EchoTip G31520Recalls & reported problems
Cook Ireland Ltd · Echotip Ultra, Endoscopic Ultrasound Needle · Class 2
All models of EchoTip (7)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Cook EchoTip ultrasound biopsy needle is used to collect tissue samples during endoscopic ultrasound procedures. The 84 reports mostly describe the needle breaking, kinking, or bending while in use, often needing to be pulled out and a new one used; there are no deaths and 9 injuries. One open recall, begun in April 2024, covers devices that may contain elevated levels of bacterial endotoxin.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 84
- Deaths reported
- 0
- Injuries reported
- 9
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 84 adverse event reports that list this exact device (DI 00827002315204), received between Feb 16, 2016 and Aug 11, 2025 — classified as 9 injuries, 75 malfunctions.
The most frequently reported specific device problem is “Break” (38 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Apr 1, 2024Class IIOpen, Classified
Devices may contain elevated levels of bacterial endotoxin.
Z-1833-2024 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
Most reported problems
Device problems
- Break38
- Retraction Problem15
- Material Twisted/Bent13
- Material Deformation10
- Material Perforation5
- Fracture4
- Kinked4
- Bent3
Patient problems
- No Known Impact Or Consequence To Patient51
- No Consequences Or Impact To Patient34
- Foreign Body In Patient2
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 25 most recent of 84 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 11, 2025Fracture; Fracture
THE CASE WAS DETERMINED TO BE REPORTABLE ON 18 JUL 2025 BASED ON THE FOLLOWING ENGINEERING INPUT AFTER THE LAB EVALUATION: "IF A DISTAL KINK OR BREAK HAD BEEN PRESENT, THE ISSUE COULD HAVE BEEN ATTRIBUTED TO USE ERROR. HOWEVER, SINCE ONLY A PROXIMAL BREAK IS EVIDENT, THIS COMPLAINT WILL BE ASSESSED AS A PROXIMAL NEEDLE BREAK. USER WAS CONDUCTING EUS PANCREAS, PUNCTURE PANCREAS BODY FOR FIRST ATTEMPT, THEN SLOWLY RETRACTED THE NEEDLE BUT FOUND OUT NO CELL OBTAINED. USER RETRACTED THE NEEDLE FROM ENDOSCOPE WORKING CHANNEL, THEN RE-INSTALLED THE NEEDLE ONTO ENDOSCOPE WHILE ENCOUNTERED RESISTANCE DURING INSERTING INTO ENDOSCOPE. USER FOUND OUT THE NEEDLE KINK AND CHANGED TO ANOTHER ONE TO COMPLETE THE PROCEDURE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE. 1. CAN THE FOLLOWING BE CONFIRMED WITH THE CUSTOMER? COULD YOU PLEASE CLARIFY WHETHER THE NEEDLE WAS KINKED AT THE PROXIMAL END, THE DISTAL END, OR BOTH? THE IMAGES SUGGEST THAT KINKS MAY BE PRESENT AT BOTH LOCATIONS. PROXIMAL E
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22769653 ↗MalfunctionFeb 14, 2020Fracture; Fracture
