Skip to content

Assembly, Array Guide TMINI, REF: 109101, part of the TMINI… recallZ-1318-2024 · Think Surgical, Inc. · initiated Feb 12, 2024

Class II Open, Classified All recalls posted in March 2024

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Think Surgical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

There is a galling effect (cold-weld) caused by friction between the array fixation pins and the TMINI array guide holes due to insufficient clearance between the outer diameter of the fixation pins and the inner diameter of the TMINI array guide holes. As a result, the array guide has the possibility of binding to the array fixation pins during drilling.

FDA root cause category: Device Design

What the company is doing

On 2/12/24, recall notices were mailed and emailed to customers who were asked to do the following: 1) Examine your inventory, locate any unused devices and quarantine them immediately. Your clinical representative(s) will assist with the recall process and provide guidance to your surgeon(s) for positioning the array fixation pins using the array bracket. 2) Do not use any affected product and return any used or unused product. 3) Complete and return the acknowledgment form to [email protected] Customers with questions can contact the firm via phone at 510-249-2300, Option 1.

Recall details

Recall number
Z-1318-2024
Recall class
Class II
Status
Open, Classified
Initiated by the company
Feb 12, 2024
Posted by the FDA
Mar 20, 2024
Recalling firm
Think Surgical, Inc.
Quantity
16
Product code
OLO
FDA submissions
K230202
Distribution
US Nationwide distribution in the states of AL, OH, NV, TX, UT, CA, MO.

Product description

Assembly, Array Guide TMINI, REF: 109101, part of the TMINI Knee Instrument Set, used with the TMINI Miniature Robotic System

Code information (models, lots, serial numbers)

UDI-DI: M9461091010. Lot Numbers: 2316701, 2232601, 23030, 2228501, 2327802, 22152

Other devices under the same FDA submission

5 devices cleared under K230202 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page