Skip to content

Ivenix Infusion System 030-0028-07Recalls & reported problems

Fresenius Kabi AG · Large Volume Pump (LVP) Software (order 1 per pump) · Class 2

All models of Ivenix Infusion System (5)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00811505030436), received between Oct 30, 2024 and Nov 6, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Computer Software Problem” (1 report).

25 other recalls cover devices cleared under the same FDA submission (K183311); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • No Apparent Adverse Event2
  • Computer Software Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 6, 2024Computer Software Problem; No Apparent Adverse Event; Computer Software Problem; No Apparent Adverse Event

    THE DEVICE TRIGGERED A PUMP PROBLEM ALARM INSTEAD OF A TUBING SET MISLOADED ALERT. THE COMBINATION OF RAPID UNLOADING AND RELOADING OF AN ADMIN SET, WITH THE SET MISLOADED IN SUCH A WAY THAT THE RESISTOR IS UNABLE TO COUPLE, CAN LEAD TO THE START BUTTON NO LONGER BEING GRAYED OUT. PROGRAMMING AN INFUSION AND PRESSING ON THE START BUTTON AS THE PUMP IS TRYING TO COUPLE WITH THE ADMIN SET CAN RESULT IN A COHERENCE FAILURE, TRIGGERING A PUMP PROBLEM ALARM. THE DEFECT RISK ASSESSMENT DETERMINED THE RISK AND SEVERITY TO BE A 1 - INCONVENIENCE. NO FURTHER ACTION IS PLANNED AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20623048 ↗
  • MalfunctionOct 30, 2024No Apparent Adverse Event; No Apparent Adverse Event

    REVIEWING THE SOURCE CODE SHOWED THAT THE ROOT CAUSE OF THE PROBLEM WAS A TIMING ISSUE THAT COULD OCCUR IF THE USER WERE TO START THE SECONDARY INFUSION JUST AS THE PRIMARY INFUSION WAS TRANSITIONING TO KVO. A FIX WAS IMPLEMENTED THAT ENSURES THAT THE PRIMARY IS NOT SET TO FOLLOW IF IT TRANSITIONS INTO KVO AND FOLLOW UP TESTING HAS CONFIRMED THAT THIS FIX WAS SUCCESSFUL. THIS WILL BE A PART OF THE NEXT SW RELEASE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20571410 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K183311. One submission often covers many models, so check the product and lot details in each record.

  • Z-2224-2026 · May 6, 2026 · Open, Classified
    Ivenix Infusion System (IIS), Large Volume Pump, Model Number: LVP-0004
  • Z-2223-2026 · May 6, 2026 · Open, Classified
    LVP software of the Ivenix Infusion System (IIS); Product Code LVP-SW-0005.
  • Z-2274-2026 · Feb 4, 2026 · Open, Classified
    The Large Volume Pump (LVP) of the Ivenix Infusion System (IIS). Model Number: LVP-0004.
  • Z-1048-2026 · Nov 21, 2025 · Open, Classified
    Ivenix Infusion System (IIS), LVP Software: LVP-SW-0005.
  • Z-0885-2026 · Nov 14, 2025 · Open, Classified
    Ivenix Infusion System (IIS), LVP Software LVP-SW-0005. Software for the controlled administration of fluids to patients.
  • Z-0586-2026 · Nov 3, 2025 · Open, Classified
    IVENIX INFUSION SYSTEM, LVP Primary Administration Set, Dual-Inlet, Low-Sorbing, Needle-Free Port, Dual Y-Site (Qty 25). Large Volume Infusion Pump Administration Set.
  • Z-0848-2026 · Nov 3, 2025 · Open, Classified
    Ivenix Infusion System (IIS), Large Volume Pump. Model Number: LVP-0004.
  • Z-0378-2026 · Sep 3, 2025 · Open, Classified
    Ivenix Infusion System (IIS), Large Volume Pump. Model Number: LVP-0004.
  • Z-2232-2025 · Jun 24, 2025 · Open, Classified
    LVP Primary Administration Set, Dual-Inlet, Low-Sorbing, Needle-Free Port, Y-Site (Qty 25). Product Code: SET-0013-25.
  • Z-2235-2025 · Jun 24, 2025 · Open, Classified
    LVP Primary Administration Set, Dual-Inlet, Low-Sorbing, Needle-Free Port, Dual Y-Site (Qty 25). Product Code: SET-0032-25.
  • Z-2234-2025 · Jun 24, 2025 · Open, Classified
    LVP Primary Administration Set, Single Outlet Low Sorbing (Qty 25). Product Code: SET-0021-25.
  • Z-2233-2025 · Jun 24, 2025 · Open, Classified
    LVP Blood Products Administration Set Dual-Inlet, Low-Sorbing, Y-Site, Mesh Filter (Qty 20). Product Code: SET-0014-20.
  • Z-1869-2025 · May 12, 2025 · Open, Classified
    IVENIX INFUSION SYSTEM LVP Blood Products Administration Set, Dual-Inlet, Low-Sorbing, Y-Site, Mesh Filter, 15 mL. Intravascular Administration Set.
  • Z-1019-2025 · Jan 10, 2025 · Open, Classified
    Ivenix Infusion System (IIS), LVP Software LVP-SW-0005. Indicated for use in a hospital and in outpatient care environments for the controlled administration of fluids through cli
  • Z-0763-2025 · Dec 6, 2024 · Open, Classified
    Ivenix Infusion System (IIS), Large Volume Pump, Model Number: LVP-0004

…and 10 more.