18 OUS
Product code
JJE
FDA submissions
K031924
Distribution
Worldwide - US Nationwide distribution in the states of AZ, CA, CT, FL, IA, ID, IL, IN, KY, LA, MO, NE, NY, OH, OR, SC, SD, TX, WA, WI, WV and the countries of Australia, Brazil, Canada, China, Colombia, India, Mexico, Singapore, United Kingdom, Italy, Spain, Russia, France, and Belgium.

Product description

VITROS 4600 Chemistry System Catalog Nos. 6802445 (regular) and 6900440 (refurbished)

Code information (models, lots, serial numbers)

VITROS System Software Version 3.8.1 Regular UDI-DI 10758750012343 Refurbished UD-DI 10758750033201 Serial Numbers US: J46000240 J46000898 J46001030 J46001293 OUS: J46000150 J46001041 J46000190 J46001042 J46000499 J46001186 J46000564 J46001187 J46000801 J46001210 J46000821 J46001217 J46000910 J46001289 J46000933 J46001639 J46001029 J46001645

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 93925).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page