ACUSON Juniper Diagnostic Ultrasound System, REF: 11335791… recallZ-1599-2024 · Siemens Medical Solutions USA, Inc. · initiated Feb 23, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Medical Solutions USA, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
On ultrasound systems, when Cardiac DICOM SR feature is configured to display either minimum or maximum measured value, and multiple cardiac region measurements are made, and results are exported into the SR feature, then The SR viewer will display the LAST measured value, not min or max, which could contribute to patient condition misdiagnosis or negatively influence patient management decisions.
FDA root cause category: Software design
What the company is doing
On 2/23/24, correction notices were sent to customers informing them of the following: This issue is specific to and limited to the use of the Cardiac DICOM SR feature: - When a user uses and configures the Cardiac DICOM SR feature to display either the MINIMUM (Min) or the MAXIMUM (Max) measured value, AND - Makes multiple measurements of the cardiac region during an examination, AND - Exports the results into the Cardiac DICOM SR feature, THEN - The Cardiac DICOM SR viewer will NOT display the MINIMUM (Min) or MAXIMUM (Max) value and will instead display the LAST measured value. In that case, the displayed measurement in the Cardiac DICOM SR viewer may NOT be the Minimum (Min) or MAXIMUM (Max) value measured during the examination. To avoid potential misdiagnosis by an over- or underrepresented measured value, it is recommended to refer to the measurement values displayed on the Ultrasound System Final Report when the configuration is to display the Min or Max value. Firm recommends a review of previous cardiac ultrasound examination results where the clinical assessment was reviewed using a DICOM SR viewer. Your Customer Service Engineer from will contact you to schedule a facility visit to update the system or inform you of a remote update when the software update is available. The software update is currently under development and estimated to be available by spring/summer of 2024. Please ensure that all users of the affected devices within your organization, and others who may need to be informed, receive the safety relevant information provided with this notice and take the actions specified herein. If you have further questions, please contact the firm's Ultrasound Service Customer Care Center at 1-800-888-7436. Not Affected: -ACUSON Juniper Ultrasound System, product version 2.5, software version VB30D. - ACUSON Redwood system, product version 2.0.
Recall details
- Recall number
- Z-1599-2024
- Status
- Open, Classified
- Initiated by the company
- Feb 23, 2024
- Posted by the FDA
- Apr 19, 2024
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Quantity
- 5090
- Product code
- IYN
- FDA submissions
- K201130
- Distribution
- Worldwide distribution - US Nationwide including in the states of AL, AR, CA, CO , CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA , MD, MI, MN, MO, MS, NC, ND, NE, NJ, NM, NY, OH, OK, OR, PA, PR, RI, SD, TN, TX, UT, VA, WA, WI, WY and the countries of BG, BW, BA, TR, AR, CL, CN, FR, SK, CD, BH, PH, AM, CR, AF, BE, AE, AZ, AO, KR, RO, HU, HR, DK, AU, DZ, CY, PL, BO, AL, NL, CI, BR, CZ, DE, AT, ES, CO, BD, CH, IT, CA, CM, DO, EC, EG, ER, ET, FI, FO, GB, GE, GR, GT, GY, HK, HN, ID, IE, IL, IN, IQ, IR, JO, JP, KE, KW, KZ, LB, LT, LU, LV, LY, MA, MD, ME, MK, MM, MN, MU, MW, MX, MY, NG, NO, NP, NZ, OM, PA, PE, PG, PK, PT, PY, QA, RS, RU, SA, SE, SG, SI, SM, SN, SV, SZ, TH, TN, TT, TW, TZ, UA, UG, UY, UZ, VN, XK, YE, ZA, ZM, ZW.
