Skip to content

CADD-Solis Ambulatory Infusion Pumps, Model 2100. Intended… recallZ-2419-2024 · Smiths Medical ASD Inc. · initiated Feb 27, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Smiths Medical ASD Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

CADD-Solis Ambulatory Infusion Pumps, with software versions before v4.3, may have the following issues:1. Upstream Occlusion, 2. Stop and Power Keys Unresponsive, 3. Manual Mode Air Detector, 4. Single Bubble Air Detection, 5. Error Codes Not Displayed at Power Up, 6. Audible Alarm, 7. Low Sensitivity Air in Line Detection Threshold, 8. PharmGuard Server Password. Smiths Medical corrected many of the issues included in this notification in previous software updates and the corrections were carried forward into all subsequent software releases. Please ensure you have the most recent CADD software (v4.3 - 2023) installed on your pumps.

FDA root cause category: Software Manufacturing/Software Deployment

What the company is doing

Smiths medical issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 02/27/2024 via letter distributed using traceable means. The notice explained the issues affecting the devices, potential risks, and requested the following: "Required Actions for Users: 1. Locate all affected pumps in your possession and ensure all users or potential users of these devices are immediately made aware of this notification and proposed mitigations. 2. Complete and return the attached Response Form to [email protected] within ten days of receipt to acknowledge your understanding of this notification, even if you do not have the affected product. 3. DISTRIBUTORS: If you have distributed potentially affected products to your customers, please immediately forward this notice to them. Request that they complete the response form and return it to [email protected] Follow up Actions by Smiths Medical: Smiths Medical is sending this notification to all affected CADD-Solis customers and is addressing the issues described in this notice through software updates. Smiths Medical will contact you to schedule the implementation of the software updates." For questions contact: 1. Additional information or technical assistance: [email protected]; 1-800-241-4002, option 3, 4 (M-F, 8:00 am 6:00 pm CT) 2. Questions about this Field Correction Notice: icumed.custhelp.com/app/market-action

Recall details

Recall number
Z-2419-2024
Status
Open, Classified
Initiated by the company
Feb 27, 2024
Posted by the FDA
Aug 3, 2024
Recalling firm
Smiths Medical ASD Inc.
Quantity
9284 units
Product code
MEA
FDA submissions
K072144
Distribution
Worldwide

Product description

CADD-Solis Ambulatory Infusion Pumps, Model 2100. Intended for the following uses: Indicated for the following uses: Intra-arterial, subcutaneous, intraperitoneal, in close proximity to nerves, into an intraoperative site, epidural space or subarachnoid space. Therapies that require a continuous rate of infusion, intermittent bolus, and/or with patient-controlled demand doses.

Code information (models, lots, serial numbers)

Show the full code information (85k characters)

REF List Number: 21-2101-0200-02; UDI/DI 30610586035337, Serial Numbers: 1006964, 1006966, 1006969, 1008404, 1008405, 1009451, 1010631, 1014398, 1010689, 1010420, 1010442, 1010443, 1010444, 1010445, 1005368, 1005365, 1006283, 1006298, 1008409, 1009427, 1009428, 1009429, 1009431, 1009433, 1010419, 1011448, 1011450, 1011451, 1011452, 1006964, 1006966, 1008404, 1009451, 1011447, 1011449, 1010432, 1009715; REF List Number: 21-2101-00; UDI/DI 10610586032950, Serial Numbers: 1002001, 1002001, 1002006, 1002009, 1002010, 1002014, 1002016, 1002017, 1002018, 1002023, 1002025, 1002026, 1002028, 1002031, 1002033, 1002034, 1002036, 1002037, 1002039, 1002040, 1002041, 1002042, 1002043, 1002045, 1002051, 1002053, 1002054, 1002055, 1002058, 1002059, 1002060, 1002063, 1002064, 1002065, 1002067, 1002068, 1002070, 1002072, 1002074, 1002077, 1002078, 1002081, 1002082, 1002085, 1002086, 1002088, 1002090, 1002094, 1002095, 1002097, 1002101, 1002103, 1002105, 1002106, 1002107, 1002108, 1002109, 1002110, 1002111, 1002113, 1002114, 1002117, 1002118, 1002119, 1002120, 1002121, 1002122, 1002123, 1002125, 1002259, 1002392, 1002406, 1002407, 1002408, 1002409, 1002410, 1002412, 1002414, 1002415, 1002416, 1002417, 1002418, 1002419, 1002420, 1002421, 1002422, 1002423, 1002424, 1002425, 1002426, 1002427, 1002428, 1002432, 1002433, 1002434, 1002436, 1002437, 1002439, 1002440, 1002442, 1002443, 1002444, 1002445, 1002448, 1002449, 1002452, 1002453, 1002497, 1002547, 1002575, 1002581, 1002583, 1002588, 1002593, 1002607, 1002620, 1002627, 1002658, 1002662, 1002753, 1002766, 1002781, 1002800, 1002815, 1002821, 1002824, 1002828, 1002831, 1002832, 1002833, 1002836, 1002838, 1002842, 1002845, 1002848, 1002868, 1002877, 1002880, 1002899, 1002952, 1002984, 1002987, 1003161, 1003170, 1003172, 1003183, 1003192, 1003632, 1003885, 1004654, 1005061, 10012441, 1005352, 1005445, 1006255, 1006625, 1006626, 1006627, 1006628, 1006629, 1006630, 1006631, 1006634, 1006635, 1006636, 1006709, 1007033, 1007509, 1007546, 1007593, 1007608, 1007627, 1007842, 1009082, 1011821, 1010639, 1010640, 1010642, 1010645, 1010648, 1010649, 1010650, 1010655, 1010658, 1010660, 1010662, 1010717, 1010805, 1011134, 1011140, 1011143, 1011290, 1011486, 1011782, 1011816, 1011818, 1011823, 1012047, 1012080, 1012244, 1002003, 1002007, 1002024, 1002413, 1002455, 1002553, 1002817, 1006633, 1002013, 1002050, 1002052, 1002056, 1002066, 1002022, 1010863, 1009085, 1002411, 1002080; REF List Number: 21-2101-0100-50; UDI/DI 10610586035548, Serial Numbers: 1011545, 1009720, 1010437, 1010634, 1010635, 1010666, 1010667, 1010668, 1010669, 1010670, 1010671, 1010672, 1010673, 1010674, 1010675, 1010676, 1010679, 1010682, 1010684, 1010685, 1010686, 1010687, 1010688, 1011520, 1011522, 1011528, 1011537, 1011538, 1011523, 1011542, 1011549, 1011550, 1011535, 1011556, 1011995, 1011996, 1010677, 1010680, 1010681, 1012342, 1012343, 1012344, 1012340, 1012341, 1011994, 1011515, 1011516, 1011517, 1011518, 1011519, 1011521, 1011524, 1011525, 1011526, 1011527, 1011529, 1011531, 1011533, 1011534, 1011536, 1011539, 1011540, 1011541, 1011543, 1011544, 1011546, 1011547, 1011548, 1011551, 1011552, 1011553, 1011554, 1011555, 1011557, 1011558, 1011559, 1011560, 1011561, 1011562, 1011563, 1011532, 1011564; REF List Number: 21-2101-0200-03; UDI/DI 30610586035344, Serial Numbers: 1008473, 1009307, 1009483, 1009484, 1009485, 1009486, 1009487, 1009488, 1009489, 1009490, 1009507, 1009508, 1009514, 1009520, 1009522, 1009524, 1009526, 1009527, 1009792, 1009795, 1009797, 1009800, 1009801, 1009802, 1009803, 1009804, 1009806, 1009833, 1009834, 1009835, 1009836, 1009837, 1009838, 1009839, 1009840, 1009841, 1010370, 1010372, 1010374, 1010375, 1010381, 1010386, 1010390, 1010411, 1010423, 1010427, 1010429, 1010467, 1010474, 1010514, 1010573, 1010574, 1010580, 1010581, 1010582, 1010583, 1010585, 1010586, 1010587, 1010589, 1010590, 1010591, 1010592, 1010593, 1010594, 1010595, 1010596, 1010597, 1010599, 1010600, 1010601, 1010602, 1010603, 1010605, 1010606, 1010611, 1010613, 1010614, 1010628, 1010630, 1010384, 1010376, 1010377, 1010378, 1010383, 1010380, 1009796, 1010382, 1010387, 1010588, 1008468, 1010541, 1010584, 1008467, 1009832, 1010385, 1009493, 1009494, 1009495, 1009497, 1009499, 1009500, 1009501, 1009503, 1009504, 1009502, 1009452, 1009453, 1009454, 1009455, 1009456, 1009457, 1009458, 1009459, 1009460, 1009462, 1009463, 1009464, 1009465, 1009466, 1009467, 1009468, 1009470, 1009471, 1009472, 1009473, 1009474, 1009475, 1009476, 1009478, 1009479, 1009480, 1009498, 1009515, 1009516, 1009518, 1009519, 1009521, 1010607, 1010608, 1010609, 1010610, 1009523, 1010371, 1009827, 1009828, 1009829, 1009830, 1009831, 1009509, 1009510, 1009511, 1009512, 1009513, 1010388, 1010389, 1010405, 1009807, 1009808, 1009809, 1009810, 1009811, 1009812, 1009813, 1009814, 1009815, 1009816, 1009817, 1009818, 1009819, 1009820, 1009821, 1009822, 1009823, 1009824, 1009825, 1009826, 1010379, 1009492, 1009491, 1009505, 1009506, 1009525, 1009798, 1009528, 1009461; REF List Number: 21-2101-0200-06; UDI/DI 10610586035357, Serial Numbers: 1003934, 1006600, 1006601, 1006602, 1009591, 1009592, 1009593, 1009594, 1009595, 1009596, 1009597, 1009598, 1009599, 1009600, 1009601, 1009602, 1009603, 1009604, 1009605, 1009606, 1009607, 1009608, 1009609, 1009610, 1009611, 1009612, 1009613, 1009614, 1009615, 1009616, 1009617, 1009618, 1009619, 1009620, 1009621, 1009622, 1009623, 1009624, 1009625, 1009626, 1009627, 1009628, 1009629, 1009630, 1009631, 1009632, 1009633, 1009634, 1009635, 1009636, 1009637, 1009638, 1009639, 1009640, 1009641, 1009642, 1009643, 1009644, 1009645, 1009646, 1009647, 1009648, 1009649, 1009650, 1009651, 1009305, 1009588, 1009589, 1009590; REF List Number: 21-2101-0200-07; UDI/DI 30610586035368, Serial Numbers: 1010450, 1010451, 1010452, 1010453, 1010454, 1010455, 1010456, 1010457, 1010458, 1010459, 1010460, 1010462, 1010463, 1010464, 1010616, 1010617, 1010618, 1010619, 1010620, 1010621, 1010622, 1010623, 1010624, 1010625, 1010626, 1025383, 1010447, 1010449, 1010448, 1010446; REF List Number: 21-2101-0200-12; UDI/DI 10610586035395, Serial Numbers: 1009584, 1009585, 1009586, 1009587, 1009582, 1009583, 1009581, 1009444; REF List Number: 21-2101-0200-14; UDI/DI 10610586035401, Serial Numbers: 1009543, 1009545, 1009547, 1009548, 1009552, 1009549, 1009550, 1009551, 1009443, 1009544, 1009546, 1009553; REF List Number: 21-2101-0200-17; UDI/DI 10610586035418, Serial Numbers: 1009532, 1009529, 1009542, 1009535, 1009536, 1009537, 1009538, 1009539, 1009540, 1009541; REF List Number: 21-2101-0200-232; UDI/DI 10610586035425, Serial Numbers: 1001841 REF 21-21-1-0200-236; UDI/DI 10610586035449, Serial Numbers: 1001870; REF List Number: 21-2101-0200-237; UDI/DI 10610586035456, Serial Numbers: 1001844; REF List Number: 21-2101-0200-238; UDI/DI 10610586035463, Serial Numbers: 1001893; REF List Number: 21-2101-0200-242; UDI/DI 10610586035487, Serial Numbers: 1001838; REF List Number: 21-2101-0200-244; UDI/DI 10610586035494, Serial Numbers: 1001888 REF List Number: 21-2101-249, 21-2111-0402-02, 21-2120-0103-02, 21-2120-0105-15L, 21-2111-0402-09JP, 21-2111-0400-78, 21-2120-0105-06L, 21-2120-0105-50, 21-2120-0105-08L, 21-2120-0103-06, 21-2120-0103-06L, 21-2120-0105-02L, 21-2120-0105-07L, 22-4095-0203-51, UDI/DI None, Serial Numbers: 000688, 003836, 003839, 1000681, 1001701, 1001702, 1001703, 1001704, 1001705, 1001706, 1001707, 1001708, 1001709, 1001710, 1001711, 1001712, 1001713, 1001714, 1001715, 1001716, 1001717, 1001718, 1001720, 1001721, 1001722, 1001724, 1001725, 1001726, 1001727, 1001728, 1001729, 1001730, 1001732, 1001733, 1001735, 1001736, 1001737, 1001738, 1001739, 1001740, 1001742, 1001743, 1001744, 1001746, 1001748, 1001750, 1001752, 1001753, 1001754, 1001755, 1001756, 1001757, 1001758, 1001759, 1001761, 1001762, 1001763, 1001765, 1001766, 1001770, 1001771, 1001772, 1001773, 1001774, 1001775, 1001777, 1001778, 1001779, 1001780, 