da Vinci Surgical SystemRecalls & reported problems across all models
Intuitive Surgical, Inc. · 9 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
3 open recalls name at least one da Vinci Surgical System model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1,459
- Deaths reported
- 13
- Injuries reported
- 336
- Recalls
- 5
We found 9 models of da Vinci Surgical System by Intuitive Surgical, Inc. in the FDA's device registry (GUDID). Together they are named in 1,459 adverse event reports in the FDA's MAUDE database, received between Apr 9, 2020 and Sep 30, 2026 — classified as 13 deaths, 336 injuries, 1,109 malfunctions.
The most frequently reported specific device problem is “Visual Prompts will not Clear” (235 reports).
5 FDA recall records name at least one of these models by its device identifier, 3 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
da Vinci Surgical System recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Apr 22, 2026Open, Classified
Operative components of surgical systems potentially distributed with E-brake retainer pin, so in event of a structural column failure, the E-brake pin would prevent E-brake engagement and may result in uncontrolled motion. In addition, if it moves/backs out it may cause restricted column range of motion and/or uncontrolled motion, which may cause patient tissue injury/crushing injury/bleeding.
Z-2521-2026 · Root cause: Process control · Recall details · FDA record ↗
- Mar 13, 2024Open, Classified
Surgical system may have inadequately welded top and/or bottom belts in the link 2 of the entry guide manipulator assembly/instrument arm joints, that may cause the following, 1) Top Failure: internal tissue injuries, body wall injury or emergent conversion to open surgery, 2) Bottom failure: pinching injury to the user with no harm to the patient.
Z-1518-2024 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
- Sep 20, 2022Open, Classified
Inadvertent energy delivery from surgical system instrument if 1) Force bipolar and bipolar instruments installed on system 2) Force bipolar connected to Force Triad generator, bipolar not connected to generator 3) Yellow pedal associated with force bipolar is pressed, released with head in surgeon console 4) Blue pedal associated with bipolar is pressed, resulting in force bipolar energy delivery
Z-0133-2023 · Root cause: Software Design Change · names 2 of these models · Recall details · FDA record ↗
- Jun 22, 2022Completed
Due to preventative maintenance (PM) not being performed on E-100 generators which is a bipolar electrosugical unit designed to be used with the da Vinci Xi and X systems.
Z-1484-2022 · Root cause: Process design · Recall details · FDA record ↗
- Mar 3, 2022Completed
Placing excessive tissue in the instrument jaws of the devices prior to sealing and transection can result in an insufficient seal.
Z-1210-2022 · Root cause: Use error · Recall details · FDA record ↗
Most reported da Vinci Surgical System problems
Device problems
- Visual Prompts will not Clear235
- Output Problem201
- Adverse Event Without Identified Device or Use Problem103
- Failure to Deliver Energy90
- Detachment of Device or Device Component89
- No Device Output85
- Energy Output Problem77
- Difficult to Open or Close63
Patient problems
- No Clinical Signs, Symptoms or Conditions1,073
- Insufficient Information132
- Unspecified Tissue Injury84
- Hemorrhage/Blood Loss/Bleeding53
- Foreign Body In Patient42
- Burn(s)15
- Hematoma9
- No Known Impact Or Consequence To Patient9
da Vinci Surgical System models
9 models, most reported first. Each model has its own page with its reports, recalls and device details.