Product description
ACUSON Juniper Diagnostic Ultrasound System, REF: 11335791, and ACUSON Juniper Select Diagnostic Ultrasound System, REF: 11653093, running software version VA10x with DICOM SR Cardiac option, from VB10A to VB10G, VB11A, VB11B, VB30B and VB30C
Code information (models, lots, serial numbers)
Show the full code information (34k characters)
REF/UDI-DI: 11335791/04056869152059, and 11653093/04056869957357. Serial Numbers: 571063 556506 555717 555502 555450 555448 552655 552659 550763 556982 556990 556996 556998 557000 557001 557002 557003 555641 555631 553876 553877 556233 557117 557122 557123 556083 556082 556080 556081 556043 556044 556045 556046 556047 555714 555285 555281 555282 555283 555286 555287 555288 555290 555289 555284 550462 550460 550461 550915 550878 573362 573314 570015 573313 573132 570258 570332 575056 570033 556390 556738 556734 556545 555666 555746 552723 553173 552986 553174 550198 550541 550310 550214 550052 550051 573202 575041 571545 570335 570288 570227 575050 575051 556510 575011 570057 570058 570028 556973 557069 556967 556974 556920 555875 556598 556514 556399 556004 555693 555107 555719 555545 555477 555429 555197 555167 555215 555144 555124 555155 553043 553161 552998 553829 553542 553435 555033 553437 553438 553302 552775 551895 551967 551970 550940 555364 555368 555622 555966 556295 556334 556462 556463 556464 556490 556724 556727 556735 556812 556885 557057 557063 557068 557127 557306 557393 570031 570118 570196 571568 573178 573343 573439 576207 576231 576245 578060 557401 557372 555951 573052 555729 557113 556508 556052 555514 555530 555166 570067 573422 578089 573392 573393 570174 573044 570066 556783 556784 556650 555611 556358 556008 556030 555109 555985 555100 555121 555793 555701 555625 555544 555390 555202 555171 552907 553037 552729 553468 552594 552937 551775 550993 550834 550294 550038 556557 556271 555967 555607 555311 552708 551574 550029 557114 557252 557364 557370 555671 550775 570349 557300 570379 570357 570364 570373 570381 557299 572027 576011 576012 570384 570351 570359 570367 570368 570387 570389 570358 570391 570352 570366 570378 570386 570388 570390 570218 570338 570353 570365 570377 570382 570383 570348 570354 570355 570356 570362 570363 570371 570372 570374 570375 570380 556997 557294 570350 570341 557292 557293 557014 557015 557239 557240 557238 557013 557236 557233 557234 557235 557237 557036 556829 556848 556827 556852 556870 556803 556869 556816 556797 556704 556703 556702 556678 556743 556744 556118 556059 555961 555963 555964 556027 556102 556006 555930 555750 555702 550365 550363 553208 553213 550303 551927 551928 551660 550325 550323 550366 550327 550283 555394 556260 557011 557012 557243 557326 570344 570345 570346 570361 570385 572029 572031 572032 572033 572043 573459 577000 556958 556928 556899 556706 556616 556110 556278 556218 555437 555902 551866 555660 556953 556656 550822 552635 550751 550562 573336 571058 556196 556204 556202 556201 556203 556213 556156 555899 555907 550614 550483 550095 610026 610089 610090 610029 610081 610056 610059 610046 610047 610048 610049 610070 552189 610052 557201 557408 610069 610027 610057 610068 610060 610061 610038 610054 610055 557203 557429 557428 557174 610042 610058 557417 557414 557430 551817 552183 610025 610045 610044 610041 610043 557412 610023 610040 610024 610033 553372 553869 610005 610034 610035 610036 551821 610028 610019 610018 610011 610012 610020 610017 610014 610015 610010 610016 557163 557394 610013 610008 610009 610006 610007 557427 557200 557421 557422 557191 557407 557197 557406 557195 557396 557378 557382 557383 557387 557388 557175 553714 557373 557173 557170 557199 557156 557194 557196 553712 553722 557155 557168 557169 557172 557202 557007 557035 553716 553717 551547 553720 553730 553733 553868 553898 553731 553713 551083 554063 550991 553954 553955 553718 553692 553688 551080 553736 553732 553880 553881 553719 553737 553651 553709 553726 553735 553371 551051 553860 553710 553715 551840 551841 553734 553661 553660 553711 553669 553672 550935 552534 551823 551064 553517 551073 550942 552503 551849 551090 551075 551050 553349 551089 553249 553346 553343 551079 551081 551066 551688 552223 551078 550928 550987 551058 552506 552993 552510 551514 551555 551045 551054 551801 552532 552224 550968 551095 550966 550982 550962 550466 550996 552522 551099 551695 551043 550927 551055 551061 551022 550990 551062 550986 551096 551052 551048 550995 551538 551097 551082 550938 551025 551687 550937 550965 551076 550866 550868 550869 550870 550871 550880 550926 550939 551060 553351 557190 610001 610002 610003 610004 610032 610039 610050 610063 610082 610086 610087 610088 610091 610092 610093 557297 557392 572024 572022 572021 576014 557391 556579 572018 557296 557390 570347 570204 570400 575076 575075 570211 557369 570241 570256 570212 570213 570210 570217 557367 557083 555537 555815 556776 557263 557264 557265 557266 557084 557085 555814 