1001781, 1001782, 1001784, 1001785, 1001786, 1001787, 1001788, 1001789, 1001790, 1001895, 1001901, 1001902, 1001906, 1001909, 1001911, 1001912, 1001913, 1001914, 1001915, 1001916, 1001917, 1001918, 1001919, 1001921, 1001922, 1001924, 1001925, 1001926, 1001927, 1001928, 1001929, 1001930, 1001931, 1001932, 1001933, 1001934, 1001935, 1001936, 1001937, 1001938, 1001939, 1001940, 1001941, 1001942, 1001943, 1001944, 1001945, 1001946, 1001947, 1001948, 1001949, 1001950, 1001951, 1001952, 1001953, 1002002, 1002005, 1002008, 1002035, 1002038, 1002044, 1002047, 1002048, 1002073, 1002093, 1002100, 1002431, 1002435, 1002441, 1002451, 1002454, 1002570, 1002571, 1002600, 1002604, 1002614, 1002623, 1002705, 1002812, 1002816, 1002818, 1002822, 1002835, 1003010, 1003110, 1003119, 1003128, 1003153, 1003156, 1003459, 1003524, 1003595, 1003638, 1003644, 1003677, 1003682, 1003760, 1003770, 1003794, 1003810, 1003822, 1003825, 1003826, 1003828, 1003839, 1003851, 1003857, 1003863, 1003903, 1004023, 1004068, 1004069, 1004070, 1004071, 1004072, 1004650, 1005607, 1005908, 1005909, 1005910, 1005911, 1005913, 1005914, 1005915, 1005916, 1005917, 1005921, 1005923, 1005925, 1005926, 1005927, 1006153, 1006167, 1006391, 1006632, 1006896, 1006897, 1006898, 1006899, 1007029, 1007101, 1007156, 1007217, 1007506, 1007603, 1007745, 1007903, 1007974, 1007975, 1008186, 1008469, 1008474, 1008487, 1009142, 1009170, 1009175, 1009181, 1009183, 1009430, 1010417, 1010636, 1010643, 1010647, 1010653, 1010663, 1010665, 1010932, 1011034, 1011289, 1011814, 1011815, 1011817, 1011819, 1011822, 1011852, 1012859, 1012875, 1012886, 1012887, 1012897, 1012898, 1012904, 1012910, 1013044, 1013116, 1013119, 1013129, 1013130, 1013134, 1013135, 1013136, 1013137, 1013138, 1013139, 1013141, 1013142, 1013143, 1013152, 1013161, 1013164, 1013793, 1013833, 1013905, 1014047, 1014058, 1014059, 1014077, 1014093, 1014164, 1014168, 1014171, 1014177, 1014181, 1014183, 1014185, 1014223, 1014224, 1014229, 1014231, 1014232, 1014233, 1014236, 1014250, 1014251, 1014585, 1014586, 1014591, 1014600, 1014601, 1014602, 1014603, 1014609, 1014614, 1014616, 1014621, 1016236, 1016237, 1016238, 1016240, 1016243, 1016395, 1019369, 1021955, 1022089, 1022117, 1022358, 1022360, 1022363, 1022709, 1022830, 1022936, 1022965, 1022974, 1023020, 1023021, 1023029, 1023033, 1023049, 1023050, 1023054, 1023059, 1023062, 1023066, 1023070, 1024086, 1001783, 1003668, 1003670, 1003671, 1003674, 1003680, 1003681, 1003687, 1001767, 1001745, 1001769, 1003883, 1001747, 1001751, 1001760, 1001741, 1001719, 1001838A, 1001841A, 1001844A, 1001870A, 1001888A, 1001893A, 1003836A, 1008476A; REF List Number: 21-2101-51; UDI/DI 10610586031854, Serial Numbers: 691953, 691954, 694918, 694919, 694964, 1001814, 1001816, 1001817, 1001818, 1001819, 1001820, 1001821, 1001822, 1001823, 1001824, 1001825, 1001826, 1001827, 1001828, 1001829, 1001830, 1001831, 1001832, 1001833, 1001834, 1001835, 1001836, 1001837, 1001839, 1001840, 1001842, 1001843, 1001845, 1001846, 1001847, 1001848, 1001849, 1001850, 1001851, 1001852, 1001853, 1001854, 1001855, 1001856, 1001857, 1001858, 1001859, 1001860, 1001861, 1001862, 1001863, 1001864, 1001865, 1001866, 1001REF List867, 1001868, 1001869, 1001871, 1001872, 1001873, 1001874, 1001875, 1001876, 1001877, 1001878, 1001879, 1001880, 1001881, 1001882, 1001883, 1001884, 1001885, 1001886, 1001889, 1001890, 1001891, 1001892, 1002126, 1002127, 1002129, 1002130, 1002137, 1002138, 1002139, 1002142, 1002143, 1002144, 1002145, 1002146, 1002147, 1002148, 1002149, 1002150, 1002151, 1002153, 1002154, 1002156, 1002158, 1002160, 1002161, 1002162, 1002163, 1002164, 1002165, 1002167, 1002169, 1002170, 1002171, 1002172, 1002174, 1002175, 1002176, 1002177, 1002179, 1002180, 1002182, 1002183, 1002184, 1002185, 1002188, 1002190, 1002191, 1002192, 1002194, 1002195, 1002196, 1002198, 1002199, 1002200, 1002203, 1002204, 1002207, 1002209, 1002210, 1002211, 1002213, 1002214, 1002216, 1002217, 1002218, 1002219, 1002221, 1002222, 1002225, 1002227, 1002228, 1002230, 1002231, 1002234, 1002235, 1002237, 1002238, 1002239, 1002240, 1002241, 1002243, 1002244, 1002248, 1002250, 1002253, 1002254, 1002256, 1002257, 1002258, 1002261, 1002262, 1002263, 1002264, 1002266, 1002269, 1002271, 1002273, 1002274, 1002278, 1002279, 1002280, 1002283, 1002285, 1002287, 1002292, 1002293, 1002295, 1002300, 1002301, 1002302, 1002305, 1002306, 1002308, 1002310, 1002311, 1002312, 1002313, 1002314, 1002315, 1002316, 1002320, 1002321, 1002322, 1002323, 1002324, 1002329, 1002333, 1002334, 1002335, 1002338, 1002342, 1002343, 1002349, 1002352, 1002357, 1002360, 1002361, 1002366, 1002369, 1002371, 1002373, 1002376, 1002380, 1002381, 1002382, 1002383, 1002385, 1002386, 1002393, 1002395, 1002396, 1002397, 1002399, 1002402, 1002403, 1002405, 1002456, 1002457, 1002459, 1002460, 1002461, 1002462, 1002464, 1002465, 1002467, 1002474, 1002475, 1002477, 1002478, 1002480, 1002482, 1002484, 1002489, 1002492, 1002493, 1002495, 1002496, 1002499, 1002500, 1002502, 1002506, 1002508, 1002509, 1002512, 1002513, 1002516, 1002519, 1002521, 1002525, 1002526, 1002527, 1002532, 1002533, 1002536, 1002538, 1002542, 1002545, 1002548, 1002549, 1002554, 1002555, 1002557, 1002558, 1002560, 1002561, 1002562, 1002563, 1002564, 1002565, 1002566, 1002568, 1002569, 1002572, 1002577, 1002578, 1002580, 1002582, 1002586, 1002587, 1002590, 1002592, 1002598, 1002610, 1002613, 1002616, 1002618, 1002619, 1002624, 1002625, 1002626, 1002629, 1002631, 1002632, 1002633, 1002636, 1002643, 1002644, 1002645, 1002648, 1002649, 1002650, 1002651, 1002653, 1002655, 1002656, 1002660, 1002661, 1002663, 1002665, 1002666, 1002668, 1002670, 1002671, 1002672, 1002673, 1002678, 1002679, 1002680, 1002681, 1002685, 1002692, 1002694, 1002696, 1002698, 1002700, 1002703, 1002704, 1002707, 1002709, 1002711, 1002713, 1002714, 1002720, 1002722, 1002727, 1002728, 1002730, 1002732, 1002735, 1002736, 1002737, 1002738, 1002741, 1002743, 1002744, 1002745, 1002747, 1002749, 1002751, 1002752, 1002754, 1002756, 1002757, 1002759, 1002760, 1002761, 1002765, 1002767, 1002769, 1002771, 1002792, 1002796, 1002797, 1002809, 1002820, 1002825, 1002827, 1002829, 1002830, 1002839, 1002840, 1002846, 1002847, 1002850, 1002852, 1002857, 1002860, 1002862, 1002863, 1002865, 1002866, 1002869, 1002870, 1002871, 1002873, 1002874, 1002878, 1002882, 1002885, 1002886, 1002887, 1002888, 1002890, 1002891, 1002893, 1002895, 1002898, 1002900, 1002901, 1002902, 1002903, 1002905, 1002908, 1002910, 1002912, 1002917, 1002919, 1002923, 1002927, 1002932, 1002935, 1002941, 1002942, 1002943, 1002946, 1002948, 1002962, 1002964, 1002966, 1002968, 1002969, 1002970, 1002978, 1002990, 1002991, 1002994, 1003005, 1003011, 1003017, 1003018, 1003019, 1003021, 1003022, 1003024, 1003025, 1003026, 1003027, 1003029, 1003032, 1003034, 1003037, 1003038, 1003039, 1003040, 1003041, 1003042, 1003043, 1003045, 1003048, 1003050, 1003051, 1003052, 1003053, 1003054, 1003055, 1003056, 1003059, 1003066, 1003071, 1003076, 1003087, 1003088, 1003099, 1003100, 1003101, 1003103, 1003104, 1003111, 1073912, 1003114, 1003115, 1003116, 1003121, 1003122, 1003123, 1003124, 1003125, 1003126, 1003127, 1003129, 1003132, 1003134, 1003138, 1003139, 1003151, 1003154, 1003164, 1003165, 1003167, 1003168, 1003169, 1003171, 1003173, 1003176, 1003177, 1003179, 1003180, 1003181, 1003182, 1003184, 1003186, 1003189, 1003193, 1003194, 1003195, 1003199, 1003202, 1003206, 1003208, 1003209, 1003210, 1003211, 1003212, 1003213, 1003215, 1003216, 1003218, 1003219, 1003222, 1003223, 1003226, 1003227, 1003229, 1003231, 1003232, 1003241, 1003243, 1003244, 1003246, 1003248, 1003249, 1003251, 1003252, 1003253, 1003254, 1003256, 1003257, 1003260, 1003262, 1003263, 1003264, 1003266, 1003272, 1003273, 1003275, 1003277, 1003279, 1003280, 1003281, 1003282, 1003284, 1003287, 1003289, 1003290, 1003291, 1003292, 1003293, 1003295, 1003296, 1003299, 1003300, 1003303, 1003306, 1003307, 1003311, 1003312, 1003313, 1003318, 1003319, 1003321, 1003323, 1003328, 1003329, 1003330, 1003332, 1003334, 1003335, 1003336, 1003337, 1003338, 1003340, 1003342, 1003346, 1003347, 1003349, 1003351, 1003352, 1003353, 1003354, 1003355, 1003356, 1003357, 1003359, 1003360, 1003361, 1003362, 1003363, 1003364, 1003365, 1003367, 1003368, 1003370, 1003372, 1003374, 1003376, 1003379, 1003381, 1003383, 1003385, 1003386, 1003387, 1003388, 1003391, 1003392, 1003393, 1003394, 1003395, 1003396, 1003398, 1003399, 1003403, 1003404, 1003405, 1003406, 1003407, 1003408, 1003409, 1003410, 1003414, 1003415, 1003420, 1003422, 1003423, 1003424, 1003425, 1003427, 1003428, 1003430, 1003431, 1003432, 1003433, 1003434, 1003438, 1003439, 1003440, 1003441, 1003442, 1003444, 1003447, 1003448, 1003455, 1003458, 1003460, 1003464, 1003467, 1003468, 1003469, 1003470, 1003471, 1003473, 1003474, 1003475, 1003476, 1003479, 1003480, 1003481, 1003482, 1003484, 1003485, 1003487, 1003488, 1003491, 1003492, 1003493, 1003494, 1003495, 1003496, 1003497, 1003498, 1003499, 1003500, 1003501, 1003506, 1003507, 1003509, 1003511, 1003512, 1003514, 1003515, 1003518, 1003522, 1003523, 1003527, 1003528, 1003529, 1003530, 1003531, 1003533, 1003537, 1003539, 1003541, 1003543, 1003544, 1003549, 1003551, 1003552, 1003553, 1003556, 1003557, 1003558, 1003559, 1003560, 1003563, 1003564, 1003565, 1003566, 1003567, 1003568, 1003569, 1003570, 1003574, 1003575, 1003577, 1003579, 1003580, 1003581, 1003582, 1003583, 1003584, 1003586, 1003587, 1003588, 1003590, 1003591, 1003594, 1003596, 1003598, 1003599, 1003600, 1003601, 1003603, 1003605, 1003606, 1003607, 1003608, 1003609, 1003612, 1003613, 1003615, 1003616, 1003625, 1003630, 1003633, 1003636, 1003639, 1003658, 1003666, 1003669, 1003697, 1003698, 1003699, 1003700, 1003701, 1003703, 1003704, 1003705, 1003708, 1003710, 1003711, 1003712, 1003713, 1003714, 1003721, 1003724, 1003725, 1003733, 1003735, 1003736, 1003738, 1003739, 1003743, 1003744, 1003745, 1003748, 1003751, 1003753, 1003755, 1003756, 1003757, 1003766, 1003767, 1003780, 1003784, 1003789, 1003803, 1003815, 1003818, 1003821, 1003823, 1003824, 1003827, 1003829, 1003830, 1003831, 1003832, 1003833, 1003834, 1003835, 1003838, 1003840, 1003842, 1003843, 1003844, 1003846, 1003847, 1003848, 1003849, 1003850, 1003852, 1003854, 1003856, 1003858, 1003859, 1003860, 1003861, 1003862, 1003864, 1003866, 1003867, 1003868, 1003869, 1003870, 1003871, 1003872, 1003876, 1003879, 1003882, 1003887, 1003888, 1003889, 1003892, 1003894, 1003897, 1003898, 1003901, 1003904, 1003908, 1003913, 1003915, 1003917, 1003918, 1003919, 1003920, 1003925, 1003926, 1003928, 1003929, 1003930, 1003935, 1003937, 1003938, 1003941, 1003942, 1003943, 1003944, 1003945, 1003946, 1003947, 1003948, 1003950, 1003953, 1003954, 1003957, 1003958, 1003959, 1003960, 1003962, 1003966, 1003968, 1003969, 1003970, 1003971, 1003972, 1003973, 1003975, 1003976, 1003977, 1003978, 1003979, 1003980, 1003981, 1003982, 1003983, 1003984, 1003985, 1003986, 1003988, 1003990, 1003991, 1003992, 1003993, 1003994, 1003995, 1003998, 1003999, 1004002, 1004005, 1004006, 1004008, 1004009, 1004010, 1004011, 1004012, 1004055, 1004014, 