COMPLETED CUSTOMER REPORTING FORM AND ADDITIONAL INFORMATION WAS RECEIVED ON 17-JAN-2020 INDICATING NEEDLE BREAKAGE. THE COMPLAINT DEVICE WAS EVALUATED ON (B)(6) 2020 AND THE NEEDLE WAS CONFIRMED TO BE BROKEN BELOW THE SHEATH EXTENDER. DURING DECOMPRESSION OF THE CYST, THE NEEDLE BROKE AFTER PASSING THROUGH THE CANAL AND TRYING TO OPEN THE NEEDLE IN THE STOMACH. THE NEEDLE WAS NO LONGER HIDDEN IN THE COVER.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9710412 ↗MalfunctionJan 24, 2020Break; Fracture; No Apparent Adverse Event; Break; Fracture; No Apparent Adverse Event
DURING ECHO ENDOSCOPY WITH BIOPSY, WHEN STAFF WANTED TO MAKE A PONCTION, DIFFICULTY TO GET THE NEEDLE OUT OF THE ECHOENDO. NEEDLE BROKE OFF INTO ECHOENDOSCOPE. WHEN NEEDLE WAS RETRIEVED STAFF NOTICED A LACK OF BEVELLED EDGE. CONSEQUENCES: LONGER ANESTHESIA. USE OF NEW NEEDLE. EXTRA FINANCIAL COST.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 9625563 ↗MalfunctionJan 16, 2020Material Twisted/Bent; Material Twisted/Bent
THEY CHECKED THAT MASS LOCATION WAS BODY OF PANCREAS BEFORE THE PROCEDURE IN EUS. THEY ADVANCED ECHO-19 TO MASS IN THROUGH THE SCOPE. HOWEVER, THE NEEDLE DIDN'T FUNCTURE TO MASS IN EUS. CONSEQUENTLY, THE NEEDLE REMOVED OUT OF THE SCOPE.SO THEY USED ANOTHER ECHO-19. IT WAS INDICATED THAT THE DISTAL END OF THE NEEDLE WAS KINKED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9596216 ↗MalfunctionDec 19, 2019Material Twisted/Bent; Material Twisted/Bent
DISTAL TIP OF NEEDLE DEFORMED OBSERVED DURING THE LAB EVALUATION OF (B)(4). RELATED TO MDR REF#: 3001845648-2019-00594.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9494096 ↗MalfunctionNov 25, 2019Break; Retraction Problem; Break; Retraction Problem
ON (B)(6) 2019 THE PATIENT UNDERWENT PUNCTURE ECHOENDOSCOPE DUE TO OESOPHAGEAL SUB EPITHELIAL LESION. THE ECHO NEEDLE WAS INTRODUCED THROUGH THE WORKING CHANNEL OF THE DEVICE AND EXPOSED TO THE SHEATH POSITIONED FOR THE ECHO GUIDED PUNCTURE, AND WHEN THE NEEDLE WAS INSERTED INTO THE LESION, A CLICK WAS HEARD AND THE NEEDLE DID NOT PROGRESS PROPERLY. AN ATTEMPT WAS MADE TO RETRACT THE NEEDLE INTO THE SHEATH WITHOUT SUCCESS. THE NEEDLE WAS REMOVED AND ANOTHER ONE WITH THE SAME SPECIFICATIONS WAS USED TO COMPLETE THE PROCEDURE UNEVENTFULLY. AFTER EXAMINATION, THE NEEDLE WAS EXAMINED AND IT WAS FOUND THAT THE SHEATH HAD BROKEN IN ITS PROXIMAL PORTION AND THAT THE MOVEMENTS OF THE NEEDLE HANDLE DID NOT MOVE IT FORWARD OR BACKWARD.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9368352 ↗MalfunctionNov 14, 2019Break; Retraction Problem; Activation, Positioning or Separation Problem; Break; Retraction Problem; Activation, Positioning or Separation Problem
DURING THE PROCEDURE THE DOCTOR NOTICED THE 19G NEEDLE WAS RATHER STIFF (DIDN`T MOVE EASILY IN AND OUT OF THE SHEET). WHILE PUNCTURING (2ND PASS) SOMETHING SNAPPED AND ON THE EUS SCREEN THE NOTICED THE NEEDLE SEEMED TO HAVE BROKEN. THEY REMOVED THE NEEDLE AND EXCHANGED IT FOR A 3-22. THEY FINISHED THE PROCEDURE WITHOUT ANY PROBLEMS.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 9322055 ↗MalfunctionNov 13, 2019Material Twisted/Bent; Material Twisted/Bent