556960 556578 555663 555665 555816 555805 555667 556636 556480 556479 556285 555813 555037 555931 555632 552601 551676 550689 550632 550631 550864 550629 550226 556245 557389 570205 570302 570549 573349 573357 573358 556845 557241 557133 557090 557091 556669 555811 555519 555533 555534 555535 555538 555554 556343 556588 556601 557134 557242 570236 570538 570540 570541 570081 570082 572017 557111 557231 557115 556778 556666 556628 556312 556216 555921 555710 555571 555466 555475 555476 555482 555484 555485 555486 555471 555214 553158 552874 552165 551757 551633 550513 550231 557192 555725 556985 556672 550430 555935 555913 555473 555133 553784 553390 559004 570062 573418 573419 571549 573384 570072 570046 570021 570026 557280 557271 559013 570001 555877 551870 555876 555888 556538 556522 556523 556515 556517 556518 556487 556324 555604 555734 556103 556173 556125 555111 555123 555040 550141 550155 555048 555051 555011 553564 553778 553473 552728 552697 552798 552593 551532 552736 552581 551999 552002 551865 551932 551632 551558 551098 550204 550772 550767 550200 550202 550712 550745 550672 550174 550302 550544 550290 550217 550230 550212 550130 550131 573218 575016 575017 573271 557354 557355 557230 557095 557096 556661 556663 556611 555704 555559 556434 557270 557269 556231 556254 555568 555273 551563 556540 570521 570504 570420 570418 575105 557361 556883 557061 553265 556625 556613 556609 556617 556622 555643 555581 555652 555909 553256 552939 551939 551940 552938 552671 552579 551972 552684 550832 550378 550142 550035 550025 551941 557214 578059 573144 573157 573158 557322 557029 556728 556503 556440 556441 556214 553900 551923 550451 551981 550337 570248 570261 573146 557342 557253 557254 556539 556387 556396 555871 555042 557359 557360 570135 570136 570137 570138 573102 573430 573307 573287 552752 573217 570059 573092 573107 573101 573103 573104 570274 570279 571513 556131 573034 556859 570300 570308 570298 570234 570245 556494 570340 556860 570189 556572 570166 570167 570144 570145 570124 570104 570084 570054 556513 556930 556171 570023 570037 570045 556524 556136 557098 557274 557210 556526 570034 570030 570032 557137 557092 557093 557094 557099 557108 557059 557023 556963 556976 556977 556168 556167 556170 555775 556925 556926 556927 556732 556731 555879 556733 555819 556737 555820 555872 555880 556647 556606 556603 555868 556584 555870 556505 556138 556491 556495 556466 556445 556409 556359 556300 556303 556270 556296 556079 556272 556200 556205 556145 555041 556072 556078 556089 555444 555662 556066 555039 556037 556029 555099 555097 555803 555972 555950 555922 555923 555101 555919 555098 555824 555767 555773 555774 555741 555555 555024 555008 555623 555431 555547 555010 555539 555540 555493 555495 555417 555474 555461 553232 555359 555401 555389 555332 555369 553774 555336 553773 555362 555363 552909 555335 555276 555083 555085 555086 555212 555217 555115 555118 555172 553592 555116 555117 555119 555082 555084 552730 552910 553772 552898 553407 553495 552509 553404 553141 551677 552813 553084 551879 552597 552625 552819 553031 551898 551616 551753 552739 551782 552812 551641 551749 551530 551105 552714 551618 552121 552174 551006 550904 550299 551984 551855 550768 551057 551682 550171 550308 550877 550168 550780 550011 550020 550668 550670 550555 550412 573091 573095 573194 555944 552693 550090 573227 573237 573251 573315 570187 578074 573229 573250 575054 573225 578094 578095 573414 578096 573401 573404 573407 570271 570185 578077 578093 578097 573124 573249 573236 573383 573403 578063 573391 578088 573385 578071 573355 573366 573367 573316 573350 573351 576224 578058 578056 573280 573284 573310 573311 573320 573321 573322 573323 573324 573328 578010 570078 573270 573165 573228 573230 573234 573252 573283 573296 576256 578007 578045 578046 573120 573149 573199 573220 573221 573226 573265 573278 573305 578031 578032 573121 573219 578024 578025 578026 553590 573122 573145 578023 578029 573118 578042 578043 573268 573223 573235 573253 573304 578055 573185 578033 570003 573171 573224 573266 573267 573269 573274 578030 573119 573125 578018 578049 578050 573141 578044 573123 573187 576246 576247 576260 578000 578041 578012 578015 573069 578013 578019 573282 571520 571532 571547 573116 578011 573093 578009 576255 573075 573073 573090 576251 573051 578002 578005 578006 576228 578008 576236 576238 576240 578003 578004 576249 578001 573054 576239 576241 576242 576243 576244 576257 576258 576259 556971 556972 556980 576233 576250 576253 576254 576234 570161 573039 576230 576232 576235 571548 571540 573040 576226 576227 570077 571511 573037 573049 576229 571506 570284 571526 556981 571541 576212 571525 573074 576223 571562 571563 571566 571564 571565 576215 571522 571551 571555 571557 571558 571559 576218 571527 571556 576216 571550 571552 571553 571554 576214 576217 571524 575042 576213 570281 571528 571529 576220 576221 571531 571521 571510 576211 570309 576005 556135 