Supple Peri-Guard recallZ-1378-2024 · Baxter Healthcare Corporation · initiated Feb 28, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Baxter Healthcare Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
A new contraindication was added to the IFU which states that Peri-Guard Repair Patch should not be used in neurosurgery since the product endotoxin level may be higher than the allowable limit for cerebrospinal fluid-contacting devices and Supple Peri-Guard Repair Patch should not be used in neurosurgery since the product endotoxin level may be higher than the allowable limit for cerebrospinal fluid-contacting devices.
FDA root cause category: Labeling Change Control
What the company is doing
Baxter issued a "SAFETY ALERT" to its consignees on 02/28/2024 via USPS first class mail. the notice explained the problem, potential hazard involved with off-label use of the product and the actions to be taken. The firm requested that the notice be forwarded all affected facilities or departments with institutions. Distributors are directed to notify their customers. For questions, contact Baxter Healthcare Center for Service at 888-229-0001, between the hours of 7:00 am and 6:00 pm Central Time, Monday through Friday.
Recall details
- Recall number
- Z-1378-2024
- Status
- Open, Classified
- Initiated by the company
- Feb 28, 2024
- Posted by the FDA
- Mar 26, 2024
- Recalling firm
- Baxter Healthcare Corporation
- Quantity
- 9549 units
- Product code
- FTM
- FDA submissions
- K983162
- Distribution
- US Distribution to states of: CT, VA, TX, PA and IL; and OUS (foreign) countries of: Austria, Finland, Germany, Greece, Ireland, Italy, Poland, Spain, Sweden, Switzerland, and United Kingdom.
Product description
Baxter Supple Peri-Guard Repair Patch (legacy product codes): a) PC0404SNBIO, b) PC0608NSBIO, c) PC0814SNBIO, d) PC1016SNBIO
Code information (models, lots, serial numbers)
a) PC0404SNBIO, UDI/DI 00085412530642; b) PC0608SNBIO, UDI/DI 00085412530673; c) PC0814SNBIO, UDI/DI 00085412530697; d) PC1016SNBIO, UDI/DI 00085412530710; ALL LOTS WITHIN EXPIRY
Devices named in this recall
4 devices in the FDA device registry whose identifier appears in the recall's code information.
- Supple Peri-Guard PC0404SNSynovis Life Technologies, Inc. · Mesh, surgical1 recall0 reports
- Supple Peri-Guard PC0608SNSynovis Life Technologies, Inc. · Mesh, surgical, collagen, diaphragmatic hernia1 recall0 reports
- Supple Peri-Guard PC0814SNSynovis Life Technologies, Inc. · PATCH, PLEDGET AND INTRACARDIAC, PETP, PTFE, POLYPROPYLENE1 recall0 reports
- Supple Peri-Guard PC1016SNSynovis Life Technologies, Inc. · Mesh, surgical, collagen, diaphragmatic hernia1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 94152).
- Z-1377-2024 · Open, ClassifiedBaxter Peri-Guard Repair Patch (legacy product codes): a) PC0404NBIO, b) PC0608NBIO, c) PC0814NBIO, d) PC1016NBIO
- Z-1379-2024 · Open, ClassifiedBaxter Peri-Guard Repair Patch (new product codes): a) PG0404, b) PG0608, c) PG0814, d) PG1016, e) PG1225
- Z-1380-2024 · Open, ClassifiedBaxter Supple Peri-Guard Repair Patch (new product codes): a) SPG0404, b) SPG0406, c) SPG0608, d) SPG0814, e) SPG1016
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page