Ivenix, Inc. recallZ-1282-2024 · Fresenius Kabi USA, LLC · initiated Sep 16, 2022
Completed
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Fresenius Kabi USA, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Retroactively reported correction from 2022: A software defect may cause an incorrect (Fail-Stop) Alarm when an administration set is loaded with an upstream occlusion present, resulting in a non-recoverable (Fail-Stop) Pump Problem alarm. This may lead to delay of therapy, and patient may experience from no to minimal to serious harm. This issue was resolved in all fielded product in software version 5.2.1, which was installed in September 2022.
FDA root cause category: Software design
What the company is doing
The issue was resolved through an update of the software to Version 5.2.1, which was installed at customer sites in September 2022. The firm did not issue a customer letter at the time. According to the firm, this issue has been resolved in all fielded product. There have been two subsequent software upgrades since Version 5.2.1 was released. Consignees are now on v 5.8.0.
Recall details
- Recall number
- Z-1282-2024
- Status
- Completed
- Initiated by the company
- Sep 16, 2022
- Posted by the FDA
- Mar 11, 2024
- Recalling firm
- Fresenius Kabi USA, LLC
- Quantity
- 3
- Product code
- FRN
- FDA submissions
- K183311
- Distribution
- US Nationwide distribution in the states of WI and NJ.
Product description
LVP software of the Ivenix Infusion System (IIS), LVP-SW-0004
Code information (models, lots, serial numbers)
UDI-DI 00811505030122 Software versions 5.0.1 and earlier.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Ivenix, Inc. 2.0Fresenius Kabi AG · Pump, Infusion10 recalls0 reports
Product lines
Other devices under the same FDA submission
4 devices cleared under K183311 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Ivenix Infusion System LVP-0004Fresenius Kabi AG · Pump, Infusion7 recalls2 deaths4,328 reports
- Ivenix Infusion System 030-0028-08Fresenius Kabi AG · Pump, Infusion7 reports
- Ivenix Infusion System 030-0027-06Fresenius Kabi AG · Infusion Safety Management Software3 reports
- Ivenix Infusion System 030-0028-07Fresenius Kabi AG · Pump, Infusion2 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page