570285 557339 557340 570262 575057 556898 570259 570267 570266 575053 556357 570257 570282 575052 570147 570231 575046 556858 570142 570230 570182 570216 570184 556571 570192 570171 575039 570191 570193 575021 570169 557385 557386 570180 570181 570172 556568 570186 570179 570162 570163 575040 570208 575033 570131 570129 557366 570070 570093 570088 557291 570080 556761 570148 570149 570139 570140 575014 575015 575020 557363 557365 570095 570130 575018 570063 570068 570069 570071 575013 570011 570048 570064 556615 570061 557128 570051 557064 570087 575012 557232 570005 570006 570076 575005 556395 570050 570053 557289 557272 575006 570035 575008 557208 557217 557218 570014 575009 557121 570008 570009 570013 575003 570004 570007 570010 557209 570012 557120 557109 557223 557050 557107 557153 556468 557125 557129 557142 557143 557152 556975 557102 557126 557022 557044 557045 557046 557124 557119 570002 557049 556986 557047 557058 557005 556991 556999 557130 557100 557101 556931 555926 556864 557043 557048 556919 557004 556810 556836 556026 556832 556020 556811 556901 556804 556892 556834 556835 556787 555506 556893 556894 556866 556861 556863 556785 556902 556838 556839 556840 556862 556895 556820 556841 556842 556843 556780 555832 555892 556639 556641 556748 556767 556745 556759 556760 556685 556667 556592 556638 556693 556634 556643 556591 555869 556690 556692 556695 556700 556681 556683 556684 556629 556541 556589 556586 556621 556542 556486 556562 556437 556459 556497 556496 556429 556376 556417 556488 556529 556471 556321 556512 556519 556484 556498 556428 556431 556432 556509 556326 556416 556276 556394 556193 556239 556306 556333 556379 556477 556388 556473 556183 556392 556460 556438 556439 555787 556430 556322 556352 556232 556378 556381 556382 556383 556384 556385 556389 556179 556369 555737 556263 556302 556368 556180 556282 556301 555113 556019 556178 556237 555503 556236 556319 556329 556191 556320 556269 556187 556248 556279 556309 556146 556186 556224 556310 556120 556227 556234 556238 556246 556287 556182 555513 556188 556054 556094 556098 556148 556151 556226 556184 556252 556253 556116 556149 556176 556177 556185 556249 555319 555454 556225 556228 556206 556194 556195 556197 556055 556189 556192 556144 556147 556150 556119 555411 556129 555893 556039 556099 556115 556117 556111 556112 556113 556114 555140 555317 555998 556053 556090 555979 556012 551068 556011 556021 555999 555989 556002 556003 555982 555991 555975 556040 556064 556070 555993 555577 555965 556036 556041 555917 555949 555976 556022 556025 556028 556031 556032 555261 555936 555860 555848 555947 555912 555445 555843 555321 555842 555915 555916 555579 555841 555844 555852 555895 555851 555327 555840 555927 555929 555106 555821 555891 555894 555900 555849 555762 555861 555047 555782 555806 555035 555720 555794 555016 555105 555766 555711 555754 555759 555765 555770 555771 555772 555779 555023 555736 555699 555017 555718 555516 555761 555763 555764 555708 555628 555698 555703 555575 555617 555669 555661 555654 555602 555574 555566 555610 555569 555565 551639 555027 555419 555452 555026 555497 555449 555562 551550 555447 553230 553270 553562 553742 555469 555472 555468 555470 555532 555451 555509 555494 555315 553589 555412 555334 555406 555407 555403 555408 555434 555435 555320 555415 555418 555365 555366 555367 555409 555307 552997 553743 555388 555316 555381 555306 555302 555339 555341 555309 555338 555232 555314 555337 555326 555330 555301 555298 555300 555061 555069 555074 555228 555280 555225 555258 555263 555223 555229 555230 553214 555200 555077 555146 555220 555222 555183 555184 555201 555068 553565 553563 555224 555226 555227 555231 555136 555151 555153 553750 553839 555090 553044 555091 555185 555067 555135 555147 555160 555161 555162 555163 555164 555063 555089 555078 555137 555138 555139 555145 555148 555149 555150 555152 555154 555156 555087 555088 555092 555093 555094 555065 555066 555071 555072 555073 555075 555076 555079 555080 555055 555056 555057 555058 555060 555062 555064 553771 552979 553761 553749 553614 552914 553593 553721 553685 552903 552915 553670 553662 553645 553649 552990 553639 553604 553623 553624 553625 553035 553050 553581 553046 553574 553576 552699 553541 552731 552989 553455 553457 552690 553478 553465 553440 553444 553397 553379 553380 553375 552080 553360 553354 553356 553332 552091 553307 553315 553289 553292 553218 553231 553224 553234 553227 552981 553170 553083 553151 551566 553104 553105 553095 551882 553060 553056 552965 552873 552974 552972 552964 552823 552783 552849 552822 552726 552782 552785 552743 552744 552753 552724 551758 551604 552576 552712 551582 551559 552193 552586 552077 552086 551007 551034 552160 552175 552177 552179 551011 550929 550851 550854 552009 552010 550598 551118 551978 551996 550919 550811 550979 551953 551922 551924 551905 551864 550068 551790 551815 551780 551769 551774 551771 551773 550843 551620 551681 551686 551501 