1004015, 1004019, 1004020, 1004021, 1004024, 1004025, 1004027, 1004028, 1004029, 1004030, 1004031, 1004032, 1004033, 1004034, 1004035, 1004036, 1008312, 1004038, 1008314, 1008316, 1004041, 1004042, 1004043, 1008317, 1004045, 1008319, 1004047, 1008321, 1004049, 1004050, 1004051, 1004053, 1008322, 1008452, 1004060, 1004061, 1004062, 1004063, 1004064, 1004065, 1004073, 1004074, 1004075, 1004076, 1004077, 1004079, 1004081, 1004082, 1004084, 1004086, 1004087, 1004088, 1004089, 1004090, 1004091, 1004093, 1004094, 1004095, 1004096, 1004097, 1004098, 1004099, 1004100, 1004102, 1004103, 1004104, 1004105, 1004106, 1004109, 1004113, 1004114, 1004115, 1004116, 1004117, 1004118, 1004119, 1004120, 1004121, 1004122, 1004126, 1004127, 1004129, 1004130, 1004131, 1004132, 1004133, 1004135, 1004136, 1004137, 1004138, 1004139, 1004141, 1004142, 1004144, 1004145, 1004149, 1004150, 1004151, 1004152, 1004154, 1004155, 1004157, 1004158, 1004160, 1004161, 1004162, 1004163, 1004169, 1004171, 1004172, 1004175, 1004177, 1004179, 1004180, 1004182, 1004183, 1004184, 1004185, 1004187, 1004188, 1004189, 1004190, 1004191, 1004192, 1004194, 1004196, 1004197, 1004198, 1004199, 1004200, 1004201, 1004204, 1004205, 1004206, 1004207, 1004208, 1004212, 1004213, 1004214, 1004215, 1004216, 1004217, 1004218, 1004219, 1004221, 1004224, 1004225, 1004226, 1004227, 1004228, 1004231, 1004232, 1004234, 1004235, 1004236, 1004237, 1004238, 1004239, 1004240, 1004241, 1004242, 1004243, 1004244, 1004245, 1004248, 1004249, 1004250, 1004251, 1004252, 1004253, 1004255, 1004256, 1004257, 1004259, 1004260, 1004261, 1004263, 1004264, 1004265, 1004267, 1004268, 1004270, 1004271, 1004272, 1004273, 1004276, 1004277, 1004278, 1004280, 1004282, 1004283, 1004285, 1004286, 1004287, 1004288, 1004289, 1004290, 1004292, 1004293, 1004294, 1004295, 1004296, 1004297, 1004299, 1004300, 1004301, 1004302, 1004303, 1004304, 1004305, 1004306, 1004307, 1004309, 1004310, 1004311, 1004313, 1004314, 1004315, 1004316, 1004317, 1004318, 1004319, 1004320, 1004321, 1004322, 1004323, 1004324, 1004327, 1004328, 1004329, 1004330, 1004331, 1004332, 1004333, 1004334, 1004335, 1004336, 1004337, 1004338, 1004339, 1004340, 1004341, 1004342, 1004343, 1004344, 1004349, 1004353, 1004355, 1004358, 1004359, 1004360, 1004361, 1004364, 1004365, 1004366, 1004367, 1004369, 1004371, 1004374, 1004375, 1004376, 1004377, 1004379, 1004383, 1004385, 1004387, 1004388, 1004389, 1004390, 1004403, 1004405, 1004407, 1004409, 1004413, 1004423, 1004428, 1004430, 1004434, 1004439, 1004440, 1004441, 1004442, 1004444, 1004446, 1004450, 1004452, 1004454, 1004456, 1004457, 1004458, 1004459, 1004462, 1004466, 1004467, 1004469, 1004472, 1004474, 1004476, 1004477, 1004478, 1004481, 1004484, 1004485, 1004486, 1004487, 1004491, 1004492, 1004496, 1004498, 1004500, 1004501, 1004503, 1004504, 1004507, 1004508, 1004510, 1004512, 1004514, 1004515, 1004516, 1004519, 1004520, 1004523, 1004525, 1004527, 1004528, 1004531, 1004532, 1004533, 1004536, 1004537, 1004539, 1004541, 1004542, 1004545, 1004547, 1004548, 1004549, 1004550, 1004551, 1004552, 1004557, 1004559, 1004562, 1004564, 1004567, 1004568, 1004569, 1004570, 1004572, 1004574, 1004575, 1004586, 1004592, 1004595, 1004598, 1004599, 1004601, 1004602, 1004603, 1004609, 1004611, 1004615, 1004617, 1004618, 1004619, 1004620, 1004630, 1004634, 1004635, 1004638, 1004640, 1004641, 1004643, 1004659, 1004661, 1004662, 1004663, 1004674, 1004678, 1004681, 1004683, 1004685, 1004687, 1004688, 1004691, 1004692, 1004693, 1004695, 1004697, 1004699, 1004700, 1004701, 1004704, 1004708, 1004709, 1004710, 1004712, 1004716, 1004717, 1004720, 1004725, 1004727, 1004729, 1004730, 1004734, 1004735, 1004736, 1004737, 1004741, 1004742, 1004743, 1004746, 1004750, 1004753, 1004755, 1004758, 1004759, 1004761, 1004764, 1004768, 1004770, 1004772, 1004775, 1004779, 1004781, 1004782, 1004788, 1004792, 1004797, 1004798, 1004802, 1004803, 1004809, 1004810, 1004811, 1004816, 1004817, 1004819, 1004821, 1004822, 1004823, 1004825, 1004826, 1004827, 1004830, 1004831, 1004833, 1004834, 1004835, 1004837, 1004841, 1004842, 1004844, 1004846, 1004850, 1004853, 1004856, 1004860, 1004861, 1004862, 1004863, 1004864, 1004865, 1004871, 1004875, 1004876, 1004877, 1004878, 1004879, 1004881, 1004884, 1004885, 1004894, 1004897, 1004898, 1004900, 1004901, 1004902, 1004905, 1004906, 1004908, 1004910, 1004911, 1004912, 1004913, 1004916, 1004917, 1004921, 1014442, 1004923, 1004928, 1004931, 1004932, 1004934, A76680, 691951, 1004939, 1004940, 1004943, 1006430, 1006428, 1004946, 1004947, 1006427, 1004949, 1004950, 1006242, 1004952, 1004953, 1006241, 1004955, 1006240, 1004957, 1006239, 1006238, 1006237, 1006236, 1006235, 1004963, 1006234, 1004965, 1004966, 1004967, 1004968, 1004969, 1006233, 1004971, 1006232, 1004973, 1006231, 1004975, 1006414, 1004977, 1004978, 1004979, 1004980, 1006413, 1006410, 1006408, 1006450, 1006449, 1004986, 1006448, 1006446, 1006445, 1004990, 1006443, 1004992, 1006441, 1006440, 1004995, 1006439, 1006438, 1004998, 1004999, 1005000, 1005001, 1006437, 1006436, 1005004, 1006435, 1006434, 1005008, 1005009, 1005010, 1005013, 1005014, 1005015, 1005022, 1005023, 1005026, 1005029, 1005030, 1005039, 1005054, 1005071, 1005077, 1005079, 1005081, 1005084, 1005086, 1005109, 1005112, 1005114, 1005116, 1005117, 1005123, 1005124, 1014548, 1005136, 1005138, 1005141, 1005150, 1005153, 1005154, 1005159, 1005162, 1005164, 1005165, 1005168, 1005170, 1005171, 1005172, 1005174, 1005176, 1005179, 1005180, 1005199, 1005200, 1005201, 1005202, 1005203, 1005204, 1005205, 1005206, 1005207, 1005208, 1005211, 1005219, 1005228, 1005230, 1005231, 1002115, 003574, 1005241, 1005245, 1006421, 1006326, 1006330, 1006331, 1006332, 1005252, 1006333, 1006334, 1005255, 1006335, 1006336, 1006337, 1005272, 1006342, 1006347, 1006200, 1006201, 1006206, 1006207, 1006208, 1005281, 1005282, 1005283, 1006309, 1005285, 1006311, 1006312, 1005288, 1006313, 1006314, 1005291, 1005292, 1005295, 1005296, 1005297, 1005298, 1005299, 1005302, 1005303, 1005305, 1006315, 1005307, 1005308, 1005309, 1005312, 1005313, 1006316, 1005316, 1005317, 1005319, 1005322, 1005326, 1005328, 1005329, 1005330, 1005331, 1006317, 1005333, 1006318, 1006319, 1005337, 1005338, 1005339, 1006321, 1005342, 1005343, 1005344, 1005345, 1005346, 1005347, 1005349, 1005350, 1005351, 1005353, 1005354, 1005355, 1006310, 1005362, 1005363, 1005367, 1005369, 1005371, 1005373, 1005375, 1005376, 1005378, 1005379, 1005380, 1005381, 1005881, 1005383, 1005912, 1005931, 1005985, 1006015, 1005390, 1005391, 1005392, 1006037, 1006063, 1006083, 1006084, 1006095, 1006099, 1006104, 1006140, 1006176, 1006188, 1005403, 1005404, 1006193, 1005406, 1006194, 1006195, 1006181, 1006182, 1006183, 1006184, 1005413, 1006185, 1005415, 1005416, 1006186, 1006113, 1006126, 1006130, 1006133, 1006134, 1006165, 1006174, 1006179, 1005181, 1005573, 1005666, 1005951, 1005430, 1005971, 1006014, 1006017, 1005434, 1005435, 1006020, 1006024, 1005438, 1005877, 1006007, 1005590, 1005605, 1005443, 1005783, 1005902, 1005904, 1005905, 1005449, 1005920, 1005939, 1005941, 1005948, 1005456, 1005950, 1005458, 1005952, 1005460, 1005461, 1005957, 1005958, 1005960, 1005961, 1005962, 1005968, 1005973, 1005984, 1005470, 1005988, 1005991, 1005473, 1005474, 1006001, 1006004, 1006006, 1006003, 1005155, 1005332, 1005358, 1005370, 1005372, 1005395, 1005405, 1005521, 1005670, 1005828, 1005489, 1005865, 1005875, 1005876, 1005894, 1005496, 1005898, 1005900, 1005499, 1005500, 1004852, 1005502, 1005504, 1005427, 1005497, 1005509, 1005512, 1005515, 1005511, 1005517, 1005522, 1005548, 1005808, 1005809, 1005810, 1005811, 1005812, 1005813, 1005814, 1005815, 1005816, 1005524, 1005817, 1005526, 1005818, 1005819, 1005820, 1005530, 1005532, 1005533, 1005534, 1005821, 1005536, 1005537, 1005538, 1005540, 1005542, 1005543, 1005544, 1005822, 1005546, 1005823, 1005824, 1005825, 1005826, 1005551, 1005552, 1005827, 1005554, 1005829, 1005830, 1005557, 1005558, 1005831, 1005832, 1005561, 1005836, 1005837, 1005839, 1005565, 1005566, 1005567, 1005840, 1005843, 1005847, 1005571, 1005852, 1005857, 1005860, 1005575, 1005861, 1005862, 1005578, 1005579, 1005580, 1005869, 1005582, 1005583, 1005584, 1005585, 1005586, 1005871, 1005588, 1005589, 1005873, 1005591, 1005592, 1005593, 1005879, 1005882, 1005596, 1005597, 1005598, 1005599, 1005883, 1005884, 1005602, 1005603, 1005604, 1005885, 1005606, 1005608, 1005886, 1005892, 1005896, 1005612, 1005613, 1005614, 1005615, 1005897, 1005846, 1005853, 1005621, 1005622, 1005623, 1005624, 1005493, 1005626, 1005716, 1005628, 1005718, 1005630, 1005631, 1005727, 1005633, 1005634, 1005635, 1005640, 1005732, 1005648, 1005743, 1005650, 1005764, 1005765, 1005767, 1005768, 1005656, 1005657, 1005658, 1005769, 1005770, 1005771, 1005773, 1005774, 1005775, 1005776, 1005777, 1005667, 1005778, 1005779, 1005781, 1005782, 1005784, 1005785, 1005674, 1005787, 1005676, 1005788, 1005789, 1005679, 1005680, 1005790, 1005682, 1005791, 1005685, 1005792, 1005687, 1005688, 1005689, 1005793, 1005691, 1005692, 1005794, 1005795, 1005796, 1005798, 1005697, 1005799, 1005800, 1005802, 1005803, 1005702, 1005804, 1005806, 1005807, 1005706, 1005374, 1005377, 1005382, 1005710, 1005711, 1005384, 1005713, 1005714, 1005386, 1005389, 1005393, 1005394, 1005396, 1005397, 1005398, 1005399, 1005400, 1005401, 1005725, 1005407, 1005121, 1005728, 1005143, 1005166, 1005178, 1005182, 1005733, 1005191, 1005210, 1005736, 1005217, 1005336, 1005739, 1005340, 1005741, 1005356, 1005357, 1005359, 1005360, 1005746, 1005361, 1005388, 1005402, 1005433, 1005452, 1005453, 1005464, 1005465, 1005475, 1005756, 1005486, 1005758, 1005545, 1005547, 1005549, 1005550, 1005553, 1005555, 1005556, 1005559, 1005560, 1005562, 1005564, 1005568, 1005772, 1005569, 1005570, 1005572, 1005574, 1005576, 1005577, 1005581, 1005780, 1005587, 1005594, 1005595, 1005600, 1005601, 1005786, 1005609, 1005610, 1005611, 1005616, 1005617, 1005619, 1005620, 1005625, 1005627, 1005629, 1005797, 1005632, 1005708, 1005709, 1005801, 1005712, 1005715, 1005717, 1005805, 1005719, 1005720, 1005721, 1005722, 1005723, 1005724, 1005726, 1005729, 1005730, 1005731, 1005734, 1005735, 1005737, 1005738, 1005740, 1005742, 1005744, 1005745, 1005747, 1005748, 1005749, 1005750, 1005751, 1005752, 1005753, 1005754, 1005755, 1005833, 1005834, 1005835, 1005757, 1005759, 1005838, 1005761, 1005762, 1005841, 1005842, 1005763, 1005844, 1005845, 1005766, 1005563, 1005848, 1005849, 1005850, 1005851, 1005535, 1005115, 1005854, 1005855, 1005856, 1005120, 1005858, 1005859, 1005128, 1005531, 1005276, 1005863, 1005864, 1005287, 1005866, 1005867, 1005868, 1005408, 1005870, 1005409, 1005872, 1005410, 1005411, 1005412, 1005878, 1005414, 1005880, 1005417, 1005418, 1005419, 1005420, 1005421, 1005422, 1005887, 1005888, 1005889, 1005891, 1005423, 1005893, 1005424, 1005895, 1005425, 1005426, 1005428, 1005899, 1005429, 1005901, 1005431, 1005903, 1005432, 1005436, 1005906, 1005907, 1005437, 1005918, 1005439, 1005922, 1005924, 1005929, 1005930, 1005440, 1005932, 1005933, 1005934, 1005935, 1005936, 1005937, 1005938, 1005441, 1005940, 1005442, 