Ivenix, Inc. recallZ-1283-2024 · Fresenius Kabi USA, LLC · initiated May 10, 2023
Completed
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Fresenius Kabi USA, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Retroactively reported corrections from 2023: 1) A software defect may cause an incorrect (Fail-Stop) alarm when an administration set is loaded or coupled while the pump is executing the power-up sequence. May lead to delay in therapy. 2) Alert is not annunciated informing the clinician that the bolus cannot be delivered when the entered bolus dose exceeds the Care Profile Hard Rate Max limit and Rapid Bolus is selected. May lead to over infusion. Both issues were resolved in all fielded product in software version 5.8.0, which was installed in affected units May thru August 2023.
FDA root cause category: Software design
What the company is doing
Both issues were resolved through an update of the software to Version 5.8.0, which was installed at customer sites between May and August 2023. The firm did not issue a customer letter at the time. According to the firm, this issue has been resolved in all fielded product.
Recall details
- Recall number
- Z-1283-2024
- Status
- Completed
- Initiated by the company
- May 10, 2023
- Posted by the FDA
- Mar 11, 2024
- Recalling firm
- Fresenius Kabi USA, LLC
- Quantity
- 6 US
- Product code
- FRN
- FDA submissions
- K183311
- Distribution
- US Nationwide distribution to the states of CA, CO, NJ, WI, UT.
Product description
LVP software of the Ivenix Infusion System (IIS); LVP-SW-0004
Code information (models, lots, serial numbers)
UDI-DI 00811505030122 Software versions 5.2.1/5.2.2
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Ivenix, Inc. 2.0Fresenius Kabi AG · Pump, Infusion10 recalls0 reports
Product lines
Other devices under the same FDA submission
4 devices cleared under K183311 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Ivenix Infusion System LVP-0004Fresenius Kabi AG · Pump, Infusion7 recalls2 deaths4,328 reports
- Ivenix Infusion System 030-0028-08Fresenius Kabi AG · Pump, Infusion7 reports
- Ivenix Infusion System 030-0027-06Fresenius Kabi AG · Infusion Safety Management Software3 reports
- Ivenix Infusion System 030-0028-07Fresenius Kabi AG · Pump, Infusion2 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