Vercise Genus recallZ-1890-2024 · Boston Scientific Neuromodulation Corporation · initiated Apr 18, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Boston Scientific Neuromodulation Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Deep Brain Stimulation (DBS) Implantable Pulse Generator (IPG) may experience routine system check during IPG charging, which may cause device reset. The device reset could lead transient loss of stimulation; patients may experience undesired sensations, transient worsening of movement disorder symptoms, which may lead patient to request surgical intervention for replacement or revision.
FDA root cause category: Device Design
What the company is doing
On 4/18/24, correction notices were mailed to Physicians/Surgeons, Hospitals, Healthcare Professionals who were asked to do the following: 1) Review any patient reports of undesired sensations indicative of potential device reset behavior during IPG charging and report these observations accordingly to the firm. 2) After firm verifies this device reset behavior has occurred during IPG charging, an in-field IPG firmware update can be scheduled for the patient s device. This will eliminate the possibility of a coincident routine device system check during IPG charging, thus preventing a potential system reset. Any adverse events or quality concerns associated with use of this product should be reported to firm via email [email protected]
Recall details
- Recall number
- Z-1890-2024
- Status
- Open, Classified
- Initiated by the company
- Apr 18, 2024
- Posted by the FDA
- May 24, 2024
- Recalling firm
- Boston Scientific Neuromodulation Corporation
- Quantity
- 14,977
- Product code
- NHL
- FDA submissions
- P150031
- Distribution
- Worldwide - US Nationwide distribution including in the states of TX, WA, MT, GA, TN, MA, OR, ID, CA, FL, WI, UT, DE, AZ, PA, DC, LA, KY, VA, MN, NC, OH, NY, MI, CO, IN, MD, NV, AL, IL, AR, SC, OK, MO, MS, NJ, ME, Hi, KS, SD, WV, NM, AK, CT, WY, NE, ND and the countries of Canada, Argentina, Chile, Colombia, Brazil, Algeria, Austria, Belgium, Bulgaria, Croatia, Denmark, Finland, France, Germany, Great Britain, Greece, Hungary, Iceland, Ireland, Italy, Jordan, Kazakhstan, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Saudi Arabia, Serbia, Slovakia, Spain, Sweden, Switzerland, Turkey, Ukraine, United Arab Emirates, Australia, Thailand, Singapore, Hong Kong, Korea, India, Japan.
Product description
Vercise Genus Deep Brain Stimulation (DBS) Implantable Pulse Generator IPG: R16 IPG KIT, REF: M365DB12160, and R32 IPG KIT, REF: M365DB12320
Code information (models, lots, serial numbers)
REF/UDI-DI(GTIN)/Serial Number Range/Expiration Date Range: M365DB12160/8714729985044/100209 - 753347/09-OCT-2020 through 26-MAR-2026; M365DB12320/8714729985051/100104 - 753200/09-OCT-2020 through 21-MAR-2026
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- Vercise Genus™ DB-1216Boston Scientific Neuromodulation Corporation · Stimulator, electrical, implanted, for essential tremor2 recalls23 deaths584 reports
- Vercise Genus™ DB-1232Boston Scientific Neuromodulation Corporation · Stimulator, electrical, implanted, for parkinsonian symptoms2 recalls2 deaths48 reports
Product lines
Other devices under the same FDA submission
32 devices cleared under P150031 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Vercise™ Cartesia™ DB-2202-45Boston Scientific Neuromodulation Corporation · STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS18 deaths718 reports
- Vercise Genus™ DB-1416Boston Scientific Neuromodulation Corporation · Stimulator, electrical, implanted, for parkinsonian tremor6 deaths240 reports
- NA NM-3138-55Boston Scientific Neuromodulation Corporation · STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS198 reports
- SureTek™ DB-4600-CBoston Scientific Neuromodulation Corporation · Stimulator, electrical, implanted, for essential tremor1 recall138 reports
- Vercise Gevia™ DB-1200-SBoston Scientific Neuromodulation Corporation · STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS5 deaths170 reports
- Vercise™ DB-1110-CBoston Scientific Neuromodulation Corporation · STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS98 reports
- Vercise™ Cartesia™ DB-2202-30Boston Scientific Neuromodulation Corporation · STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS1 death95 reports
- SureTek™ DB-4605-CBoston Scientific Neuromodulation Corporation · STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS1 recall5 reports
- NA DB-2201-45DCBoston Scientific Neuromodulation Corporation · STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS2 deaths53 reports
- NA DB-3128-55BBoston Scientific Neuromodulation Corporation · Stimulator, electrical, implanted, for essential tremor38 reports
- Vercise™ Cartesia™ X DB-2203-45Boston Scientific Neuromodulation Corporation · Stimulator, electrical, implanted, for parkinsonian symptoms3 deaths30 reports
- Vercise™ DB-9218-15Boston Scientific Neuromodulation Corporation · Stimulator, electrical, implanted, for essential tremor25 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page