1005942, 1005943, 1005945, 1005946, 1005947, 1005444, 1005949, 1005446, 1005447, 1005448, 1005953, 1005954, 1005955, 1005956, 1005450, 1005454, 1005457, 1005459, 1005462, 1005963, 1005964, 1005965, 1005966, 1005967, 1005463, 1005969, 1005970, 1005466, 1005972, 1005467, 1005974, 1005975, 1005976, 1005977, 1005978, 1005979, 1005980, 1005981, 1005982, 1005983, 1005468, 1005469, 1005986, 1005987, 1005471, 1005989, 1005990, 1005472, 1005992, 1005993, 1005994, 1005995, 1005996, 1005997, 1005998, 1005999, 1006000, 1005476, 1006002, 1005477, 1005478, 1006008, 1006011, 1006012, 1005479, 1005480, 1006016, 1005481, 1006018, 1006019, 1005482, 1006021, 1006022, 1006023, 1005483, 1006026, 1006029, 1006034, 1005484, 1006039, 1006041, 1006044, 1006047, 1006054, 1005485, 1006065, 1006066, 1006067, 1006068, 1006069, 1006070, 1006071, 1006072, 1006073, 1006074, 1006075, 1006076, 1006077, 1006078, 1006079, 1006080, 1006081, 1006082, 1005487, 1005488, 1006085, 1006086, 1006087, 1006088, 1006089, 1006090, 1006091, 1006092, 1006093, 1006094, 1005491, 1006096, 1006097, 1006098, 1005492, 1006100, 1006101, 1006102, 1006103, 1005494, 1006105, 1006106, 1006107, 1006108, 1006109, 1006110, 1006111, 1006112, 1005495, 1006114, 1006115, 1006116, 1006117, 1006118, 1006119, 1006120, 1006121, 1006122, 1006123, 1006124, 1006125, 1005498, 1006127, 1006128, 1006129, 1005501, 1006131, 1006132, 1005503, 1005505, 1006135, 1006136, 1006137, 1006138, 1006139, 1005506, 1006141, 1006142, 1006143, 1006144, 1006145, 1006146, 1006147, 1006148, 1006149, 1006150, 1006151, 1006152, 1006155, 1006156, 1006157, 1006158, 1006159, 1006160, 1006161, 1006162, 1006163, 1006164, 1005507, 1006166, 1006168, 1006169, 1006170, 1006171, 1006172, 1006173, 1005508, 1006175, 1005510, 1006177, 1006178, 1005513, 1006180, 1006187, 1005514, 1006189, 1006190, 1006191, 1006192, 1005516, 1005518, 1005519, 1006196, 1006197, 1006198, 1006199, 1005520, 1005523, 1006202, 1006203, 1006204, 1006205, 1005525, 1005527, 1005528, 1006210, 1006211, 1006212, 1006213, 1006214, 1006215, 1006216, 1006217, 1006218, 1006219, 1006220, 1006221, 1006222, 1006223, 1006224, 1006225, 1006226, 1006227, 1006228, 1006229, 1005529, 1005668, 1005669, 1005671, 1005672, 1005673, 1005675, 1005677, 1005678, 1005681, 1005683, 1005686, 1006243, 1006244, 1006245, 1006246, 1006247, 1006248, 1006249, 1006250, 1006251, 1006252, 1006253, 1006254, 1006256, 1006257, 1006258, 1006259, 1006260, 1006261, 1006262, 1006263, 1006264, 1006265, 1006266, 1006267, 1006268, 1006269, 1006270, 1006271, 1006272, 1006273, 1006274, 1006275, 1006276, 1006277, 1006279, 1006280, 1006281, 1006282, 1006284, 1006287, 1006288, 1006289, 1006290, 1006291, 1006292, 1006293, 1006294, 1006295, 1006296, 1006297, 1006300, 1006301, 1006302, 1006303, 1006304, 1006305, 1006306, 1006307, 1006308, 1005690, 1005693, 1005694, 1005695, 1005696, 1005698, 1005699, 1005700, 1005701, 1005703, 1006320, 1005704, 1006322, 1006323, 1006325, 1005705, 1006328, 1006329, 1005707, 1005647, 1005649, 1005651, 1005652, 1005653, 1005655, 1005659, 1006338, 1006339, 1006340, 1006341, 1005660, 1006343, 1006344, 1006345, 1006346, 1005661, 1006348, 1006349, 1006350, 1006351, 1006353, 1005662, 1006363, 1005663, 1005664, 1005665, 1005160, 1005161, 1005167, 1005169, 1006383, 1005173, 1005158, 1006390, 1006397, 1005006, 1006399, 1006400, 1006401, 1006402, 1006403, 1006404, 1006405, 1006406, 1006407, 1005028, 1006409, 1005106, 1006411, 1006412, 1005107, 1005108, 1006415, 1006416, 1006417, 1006418, 1006419, 1006420, 1005110, 1006423, 1006424, 1006425, 1006426, 1005111, 1005113, 1006429, 1006431, 1006432, 1005118, 1005119, 1005122, 1005125, 1005126, 1005127, 1005129, 1005130, 1006442, 1005131, 1006444, 1005132, 1005133, 1006447, 1005134, 1005135, 1006451, 1006452, 1006453, 1006454, 1006455, 1006456, 1006457, 1006460, 1006461, 1006462, 1006463, 1006464, 1006465, 1006466, 1006469, 1006470, 1006471, 1006472, 1006473, 1006475, 1006476, 1006478, 1006479, 1006480, 1006481, 1006482, 1006483, 1006484, 1006485, 1006486, 1006487, 1006488, 1006489, 1006490, 1006491, 1006492, 1006493, 1006494, 1006495, 1006496, 1006497, 1006498, 1006499, 1006500, 1006501, 1006502, 1006503, 1006504, 1006505, 1006506, 1006507, 1006508, 1006509, 1006510, 1006511, 1006512, 1006513, 1006514, 1006515, 1006516, 1006517, 1006518, 1006519, 1006520, 1006521, 1006522, 1006523, 1006524, 1006525, 1006526, 1006527, 1006528, 1006529, 1006530, 1006531, 1006532, 1006533, 1006534, 1006535, 1006536, 1006537, 1006538, 1006539, 1006541, 1006542, 1006544, 1006545, 1006548, 1006549, 1006550, 1006551, 1006552, 1006553, 1006554, 1006555, 1006556, 1006557, 1006558, 1006559, 1006560, 1006561, 1006562, 1006563, 1006564, 1006565, 1006566, 1006567, 1006568, 1006569, 1006570, 1006571, 1006572, 1006573, 1006574, 1006575, 1006576, 1006577, 1006578, 1006579, 1006580, 1006581, 1006583, 1006584, 1006585, 1006586, 1006587, 1006588, 1006589, 1006590, 1006591, 1006592, 1006593, 1006594, 1006595, 1006596, 1006597, 1006598, 1006599, 1006604, 1006606, 1006607, 1006608, 1006609, 1006610, 1006611, 1006612, 1006613, 1006614, 1006615, 1006616, 1006617, 1006618, 1006619, 1006620, 1006621, 1006622, 1006623, 1006624, 1006637, 1006638, 1006639, 1006640, 1006641, 1006642, 1006643, 1006644, 1006645, 1006646, 1006647, 1006648, 1006649, 1006650, 1006651, 1006652, 1006653, 1006654, 1006655, 1006656, 1006657, 1006658, 1006659, 1006660, 1006661, 1006662, 1006663, 1006664, 1006665, 1006666, 1006667, 1006668, 1006670, 1006671, 1006672, 1006673, 1006674, 1006675, 1006676, 1006677, 1006678, 1006679, 1006680, 1006681, 1006682, 1006683, 1006684, 1006685, 1006686, 1006687, 1006688, 1006689, 1006690, 1006691, 1006692, 1006693, 1006694, 1006695, 1006696, 1006697, 1006698, 1006699, 1006700, 1006701, 1006702, 1006703, 1006704, 1006705, 1006706, 1006707, 1006708, 1006711, 1006712, 1006713, 1006714, 1006715, 1006717, 1006718, 1006719, 1006720, 1006721, 1006722, 1006723, 1006724, 1006725, 1006726, 1006727, 1006734, 1006735, 1006736, 1006737, 1006738, 1006739, 1006740, 1006741, 1006743, 1006744, 1006745, 1006746, 1006747, 1006748, 1006749, 1006750, 1006752, 1006754, 1006761, 1006763, 1006764, 1006767, 1006770, 1006771, 1006773, 1006774, 1006775, 1006776, 1006777, 1006778, 1006779, 1006780, 1006782, 1006783, 1006784, 1006785, 1006786, 1006787, 1006789, 1006790, 1006791, 1006792, 1006793, 1006794, 1006795, 1006796, 1006797, 1006798, 1006799, 1006800, 1006801, 1006802, 1006803, 1006805, 1006809, 1006814, 1006816, 1006818, 1006819, 1006835, 1006837, 1006840, 1006841, 1006842, 1006843, 1006844, 1006845, 1006846, 1006847, 1006848, 1006849, 1006850, 1006851, 1006852, 1006853, 1006854, 1006855, 1006861, 1006863, 1006864, 1006865, 1006866, 1006867, 1006868, 1006869, 1006870, 1006871, 1006872, 1006873, 1006874, 1006875, 1006876, 1006877, 1006878, 1006879, 1006880, 1006881, 1006882, 1006883, 1006884, 1006885, 1006886, 1006887, 1006888, 1006889, 1006890, 1006891, 1006892, 1006893, 1006895, 1006900, 1006901, 1006903, 1006904, 1006905, 1006906, 1006907, 1006908, 1006909, 1006910, 1006911, 1006912, 1006913, 1006914, 1006915, 1006916, 1006917, 1006918, 1006919, 1006920, 1006921, 1006922, 1006924, 1006925, 1006926, 1006927, 1006928, 1006929, 1006930, 1006931, 1006932, 1006933, 1006934, 1006935, 1006936, 1006937, 1006938, 1006939, 1006940, 1006941, 1006942, 1006943, 1006944, 1006945, 1006946, 1006947, 1006948, 1006961, 1006963, 1006973, 1006974, 1006975, 1006976, 1006978, 1006979, 1006981, 1006982, 1006983, 1006984, 1006985, 1006986, 1006987, 1006988, 1006989, 1006990, 1006991, 1006992, 1006993, 1006994, 1006995, 1006996, 1006997, 1006998, 1006999, 1007000, 1007001, 1007002, 1007003, 1007004, 1007005, 1007006, 1007007, 1007008, 1007010, 1007011, 1007012, 1007013, 1007014, 1007015, 1007016, 1007017, 1007018, 1007019, 1007020, 1007021, 1007022, 1007023, 1007024, 1007025, 1007026, 1007027, 1007028, 1007030, 1007031, 1007032, 1007034, 1007036, 1007037, 1007038, 1007039, 1007040, 1007041, 1007042, 1007043, 1007044, 1007045, 1007046, 1007047, 1007048, 1007049, 1007050, 1007052, 1007054, 1007055, 1007056, 1007058, 1007061, 1007062, 1007063, 1007064, 1007065, 1007066, 1007067, 1007068, 1007069, 1007070, 1007071, 1007072, 1007073, 1007074, 1007075, 1007076, 1007077, 1007078, 1007079, 1007080, 1007081, 1007083, 1007086, 1007089, 1007090, 1007095, 1007096, 1007097, 1007098, 1007099, 1007102, 1007106, 1007107, 1007108, 1007109, 1007111, 1007112, 1007113, 1007114, 1007115, 1007116, 1007118, 1007119, 1007120, 1007121, 1007124, 1007125, 1007126, 1007127, 1007128, 1007129, 1007130, 1007131, 1007132, 1007135, 1007136, 1007137, 1007138, 1007139, 1007141, 1007144, 1007145, 1007146, 1007147, 1007148, 1007149, 1007150, 1007151, 1007152, 1007153, 1007154, 1007155, 1007157, 1007158, 1007159, 1007160, 1007161, 1007162, 1007163, 1007164, 1007165, 1007166, 1007167, 1007168, 1007169, 1007170, 1007171, 1007172, 1007173, 1007174, 1007175, 1007176, 1007177, 1007178, 1007179, 1007180, 1007181, 1008548, 1007183, 1008549, 1007185, 1007186, 1008550, 1007188, 1007189, 1005137, 1007192, 1007193, 1007194, 1007195, 1007196, 1007197, 1007198, 1007199, 1007200, 1007201, 1007202, 1007204, 1007205, 1007208, 1007213, 1007218, 1007219, 1007220, 1007221, 1007222, 1007223, 1007224, 1007225, 1007226, 1007227, 1007228, 1007229, 1007230, 1007231, 1007232, 1007233, 1007234, 1007235, 1007236, 1007237, 1007238, 1007239, 1007240, 1007241, 1007242, 1007243, 1007244, 1007245, 1007246, 1007247, 1007249, 1007250, 1007251, 1007252, 1007253, 1007254, 1007255, 1007256, 1007257, 1007258, 1007259, 1007260, 1007261, 1007262, 1007263, 1007264, 1007265, 1007266, 1007267, 1007268, 1007269, 1007270, 1007271, 1007272, 1007273, 1007274, 1007275, 1007277, 1007278, 1007279, 1007280, 1007281, 1007282, 1007283, 1007284, 1007285, 1007286, 1007287, 1007288, 1007289, 1007290, 1007291, 1007292, 1007293, 1007294, 1007295, 1007296, 1007297, 1007298, 1007299, 1007300, 1007301, 1007302, 1007303, 1007304, 1007305, 1007306, 1007307, 1007308, 1007309, 1007310, 1007312, 1007313, 1007314, 1007315, 1007316, 1007318, 1007319, 1007320, 1007321, 1007322, 1007323, 1007324, 1007325, 1007326, 1007327, 1007328, 1007329, 1007330, 1007331, 1007332, 1007333, 1007334, 1007335, 1007336, 1007337, 1007338, 1007339, 1007340, 1007341, 1007342, 1007343, 1007344, 1007345, 1007346, 1007347, 1007348, 1007349, 1007350, 1007351, 1007352, 1007353, 1007354, 1007355, 1007356, 1007357, 1007358, 1007359, 1007360, 1007361, 1007362, 1007363, 1007364, 1007365, 1007366, 1007367, 1007368, 1007369, 1007370, 1007371, 1007372, 1007373, 1007374, 1007375, 1007376, 1007377, 1007378, 1007379, 1007380, 1007381, 1007382, 1007383, 1007384, 1007385, 1007386, 1007387, 1007388, 1007389, 1007390, 1007391, 1007392, 1007393, 1007394, 1007395, 1007396, 1007397, 1007398, 1007399, 1007400, 1007401, 1007402, 1007403, 1007404, 1007405, 1007406, 1007407, 1007408, 1007409, 1007410, 1007411, 1007412, 1007413, 1007422, 1007425, 1007429, 1007438, 1007445, 1007447, 1007448, 1007449, 1007450, 1007451, 1007452, 1007453, 1007455, 1007456, 1007458, 1007459, 1007460, 1007461, 1007462, 