Palindrome recallZ-1634-2024 · Covidien LP · initiated Mar 15, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Covidien LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Products may not be heparin coated as indicated on the label-leads to a potential loss of benefit due to the absence of the heparin coating
FDA root cause category: Process control
What the company is doing
Medtronic issued Urgent Medical Device Recall Letter on 3/15/24. Letter states reason for recall, health risk and action to take: 1.Immediately quarantine and discontinue use of unused affected product (see Attachment A). 2. Return all unused products from your inventory to Medtronic. Please contact: [email protected] for the Return Good Authorization (RGA). a.Credit for the returned affected product will be issued based on the RGA number. b. If purchased from a distributor, contact your distributor directly to arrange for the return of the product back to your distributor. 3. In addition, please complete and return the enclosed Customer Confirmation Form to [email protected] even if you do not have unused inventory. 4. This notice should be passed on to those who need to be aware within your organization or to any organization including, but not limited to, nephrologists, intensivists, implanting and managing physicians, renal nurses, critical care nurses, or other dialysis staff where the potentially affected devices have been transferred. Please maintain a copy of this notice in your records. If you have questions regarding this communication, please contact your local representative or Customer Service at 800-962-9888, option 2 (Monday Friday 8:00 a.m. to 6:30 p.m.).
Recall details
- Recall number
- Z-1634-2024
- Status
- Open, Classified
- Initiated by the company
- Mar 15, 2024
- Posted by the FDA
- Apr 24, 2024
- Recalling firm
- Covidien LP
- Quantity
- 305 units
- Product code
- MSD
- FDA submissions
- K123196
- Distribution
- Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canary Islands, Denmark, Faroe Islands, France, Germany, Greece, Iceland, Ireland, Italy, Mayotte, Norway, Poland, Reunion, South Africa, Spain, Sweden, United Kingdom.
Product description
Palindrome Precision HSI Chronic Catheter Sport Pack 14.5 Fr/Ch (4.8mm) x 19 cm (Symmetrical Tip, Heparin Coating, Silver Ion Sleeve and Tal VenaTrac Stylet- Intended for acute and chronic hemodialysis, apheresis, and infusion. Model Number (CFN): 8888119369P
Code information (models, lots, serial numbers)
GTIN: 20884521157733 Lot Number: 2318500118
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Palindrome 8888119369PMozarc Medical US LLC · Catheter, hemodialysis, implanted3 recalls1 death14 reports
Product lines
Other devices under the same FDA submission
69 devices cleared under K123196 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Palindrome 8888145044CPMozarc Medical US LLC · Catheter, hemodialysis, implanted, coated2 recalls2 deaths30 reports
- Palindrome 8888119364PMozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls1 death17 reports
- Palindrome 8888123404PMozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls1 death17 reports
- Palindrome 8888123409PMozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls1 death12 reports
- Palindrome 8888145014PMozarc Medical US LLC · Catheter, hemodialysis, implanted1 recall62 reports
- Palindrome 8888123410PMozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls9 reports
- Palindrome 8888145040PMozarc Medical US LLC · Catheter, hemodialysis, implanted1 recall1 death58 reports
- Palindrome 8888123400PMozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls5 reports
- Palindrome 8888128454PMozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls5 reports
- Palindrome 8888133504PMozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls3 reports
- Palindrome 8888119365PMozarc Medical US LLC · Catheter, hemodialysis, implanted, coated2 recalls2 reports
- Palindrome 8888123405PMozarc Medical US LLC · Catheter, hemodialysis, implanted, coated2 recalls2 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 94266).
- Z-1635-2024 · Open, ClassifiedPalindrome Precision H Chronic Catheter Kit 14.5 Fr/Ch (4.8mm) x 23 cm (Symmetrical Tip, Heparin Coating and Tal VenaTrac Stylet- Intended for acute and chronic hemodialysis, aph…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page