VITROS recallZ-1226-2024 · Ortho-Clinical Diagnostics, Inc. · initiated Jan 26, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Ortho-Clinical Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
A software defect affecting VITROS Systems running VITROS Software Versions 3.8.1 causes QC baseline statistics to not update as expected when changed by the user. This may cause erroneous patient results to be reported.
FDA root cause category: Software Design Change
What the company is doing
On January 26, 2024, the firm sent a customer letter to all customers who have V3.8.1 installed on their VITROS Analyzer. Customers were informed of this issue and instructed to follow the instructions when using the VITROS System onboard QC package to update the QC baseline. The software will be updated to mitigate both issues within the VITROS Software QC feature. The estimated availability is 2Q2024. If you have further questions, please contact Global Services Organization at 1-800-421-3311.
Recall details
- Recall number
- Z-1226-2024
- Status
- Open, Classified
- Initiated by the company
- Jan 26, 2024
- Posted by the FDA
- Mar 1, 2024
- Recalling firm
- Ortho-Clinical Diagnostics, Inc.
- Quantity
- 30 US: 38 OUS
- Product code
- JJE
- FDA submissions
- K081543
- Distribution
- Worldwide - US Nationwide distribution in the states of AZ, CA, CT, FL, IA, ID, IL, IN, KY, LA, MO, NE, NY, OH, OR, SC, SD, TX, WA, WI, WV and the countries of Australia, Brazil, Canada, China, Colombia, India, Mexico, Singapore, United Kingdom, Italy, Spain, Russia, France, and Belgium.
Product description
VITROS 5600 Integrated System Catalog Nos. 6802413 (regular) and 6802915 (refurbished)
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
VITROS System Software Version 3.8.1 Regular UDI-DI 10758750002740 Refurbished UD-DI 10758750007110 Serial Numbers US: J56001885 J56003484 J56001886 J56003541 J56001910 J56003570 J56002308 J56003583 J56002461 J56003646 J56002474 J56003669 J56002608 J56003670 J56002698 J56004068 J56002704 J56004137 J56003342 J56004174 J56003356 J56004491 J56003377 J56004506 J56003433 J56004549 J56003477 J56004551 J56003478 J56004577 OUS: J56000425 J56002518 J56000512 J56002615 J56000533 J56002625 J56000674 J56002637 J56000680 J56002926 J56000719 J56002929 J56000809 J56003061 J56000831 J56003260 J56000836 J56003285 J56001051 J56003401 J56001111 J56003464 J56001114 J56003508 J56001208 J56003523 J56001731 J56003538 J56001765 J56004346 J56001826 J56004355 J56001853 J56004415 J56002085 J56004531 J56002116 J56004695
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- VITROS 6802413Ortho-Clinical Diagnostics, Inc. · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE12 recalls69 reports
- VITROS 6802915Ortho-Clinical Diagnostics, Inc. · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE8 recalls3 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 93925).
- Z-1224-2024 · Open, ClassifiedVITROS 3600 Immunodiagnostic System Catalog Nos. 6802783 (regular) and 6802914 (refurbished)
- Z-1225-2024 · Open, ClassifiedVITROS 4600 Chemistry System Catalog Nos. 6802445 (regular) and 6900440 (refurbished)
- Z-1227-2024 · Open, ClassifiedVITROS XT 3400 Chemistry System, Catalog No. 6844458
- Z-1228-2024 · Open, ClassifiedVITROS XT 7600 Integrated System, Catalog No. 6844461
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page