AS REPORTED TO CUSTOMER RELATIONS ON (B)(6) 2019: "EUS POSSIBLE DRAINAGE OR ACCESS RENDEZVOUS." ADDITIONAL INFORMATION PROVIDED BY DM ON 16-OCT-2019: 'PATIENT (B)(6) FEMALE WITH SEVERE CASE OF PANCREATITIS AND COMPARTMENT SYNDROME PRESENTED WITH PANCREATIC SUDO CYST. PATIENT WAS SCHEDULED FOR AN EUS GUIDED CYSTGASTROSTOMY. PER USUAL DR (B)(6) WILL USE THE 19G ECHO TIP COOK NEEDLE FOR ACCESS OF THE CYST UNDER EUS GUIDANCE. DR (B)(6) PASSED THE SCOPE AND QUICKLY IDENTIFIED THE CYST. WE REMOVED THE NEEDLE FROM ITS PACKAGING WITH NO FORCE AND IT CAME OUT EASILY. I SET THE SHEATH TO DR (B)(6) PREFERRED SETTING AND TIGHTEN THE KNOBS AS I USUALLY DO AND HANDED IT TO HIM. THE NEEDLE ATTACHED TO OUR GFUC 180 OLYMPUS LINEAR SCOPE WITH NO TROUBLE. AS DR (B)(6) WAS LINING UP TO STICK THE CYST HE HAD MORE RESISTANCE THAN USUAL SO HE HAD ME CHECK THE NEEDLE AS IT USUALLY WORKS WITH MORE EASE. AS DR (B)(6) ATTEMPTS TO BRING THE NEEDLE BACK INTO THE PROTECTIVE SHEATH HE FELT A SNAP SO HE QUICKLY REMOVED THE CATHETER FROM THE SCOPE. WE ALL NOTICED THE NEEDLE WAS STILL EXPOSED AND HAD NEVER RETRACTED INTO THE SHEATH. WE ATTEMPTED TO BRING THE NEEDLE BACK INTO TO SHEATH WITH NO LUCK. WE WERE ABLE TO
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9315704 ↗MalfunctionNov 12, 2019Material Twisted/Bent; Material Twisted/Bent
THEY CHECKED THAT MASS LOCATION WAS BODY OF PANCREAS BEFORE THE PROCEDURE IN EUS. THEY ADVANCED ECHO-19 TO MASS IN THROUGH THE SCOPE. HOWEVER, THE NEEDLE DIDN'T PUNCTURE TO MASS IN EUS. CONSEQUENTLY, THE NEEDLE REMOVED OUT OF THE SCOPE. SO THEY USED ANOTHER ECHO-19.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9308979 ↗MalfunctionSep 10, 2019Material Perforation; Material Perforation
THE DEVICE WAS USED FOR EUS-FNA OF THE PANCREAS TUMOR SINCE PANCREATIC CANCER WAS SUSPECTED. APPROACH WAS GAINED FROM THE STOMACH TO THE PANCREATIC BODY. THE APPROACH, FROM THE STOMACH TO THE PANCREATIC BODY, WAS COMPARATIVELY SIMPLE, SO THE ENDOSCOPE WAS ALMOST NOT IN AN ANGLED POSITION. ONE ECHO-19/ LOT# C1332149 WAS INSERTED THROUGH THE ENDOSCOPE. WHEN THE PHYSICIAN MOVED THE HANDLE TO RETRACT THE NEEDLE AFTER NEEDLE ADVANCEMENT TO THE TARGET SITE, THE NEEDLE DID NOT MOVE AND COULD NOT BE RETRACTED INTO THE SHEATH ALTHOUGH THE HANDLE COULD BE MOVED. THEREFORE, THE DEVICE WAS REMOVED THROUGH THE ENDOSCOPIC CHANNEL WITH THE NEEDLE PROTRUDING OUT OF THE SHEATH AND REPLACED WITH ECHO-19/ LOT# C1483682. (B)(4) ONE ECHO-19/ LOT# C1483682 WAS INSERTED THROUGH THE ENDOSCOPE. WHEN THE PHYSICIAN ATTEMPTED PUNCTURE OF THE TARGET SITE, THE NEEDLE DID NOT MOVE AND COULD NOT ADVANCE OUT OF THE SHEATH. HE REPEATED MOVEMENT OF WITHDRAWAL AND ADVANCEMENT OF THE DEVICE THROUGH THE ENDOSCOPE SEVERAL TIMES, BUT THE NEEDLE WOULD NOT ADVANCE. HE REMOVED THE ENTIRE DEVICE THROUGH THE ENDOSCOPIC CHANNEL, THEN NOTICED THAT THE NEEDLE PERFORATED THE SHEATH AT A CERTAIN POINT. ALSO, THE TIP OF THE SHEATH
