Patient Cart, Part: 380601, a component of the da Vinci SP… recallZ-1518-2024 · Intuitive Surgical, Inc. · initiated Mar 13, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Intuitive Surgical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Surgical system may have inadequately welded top and/or bottom belts in the link 2 of the entry guide manipulator assembly/instrument arm joints, that may cause the following, 1) Top Failure: internal tissue injuries, body wall injury or emergent conversion to open surgery, 2) Bottom failure: pinching injury to the user with no harm to the patient.

FDA root cause category: Nonconforming Material/Component

What the company is doing

On 3/13/24, Field Safety Notices were mailed and emailed to customers who were asked to do the following: 1) To prevent a pinch injury, please use two people when attempting to move the EGM in an unpowered state. 2) Inform necessary hospital personnel about this issue and affected personnel when the requested actions have been completed. 3) Complete and return the response form via email to [email protected] A firm representative will schedule a site visit to perform the inspection of the Link 2 EGM belts of the affected population of SP systems. If you need further information or support concerning this Medical Device Notification, please contact your Clinical Sales Representative or contact Firm Customer Service: - North and South America: (800) 876-1310, Option 3 (4 AM to 5 PM PST) or mail: [email protected]. - Europe, Middle East, Asia and Africa: +800 0821 2020 or +41 21 821 2020 (8 AM to 6 PM CET) or [email protected] - South Korea: 02-3271-3200 (9 AM to 6 PM KSTJ) or [email protected] - Japan: 0120-56-5635 or 03-5575-1362 (9 AM to 6 PM JST) or [email protected]

Recall details

Recall number
Z-1518-2024
Status
Open, Classified
Initiated by the company
Mar 13, 2024
Posted by the FDA
Apr 11, 2024
Recalling firm
Intuitive Surgical, Inc.
Quantity
53
Product code
NAY
FDA submissions
K173906, K182371
Distribution
US: VA, CA, NH, FL, NY, CO, AL, KY, MD, NC, NJ, OH, SC, SD, TN, WA, WI, WV, MS, TX, KS. OUS: China, France, Japan, South Korea

Product description

Patient Cart, Part: 380601, a component of the da Vinci SP Surgical System, Model: SP1098

Code information (models, lots, serial numbers)

Part/UDI-DI: 380601/00886874114605. SP1098 Serial Numbers: SP0202, SP0253, SP0254, SP0219, SP0228, SP0231, SP0238, SP0273, SP0208, SP0226, SP0240, SP0247, SP0249, SP0256, SP0257, SP0268, SP0270, SP0272, SP0221, SP0223, SP0225, SP0229, SP0230, SP0232, SP0233, SP0235, SP0236, SP0237, SP0239, SP0241, SP0242, SP0243, SP0244, SP0245, SP0248, SP0250, SP0251, SP0252, SP0255, SP0258, SP0259, SP0260, SP0261, SP0262, SP0263, SP0264, SP0265, SP0266, SP0269, SP0274, SP0278, SP0280, SP0282