Skip to content

Vercise Genus recallZ-1890-2024 · Boston Scientific Neuromodulation Corporation · initiated Apr 18, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Boston Scientific Neuromodulation Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Deep Brain Stimulation (DBS) Implantable Pulse Generator (IPG) may experience routine system check during IPG charging, which may cause device reset. The device reset could lead transient loss of stimulation; patients may experience undesired sensations, transient worsening of movement disorder symptoms, which may lead patient to request surgical intervention for replacement or revision.

FDA root cause category: Device Design

What the company is doing

On 4/18/24, correction notices were mailed to Physicians/Surgeons, Hospitals, Healthcare Professionals who were asked to do the following: 1) Review any patient reports of undesired sensations indicative of potential device reset behavior during IPG charging and report these observations accordingly to the firm. 2) After firm verifies this device reset behavior has occurred during IPG charging, an in-field IPG firmware update can be scheduled for the patient s device. This will eliminate the possibility of a coincident routine device system check during IPG charging, thus preventing a potential system reset. Any adverse events or quality concerns associated with use of this product should be reported to firm via email [email protected]

Recall details

Recall number
Z-1890-2024
Status
Open, Classified
Initiated by the company
Apr 18, 2024
Posted by the FDA
May 24, 2024
Recalling firm
Boston Scientific Neuromodulation Corporation
Quantity
14,977
Product code
NHL
FDA submissions
P150031
Distribution
Worldwide - US Nationwide distribution including in the states of TX, WA, MT, GA, TN, MA, OR, ID, CA, FL, WI, UT, DE, AZ, PA, DC, LA, KY, VA, MN, NC, OH, NY, MI, CO, IN, MD, NV, AL, IL, AR, SC, OK, MO, MS, NJ, ME, Hi, KS, SD, WV, NM, AK, CT, WY, NE, ND and the countries of Canada, Argentina, Chile, Colombia, Brazil, Algeria, Austria, Belgium, Bulgaria, Croatia, Denmark, Finland, France, Germany, Great Britain, Greece, Hungary, Iceland, Ireland, Italy, Jordan, Kazakhstan, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Saudi Arabia, Serbia, Slovakia, Spain, Sweden, Switzerland, Turkey, Ukraine, United Arab Emirates, Australia, Thailand, Singapore, Hong Kong, Korea, India, Japan.

Product description

Vercise Genus Deep Brain Stimulation (DBS) Implantable Pulse Generator IPG: R16 IPG KIT, REF: M365DB12160, and R32 IPG KIT, REF: M365DB12320

Code information (models, lots, serial numbers)

REF/UDI-DI(GTIN)/Serial Number Range/Expiration Date Range: M365DB12160/8714729985044/100209 - 753347/09-OCT-2020 through 26-MAR-2026; M365DB12320/8714729985051/100104 - 753200/09-OCT-2020 through 21-MAR-2026

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

32 devices cleared under P150031 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.