1007464, 1007466, 1007467, 1007468, 1007469, 1007470, 1008553, 1007472, 1007473, 1007474, 1007475, 1007476, 1007477, 1007478, 1007479, 1007480, 1007481, 1007483, 1007484, 1007485, 1007486, 1007487, 1007489, 1007490, 1007492, 1007493, 1007494, 1007495, 1007496, 1007497, 1007498, 1007499, 1007500, 1007501, 1007502, 1007503, 1007504, 1007505, 1007507, 1007508, 1007510, 1007511, 1007512, 1007513, 1007514, 1007515, 1007516, 1007517, 1007518, 1007519, 1007520, 1007521, 1007522, 1007523, 1007524, 1007526, 1007527, 1007528, 1007529, 1007530, 1007531, 1007532, 1007533, 1007534, 1007535, 1007537, 1007538, 1007539, 1007540, 1007541, 1007542, 1007543, 1007544, 1007547, 1007548, 1007549, 1007550, 1007551, 1007552, 1007553, 1007554, 1007555, 1007556, 1007557, 1007558, 1007559, 1007560, 1007561, 1007562, 1007563, 1007564, 1007565, 1007566, 1007567, 1007568, 1007569, 1007570, 1007571, 1007572, 1007573, 1007574, 1007575, 1007576, 1007577, 1007578, 1007579, 1007580, 1007581, 1007582, 1007583, 1007584, 1007585, 1007586, 1007587, 1007588, 1007589, 1007590, 1007591, 1007594, 1007595, 1007596, 1007597, 1007598, 1007599, 1007600, 1007601, 1007602, 1007604, 1007605, 1007606, 1007607, 1007609, 1007610, 1007611, 1007612, 1007613, 1007614, 1007615, 1007616, 1007617, 1007618, 1007619, 1007620, 1007621, 1007622, 1007623, 1007624, 1007625, 1007626, 1007628, 1007629, 1007630, 1007631, 1007632, 1007633, 1007634, 1007635, 1007636, 1007637, 1007638, 1007639, 1007640, 1007641, 1007642, 1007643, 1007644, 1007645, 1007646, 1007647, 1007648, 1007649, 1007650, 1007651, 1007652, 1007653, 1007654, 1007655, 1007656, 1007657, 1007658, 1007659, 1007660, 1007661, 1007662, 1007663, 1007664, 1007665, 1007666, 1007667, 1007668, 1007669, 1007670, 1007671, 1007672, 1007673, 1007674, 1007675, 1007676, 1007677, 1007678, 1007679, 1007680, 1007681, 1007682, 1007683, 1007684, 1007685, 1007686, 1007687, 1007688, 1007689, 1007690, 1007691, 1007692, 1007693, 1007694, 1007695, 1007696, 1007697, 1007698, 1007699, 1007700, 1007701, 1007702, 1007703, 1007704, 1007705, 1007706, 1007707, 1007708, 1007709, 1007711, 1007712, 1007713, 1007714, 1007715, 1007716, 1007717, 1007718, 1007719, 1007720, 1007721, 1007722, 1007723, 1007724, 1007725, 1007726, 1007727, 1007728, 1007729, 1007730, 1007731, 1007732, 1007733, 1007734, 1007735, 1007736, 1007737, 1007738, 1007739, 1007740, 1007744, 1007746, 1007759, 1007760, 1007761, 1007762, 1007763, 1007764, 1007765, 1007766, 1007767, 1007768, 1007769, 1007770, 1007771, 1007772, 1007773, 1007774, 1007775, 1007776, 1007777, 1007778, 1007779, 1007780, 1007781, 1007782, 1007783, 1007784, 1007785, 1007786, 1007787, 1007788, 1007789, 1007790, 1007791, 1007792, 1007793, 1007794, 1007795, 1007796, 1007797, 1007798, 1007799, 1007800, 1007802, 1007803, 1007804, 1007805, 1007806, 1007807, 1007808, 1007809, 1007810, 1007811, 1007812, 1007813, 1007814, 1007815, 1007817, 1007818, 1007819, 1007820, 1007821, 1007822, 1007823, 1007824, 1007825, 1007826, 1007827, 1007828, 1007829, 1007830, 1007831, 1007832, 1007833, 1007834, 1007835, 1007836, 1007837, 1007838, 1007839, 1007840, 1007841, 1007843, 1007844, 1007845, 1007846, 1007847, 1007848, 1007849, 1007850, 1007851, 1007852, 1007853, 1007854, 1007855, 1007856, 1007857, 1007858, 1007859, 1007861, 1007862, 1007863, 1007864, 1007865, 1007866, 1007867, 1007868, 1007869, 1007870, 1007871, 1007872, 1007873, 1007874, 1007875, 1007876, 1007877, 1007878, 1007879, 1007880, 1007881, 1007882, 1007883, 1007884, 1007885, 1007886, 1007887, 1007888, 1007889, 1007890, 1007891, 1007892, 1007893, 1007894, 1007896, 1007897, 1007898, 1007899, 1007900, 1007901, 1007902, 1007904, 1007905, 1007906, 1007907, 1007908, 1007909, 1007910, 1007911, 1007912, 1007913, 1007914, 1007916, 1007917, 1007918, 1007919, 1007920, 1007921, 1007922, 1007923, 1007924, 1007925, 1007926, 1007927, 1007928, 1007929, 1007930, 1007931, 1007932, 1007933, 1007934, 1007935, 1007936, 1007937, 1007938, 1007939, 1007940, 1007941, 1007942, 1007944, 1007945, 1007946, 1007947, 1007948, 1007949, 1007950, 1007951, 1007952, 1007953, 1007954, 1007955, 1007956, 1007957, 1007958, 1007959, 1007960, 1007961, 1007962, 1007963, 1007964, 1007965, 1007966, 1007967, 1007968, 1007969, 1007970, 1007971, 1007972, 1007976, 1007977, 1007978, 1007979, 1007980, 1007981, 1007982, 1007983, 1007984, 1007985, 1007986, 1007987, 1007988, 1007989, 1007990, 1007991, 1007992, 1007993, 1007994, 1007995, 1007996, 1007997, 1007998, 1007999, 1008000, 1008001, 1008002, 1008003, 1008004, 1008005, 1008006, 1008007, 1008008, 1008009, 1008010, 1008011, 1008012, 1008013, 1008014, 1008015, 1008016, 1008017, 1008018, 1008019, 1008020, 1008021, 1008022, 1008023, 1008024, 1008025, 1008026, 1008027, 1008028, 1008029, 1008030, 1008031, 1008032, 1008554, 1008557, 1008058, 1008059, 1008060, 1008061, 1008062, 1008063, 1008064, 1008065, 1008066, 1008067, 1008068, 1008069, 1008070, 1008071, 1008072, 1008073, 1008074, 1008075, 1008076, 1008077, 1008078, 1008079, 1008080, 1008081, 1008082, 1008083, 1008084, 1008085, 1008086, 1008087, 1008088, 1008089, 1008090, 1008091, 1008092, 1008103, 1008104, 1008105, 1008106, 1008107, 1008108, 1008109, 1008110, 1008111, 1008112, 1008113, 1008114, 1008115, 1008116, 1008117, 1008118, 1008119, 1008120, 1008121, 1008122, 1008123, 1008124, 1008125, 1008126, 1008127, 1008128, 1008129, 1008130, 1008131, 1008132, 1008133, 1008134, 1008135, 1008136, 1008137, 1008138, 1008139, 1008140, 1008141, 1008142, 1008143, 1008144, 1008145, 1008146, 1008147, 1008148, 1008149, 1008150, 1008151, 1008152, 1008153, 1008154, 1008155, 1008156, 1008157, 1008158, 1008159, 1008160, 1008161, 1008162, 1008163, 1008164, 1008165, 1008166, 1008167, 1008168, 1008169, 1008170, 1008171, 1008172, 1008173, 1008174, 1008175, 1008176, 1008177, 1008178, 1008179, 1008180, 1008181, 1008182, 1008183, 1008184, 1008185, 1008187, 1008188, 1008189, 1008191, 1008192, 1008193, 1008194, 1008195, 1008196, 1008197, 1008198, 1008199, 1008200, 1008201, 1008202, 1008203, 1008204, 1008205, 1008206, 1008207, 1008208, 1008209, 1008210, 1008211, 1008212, 1008213, 1008214, 1008215, 1008216, 1008217, 1008218, 1008219, 1008220, 1008221, 1008222, 1008223, 1008224, 1008225, 1008226, 1008227, 1008228, 1008229, 1008231, 1008232, 1008233, 1008234, 1008235, 1008236, 1008237, 1008238, 1008239, 1008240, 1008241, 1008242, 1008243, 1008244, 1008245, 1008247, 1008248, 1008249, 1008250, 1008251, 1008252, 1008253, 1008254, 1008255, 1008256, 1008257, 1008258, 1008259, 1008260, 1008261, 1008262, 1008263, 1008264, 1008265, 1008266, 1008267, 1008269, 1008270, 1008271, 1008272, 1008273, 1008274, 1008275, 1008276, 1008277, 1008278, 1008279, 1008280, 1008281, 1008282, 1008283, 1008284, 1008285, 1008286, 1008287, 1008288, 1008289, 1008290, 1008291, 1008292, 1008293, 1008295, 1008297, 1008298, 1008299, 1008300, 1008301, 1008302, 1008303, 1008304, 1008305, 1008306, 1008308, 1008310, 1008311, 1008313, 1008315.1008318, 1008320, 1008323, 1008325, 1008326, 1008327, 1008328, 1008329, 1008330, 1008331, 1008332, 1008333, 1008334, 1008335, 1008336, 1008337, 1008338, 1008339, 1008340, 1008341, 1008342, 1008343, 1008344, 1008345, 1008348, 1008349, 1008350, 1008351, 1008353, 1008354, 1008355, 1008359, 1008360, 1008361, 1008362, 1008363, 1008364, 1008365, 1008366, 1008368, 1008370, 1008371, 1008372, 1008373, 1008375, 1008376, 1008377, 1008378, 1008379, 1008380, 1008381, 1008382, 1008383, 1008384, 1008386, 1008387, 1008388, 1008389, 1008390, 1008391, 1008392, 1008394, 1008395, 1008396, 1008397, 1008398, 1008399, 1008400, 1008401, 1008402, 1008403, 1008410, 1008411, 1008412, 1008416, 1008417, 1008418, 1008420, 1008421, 1008422, 1008425, 1008426, 1008430, 1008431, 1008432, 1008433, 1008434, 1008435, 1008438, 1008440, 1008444, 1008445, 1008446, 1008447, 1008448, 1008450, 1008451, 1008453, 1008454, 1008455, 1008456, 1005139, 1005140, 1008516, 1008517, 1008518, 1008519, 1008520, 1008521, 1008522, 1008523, 1008524, 1008525, 1008526, 1008527, 1008528, 1008529, 1008530, 1008532, 1008533, 1008534, 1008535, 1008536, 1008537, 1008538, 1008539, 1008540, 1008541, 1008542, 1008543, 1008544, 1008545, 1008546, 1008547, 1008551, 1008552, 1008555, 1008556, 1008558, 1008559, 1008560, 1008561, 1008562, 1008563, 1008564, 1008565, 1008566, 1008567, 1008568, 1008569, 1008570, 1008571, 1008572, 1008573, 1008574, 1008575, 1008576, 1008577, 1008578, 1008579, 1008580, 1008581, 1008582, 1008583, 1008584, 1008585, 1008586, 1008587, 1008588, 1008589, 1008590, 1008591, 1008592, 1008593, 1008594, 1008595, 1008596, 1008598, 1008599, 1008600, 1008601, 1008602, 1008603, 1008604, 1008605, 1008606, 1008607, 1008608, 1008609, 1008610, 1008611, 1008612, 1008613, 1008614, 1008615, 1008616, 1008617, 1008618, 1008619, 1008620, 1008621, 1008622, 1008623, 1008624, 1008625, 1008626, 1008627, 1008628, 1008629, 1008630, 1008631, 1008632, 1008633, 1008634, 1008635, 1008636, 1008637, 1008638, 1008639, 1008640, 1008641, 1008642, 1008643, 1008644, 1008645, 1008646, 1008647, 1008648, 1008649, 1008650, 1008651, 1008652, 1008653, 1008654, 1008655, 1008656, 1008657, 1008658, 1008659, 1008660, 1008661, 1008662, 1008663, 1008664, 1008665, 1008666, 1008667, 1008668, 1008669, 1008670, 1008672, 1008673, 1008674, 1008675, 1008676, 1008677, 1008678, 1008679, 1008680, 1008681, 1008682, 1008683, 1008684, 1008685, 1008686, 1008687, 1008688, 1008690, 1008691, 1008692, 1008694, 1008696, 1008697, 1008698, 1008699, 1008700, 1008701, 1008702, 1008703, 1008704, 1008705, 1008706, 1008707, 1008708, 1008709, 1008710, 1008712, 1008713, 1008714, 1008716, 1008717, 1008718, 1008719, 1008720, 1008721, 1008722, 1008723, 1008724, 1008725, 1008726, 1008727, 1008728, 1008730, 1008731, 1008733, 1008734, 1008735, 1008736, 1008737, 1008738, 1008739, 1008740, 1008742, 1008744, 1008745, 1008746, 1008747, 1008748, 1008749, 1008750, 1008751, 1008752, 1008753, 1008755, 1008757, 1008758, 1008759, 1008760, 1008761, 1008762, 1008763, 1008764, 1008765, 1008766, 1008767, 1008768, 1008769, 1008770, 1008771, 1008773, 1008774, 1008775, 1008776, 1008777, 1008778, 1008779, 1008780, 1008781, 1008782, 1008784, 1008785, 1008786, 1008787, 1008788, 1008789, 1008790, 1008791, 1008792, 1008793, 1008794, 1008795, 1008796, 1008797, 1008798, 1008799, 1008801, 1008802, 1008804, 1008806, 1008807, 1008808, 1008809, 1008811, 1008812, 1008813, 1008814, 1008815, 1008816, 1008817, 1008818, 1008819, 1008820, 1008821, 1008822, 1008823, 1008824, 1008825, 1008826, 1008827, 1008828, 1008829, 1008830, 1008831, 1008832, 1008833, 1008834, 1008835, 1008837, 1008838, 1008839, 1008840, 1008841, 1008842, 1008843, 1008844, 1008845, 1008846, 1008847, 1008848, 1008849, 1008850, 1008851, 1008852, 1008853, 1008854, 1008855, 1008857, 1008858, 1008859, 1008860, 1008861, 1008863, 1008864, 1008865, 1008866, 1008867, 1008868, 1008870, 1008872, 1008873, 1008874, 1008875, 1008876, 1008877, 1008878, 1008879, 1008880, 1008881, 1008882, 1008883, 1008884, 1008886, 1008887, 1008888, 1008889, 1008890, 1008891, 1008892, 1008893, 1008894, 1008895, 1008897, 1008898, 1008899, 1008900, 1008901, 1008902, 1008903, 1008904, 1008905, 1008906, 1008907, 1008908, 