Biomet SpinalPak / OrthoPak recallZ-1618-2024 · Ebi, LLC · initiated Mar 22, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Ebi, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Damage to the lead wire sheath (black cable) component and causing exposure of the wires, potentially lead to no therapeutic treatment signal being delivered to the electrodes, and result in delay of treatment
FDA root cause category: Device Design
What the company is doing
ZimVie issued Urgent Medical Device Correction letter to Patients, Physicians, Sales Representatives on 3/22/24 . Letter states reason for recall, health risk and action to take: 1. Review this notification for awareness of the contents. 2. Before each use, inspect your device s lead wires (black cable) for damage. If found to be damaged, contact Customer Service at 1-800-526-2579 x 6000 for a complimentary replacement. 3. Complete the Acknowledgement and Receipt Form either by scanning the QR code on the postcard included with this letter and completing the online form or mailing back the completed postcard. To complete your Acknowledgement, you will need to include your Unique Patient Reference Number included on the top of the first page of this letter. This acknowledgement should be completed even if you do not have affected product. 4. If you have further questions or concerns after reviewing this notice, please call customer service at 1-800-526-2579 x 6000. Customer Service Representatives are available from 8:30 am to 5:00 pm (EST), Monday through Friday. At other times, please leave a clear message for a return call by the next business day. Alternatively, your questions may be emailed to [email protected].
Recall details
- Recall number
- Z-1618-2024
- Status
- Open, Classified
- Initiated by the company
- Mar 22, 2024
- Posted by the FDA
- Apr 23, 2024
- Recalling firm
- Ebi, LLC
- Quantity
- 829 units (2 lead wires assembly)
- Product code
- LOF
- FDA submissions
- P850022
- Distribution
- US Nationwide distribution.
Product description
20" Lead Wires, Replacement Part Number: 1067724-2
Code information (models, lots, serial numbers)
UDI-DI: 00812301020300 All devices distributed since May 1, 2023
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Biomet SpinalPak / OrthoPak 1067724-2Ebi, LLC · Stimulator, Bone Growth, Non-Invasive1 recall4 reports
Product lines
Other devices under the same FDA submission
11 devices cleared under P850022 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Biomet SpinalPak Non-invasive Spine Fusion Stimulator System 1067716Ebi, LLC · Stimulator, Bone Growth, Non-Invasive1 recall1 death262 reports
- Biomet SpinalPak / OrthoPak 106130-20Ebi, LLC · Stimulator, Bone Growth, Non-Invasive1 death188 reports
- Biomet SpinalPak / OrthoPak 106130-22Ebi, LLC · Stimulator, Bone Growth, Non-Invasive101 reports
- Biomet OrthoPak Non-invasive Bone Growth Stimulator System 1067718Ebi, LLC · Stimulator, Bone Growth, Non-Invasive1 recall41 reports
- Biomet SpinalPak / OrthoPak 106130-17Ebi, LLC · Stimulator, Bone Growth, Non-Invasive52 reports
- Biomet SpinalPak / OrthoPak 1067724-4Ebi, LLC · Stimulator, Bone Growth, Non-Invasive1 recall2 reports
- Biomet SpinalPak Non-invasive Spine Fusion Stimulator System 1067717Ebi, LLC · Stimulator, Bone Growth, Non-Invasive22 reports
- Biomet SpinalPak / OrthoPak 106130-12Ebi, LLC · Stimulator, Bone Growth, Non-Invasive13 reports
- Biomet OrthoPak Non-invasive Bone Growth Stimulator System 1067719Ebi, LLC · Stimulator, Bone Growth, Non-Invasive1 report
- Biomet SpinalPak / OrthoPak 1067724-5Ebi, LLC · Stimulator, Bone Growth, Non-Invasive1 report
- Biomet SpinalPak / OrthoPak 1067728-1Ebi, LLC · Stimulator, Bone Growth, Non-Invasive1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 94291).
- Z-1616-2024 · Open, ClassifiedBiomet SpinalPak Non-invasive Spine Fusion Stimulator System- Indicated for the treatment of an established nonunion acquired secondary to trauma, excluding vertebrae and all flat …
- Z-1617-2024 · Open, ClassifiedBiomet OrthoPak Non-invasive Bone Growth Stimulator System-Indicated for the treatment of an established nonunion acquired secondary to trauma, excluding vertebrae and all flat bon…
- Z-1619-2024 · Open, Classified48" Lead Wires, Replacement Part Number: 1067724-4
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page