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8984284 ↗InjurySep 10, 2019Break; Break
THE NEEDLE BROKE DURING THE PROCEDURE. A CORRECTION REPORT IS BEING SUBMITTED TO CLASSIFY THIS REPORT UNDER SERIOUS INJURY. ADOPTING A CAUTIOUS APPROACH, SERIOUS INJURY REPORTING REQUIRED AS THE NEEDLE BROKE INSIDE THE PATIENT'S BODY AND THERE IS POTENTIALLY A REQUIREMENT FOR SURGICAL INTERVENTION. THE RISK OF THIS COMPLAINT WAS ASSESSED BASED ON A SEVERITY OF +3 (HARM REQUIRING MEDICAL INTERVENTION). IT SHOULD BE NOTED THAT THESE DETAILS DO NOT IMPACT THE EVALUATION CONCLUSIONS PREVIOUSLY SUBMITTED IN RELATION TO THIS EVENT.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 8984217 ↗MalfunctionSep 10, 2019Retraction Problem; Retraction Problem
THE DEVICE WAS USED FOR EUS-FNA OF THE PANCREAS TUMOR SINCE PANCREATIC CANCER WAS SUSPECTED. APPROACH WAS GAINED FROM THE STOMACH TO THE PANCREATIC BODY. THE APPROACH, FROM THE STOMACH TO THE PANCREATIC BODY, WAS COMPARATIVELY SIMPLE, SO THE ENDOSCOPE WAS ALMOST NOT IN AN ANGLED POSITION. ONE ECHO-19/ LOT# C1332149 WAS INSERTED THROUGH THE ENDOSCOPE. WHEN THE PHYSICIAN MOVED THE HANDLE TO RETRACT THE NEEDLE AFTER NEEDLE ADVANCEMENT TO THE TARGET SITE, THE NEEDLE DID NOT MOVE AND COULD NOT BE RETRACTED INTO THE SHEATH ALTHOUGH THE HANDLE COULD BE MOVED. THEREFORE, THE DEVICE WAS REMOVED THROUGH THE ENDOSCOPIC CHANNEL WITH THE NEEDLE PROTRUDING OUT OF THE SHEATH AND REPLACED WITH ECHO-19/ LOT# C1483682. (B)(4) - THIS COMPLAINT) ONE ECHO-19/ LOT# C1483682 WAS INSERTED THROUGH THE ENDOSCOPE. WHEN THE PHYSICIAN ATTEMPTED PUNCTURE OF THE TARGET SITE, THE NEEDLE DID NOT MOVE AND COULD NOT ADVANCE OUT OF THE SHEATH. HE REPEATED MOVEMENT OF WITHDRAWAL AND ADVANCEMENT OF THE DEVICE THROUGH THE ENDOSCOPE SEVERAL TIMES, BUT THE NEEDLE WOULD NOT ADVANCE. HE REMOVED THE ENTIRE DEVICE THROUGH THE ENDOSCOPIC CHANNEL, THEN NOTICED THAT THE NEEDLE PERFORATED THE SHEATH AT A CERTAIN POINT. ALSO, THE
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8984216 ↗MalfunctionAug 29, 2019Failure to Advance; Failure to Advance
NEEDLE CANNOT BE ADVANCED IN ENDOSCOPY. USER TRIED SEVERAL TIMES BUT FAILED. USER CHANGED ANOTHER SAME DEVICE TO COMPLETE THE BIOPSY. DEVICE EVALUATED AND A DISTAL KINK ON THE NEEDLE WAS OBSERVED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8944399 ↗MalfunctionAug 28, 2019Material Twisted/Bent; Material Twisted/Bent
