Palindrome recallZ-1635-2024 · Covidien LP · initiated Mar 15, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Covidien LP or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Products may not be heparin coated as indicated on the label-leads to a potential loss of benefit due to the absence of the heparin coating

FDA root cause category: Process control

What the company is doing

Medtronic issued Urgent Medical Device Recall Letter on 3/15/24. Letter states reason for recall, health risk and action to take: 1.Immediately quarantine and discontinue use of unused affected product (see Attachment A). 2. Return all unused products from your inventory to Medtronic. Please contact: [email protected] for the Return Good Authorization (RGA). a.Credit for the returned affected product will be issued based on the RGA number. b. If purchased from a distributor, contact your distributor directly to arrange for the return of the product back to your distributor. 3. In addition, please complete and return the enclosed Customer Confirmation Form to [email protected] even if you do not have unused inventory. 4. This notice should be passed on to those who need to be aware within your organization or to any organization including, but not limited to, nephrologists, intensivists, implanting and managing physicians, renal nurses, critical care nurses, or other dialysis staff where the potentially affected devices have been transferred. Please maintain a copy of this notice in your records. If you have questions regarding this communication, please contact your local representative or Customer Service at 800-962-9888, option 2 (Monday Friday 8:00 a.m. to 6:30 p.m.).

Recall details

Recall number
Z-1635-2024
Status
Open, Classified
Initiated by the company
Mar 15, 2024
Posted by the FDA
Apr 24, 2024
Recalling firm
Covidien LP
Quantity
878 units
Product code
MSD
FDA submissions
K123196
Distribution
Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canary Islands, Denmark, Faroe Islands, France, Germany, Greece, Iceland, Ireland, Italy, Mayotte, Norway, Poland, Reunion, South Africa, Spain, Sweden, United Kingdom.

Product description

Palindrome Precision H Chronic Catheter Kit 14.5 Fr/Ch (4.8mm) x 23 cm (Symmetrical Tip, Heparin Coating and Tal VenaTrac Stylet- Intended for acute and chronic hemodialysis, apheresis, and infusion. Model Number (CFN): 8888145044CP

Code information (models, lots, serial numbers)

GTIN: 20884521158082, 10884521158085 Lot Number: 2228000089, 2201700103, 2228000074

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

69 devices cleared under K123196 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.