551502 551503 551506 551515 551517 551519 551543 550586 551634 550298 551653 551654 550296 550585 551619 551615 551600 550587 551107 551541 551544 550577 551524 551528 551069 551513 551507 551511 551113 551114 551116 551117 550300 551102 551104 550112 550173 551072 551059 550891 550947 550980 550988 550178 550961 550963 550969 550944 550946 550949 550950 550933 550900 550297 550034 550875 550036 550027 550876 550881 550826 550872 550859 550803 550805 550806 550808 550753 550039 550588 550534 550537 550509 550511 550490 550491 550478 550418 550369 550403 550343 550345 550278 550320 550270 550271 550210 550219 550233 550220 550184 550185 550186 550196 550170 550175 550161 550162 550158 550163 550137 550086 550125 550126 550127 550128 550129 550120 550122 550123 550124 550088 550098 550110 550075 550077 550066 550067 550041 550057 550055 550056 550017 550018 550023 550024 550026 576225 555303 575044 571065 570206 570207 575043 575045 570141 556137 557082 557088 557089 557080 557060 556994 556995 556879 556850 556851 556794 556793 556796 556768 556689 556682 556585 556531 556534 556535 555646 556404 556427 555612 556327 556264 556265 556181 556153 556154 556155 556159 556100 555928 555901 555829 555830 555642 555808 555807 555795 555768 555747 555748 555722 555549 555601 555552 555459 555460 553637 553591 555425 553034 555331 555304 555262 552911 555142 555143 552696 552894 553076 551962 551985 553250 553251 553246 553023 551693 551881 551874 550857 550734 550595 553217 556795 556961 573380 570339 570128 557184 556465 556105 555333 556913 556914 572007 572006 572005 572008 553425 553426 553427 553428 553429 553431 553432 553433 553434 553430 550893 570159 557330 557066 557065 556932 555397 553449 553450 553396 552787 551029 551032 550482 553832 556000 571054 571055 570476 570458 570470 570471 570472 570474 570478 570479 570466 570467 570460 570468 570464 570465 570463 570469 570287 570168 556198 556199 555957 555958 555959 555490 553644 553515 553514 553022 552034 551581 550890 550760 550192 573356 570487 570485 570491 570493 570486 570249 570225 557282 570209 557333 557336 557295 556918 556954 556175 556798 556929 555995 556637 556696 555968 556627 555983 555994 555914 555937 555938 555896 555897 552694 555492 555340 555170 555173 553143 553599 552235 550626 555691 551706 555681 555672 555685 553825 552148 551710 553815 552098 552099 551724 551723 553257 553820 555673 552149 555692 555695 555697 553817 553818 552157 552124 555678 553259 555674 551719 552115 552126 555676 555683 555686 555687 555688 552151 552096 552139 551712 551718 552097 552105 552111 552134 552135 552136 552137 552143 552145 552147 552156 553253 553269 553819 553823 550022 551705 552095 552101 552102 552109 552129 552130 552140 552141 553255 552150 551716 551717 552107 552117 552118 552125 552127 552128 552131 552138 552142 552144 552146 552155 552094 552119 552114 551713 552116 552120 552106 552103 552110 552112 552113 552100 552108 551714 573299 556906 556911 556939 556163 556908 556910 551596 551587 551591 570178 555886 552600 551986 570330 550413 573409 573078 573081 573100 573338 573339 570238 573411 573415 572034 573406 571050 571561 572003 573431 573432 573433 573434 570020 572035 573435 573340 573341 573303 573329 573254 573086 556857 557212 573331 573156 573330 571073 555275 555890 556127 556134 556174 556448 556449 556451 556457 556502 556525 556527 557344 557345 557346 557347 557348 557349 557350 557351 557352 557353 570123 570125 570195 571013 571014 571036 571045 571060 573001 573046 556126 570056 573083 572011 573084 573154 573155 573094 570027 572014 572015 573085 556755 557276 573076 571560 570126 573026 573050 573009 557213 571059 571051 571052 556516 571017 556511 571016 571021 570324 571018 570301 556289 570111 570107 570110 570112 556450 570194 570197 570160 570173 570175 570150 556128 556752 570108 570106 570109 557311 570036 570040 570041 570042 557277 556865 557308 557307 557309 557310 557198 557053 556754 556962 550309 556989 557010 556849 550790 552612 555883 556747 556447 556751 556753 556756 556499 556500 553656 556433 551934 556095 555873 556679 556093 556645 556096 556314 556501 555836 556133 556398 556097 556478 556397 556436 556455 556092 556217 556219 556220 556221 556386 556467 556435 556456 553780 555834 556377 553184 556063 555644 556273 555278 556250 555637 556058 555234 555920 555245 556106 555310 555904 556057 555969 555970 555906 555032 555954 552611 555952 555777 555274 555279 555244 555246 555786 555277 555243 553116 553131 553194 553840 555237 555238 555233 555236 555235 553051 555328 555329 552610 555198 551767 553793 552060 553634 553185 552660 553126 552059 552066 551958 551750 553119 553120 553121 553123 553129 553133 553135 553057 552055 552056 552057 552058 552061 552062 552895 552906 552916 552901 552054 551747 552624 552727 552070 550079 550973 550975 551973 550276 550569 551576 551977 551537 552065 552067 552068 552069 552071 551766 551016 550883 550725 