Ivenix, Inc. recallZ-1484-2024 · Fresenius Kabi USA, LLC · initiated Mar 7, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Fresenius Kabi USA, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Software has anomalies that have the potential to cause underdose, overdose, or delay in therapy which could lead to serious patient harm or death.

FDA root cause category: Software Design Change

What the company is doing

A letter titled URGENT - Medical Device Field Correction, dated 3/7/24, was delivered by hand to consignees or by mail notifying them of this recall event. This notice details eleven software anomalies that could pose a potential risk of serious patient harm or death and the mitigations consignees should perform until software is updated. Consignees are to update their software to LVP software version 5.9.1 by reaching out to their Fresenius Kabi representative at 1-978-775-8100 to schedule installation. Consignees are to complete and return the Customer Reply Form via fax or email. Questions about the recall notice can be directed to Fresenius Kabi by email at [email protected] or by phone at 1-978-775-8100 (Monday through Friday from 8:30 am to 5:00 pm).

Recall details

Recall number
Z-1484-2024
Status
Open, Classified
Initiated by the company
Mar 7, 2024
Posted by the FDA
Apr 11, 2024
Recalling firm
Fresenius Kabi USA, LLC
Quantity
17 units
Product code
FRN
FDA submissions
K183311
Distribution
US Distribution: CA, CO, ID, MI, NE, NJ, NV, OK, TX, UT, VA, WA, & WI.

Product description

LVP Software of the Ivenix Infusion System (IIS), Version 5.8.0.

Code information (models, lots, serial numbers)

Product LVP-SW-0004; UDI-DI 00811505030122.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

4 devices cleared under K183311 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.