Covidien Signia" Small Diameter Curved Tip Intelligent… recallZ-1458-2024 · Covidien, LP · initiated Feb 14, 2024
Class II Open, Classified All recalls posted in April 2024
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Covidien, LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Under certain firing conditions, reloads were found to articulate in an uncontrolled manner potentially causing disruption to the staple line. The issue is related to components in impacted reloads that are not fully secure, which leads to uncontrolled articulation. The issue can occur using either the Endo GIA" Ultra Universal Stapler or Signia" Stapling System.
FDA root cause category: Under Investigation by firm
What the company is doing
On February 14, 2024, the firm notified domestic customers and notified international customers starting February 19, 2024 of the issue via URGENT: MEDICAL DEVICE RECALL letters. Actions to be taken by the customer: 1. Identify and quarantine all unused impacted product listed above. 2. See Attachment A for guidance on identifying affected lots. 3. Return all quarantined devices to Medtronic. Please contact [email protected] for the Return Goods Authorization (RGA). 4. Credit for the returned affected devices will be issued based on the RGA number. 5. If purchased from a distributor, contact your distributor directly to arrange for the return of the devices back to your distributor. 6. Please complete and return the enclosed Customer Confirmation Form to [email protected] even if you do not have unused inventory. 7. Pass on this notice to all those who need to be aware within your organization or to any organization where the listed affected lots may have been transferred or distributed. Adverse events or quality problems experienced with this product should be reported to the FDA and Medtronic: Online at http://www.fda.gov/Safety/MedWatch/HowToReport/default.htm, or call FDA at 1-800-FDA-1088 (1-800-332-1088) and Medtronic Technical Services at 800-255-6774, option 2 to provide information regarding those events. We regret any inconvenience this may cause. We are committed to patient safety and appreciate your prompt attention to this matter. If you have any questions regarding this communication, please contact your Medtronic Field Representative or Customer Service at 800-962-9888 Option 2.
Recall details
- Recall number
- Z-1458-2024
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Feb 14, 2024
- Posted by the FDA
- Apr 1, 2024
- Recalling firm
- Covidien, LP
- Quantity
- 3,023 units
- Product code
- GDW
- FDA submissions
- K191070
- Distribution
- Worldwide distribution - US Nationwide and the countries of France, Hong Kong, Israel and Japan.
Product description
Covidien Signia" Small Diameter Curved Tip Intelligent Reload, Model SIGSDS30CTVT
Code information (models, lots, serial numbers)
Model Number: SIGSDS30CTVT; UDI/DI: 10884521741888; Lot numbers: N3L1909UY, N3L1868UY, N3L2008UY, N3L2221UY, N3M1541UY, N3M1603UY, N3M1639UY;
Other devices under the same FDA submission
3 devices cleared under K191070 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Signia SIGSDS30CTVTCovidien LP · Staple, implantable2 recalls1 death152 reports
- Signia SIGSDS30CTVCovidien LP · Staple, implantable1 recall1 death80 reports
- Signia SIGSDL45CTVTCovidien LP · Staple, implantable2 recalls27 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 94128).
- Z-1457-2024 · Open, ClassifiedSignia" Small Diameter Curved Tip Intelligent Reload, Model SIGSDS30CTVT
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page