1008909, 1008910, 1008911, 1008912, 1008913, 1008914, 1008915, 1008917, 1008918, 1008919, 1008920, 1008921, 1008922, 1008923, 1008924, 1008925, 1008926, 1008927, 1008928, 1008929, 1008930, 1008931, 1008932, 1008933, 1008934, 1008935, 1008936, 1008937, 1008938, 1008940, 1008941, 1008942, 1008943, 1008944, 1008946, 1008947, 1008948, 1008949, 1008950, 1008956, 1008957, 1008958, 1008960, 1008961, 1008962, 1008963, 1008964, 1008965, 1008966, 1008968, 1008969, 1008970, 1008971, 1008972, 1008973, 1008974, 1008975, 1008976, 1008979, 1008980, 1008981, 1008982, 1008984, 1008986, 1008987, 1008988, 1008989, 1008991, 1008993, 1008994, 1008996, 1008997, 1008998, 1008999, 1009000, 1009002, 1009003, 1009005, 1009006, 1009007, 1009008, 1009010, 1009011, 1009012, 1009013, 1009014, 1009015, 1009016, 1009017, 1009020, 1009021, 1009022, 1009024, 1009025, 1009026, 1009028, 1009029, 1009030, 1009031, 1009032, 1009033, 1009034, 1009035, 1009036, 1009037, 1009038, 1009039, 1009040, 1009041, 1009042, 1009043, 1009044, 1009045, 1009046, 1009047, 1009048, 1009049, 1009050, 1009051, 1009052, 1009053, 1009054, 1009055, 1009056, 1009057, 1009058, 1009059, 1009060, 1009061, 1009062, 1009063, 1009064, 1009065, 1009066, 1009067, 1009068, 1009069, 1009070, 1009071, 1009072, 1009073, 1009074, 1009075, 1009076, 1009077, 1009078, 1009079, 1009080, 1009081, 1009083, 1009084, 1009086, 1009087, 1009089, 1009091, 1009092, 1009094, 1009096, 1009097, 1009099, 1009102, 1009105, 1009106, 1009107, 1009108, 1009110, 1009111, 1009112, 1009113, 1009114, 1009117, 1009118, 1009119, 1009120, 1009122, 1009123, 1009124, 1009125, 1009126, 1009127, 1009128, 1009129, 1009130, 1009131, 1009132, 1009133, 1009134, 1009135, 1009136, 1009137, 1009138, 1009139, 1009140, 1009141, 1009143, 1009144, 1009145, 1009146, 1009147, 1009148, 1009149, 1009150, 1009173, 1009174, 1009176, 1009178, 1009179, 1009182, 1009184, 1009185, 1009186, 1009187, 1009188, 1009189, 1009190, 1009191, 1009192, 1009193, 1009194, 1009195, 1009196, 1009197, 1009198, 1009199, 1009200, 1009201, 1009202, 1009203, 1009204, 1009205, 1009206, 1009207, 1009208, 1009209, 1009210, 1009211, 1009212, 1009213, 1009214, 1009215, 1009216, 1009217, 1009218, 1009219, 1009220, 1009221, 1009222, 1009223, 1009224, 1009225, 1009226, 1009227, 1009228, 1009229, 1009230, 1009231, 1009232, 1009233, 1009234, 1009235, 1009236, 1009237, 1009239, 1009240, 1009241, 1009242, 1009251, 1009252, 1009253, 1009254, 1009255, 1009256, 1009257, 1009258, 1009259, 1009260, 1009261, 1009262, 1009263, 1009264, 1009265, 1009266, 1009267, 1009268, 1009269, 1009270, 1009271, 1009272, 1009273, 1009274, 1009275, 1009276, 1009277, 1009278, 1009279, 1009280, 1009281, 1009282, 1009283, 1009284, 1009285, 1009286, 1009289, 1009290, 1009291, 1009292, 1009293, 1009295, 1009296, 1009298, 1009299, 1009300, 1009302, 1005142, 1009309, 1009310, 1009311, 1009312, 1009313, 1009314, 1009315, 1009316, 1009317, 1009318, 1009319, 1009320, 1009321, 1009323, 1009324, 1009325, 1009326, 1009327, 1009329, 1009330, 1009331, 1009332, 1009333, 1009334, 1009335, 1009336, 1009337, 1009338, 1009339, 1009341, 1009342, 1009343, 1009344, 1009345, 1009346, 1009347, 1009348, 1009349, 1009350, 1009351, 1009352, 1009353, 1009354, 1009355, 1009356, 1009357, 1009358, 1009359, 1009360, 1009361, 1009362, 1009363, 1009364, 1009365, 1009366, 1009367, 1009370, 1009371, 1009372, 1009373, 1009374, 1009377, 1009379, 1009380, 1009381, 1009382, 1009383, 1009384, 1009385, 1009386, 1009387, 1009388, 1009389, 1009390, 1009391, 1009392, 1009393, 1009394, 1009395, 1009396, 1009397, 1009399, 1005144, 1009432, 1009469, 1005145, 1009517, 1005146, 1009534, 1005147, 1005148, 1005151, 1005152, 1005156, 1005157, 1009652, 1009653, 1009654, 1009655, 1009656, 1009657, 1009658, 1009659, 1009661, 1009662, 1009663, 1009664, 1009665, 1009666, 1009667, 1009668, 1009669, 1009670, 1009671, 1009672, 1009673, 1009674, 1009675, 1009676, 1009677, 1009678, 1009679, 1009680, 1009681, 1009682, 1009683, 1009684, 1009686, 1009687, 1009688, 1009689, 1009691, 1009692, 1009693, 1009694, 1009695, 1009696, 1009697, 1009698, 1009699, 1009700, 1009701, 1009702, 1009703, 1009704, 1009705, 1009706, 1009708, 1009709, 1009710, 1009712, 1009713, 1009716, 1009717, 1009718, 1009719, 1009721, 1009722, 1009723, 1009725, 1009726, 1009727, 1009728, 1009730, 1009731, 1009732, 1009734, 1009736, 1009737, 1009738, 1009739, 1009740, 1009741, 1009742, 1009743, 1009744, 1009745, 1009746, 1009747, 1009748, 1009749, 1009750, 1005223, 1005224, 1009842, 1009843, 1009844, 1009845, 1009846, 1009847, 1009848, 1009849, 1009850, 1009851, 1009852, 1009853, 1009854, 1009855, 1009856, 1009857, 1009858, 1009859, 1009860, 1009861, 1009862, 1009863, 1009864, 1009865, 1009866, 1009867, 1009868, 1009869, 1009871, 1009872, 1009873, 1009874, 1009875, 1009876, 1009877, 1009878, 1009879, 1009880, 1009881, 1009882, 1009883, 1009884, 1009885, 1009886, 1005225, 1009888, 1009889, 1009890, 1009891, 1009893, 1009895, 1009896, 1009897, 1009898, 1009899, 1009900, 1009901, 1009902, 1009903, 1009904, 1009905, 1009906, 1009907, 1009908, 1009910, 1009911, 1009912, 1009913, 1009915, 1009916, 1009917, 1009918, 1009919, 1009920, 1009921, 1009922, 1009923, 1009924, 1009925, 1009926, 1009927, 1009928, 1009929, 1009930, 1009931, 1009933, 1009935, 1009936, 1009937, 1009938, 1009939, 1009940, 1009941, 1009943, 1009944, 1009945, 1009946, 1009947, 1009948, 1009949, 1009950, 1009951, 1009952, 1009953, 1009954, 1009955, 1009956, 1009957, 1009958, 1009959, 1009960, 1009961, 1009962, 1009963, 1009965, 1009966, 1009967, 1009968, 1009969, 1009971, 1009973, 1009974, 1009975, 1009977, 1009978, 1009980, 1009981, 1009982, 1009983, 1009984, 1009985, 1009986, 1009987, 1009988, 1009989, 1009990, 1009991, 1009992, 1009993, 1009994, 1009995, 1009996, 1009997, 1009998, 1009999, 1010000, 1010001, 1010002, 1010003, 1010004, 1010005, 1010006, 1010007, 1010009, 1010010, 1010011, 1010012, 1010013, 1010014, 1010015, 1010016, 1010017, 1010018, 1010019, 1010020, 1010021, 1010022, 1010023, 1010024, 1010025, 1010026, 1010027, 1010028, 1010029, 1010030, 1010031, 1010032, 1010033, 1010034, 1010035, 1010036, 1010037, 1010038, 1010039, 1010041, 1010042, 1010043, 1010044, 1010045, 1010046, 1010047, 1010048, 1010049, 1010050, 1010051, 1010052, 1010053, 1010054, 1010055, 1010056, 1010057, 1010058, 1010059, 1010060, 1010061, 1010062, 1010063, 1010064, 1010065, 1010066, 1010068, 1010069, 1010070, 1010071, 1010072, 1010073, 1010074, 1010075, 1010076, 1010077, 1010078, 1010079, 1010080, 1010081, 1010082, 1010083, 1010084, 1010085, 1010086, 1010087, 1010088, 1010089, 1010090, 1010091, 1010092, 1010093, 1010094, 1010095, 1010096, 1010098, 1010099, 1010100, 1010102, 1010103, 1005226, 1010105, 1010106, 1010107, 1010108, 1010109, 1010110, 1010111, 1010113, 1010114, 1010116, 1010117, 1010118, 1010119, 1010120, 1010121, 1010122, 1010123, 1010124, 1010125, 1010126, 1010127, 1010128, 1010129, 1010130, 1010131, 1010132, 1010133, 1010135, 1010136, 1010137, 1010138, 1010139, 1010140, 1010141, 1010142, 1010143, 1010145, 1010146, 1010147, 1010148, 1010149, 1010151, 1010152, 1010153, 1010155, 1010156, 1010158, 1010159, 1010161, 1010162, 1010163, 1010164, 1010165, 1010166, 1010167, 1010168, 1010169, 1010170, 1010171, 1010172, 1010173, 1010174, 1010175, 1010176, 1010177, 1010178, 1010179, 1010180, 1010181, 1010182, 1010183, 1010184, 1010185, 1010186, 1010187, 1010188, 1010189, 1010190, 1010191, 1010192, 1010193, 1010194, 1010195, 1010196, 1010198, 1010199, 1010201, 1010202, 1010203, 1010204, 1010205, 1010206, 1010207, 1010208, 1010209, 1010210, 1010211, 1010212, 1010213, 1010214, 1010215, 1010216, 1010217, 1010218, 1010219, 1010220, 1010221, 1010222, 1010223, 1010224, 1010225, 1010226, 1010227, 1010228, 1010229, 1010230, 1010231, 1010233, 1010234, 1010235, 1010236, 1010237, 1010238, 1010239, 1010240, 1010241, 1010242, 1010243, 1010244, 1010245, 1010246, 1010247, 1010248, 1010249, 1010250, 1010251, 1010252, 1010253, 1010254, 1010255, 1010256, 1010257, 1010258, 1010259, 1010260, 1010261, 1010262, 1010263, 1010264, 1010265, 1010266, 1010267, 1010268, 1010269, 1010270, 1010271, 1010272, 1010273, 1010274, 1010275, 1010276, 1010277, 1010278, 1010279, 1010280, 1010281, 1010282, 1010283, 1010285, 1010286, 1010287, 1010288, 1010289, 1010290, 1010291, 1010292, 1010293, 1010294, 1010295, 1010296, 1010297, 1010298, 1010299, 1010300, 1010301, 1010302, 1010303, 1010304, 1010305, 1010306, 1010307, 1010308, 1010309, 1010310, 1010311, 1010312, 1010313, 1010314, 1010315, 1010316, 1010317, 1010318, 1010319, 1010320, 1010321, 1010322, 1010323, 1010324, 1010325, 1010326, 1010327, 1010328, 1010329, 1010330, 1010331, 1010332, 1010333, 1010334, 1010335, 1010336, 1010337, 1010338, 1010339, 1010340, 1010341, 1010342, 1010343, 1010344, 1010345, 1010347, 1010348, 1010349, 1010350, 1010351, 1010352, 1010353, 1010354, 1010355, 1010356, 1010357, 1010358, 1010359, 1010360, 1010361, 1010362, 1010363, 1010364, 1010365, 1010366, 1010367, 1010369, 1005227, 1010373, 1005229, 1010391, 1010392, 1010393, 1010394, 1010397, 1010398, 1010399, 1010400, 1010401, 1010402, 1010404, 1010406, 1010407, 1010408, 1010409, 1010412, 1010413, 1010414, 1010415, 1010416, 1010425, 1010428, 1005232, 1005233, 1010441, 1010461, 1010466, 1010468, 1010469, 1010470, 1010472, 1010473, 1010475, 1010476, 1010477, 1010478, 1010479, 1010480, 1010481, 1010483, 1010484, 1010485, 1010486, 1010487, 1010488, 1010489, 1010491, 1010492, 1010494, 1010495, 1010497, 1010498, 1010499, 1010500, 1010501, 1010504, 1010506, 1010507, 1010508, 1010509, 1010510, 1010511, 1010512, 1010513, 1010515, 1010516, 1010518, 1010519, 1010520, 1010521, 1010522, 1010523, 1010524, 1010525, 1010526, 1010527, 1010528, 1010529, 1010530, 1010531, 1010532, 1010533, 1010534, 1010535, 1010536, 1010537, 1010538, 1010539, 1010540, 1010542, 1010543, 1010544, 1010545, 1010546, 1010547, 1010548, 1010549, 1010550, 1010551, 1010552, 1010553, 1010554, 1010555, 1010556, 1010557, 1010558, 1010559, 1010560, 1010561, 1010562, 1010563, 1010564, 1010565, 1010566, 1010567, 1010568, 1010569, 1010570, 1010571, 1010572, 1010575, 1010578, 1010579, 1010615, 1010629, 1010632, 1010633, 1010646, 1010651, 1010652, 1010661, 1010664, 1010678, 1010683, 1010690, 1010691, 1010692, 1010693, 1010694, 1010695, 1010696, 1010697, 1010698, 1010699, 1010700, 1010701, 1010702, 1010703, 1010704, 1010705, 1010706, 1010707, 1010708, 1010709, 1010710, 1010711, 1010712, 1010713, 1010714, 1010715, 1010716, 1010718, 1010719, 1010721, 1010722, 1010723, 1010724, 1010725, 1010726, 1010727, 1010728, 1010729, 1010730, 1010731, 1010732, 1010733, 1010735, 1010736, 1010737, 1010738, 1010739, 1010740, 1010741, 1010742, 1010743, 1010744, 1010745, 1010746, 1010747, 1010748, 1010749, 1010750, 1010751, 1010752, 1010753, 1010754, 1010755, 1010756, 1010757, 1010758, 1010759, 1010760, 1010761, 1010762, 1010763, 1010764, 1010765, 1010766, 1010767, 1010768, 1010769, 1010770, 1010771, 1010772, 1010774, 1010775, 1010778, 1010781, 1010783, 1010786, 1010787, 1010789, 1010790, 1010791, 1010794, 1010795, 1010797, 1010798, 1010799, 1010801, 1010802, 1010804, 