- da Vinci Energy 480440System, surgical, computer controlled instrument1 recall2 deaths493 reports
- da Vinci SP 380601System, surgical, computer controlled instrument2 recalls9 deaths286 reports
- da Vinci Xi 380721System, surgical, computer controlled instrument1 recall299 reports
- da Vinci Xi 380723System,surgical,computer controlled instrument1 recall170 reports
- da Vinci SP 380941System, surgical, computer controlled instrument133 reports
- da Vinci SP 380940System, surgical, computer controlled instrument1 death59 reports
- da Vinci Energy 378848System, surgical, computer controlled instrument1 recall3 reports
- da Vinci Energy 480522System, surgical, computer controlled instrument1 death15 reports
- da Vinci SP Endoscope, 0° 430600System, surgical, computer controlled instrument1 report
Recent da Vinci Surgical System adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Failure to Power Up; Material Split, Cut or Torn
Patient problems: Insufficient Information
MAUDE report 50953697 ↗MalfunctionSep 29, 2026Visual Prompts will not Clear
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50939505 ↗MalfunctionSep 29, 2026Output Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50939481 ↗MalfunctionSep 29, 2026Energy Output Problem
It was reported that during a da Vinci-assisted surgical procedure, a remote site representative had contacted the Intuitive Surgical, Inc. (ISI) Technical Support Engineer (TSE) on behalf of the user site to report that monopolar energy could not be activated and that the cord indicator on the energy switch module had been amber while the other indicators had been blue. Prior to the call, users had verified that all cords had been connected correctly and had replaced the energy cord and the instrument, but the issue had persisted. The TSE had asked the caller to verify that the instrument cord had been connected to the energy switch module rather than the generator, and the caller had confirmed this had already been verified based on a photo of the setup. The TSE had then advised restarting the system, and the caller had planned to relay that instruction to the user site and call back; the call had ended before additional troubleshooting details had been obtained. The procedure outcome is unknown with no reported injury.ISI followed up with the initial reporter and obtained the following additional information: The case was completed robotically but the monopolar energy wasn't wor
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50928033 ↗MalfunctionSep 29, 2026Break; Failure to Seal
It was reported that the sealing was not completed when using a Vessel Sealer Curved instrument. No known impact or patient consequence was reported.
Patient problems: Insufficient Information
MAUDE report 50927510 ↗MalfunctionSep 29, 2026Difficult to Open or Close
It was reported that during a da Vinci-assisted surgical procedure, the Vessel Sealer Curved instrument jaws would not open. The procedure was completed. No known impact or patient consequence was reported.Intuitive Surgical, Inc. (ISI) followed up with the initial reporter and obtained the following additional information: There was no injury to the patient.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50927498 ↗MalfunctionSep 28, 2026Difficult to Open or Close
It was reported that during a da Vinci-assisted surgical procedure, the Vessel Sealer Curved instrument jaws would not open. The procedure was completed. No known impact or patient consequence was reported.ISI followed up with the initial reporter and obtained the following additional information: There was no injury to the patient. The instrument was returned.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50924010 ↗MalfunctionSep 26, 2026Output Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50906735 ↗MalfunctionSep 25, 2026Visual Prompts will not Clear
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50905863 ↗MalfunctionSep 25, 2026Material Fragmentation; Unintended Movement
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50905857 ↗MalfunctionSep 25, 2026Output Problem
Patient problems: Insufficient Information
MAUDE report 50905524 ↗MalfunctionSep 25, 2026Failure to Power Up
Patient problems: Insufficient Information
MAUDE report 50904522 ↗MalfunctionSep 25, 2026No Device Output
It was reported that during a da Vinci-assisted surgical procedure, the customer reported that the bottom monopolar port on the Integrated ElectroSurgical Unit (IESU) was not functioning, displaying a question mark when instruments were attached. Different instruments and energy cords were tested but the port still failed to recognize them. The top monopolar port remained functional. The Intuitive Surgical, Inc. (ISI) Technical Support Engineer (TSE) log review showed no related errors. The procedure was completed, and no patient injury was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50893636 ↗MalfunctionSep 24, 2026No Display/Image
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50891143 ↗MalfunctionSep 24, 2026No Device Output; Material Split, Cut or Torn
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50890908 ↗
Other Intuitive Surgical, Inc. products
- EndoWrist62 models · 69,753 reports · 48 recalls
- daVinci Xi4 models · 11,477 reports · 3 recalls
- EndoWrist;daVinci Si35 models · 1,683 reports · 8 recalls
- daVinci 53 models · 1,608 reports · 5 recalls
- Ion6 models · 1,367 reports · 3 recalls
- SureForm16 models · 1,190 reports · 3 recalls
- EndoWrist SP20 models · 993 reports · 1 recall
- daVinci Si8 models · 681 reports · 1 recall
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered, combining the brand names used for this product line.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page