551752 550207 551754 551001 551002 550788 550465 550951 550955 550603 551689 550560 550738 550448 550652 550377 550188 550227 550406 550136 573378 573379 573390 573334 550244 573067 571068 550238 570291 570290 550242 556837 556746 550152 550234 550253 556507 550048 556485 555680 555910 555971 555943 550047 552074 555312 555436 555424 555383 550236 550566 553004 550304 553140 553061 553000 552664 552665 552666 552019 551979 550572 550862 550727 550687 550570 550514 550353 550315 550237 550245 550258 550259 550260 550235 550254 550046 550050 571034 570049 550719 555110 550532 557323 557324 557325 557041 557042 555884 556707 555735 555102 556726 557185 556371 556372 556373 556374 555356 553261 553241 552896 550201 557165 557186 556292 556141 555269 555270 553147 556162 556736 555740 550091 556073 556074 556075 556076 556077 556612 573200 550710 556610 555992 555721 555723 553513 553402 552913 550476 556729 556730 556824 556846 570096 570097 570115 570116 570117 570273 570278 573047 573180 556305 556788 551605 556122 556121 555603 555716 556964 556983 557038 557039 556887 556677 556581 556481 556482 556483 555812 555828 555440 553866 553648 552709 552760 552212 552217 552218 552219 551910 551911 551862 551115 551120 550898 550404 575134 570519 570370 570550 570552 570551 570536 570532 570553 570520 570539 570542 570555 570434 570440 570481 570497 570502 555405 570343 570113 570092 557284 556941 557086 557017 556938 556934 556935 556470 556363 556364 556365 556330 555385 556313 555386 555621 555653 555798 555384 555387 555453 555038 553568 553569 553570 553571 553572 553142 553068 552678 551987 551110 550787 555670 555647 555939 555940 553745 553746 572013 570246 557402 570247 557368 557357 557074 557116 557067 556375 556393 556644 556380 555810 555551 555400 570534 557332 557164 556867 555881 553480 552667 553233 550974 550583 550419 570263 555955 552021 550897 570143 557078 556668 556680 556599 556608 556596 556489 556536 556172 556169 555108 555946 555675 555889 555053 555489 555242 553156 553048 551916 552987 552988 552843 552832 552575 552578 552568 550798 552013 551554 550714 550399 550400 550401 550402 555015 553216 553220 553045 552859 552580 551915 551056 551049 550724 550421 553500 553501 570198 555924 557051 557055 578062 570296 575007 555272 552926 557244 570114 557188 556426 556425 552668 550794 576279 573203 572016 573063 573064 573065 570305 570306 570237 557405 557305 557076 557081 557021 556418 556056 555168 555933 555934 556051 556005 555323 555157 555524 553578 551005 550081 550053 550078 573402 570314 578065 578047 578048 570313 575059 570317 570170 576009 575060 575058 570151 575061 556966 556921 556922 556923 556924 556904 556868 556720 556721 556722 556619 556620 556657 555541 556297 556298 556299 556336 556338 556339 556342 556344 556346 556347 556348 556349 556350 556351 556353 556354 556355 556035 556013 556014 556038 555822 555679 555756 555769 555755 555757 555052 555658 555659 555346 555348 555343 555345 555210 555432 555344 555347 555208 555203 555204 555205 555206 552809 552984 552985 553087 555207 555211 553821 552902 553760 553687 553618 553619 553620 553621 553622 553456 553248 552890 551936 553009 552968 552891 552893 552892 552876 551730 552778 552770 552171 551697 551696 551964 551623 551748 551756 551744 551732 551608 551015 551012 550874 551038 550971 550111 550908 550028 550894 550749 550010 550411 550331 550249 550213 550145 571064 557409 557411 570122 557132 557131 557040 557104 556847 556789 556705 556671 556665 556600 556602 556605 556304 555558 556261 556262 556244 555553 555978 555948 555586 555629 555402 555267 553039 553049 555318 555129 555128 552662 552698 551960 551899 552878 552771 552741 552078 550669 550720 550211 551614 550662 557343 556956 556957 575063 575064 556800 556801 556955 556802 555898 553521 552053 551903 550179 557070 557071 557072 557073 571057 573041 573042 573183 576006 578027 573246 573162 570450 570307 570342 556936 570270 557147 557149 557056 557037 556947 556875 556819 556772 556773 555013 555014 556697 556649 556633 556635 556624 556504 556401 556461 556328 556307 556258 556259 556286 556048 555987 555988 555986 555956 555903 555867 555862 555174 555175 555176 555177 555179 555180 555239 555240 555265 555656 555583 555633 555584 555564 555543 555458 555438 555423 555422 555358 555132 555182 555294 555295 555296 553626 555095 555158 555209 555096 555181 555159 555125 555126 555127 552525 553284 555006 551850 553300 553172 552925 551053 552844 550040 550059 552631 552536 552538 551824 551649 550065 551520 551602 551092 550750 550526 550528 550535 550064 550097 550060 550030 550058 550062 557398 570086 578053 573370 573416 573405 573373 573429 573261 573372 573297 573361 573208 557249 557285 557312 573168 573169 573170 573182 573181 557314 573206 573210 573205 571515 573201 573204 573196 573272 573066 573167 570295 576222 571514 576209 570303 570304 570311 571517 571040 570188 556142 556157 570183 