Ivenix, Inc. recallZ-1283-2024 · Fresenius Kabi USA, LLC · initiated May 10, 2023

Completed

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Fresenius Kabi USA, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Retroactively reported corrections from 2023: 1) A software defect may cause an incorrect (Fail-Stop) alarm when an administration set is loaded or coupled while the pump is executing the power-up sequence. May lead to delay in therapy. 2) Alert is not annunciated informing the clinician that the bolus cannot be delivered when the entered bolus dose exceeds the Care Profile Hard Rate Max limit and Rapid Bolus is selected. May lead to over infusion. Both issues were resolved in all fielded product in software version 5.8.0, which was installed in affected units May thru August 2023.

FDA root cause category: Software design

What the company is doing

Both issues were resolved through an update of the software to Version 5.8.0, which was installed at customer sites between May and August 2023. The firm did not issue a customer letter at the time. According to the firm, this issue has been resolved in all fielded product.

Recall details

Recall number
Z-1283-2024
Status
Completed
Initiated by the company
May 10, 2023
Posted by the FDA
Mar 11, 2024
Recalling firm
Fresenius Kabi USA, LLC
Quantity
6 US
Product code
FRN
FDA submissions
K183311
Distribution
US Nationwide distribution to the states of CA, CO, NJ, WI, UT.

Product description

LVP software of the Ivenix Infusion System (IIS); LVP-SW-0004

Code information (models, lots, serial numbers)

UDI-DI 00811505030122 Software versions 5.2.1/5.2.2

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

4 devices cleared under K183311 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page