ACUSON Juniper Select Diagnostic Ultrasound System 11653093Recalls & reported problems
Siemens Medical Solutions USA, Inc. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Aug 15, 2024Class IIOpen, Classified
If ultrasound systems with software, are changed from factory default to : 1) Milliliters per second (ml/sec, mL/sec) or 2) Milliliters per minute (ml/min, mL/min); then systems will perform incorrect unit conversion of liters to milliliters using multiplier 100 instead of 1000, resulting values 10x smaller than actual, which could contribute misdiagnosis or negatively influence patient management
Z-3215-2024 · Root cause: Software design · Recall details · FDA record ↗
- Feb 23, 2024Class IIOpen, Classified
On ultrasound systems, when Cardiac DICOM SR feature is configured to display either minimum or maximum measured value, and multiple cardiac region measurements are made, and results are exported into the SR feature, then The SR viewer will display the LAST measured value, not min or max, which could contribute to patient condition misdiagnosis or negatively influence patient management decisions.
Z-1599-2024 · Root cause: Software design · Recall details · FDA record ↗
Other Transducer, ultrasonic, diagnostic devices
- Site~Rite 8 Ultrasound System 9770603Bard Access Systems, Inc. · Transducer, ultrasonic, diagnostic620 reports
- ULTRASONIC PROBE UM-S20-17SOlympus Medical Systems Corp. · Transducer, ultrasonic, diagnostic2 deaths344 reports
- Transducer X7-2t 989605414122Philips Ultrasound LLC · Transducer, ultrasonic, diagnostic6 recalls21 reports
- EPIQ Diagnostic Ultrasound System 989605488741Philips Ultrasound LLC · Transducer, ultrasonic, diagnostic3 recalls165 reports
- bk3000, bk3500, bk5000, bkActiv 2300Bk Medical Aps · Transducer, Ultrasonic, Diagnostic6 recalls1 report
- Transducer L12-3 989605412072Philips Ultrasound LLC · Transducer, ultrasonic, diagnostic5 recalls1 report
- Transducer L15-7io 989605387852Philips Ultrasound LLC · Transducer, ultrasonic, diagnostic5 recalls0 reports
- Transducer eL18-4 EMT 989605431531Philips Ultrasound LLC · Transducer, ultrasonic, diagnostic5 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page