InflammaDry recallZ-1637-2024 · Quidel Corporation · initiated Mar 5, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Quidel Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There are two windows on the test device, the control line window and the test results window. The line for the test results window were misaligned. In instances of positive test results, the positivite test line may not be visible/ hidden under the cassette of the test device thus potentially leading to a false negative result.
FDA root cause category: Process design
What the company is doing
On March 5, 2024, QuidelOrtho issued a correction notification to affected consignees. QuidelOrtho asked consignees to take the following actions: 1. Please complete and return the attached confirmation of notification form to QuidelOrtho via fax at 858.203.9297 or email at [email protected]. 2. Indicate the amount of product to be replaced on the confirmation of notification form. You are not required to take any action related to already used kit material. 3. If requesting replacement product, discontinue use, render unusable, and discard the remaining inventory of affected lot numbers once replacement product arrives. 4. If unaffected replacement product is not immediately available or you wish to continue testing, please follow these steps, as applicable, to minimize patient risk: 4a. Immediately prior to running a test, carefully inspect each unused InflammaDry test device to ensure two faint orange lines appear in the result window prior to use. Only open and inspect the test device immediately prior to use as the device should not be stored outside of the foil pouch. 4b. Do not use any of the InflammaDry test devices if two faint orange lines are not visible prior to use. Please request replacement for any unused devices using the confirmation of notification form. . 4c. Perform external controls with each new lot, each new shipment, and every 30 days, per the package insert recommendations. Ensure the positive control produces expected results. 5. To report adverse events, contact your local Technical Solutions Center. Adverse reactions or quality problems experienced with the use of this product may also be reported to the FDA s MedWatch Adverse Event Reporting program: www.fda.gov/medwatch/report.htm, 1-800-332-1088.
Recall details
- Recall number
- Z-1637-2024
- Status
- Open, Classified
- Initiated by the company
- Mar 5, 2024
- Posted by the FDA
- Apr 24, 2024
- Recalling firm
- Quidel Corporation
- Quantity
- 31,140 units
- Product code
- PFQ
- FDA submissions
- K132066
- Distribution
- US: AK AL AR AZ CA CO CT DC FL GA HI IA ID IL IN KS KY LA MA MD MI MN MO MT NC ND NE NH NJ NV NY OH OK OR PA SC SD TN TX UT VA WA WI WV OUS: United Arab Emirates, Belgium, Canada, Chile, China, Germany, Spain, United Kingdom, Ireland, Italy, South Korea, Lithuania, Qatar
Product description
QUIDEL InflammaDry MMP-9 Test, REF: PRS-ID-20, 20 Tests
Code information (models, lots, serial numbers)
UDI: 30014613336747/Lot # 162620 and 162748
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- InflammaDry RPS-ID-20Quidel Corporation · MMP-9 test system1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 94219).
- Z-1636-2024 · Open, ClassifiedQUIDEL InflammaDry MMP-9 Test, REF: PRS-ID-20-U, 20 Tests
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page