570119 570283 570121 552883 555221 557140 570090 552072 556597 556331 556332 556317 556288 556190 556240 556160 556034 556084 556016 556017 555990 555854 555780 555781 555760 555758 555505 555700 555352 553111 555650 555488 555446 555433 555416 555199 555322 555070 555247 553856 555120 555131 553632 553508 553263 553229 553110 553090 553094 553091 552940 552818 552777 552719 552720 552682 552626 552613 552595 552598 552505 552049 552018 552007 551666 551622 551531 551533 551024 550449 550292 550148 555456 555732 555778 556009 556042 556222 556223 556230 556294 556574 556575 556577 556646 556725 556741 556742 556786 556884 556891 556900 556917 556950 557112 557138 557141 557144 557145 557281 557303 557341 573211 578066 557410 556813 556814 556815 557404 557426 557397 557399 557180 557182 557181 557183 557377 556822 556823 557248 556905 557251 557006 556826 556877 556874 556933 556888 556873 556871 556872 553522 554014 553913 553891 553047 552905 557403 557420 570043 571044 555705 555706 555707 557400 557304 556424 556777 556422 556419 555194 556423 556033 556472 556229 555997 555796 555797 552663 555548 555550 555651 552643 552644 552645 552646 553364 553366 555196 555254 555256 555257 555266 555379 555219 552637 552640 553363 553365 553368 552636 553215 553758 553759 553399 552833 552834 552882 551678 551920 551994 551975 551909 551913 551684 551729 551101 551111 550475 550477 556970 555491 573130 557157 557154 556951 556952 556844 556781 556582 556446 556452 556453 556454 555605 555619 555627 555636 555724 555528 556065 556068 556067 556069 556071 555818 555414 553523 553566 555413 553834 553835 551573 551578 551946 557118 556544 556543 556161 556050 556001 555857 555856 555668 555216 555213 553613 551692 550821 550071 551725 578021 571519 570333 555268 553739 555012 553014 550273 557273 556210 556209 555726 555305 555050 551779 550855 550092 551529 550860 550301 550424 555122 555260 555613 555626 555783 555784 555785 555980 555984 556241 556537 556779 556782 556916 573389 573191 573192 573193 573108 557431 557423 557434 571069 571530 557415 570326 570327 570337 570165 557298 557246 556984 556530 556940 556493 556492 556676 556675 555299 555847 555465 553382 552615 550711 550160 552188 552199 552201 552202 552204 556290 556402 556403 556408 571023 573080 570299 571011 570255 570085 556130 556340 570079 570091 570029 552015 556699 556766 556642 556648 556640 556549 556550 556551 556552 556553 556554 556555 556556 556558 556559 556560 556561 556563 556564 556566 556567 553128 556361 556366 555518 556315 556405 556412 556247 556280 556281 556283 556284 556291 553130 556337 556341 556010 555614 555043 555517 555525 553525 555531 555342 555354 555355 555371 555372 555373 555374 555375 555376 555378 555380 553040 553059 555377 555252 553052 555218 555134 552899 553458 553538 553038 553560 553355 552203 552603 552897 553078 553459 552657 553373 553132 553134 552656 553117 553118 553122 553124 553125 553127 551935 551959 551963 551969 552661 552763 552191 552195 552196 552200 550741 552036 550793 550796 550789 550810 550703 550704 550705 550706 550707 550500 550032 550779 550628 550450 552670 570292 573061 557176 571037 556444 555478 555479 555480 553041 555361 555103 553483 551883 551884 551783 551525 550848 551647 551535 551108 550783 550785 556688 555885 556694 555878 556687 555887 555638 555112 556256 556360 556255 555324 555031 555863 555945 555527 555104 555025 555630 555567 555439 555410 555259 555264 553036 553633 553833 552912 553776 553663 553470 553424 553388 553394 553219 552639 551947 552959 552936 552846 552824 551851 551624 552692 552020 551526 551949 551918 551901 551727 551736 551698 551629 551611 551014 551017 551018 550777 550650 550473 550446 570038 557313 557278 557024 557026 557034 557025 557033 557027 556806 556807 556808 556809 556632 556631 556520 556521 556323 556391 556293 556242 556085 555981 556018 555905 555745 555799 555800 553767 555648 555649 555639 555510 555521 555511 555370 555250 555249 555251 555059 553782 553781 550601 553787 553788 553447 553264 553150 553384 552850 550774 550423 570523 557187 556547 556158 555960 555727 555728 555572 553752 553472 553408 552776 551880 551793 551655 550458 550151 573057 573062 573059 552596 573359 578039 573197 573198 578038 575031 571523 575082 570215 575032 557379 557178 557228 557110 556830 556769 556691 555498 556406 556407 556443 555501 556268 555845 555846 555749 555751 553684 555499 556946 555500 555463 555464 555462 555457 555441 555442 555426 553612 555325 552803 550584 550094 575055 573189 573427 570060 557334 556878 556362 555974 555882 555973 555850 555576 555578 555655 555526 555009 555427 555054 555428 555585 555034 555657 555573 553479 552908 550994 551943 551889 551914 550031 550021 550015 550033 550019 550016 550087 550845 550856 551595 551890 551942 551971 557335 570047 570075 570089 575073 576277 576278 578054 571070 570265 555130 555420 553603 553583 553584 553585 553586 553587 