1010806, 1010807, 1010808, 1010809, 1010810, 1010811, 1010812, 1010813, 1010815, 1010816, 1010817, 1010818, 1010819, 1010820, 1010821, 1010822, 1010823, 1010824, 1010826, 1010828, 1010830, 1010831, 1010832, 1010833, 1010834, 1010835, 1010836, 1010837, 1010838, 1010839, 1010840, 1010841, 1010843, 1010845, 1010847, 1010848, 1010849, 1010850, 1010852, 1010853, 1010856, 1010857, 1010858, 1010859, 1010860, 1010861, 1010862, 1010864, 1010866, 1010868, 1010869, 1010870, 1010871, 1010872, 1010874, 1010875, 1010876, 1010877, 1010879, 1010880, 1010882, 1010883, 1010884, 1010885, 1010887, 1010888, 1010890, 1010891, 1010893, 1010894, 1010895, 1010896, 1010897, 1010898, 1010899, 1010900, 1010901, 1010902, 1010903, 1010904, 1010905, 1010906, 1010907, 1010908, 1010909, 1010910, 1010911, 1010912, 1010913, 1010914, 1010915, 1010916, 1010917, 1010920, 1010921, 1010923, 1010924, 1010925, 1010926, 1010927, 1010928, 1010929, 1010930, 1010931, 1010933, 1010934, 1010936, 1010937, 1010939, 1010940, 1010941, 1010943, 1010944, 1010947, 1010948, 1010949, 1010952, 1010955, 1010956, 1010958, 1010959, 1010961, 1010962, 1010963, 1010966, 1010967, 1010968, 1010969, 1010970, 1010971, 1010974, 1010975, 1010976, 1010977, 1010978, 1010979, 1010980, 1010981, 1010982, 1010983, 1010984, 1010985, 1010986, 1010987, 1010988, 1010989, 1010990, 1010991, 1010992, 1010993, 1010994, 1010995, 1010996, 1010997, 1010998, 1010999, 1011000, 1011001, 1011002, 1011003, 1011004, 1011005, 1011006, 1011007, 1011008, 1011009, 1011010, 1011011, 1011012, 1011013, 1011014, 1011015, 1011016, 1011017, 1011018, 1011019, 1011020, 1011022, 1011023, 1011024, 1011025, 1011026, 1011027, 1011028, 1011029, 1011030, 1011031, 1011032, 1011033, 1011035, 1011036, 1011037, 1011038, 1011039, 1011040, 1011041, 1011042, 1011044, 1011045, 1011046, 1011048, 1011049, 1011050, 1011051, 1011052, 1011053, 1011054, 1011055, 1011056, 1011057, 1011058, 1011059, 1011061, 1011062, 1011063, 1011064, 1011065, 1011066, 1011067, 1011068, 1011069, 1011070, 1011071, 1011072, 1011073, 1011074, 1011075, 1011076, 1011077, 1011078, 1011079, 1011080, 1011081, 1011082, 1011083, 1011084, 1011085, 1011086, 1011087, 1011088, 1011089, 1011090, 1011091, 1011092, 1011093, 1011094, 1011095, 1011096, 1011097, 1011098, 1011099, 1011100, 1011101, 1011102, 1011103, 1011104, 1011105, 1011106, 1011107, 1011108, 1011109, 1011110, 1011112, 1011113, 1011114, 1011115, 1011116, 1011117, 1011120, 1011121, 1011122, 1011123, 1011124, 1011125, 1011126, 1011127, 1011128, 1011129, 1011130, 1011131, 1011132, 1011133, 1011135, 1011136, 1011137, 1011138, 1011139, 1011141, 1011144, 1011145, 1011146, 1011147, 1011148, 1011149, 1011150, 1011151, 1011152, 1011153, 1011154, 1011155, 1011156, 1011157, 1011158, 1011159, 1011160, 1011161, 1011162, 1011163, 1011164, 1011165, 1011166, 1011167, 1011168, 1011169, 1011170, 1011172, 1011173, 1011174, 1011176, 1011177, 1011178, 1011179, 1011180, 1011181, 1011182, 1011183, 1011184, 1011185, 1011186, 1011187, 1011188, 1011189, 1011190, 1011192, 1011193, 1011194, 1011195, 1011196, 1011197, 1011198, 1011199, 1011200, 1011201, 1011202, 1011203, 1011204, 1011205, 1011206, 1011207, 1011209, 1011211, 1011212, 1011213, 1011214, 1011215, 1011216, 1011217, 1011218, 1011219, 1011220, 1011221, 1011222, 1011223, 1011224, 1011225, 1011226, 1011227, 1011228, 1011229, 1011230, 1011231, 1011232, 1011233, 1011234, 1011235, 1011236, 1011237, 1011238, 1011239, 1011240, 1011241, 1011242, 1011243, 1011244, 1011245, 1011246, 1011247, 1011248, 1011249, 1011250, 1011251, 1011252, 1011253, 1011254, 1011255, 1011256, 1011257, 1011258, 1011259, 1011260, 1011261, 1011262, 1011263, 1011264, 1011265, 1011266, 1011267, 1011268, 1011269, 1011270, 1011271, 1011272, 1011273, 1011274, 1011275, 1011276, 1011277, 1011279, 1011280, 1011281, 1011282, 1011283, 1011284, 1011285, 1011286, 1011287, 1011288, 1011291, 1011292, 1011293, 1011294, 1011295, 1011296, 1011297, 1011298, 1011300, 1011301, 1011302, 1011303, 1011304, 1011305, 1011306, 1011307, 1011308, 1011309, 1011310, 1011311, 1011312, 1011313, 1011316, 1011319, 1011321, 1011323, 1011324, 1011325, 1011326, 1011327, 1011330, 1011331, 1011332, 1011333, 1011334, 1011335, 1011336, 1011337, 1011338, 1011340, 1011343, 1011344, 1011345, 1011346, 1011350, 1011351, 1011352, 1011353, 1011354, 1011357, 1011361, 1011363, 1011365, 1011367, 1011371, 1011372, 1011373, 1011375, 1011376, 1011377, 1011378, 1011379, 1011380, 1011382, 1011383, 1011384, 1011385, 1011386, 1011387, 1011388, 1011390, 1011391, 1011393, 1011394, 1011395, 1011397, 1011398, 1011400, 1011401, 1011403, 1011404, 1011405, 1011406, 1011409, 1011410, 1011411, 1011413, 1011414, 1011415, 1011416, 1011417, 1011418, 1011419, 1011421, 1011422, 1011423, 1011424, 1011425, 1011427, 1011428, 1011429, 1011430, 1011431, 1011432, 1011433, 1011434, 1011435, 1011436, 1011438, 1011439, 1011440, 1011441, 1011442, 1011443, 1005234, 1011453, 1011454, 1011455, 1011456, 1011457, 1011458, 1011460, 1011461, 1011462, 1011463, 1011464, 1011465, 1011466, 1011467, 1011468, 1011469, 1011470, 1011471, 1011472, 1011473, 1011474, 1011475, 1011476, 1011477, 1011478, 1011479, 1011480, 1011481, 1011482, 1011483, 1011484, 1011485, 1011487, 1011488, 1011489, 1011490, 1011491, 1011492, 1011493, 1011494, 1011495, 1011496, 1011497, 1011498, 1011499, 1011500, 1011501, 1011502, 1011503, 1011504, 1011505, 1011506, 1011507, 1011508, 1011509, 1011510, 1011511, 1011513, 1011514, 1007182, 1008324, 1011565, 1011566, 1011567, 1011568, 1011569, 1011570, 1011571, 1011574, 1011575, 1011576, 1011577, 1011581, 1011582, 1011583, 1011584, 1011585, 1011587, 1011588, 1011589, 1011641, 1011642, 1011643, 1011644, 1011645, 1011646, 1011647, 1011648, 1011649, 1011650, 1011651, 1011652, 1011653, 1011656, 1011658, 1011659, 1011660, 1011661, 1011664, 1011665, 1011666, 1011667, 1011668, 1011669, 1011670, 1011671, 1011672, 1011673, 1011674, 1011675, 1011676, 1011678, 1011680, 1011681, 1011682, 1011683, 1011684, 1011686, 1011687, 1011688, 1011689, 1011690, 1011691, 1011692, 1011693, 1011694, 1011696, 1011698, 1011700, 1011703, 1011707, 1011708, 1011709, 1011710, 1011712, 1011713, 1011714, 1011715, 1011716, 1011717, 1011718, 1011719, 1011720, 1011721, 1011722, 1011723, 1011724, 1011725, 1011726, 1011727, 1011728, 1011729, 1011730, 1011731, 1011732, 1011733, 1011734, 1011736, 1011737, 1011738, 1011739, 1011740, 1011741, 1011742, 1011743, 1011744, 1011745, 1011746, 1011747, 1011748, 1011749, 1011750, 1011751, 1011752, 1011753, 1011754, 1011755, 1011756, 1011757, 1011758, 1011759, 1011760, 1011761, 1011762, 1011764, 1011765, 1011766, 1011767, 1011768, 1011769, 1011770, 1011772, 1011773, 1011775, 1011776, 1011777, 1011778, 1011779, 1011780, 1011781, 1011783, 1011784, 1011785, 1011786, 1011787, 1011788, 1011789, 1011790, 1011791, 1011792, 1011793, 1011794, 1011795, 1011796, 1011797, 1011798, 1011799, 1011800, 1011801, 1011802, 1011803, 1011804, 1011805, 1011806, 1011807, 1011808, 1011809, 1011810, 1011811, 1011812, 1011824, 1011825, 1011826, 1011827, 1011828, 1011829, 1011830, 1011831, 1011832, 1011833, 1011834, 1011835, 1011836, 1011837, 1011838, 1011839, 1011840, 1011841, 1011842, 1011846, 1011847, 1011848, 1011849, 1011850, 1011851, 1011853, 1011854, 1011855, 1011856, 1011857, 1011858, 1011859, 1011860, 1011861, 1011862, 1011863, 1011864, 1011865, 1011866, 1011867, 1011868, 1011869, 1011870, 1011871, 1011872, 1011873, 1011874, 1011875, 1011876, 1011878, 1011879, 1011880, 1011881, 1011882, 1011883, 1011884, 1011885, 1011887, 1011888, 1011889, 1011890, 1011891, 1011892, 1011893, 1011894, 1011895, 1011896, 1011897, 1011898, 1011899, 1011900, 1011901, 1011902, 1011903, 1011904, 1011905, 1011906, 1011907, 1011908, 1011909, 1011910, 1011911, 1011912, 1011913, 1011914, 1011915, 1011916, 1011917, 1011918, 1011919, 1011920, 1011921, 1011922, 1011923, 1011924, 1011925, 1011927, 1011928, 1011929, 1011930, 1011931, 1011932, 1011933, 1011934, 1011935, 1011936, 1011937, 1011938, 1011939, 1011940, 1011941, 1011942, 1011943, 1011944, 1011945, 1011946, 1011947, 1011948, 1011949, 1011950, 1011951, 1011952, 1011953, 1011954, 1011955, 1011956, 1011957, 1011958, 1011959, 1011960, 1011961, 1011963, 1011964, 1011965, 1011966, 1011967, 1011968, 1011969, 1011970, 1011971, 1011972, 1011973, 1011974, 1011975, 1011978, 1011979, 1011980, 1011981, 1011982, 1011983, 1011984, 1011985, 1011986, 1011987, 1011988, 1011989, 1011991, 1011992, 1011993, 1011997, 1011998, 1011999, 1012000, 1012001, 1012002, 1012003, 1012004, 1012005, 1012006, 1012007, 1012008, 1012009, 1012010, 1012011, 1012012, 1012013, 1012014, 1012015, 1012016, 1012017, 1012018, 1012019, 1012020, 1012021, 1012022, 1012023, 1012024, 1012025, 1012027, 1012028, 1012029, 1012031, 1012032, 1012033, 1012034, 1012035, 1012036, 1012037, 1012038, 1012039, 1012040, 1012041, 1012042, 1012043, 1012044, 1012046, 1012049, 1012050, 1012051, 1012052, 1012054, 1012055, 1012057, 1012058, 1012059, 1012060, 1012061, 1012062, 1012063, 1012064, 1012065, 1012066, 1012067, 1012069, 1012070, 1012071, 1012072, 1012073, 1012074, 1012075, 1012076, 1012077, 1012078, 1012079, 1012081, 1012082, 1012083, 1012084, 1012085, 1012086, 1012087, 1012088, 1012089, 1012090, 1012091, 1012092, 1012093, 1012094, 1012095, 1012096, 1012097, 1012098, 1012099, 1012100, 1012102, 1012103, 1012104, 1012109, 1012110, 1012113, 1012115, 1012116, 1012117, 1012119, 1012120, 1012121, 1012122, 1012124, 1012125, 1012126, 1012127, 1012133, 1012135, 1012137, 1012138, 1012139, 1012141, 1012142, 1012143, 1012144, 1012145, 1012146, 1012147, 1012148, 1012150, 1012152, 1012154, 1012158, 1012161, 1012162, 1012163, 1012164, 1012165, 1012166, 1012167, 1012168, 1012169, 1012170, 1012171, 1012172, 1012173, 1012174, 1012175, 1012176, 1012177, 1012178, 1012179, 1012180, 1012181, 1012182, 1012183, 1012184, 1012185, 1012187, 1012188, 1012189, 1012190, 1012191, 1012192, 1012193, 1012194, 1012195, 1012196, 1012197, 1012198, 1012199, 1012200, 1012201, 1012202, 1012204, 1012205, 1012206, 1012207, 1012208, 1012209, 1012210, 1012211, 1012212, 1012213, 1012214, 1012215, 1012216, 1012218, 1012219, 1012220, 1012221, 1012223, 1012224, 1012225, 1012226, 1012227, 1012228, 1012229, 1012230, 1012231, 1012232, 1012233, 1012234, 1012235, 1012236, 1012237, 1012238, 1012239, 1012240, 1012241, 1012242, 1012243, 1012245, 1012246, 1012247, 1012248, 1012249, 1012250, 1012252, 1012253, 1012255, 1012256, 1012257, 1012258, 1012259, 1012260, 1012261, 1012262, 1012263, 1012264, 1012266, 1012267, 1012268, 1012269, 1012270, 1012271, 1012272, 1012273, 1012345, 1012346, 1012347, 1012349, 1012350, 1012351, 1012353, 1012354, 1012355, 1012357, 1012358, 1012359, 1012360, 1012361, 1012362, 1012363, 1012364, 1012365, 1005235, 1005236, 1005237, 1005238, 1005239, 1005240, 1005242, 1005243, 1005244, 1005246, 1005247, 1005248, 1005249, 1005250, 1005251, 1014558, 1005253, 1005254, 1005256, 1005257, 1003344, 1005183, 1005184, 1005185, 1005186, 1018804, 1005187, 1005188, 1005189, 1005190, 1005192, 1005193, 1005194, 1005195, 1005196, 1005197, 1005198, 1005209, 1005213, 1005214, 1006467, 1005215, 1005216, 1005218, 1005220, 