Skip to content

EchoTip recallZ-1833-2024 · Cook Medical Incorporated · initiated Apr 1, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Cook Medical Incorporated or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Devices may contain elevated levels of bacterial endotoxin.

FDA root cause category: Nonconforming Material/Component

What the company is doing

On April 01, 2024, Cook Ireland Ltd.(CIRL) initiated a field action for the EchoTip Ultra High Definition Ultrasound Access Needle (ECHO-HD-19-A) and EchoTip Ultra Endoscopic Ultrasound Needle (ECHO-19). Cook Ireland Ltd has identified 9 devices located outside of the United States (US) that are associated with the issue and has initiated a voluntary recall due to the potential of elevated levels of bacterial endotoxins on the devices. No US customers were impacted by this Field Safety Corrective Action (FSCA). No monitoring will be required as US consignees are not impacted by this recall. Effectiveness checks for this recall will not be necessary as US Consignees were not impacted by the distribution.

Recall details

Recall number
Z-1833-2024
Status
Open, Classified
Initiated by the company
Apr 1, 2024
Posted by the FDA
May 16, 2024
Recalling firm
Cook Medical Incorporated
Quantity
8 units OUS
Product code
FCG
FDA submissions
K210476
Distribution
International distribution in the countries of Germany, Netherlands, Switzerland, United Kingdom.

Product description

EchoTip Ultra Endoscopic Ultrasound Access Needle, ECHO-19, G31520

Code information (models, lots, serial numbers)

Catalogue Number: ECHO-19; GPN: G31520; UDI/DI: 00827002315204; Lot Numbers: C2145003 and C2144407.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 94395).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page