Skip to content

InflammaDry recallZ-1636-2024 · Quidel Corporation · initiated Mar 5, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Quidel Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

There are two windows on the test device, the control line window and the test results window. The line for the test results window were misaligned. In instances of positive test results, the positivite test line may not be visible/ hidden under the cassette of the test device thus potentially leading to a false negative result.

FDA root cause category: Process design

What the company is doing

On March 5, 2024, QuidelOrtho issued a correction notification to affected consignees. QuidelOrtho asked consignees to take the following actions: 1. Please complete and return the attached confirmation of notification form to QuidelOrtho via fax at 858.203.9297 or email at [email protected]. 2. Indicate the amount of product to be replaced on the confirmation of notification form. You are not required to take any action related to already used kit material. 3. If requesting replacement product, discontinue use, render unusable, and discard the remaining inventory of affected lot numbers once replacement product arrives. 4. If unaffected replacement product is not immediately available or you wish to continue testing, please follow these steps, as applicable, to minimize patient risk: 4a. Immediately prior to running a test, carefully inspect each unused InflammaDry test device to ensure two faint orange lines appear in the result window prior to use. Only open and inspect the test device immediately prior to use as the device should not be stored outside of the foil pouch. 4b. Do not use any of the InflammaDry test devices if two faint orange lines are not visible prior to use. Please request replacement for any unused devices using the confirmation of notification form. . 4c. Perform external controls with each new lot, each new shipment, and every 30 days, per the package insert recommendations. Ensure the positive control produces expected results. 5. To report adverse events, contact your local Technical Solutions Center. Adverse reactions or quality problems experienced with the use of this product may also be reported to the FDA s MedWatch Adverse Event Reporting program: www.fda.gov/medwatch/report.htm, 1-800-332-1088.

Recall details

Recall number
Z-1636-2024
Status
Open, Classified
Initiated by the company
Mar 5, 2024
Posted by the FDA
Apr 24, 2024
Recalling firm
Quidel Corporation
Quantity
104,300 units
Product code
PFQ
FDA submissions
K132066
Distribution
US: AK AL AR AZ CA CO CT DC FL GA HI IA ID IL IN KS KY LA MA MD MI MN MO MT NC ND NE NH NJ NV NY OH OK OR PA SC SD TN TX UT VA WA WI WV OUS: United Arab Emirates, Belgium, Canada, Chile, China, Germany, Spain, United Kingdom, Ireland, Italy, South Korea, Lithuania, Qatar

Product description

QUIDEL InflammaDry MMP-9 Test, REF: PRS-ID-20-U, 20 Tests

Code information (models, lots, serial numbers)

UDI: 30014613336754/Lot # 162621 162695 162498 162718 162749 163025

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

1 device cleared under K132066 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 94219).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page