1005221, 1005222, 1005088, 1005104, 1005271, 1005273, 1005274, 1005275, 1005277, 1005278, 1005279, 1005284, 1005286, 1005289, 1005290, 1005306, 1005258, 1005259, 1005260, 1005261, 1005262, 1021851, 1005263, 1005264, 1005265, 1005266, 1005267, 1005268, 1005269, 1005270, 1004368, 1004399, 1004411, 1004676, 1004920, 1005021, 1005035, 1005037, 1005038, 1005058, 1005070, 1005102, 1005103, 1004363, 1004373, 1004382, 1004415, 1004418, 1004429, 1004432, 1004437, 1004464, 1004470, 1004473, 1004479, 1004490, 1004494, 1004521, 1004553, 1004560, 1004561, 1004582, 1004584, 1004614, 1004625, 1004627, 1004633, 1004639, 1004648, 1004652, 1004665, 1004677, 1004682, 1004713, 1004724, 1004760, 1004769, 1004793, 1004795, 1004796, 1004799, 1004808, 1004813, 1004814, 1004815, 1004818, 1004820, 1004824, 1004828, 1004829, 1004832, 1004836, 1004838, 1004839, 1004840, 1004843, 1004847, 1004848, 1004849, 1004851, 1004854, 1004855, 1004857, 1004858, 1004859, 1004866, 1004867, 1004868, 1004869, 1004870, 1004872, 1004873, 1004874, 1004880, 1004882, 1004883, 1004886, 1004887, 1004888, 1004889, 1004890, 1004891, 1004892, 1004893, 1004895, 1004896, 1004899, 1004903, 1004904, 1004907, 1004909, 1004914, 1004915, 1004918, 1004919, 1004922, 1004924, 1004925, 1004926, 1004927, 1004929, 1004930, 1004933, 1004935, 1004936, 1004937, 1004938, 1004941, 1004942, 1004944, 1004945, 1004948, 1004951, 1004954, 1004956, 1004958, 1004959, 1004960, 1004961, 1004962, 1004964, 1004970, 1004972, 1004974, 1004976, 1004981, 1004982, 1004983, 1004984, 1004985, 1004987, 1010577, 1004988, 1009104, 1004989, 1004991, 1004993, 1004994, 1004996, 1004997, 1005002, 1005003, 1005005, 1005007, 1005011, 1008294, 1005012, 1005016, 1005017, 1005018, 1005019, 1005020, 1005024, 1005025, 1005027, 1005031, 1005032, 1005072, 1005073, 1005074, 1005075, 1005078, 1005080, 1005082, 1005083, 1005085, 1005087, 1005089, 1005090, 1005091, 1005092, 1005093, 1005094, 1005096, 1005097, 1005098, 1005101, 1004356, 1004209, 1004370, 1004381, 1004384, 1004396, 1004401, 1004412, 1004416, 1004419, 1004421, 1004427, 1004436, 1004447, 1004448, 1004449, 1004451, 1004453, 1004455, 1004460, 1004463, 1004465, 1004468, 1004471, 1004475, 1004480, 1004483, 1004489, 1004497, 1004499, 1004502, 1004506, 1004509, 1004511, 1004513, 1004517, 1004522, 1004524, 1004526, 1004530, 1004538, 1004540, 1004543, 1004544, 1004546, 1004554, 1004555, 1004558, 1004563, 1004565, 1004566, 1004571, 1004573, 1004576, 1004577, 1004578, 1004580, 1004583, 1004585, 1004587, 1004588, 1004589, 1004590, 1004591, 1004593, 1004594, 1004597, 1004600, 1004604, 1004605, 1004606, 1004607, 1004608, 1004610, 1004612, 1004613, 1004616, 1004621, 1004622, 1004623, 1004628, 1004629, 1004631, 1004632, 1004636, 1004637, 1004642, 1004645, 1004658, 1004664, 1004666, 1004667, 1004668, 1004670, 1004671, 1004672, 1006923, 1004673, 1004675, 1004679, 1004680, 1004684, 1004686, 1004689, 1004694, 1004696, 1004698, 1004702, 1004703, 1004705, 1004706, 1004707, 1004711, 1004714, 1004715, 1004718, 1004719, 1004722, 1004723, 1004728, 1040010, 1040016, 1004731, 1004732, 1004733, 1004739, 1004740, 1004745, 1004751, 1004754, 1004756, 1004762, 1004763, 1004766, 1004767, 1004773, 1004774, 1004776, 1004777, 1004778, 1004780, 1004783, 1004784, 1004785, 1004786, 1004787, 1004791, 1004800, 1004801, 1004804, 1004805, 1004845, 1004443, 1004445, 1004425, 1004426, 1004431, 1004433, 1004435, 1004352, 1004372, 1004378, 1004386, 1004391, 1004392, 1004393, 1004394, 1004397, 1004398, 1004402, 1004406, 1004414, 1004417, 1004420, 1004422, 1004380, 1003728, 1003805, 1003807, 1003884, 1003886, 1003890, 1003891, 1003893, 1003895, 1003896, 1003899, 1003900, 1003902, 1003905, 1003906, 1003907, 1003909, 1003910, 1003914, 1003952, 1003955, 1003956, 1003961, 1003965, 1003877, 1003878, 1003880, 1003881, 1003912, 1003916, 1003873, 1003874, 1003875, 1003949, 1003951, 1004140, 1004143, 1004146, 1004147, 1004148, 1004153, 1004156, 1004159, 1004165, 1004166, 1004167, 1004168, 1004170, 1004173, 1004174, 1003795, 1003813, 1004123, 1004124, 1004125, 1004128, 1004134, 1003520, 1003547, 1003585, 1003589, 1048582, 1003593, 1003617, 1003626, 1003647, 1003768, 1003769, 1003771, 1003772, 1003773, 1003774, 1003775, 1003776, 1003777, 1003778, 1003779, 1003781, 1003782, 1003783, 1003785, 1003786, 1003787, 1003788, 1003790, 1003791, 1003792, 1003793, 1003796, 1003797, 1003798, 1003799, 1003800, 1003801, 1003802, 1003804, 1003806, 1003808, 1003809, 1003811, 1003812, 1003814, 1003816, 1004052, 1004202, 1004203, 1004230, 1004246, 1004254, 1004262, 1003602, 1003452, 1003684, 1003685, 1003686, 1003689, 1003702, 1003672, 1003754, 1003763, 1003758, 1003759, 1003762, 1054577, 1003747, 1003749, 1003750, 1003746, 1003720, 1003723, 1003726, 1003727, 1003729, 1003730, 1003731, 1003732, 1003740, 1003741, 1003706, 1003707, 1003709, 1003715, 1003716, 1003717, 1003718, 1003719, 1003691, 1003692, 1003695, 1003696, 1004058, 1003667, 1003688, 1003661, 1003662, 1003655, 1003656, 1003657, 1003659, 1003660, 1008413, 1003270, 1003283, 1003286, 1003308, 1003310, 1003317, 1003326, 1003400, 1003416, 1003449, 1003503, 1003513, 1003561, 1003572, 1003454, 1003540, 1002858, 1002859, 1002861, 1002864, 1002867, 1002875, 1002881, 1002883, 1002924, 1002930, 1002992, 1003106, 1003185, 1003187, 1003188, 1003191, 1003198, 1003200, 1003201, 1003203, 1003205, 1003207, 1003214, 1003217, 1003221, 1003233, 1003234, 1003235, 1003236, 1003237, 1003238, 1003239, 1003240, 1003242, 1003623, 1003624, 1003628, 1003629, 1003640, 1003641, 1003642, 1003643, 1002851, 1002853, 1002755, 1002854, 1002920, 1002918, 1002914, 1002915, 1002634, 1002892, 1002907, 1002916, 1003610, 1002904, 1002906, 1002911, 1002913, 1003142, 1003098, 1003102, 1003108, 1003109, 1003112, 1003113, 1003117, 1003118, 1003120, 1003131, 1003987, 1003133, 1003136, 1003137, 1003140, 1003070, 1003073, 1003074, 1003075, 1003077, 1003078, 1003079, 1003080, 1003081, 1003083, 1003084, 1003086, 1003350, 1003089, 1003090, 1003091, 1003092, 1003093, 1003094, 1003095, 1003097, 1003057, 1003063, 1003064, 1003065, 1003067, 1003069, 1003320, 1002872, 1003016, 1003020, 1003028, 1003030, 1003031, 1003036, 1003044, 1003047, 1003049, 1002955, 1002596, 1002931, 1002934, 1002937, 1002939, 1002940, 1002944, 1002947, 1002949, 1002951, 1002953, 1002954, 1002763, 1002764, 1002772, 1002773, 1002758, 1002762, 1002770, 1002723, 1002724, 1002725, 1002726, 1002731, 1002742, 1002746, 1002750, 1002529, 1002531, 1002539, 1002556, 1002601, 1002646, 1008439, 1002693, 1011047, 1002463, 1002471, 1002472, 1002476, 1002479, 1002481, 1002483, 1002485, 1002486, 1002487, 1002488, 1002490, 1002494, 1002498, 1002503, 1002505, 1002504, 1002377, 1002378, 1002379, 1002384, 1002387, 1002388, 1002389, 1002390, 1002391, 1002398, 1002400, 1002401, 1002404, 1002368, 1002370, 1002372, 1002374, 1002375, 1002326, 1002327, 1002328, 1002331, 1002337, 1002339, 1002340, 1002341, 1002344, 1002345, 1007592, 1002346, 1002348, 1002350, 1002351, 1002353, 1002354, 0001502, 1002355, 1002358, 1002362, 1002363, 1002364, 1002242, 1002245, 1002246, 1002247, 1002249, 1002251, 1002252, 1002255, 1002260, 1002265, 1002267, 1002268, 1002270, 1007094, 1002272, 1002275, 1002276, 1002277, 1002281, 1004013, 1004037, 1004039, 1004040, 1004044, 1002282, 1002284, 1004046, 1004048, 1004054, 1007184, 1007187, 1008307, 1008309, 1002286, 1002288, 1002289, 1002290, 1002291, 1002294, 1002296, 1002297, 1006582, 1002298, 1002299, 1002303, 1002304, 1002307, 1002309, 1002317, 1002318, 1002319, 1002141, 1002152, 1002155, 1002168, 1002178, 1002181, 1002187, 1002193, 1002197, 1002201, 1002206, 1002215, 1002223, 1002224, 1002226, 1002229, 1002233, 1002236, 1002128, 1002131, 1002132, 1002133, 1002134, 1002135, 1002140; REF List Number: 21-2102-0200-02; UDI/DI 30610586035511, Serial Numbers: 1010433, 1010434, 1009306; REF List Number: 21-2102-0200-03; UDI/DI 30610586035528, Serial Numbers: 1009308; REF List Number: 21-2102-0200-06; UDI/DI 10610586035531, Serial Numbers: 1009304; REF List Number: 21-2102-249; UDI/DI 10610586033940, Serial Numbers: 1006710, 1006716, 1006768, 1006788, 1006804, 1006806, 1006807, 1006808, 1006810, 1006812, 1006813, 1006815, 1006817, 1006820, 1006821, 1006822, 1006823, 1006824, 1006825, 1006826, 1006828, 1006829, 1006832, 1006833, 1006834, 1006836, 1007085, 1007203, 1007206, 1007207, 1007209, 1007210, 1007211, 1007212, 1007214, 1007215, 1007216, 1006811; REF List Number: 21-2102-51; UDI/DI 10610586033933, Serial Numbers: 1006729, 1002473, 1002501, 1031846, 1003761, 1003819, 1004646, 1004647, 1004649, 1004651, 1004653, 1004655, 1004656, 1004657, 1005033, 1005034, 1005036, 1005040, 1005041, 1005042, 1005043, 1005044, 1005045, 1005046, 1005047, 1005048, 1005049, 1005050, 1005051, 1005052, 1005053, 1005055, 1005056, 1005057, 1005059, 1005060, 1005062, 1005063, 1005064, 1005065, 1005066, 1005067, 1005068, 1005069, 1005076, 1005105, 1006009, 1006728, 1006730, 1006731, 1006733, 1006742, 1006751, 1006753, 1006755, 1006756, 1006757, 1006758, 1006759, 1006760, 1006762, 1006765, 1006766, 1006769, 1006772, 1007009, 1007082, 1007084, 1007087, 1007088, 1007091, 1007093, 1007103, 1007104, 1007105, 1007140, 1007142, 1007143, 1007248, 1007276, 1007414, 1007415, 1007416, 1007417, 1007418, 1007419, 1007420, 1007423, 1007424, 1007426, 1007427, 1007428, 1007430, 1007432, 1007433, 1007434, 1007435, 1007436, 1007437, 1007439, 1007440, 1007441, 1007442, 1007443, 1007444, 1007446, 1007457, 1007463, 1007465, 1007482, 1007488, 1007741, 1007742, 1007743, 1008037, 1008040, 1008041, 1008042, 1008044, 1008045, 1008047, 1008048, 1008051, 1008055, 1008056, 1008057, 1008093, 1008094, 1008095, 1008096, 1008097, 1008098, 1008099, 1008100, 1008101, 1008102, 1008347, 1008369, 1008374, 1008385, 1008419, 1008423, 1008424, 1008427, 1008441, 1008442, 1008475, 1008491, 1008492, 1008493, 1008494, 1008495, 1008496, 1008497, 1008498, 1008499, 1008500, 1008501, 1008502, 1008503, 1008504, 1008505, 1008506, 1008507, 1008508, 1008509, 1008510, 1008511, 1008512, 1008513, 1008514, 1008515, 1008869, 1009153, 1009157, 1009160, 1009161, 1009171, 1009172, 1010368, 1010395, 1010396, 1010403, 1010576, 1010825, 1010829, 1010842, 1010844, 1010846, 1010851, 1010854, 1010855, 1010865, 1010867, 1010873, 1010878, 1010886, 1010892, 1010918, 1010919, 1010922, 1010935, 1010938, 1010942, 1010945, 1010946, 1010950, 1010951, 1010953, 1010957, 1010960, 1010964, 1010972, 1010973, 1011315, 1011317, 1011318, 1011320, 1011322, 1011328, 1011329, 1011339, 1011348, 1011349, 1011355, 1011356, 1011358, 1011359, 1011362, 1011364, 1011366, 1011368, 1011369, 1011370, 1011374, 1011381, 1011389, 1011392, 1011396, 1011399, 1011407, 1011420, 1011437, 1009156, 1010954, 1010965, 1011426, 1047547, 1007471, 1008034, 1008035, 1008036, 1008038, 1008039, 1008043, 1008046, 1008049, 1008050, 1008052, 1008053, 1008054, 1009151, 1009152, 1009154, 1009155, 1009158, 1009159, 1002468, 1002470, 10006770

Other devices under the same FDA submission

3 devices cleared under K072144 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 94225).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page