555292 555293 553610 553611 552887 552037 552038 550663 571512 573426 571539 570024 573387 570025 570220 570224 573029 573030 570242 556165 570399 570402 570219 570250 570221 570223 570222 575048 570228 557338 557286 556993 557287 557179 556594 556739 556799 557028 557054 556790 557020 556166 556762 556763 556764 556750 556765 556757 556595 556604 556607 556698 556411 556325 555081 556211 556212 555858 555859 555483 555515 555028 555557 555520 555522 555523 555029 555496 555007 555620 555624 555049 555467 552904 555399 555398 555349 553526 555351 555350 553490 553237 553054 551945 550295 551886 550847 550600 550293 550983 550291 550978 550985 550575 573423 572044 557371 551672 551673 551675 550885 550886 550887 570495 575117 570500 573292 570517 557290 570454 570494 575097 575098 570431 570484 573150 575099 575123 570505 570507 575066 570325 570328 557105 570506 570508 570433 556792 570331 575062 570321 570329 570322 570323 575065 557268 570100 570101 570103 570226 550089 557166 557106 557167 556942 556943 556944 556903 556821 553768 555866 556024 556583 555791 555792 556533 556442 555865 556023 555864 555801 555743 555742 555802 555804 555645 555640 555392 555036 553857 553290 552765 553293 553267 553266 552162 553268 553285 553286 553244 553291 553294 553304 551885 552159 552167 552168 553099 552164 552163 552166 552885 552169 552170 552029 552031 551571 551896 552161 552027 552028 552032 552158 551560 551908 551786 551562 551569 551822 551826 551827 551828 551557 551100 551536 550730 550977 550912 550896 550756 550504 550488 550700 550539 550512 550540 550436 550543 550322 553765 557301 557302 570427 570527 573288 573289 573308 578028 578051 578052 556653 556655 556889 556654 556413 556414 556400 556415 553322 553546 553321 571046 550350 550106 550605 551707 553160 900324 900325 900323 900320 900322 900319 900321 555044 555046 553601 900336 555874 556049 556311 556546 556593 556771 556770 556818 556817 556876 575001 900415 900413 900414 575002 557052 557075 557079 557103 557136 557151 557146 557135 557250 557260 557258 557259 557257 570016 557262 557275 570022 570039 557337 570044 570055 570074 575010 570099 570098 570105 570102 575023 575025 575024 570120 570127 570132 570133 570134 575019 570152 570157 570156 570154 570155 570153 570190 571001 571002 571003 571004 571006 571005 571008 571007 571010 571009 576003 576004 576001 570203 570202 570200 570199 570201 576002 570232 557226 570239 570243 570251 570244 570269 570268 570272 570276 570280 570277 570286 570293 570289 570312 570310 570315 557433 570316 570319 570320 570318 571012 570334 570336 575068 575067 575069 575071 575072 575070 570392 570396 570394 575049 570395 570397 570401 570393 570398 570407 570406 570408 570404 570403 570405 570409 576008 571061 576208 571516 576210 571538 576010 573016 573028 573023 573025 573032 573033 573031 573013 572004 573048 576237 573053 570501 570499 570457 570445 570518 570453 570451 570449 570448 570446 570444 570425 570533 570529 570503 570496 570426 573055 573058 572009 572010 576013 573071 573072 573070 573079 573096 573113 572019 573111 578014 572020 570480 570430 578017 570531 570530 570528 570498 570483 570456 570452 570447 573138 573139 573140 570443 570441 570439 570438 570437 570436 570435 570432 573135 573142 573143 573137 573153 572026 573177 573163 573176 573174 573161 573164 573175 573173 573159 573160 578035 578034 578036 578037 573212 578040 573207 573213 573209 610053 573216 570429 570428 573222 573231 573232 573233 573244 573255 573238 573242 573239 573247 573243 573240 573248 573241 573264 573256 573245 573257 573258 573259 573275 573260 573262 573263 570482 570455 570442 573281 573286 573285 573301 570526 570524 610062 610064 610065 610066 610067 573306 573312 573309 573317 573319 573318 573332 573333 573345 573344 573353 573354 573369 573365 578064 573368 573371 573375 573376 572037 578076 576271 576270 578073 578080 578086 578087 576273 578081 576274 576275 576276 573397 578084 573400 573398 578085 573408 573399 573386 573410 573413 573417 573424 573428 572041 572042 578098 573437 573440 572040 610071 610072 610073 610074 610075 610076 610077 610078 610079 610080 610083 610084 610085 573478 570473 610094 571250.
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- ACUSON Juniper Diagnostic Ultrasound System 11335791Siemens Medical Solutions USA, Inc. · Transducer, ultrasonic, diagnostic3 recalls2 reports
- ACUSON Juniper Select Diagnostic Ultrasound System 11653093Siemens Medical Solutions USA, Inc. · Transducer, ultrasonic, diagnostic2 recalls0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 94211).
- Z-1600-2024 · Open, ClassifiedACUSON Redwood 1.0 Diagnostic Ultrasound System, REF: 11503314, running software version VA10A to VA10G
- Z-1601-2024 · Open, ClassifiedACUSON Maple Diagnostic Ultrasound System, REF: 11711750, ACUSON Maple Select, REF: 11711779, running software version VA10A
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page