THEY CHECKED THAT MASS LOCATION WAS BODY OF PANCREAS BEFORE THE PROCEDURE IN EUS. THEY ADVANCED ECHO-19 TO MASS IN THROUGH THE SCOPE. HOWEVER, THE NEEDLE DIDN'T FUNCTURE TO MASS IN EUS. CONSEQUENTLY, THE NEEDLE REMOVED OUT OF THE SCOPE. SO THEY USED ANOTHER ECHO-19. ADDITIONAL INFORMATION CONFIRMED THAT THE NEEDLE WAS KINKED AT THE DISTAL END.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8939199 ↗MalfunctionAug 21, 2019Break; Break
DEVICE WAS ADVANCED TO TARGET AREA AND 3 ATTEMPTS WERE COMPLETED. USER CANNOT ADVANCE THE NEEDLE AFTER THIRD ATTEMPT, AND FOUND OUT THE NEEDLE BROKEN UNDER ENDOSCOPY VIEW. USER CHANGED ANOTHER SAME DEVICE TO COMPLETE THE PROCEDURE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8915547 ↗MalfunctionAug 1, 2019Material Twisted/Bent; Material Twisted/Bent
THEY CHECKED THAT MASS LOCATION WAS BODY OF PANCREAS BEFORE THE PROCEDURE IN EUS. THEY ADVANCED ECHO-19 TO MASS IN THROUGH THE SCOPE. HOWEVER, THE NEEDLE DIDN'T FUNCTURE TO MASS IN EUS. CONSEQUENTLY, THE NEEDLE REMOVED OUT OF THE SCOPE.SO THEY USED ANOTHER ECHO-19.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8850175 ↗MalfunctionJul 25, 2019Material Twisted/Bent; Material Twisted/Bent
THEY CHECKED THAT MASS LOCATION WAS BODY OF PANCREAS BEFORE THE PROCEDURE IN EUS. THEY ADVANCED ECHO-19 TO MASS IN THROUGH THE SCOPE. HOWEVER, THE NEEDLE DIDN'T PUNCTURE TO MASS IN EUS. CONSEQUENTLY, THE NEEDLE WAS REMOVED OUT OF THE SCOPE. SO THEY USED ANOTHER ECHO-19 DEVICE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8826586 ↗MalfunctionJul 25, 2019Material Twisted/Bent; Material Twisted/Bent
THEY CHECKED THAT MASS LOCATION WAS BODY OF PANCREAS BEFORE THE PROCEDURE IN EUS. THEY ADVANCED ECHO-19 TO MASS IN THROUGH THE SCOPE. HOWEVER, THE NEEDLE DIDN'T PUNCTURE TO MASS IN EUS. CONSEQUENTLY, THE NEEDLE WAS REMOVED OUT OF THE SCOPE.SO THEY USED COMPETITOR EUS NEEDLE. (MEDIGLOBE 19G).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8826584 ↗MalfunctionJul 25, 2019Material Twisted/Bent; Material Twisted/Bent
THEY CHECKED THAT MASS LOCATION WAS BODY OF PANCREAS BEFORE THE PROCEDURE IN EUS. THEY ADVANCED ECHO-19 TO MASS IN THROUGH THE SCOPE. HOWEVER, THE NEEDLE DIDN'T PUNCTURE TO MASS IN EUS. CONSEQUENTLY, THE NEEDLE WAS REMOVED OUT OF THE SCOPE. SO THEY USED ANOTHER ECHO-19 DEVICE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8826553 ↗MalfunctionJul 18, 2019Material Twisted/Bent; Material Twisted/Bent
THEY CHECKED THAT MASS LOCATION WAS BODY OF PANCREAS BEFORE THE PROCEDURE IN EUS. THEY ADVANCED ECHO-19 TO MASS IN THROUGH THE SCOPE. HOWEVER, THE NEEDLE DIDN'T PUNCTURE TO MASS IN EUS. CONSEQUENTLY, THE NEEDLE REMOVED OUT OF THE SCOPE. SO THEY USED ANOTHER ECHO-19. DEVICE RELATED TO THIS COMPLAINT WAS EVALUATED ON 20-JUN-2019. THE TIP OF THE NEEDLE WAS NOTED TO BE DEFORMED - OVERALL RISK ASSESSED AS CATEGORY IIB (MODERATE). NO ADVERSE EFFECTS TO THE PATIENT WAS REPORTED AS OCCURRING. FDA MDR REPORTING REQUIRED AS THE RISK IS NOT CONSIDERED TO BE 'REMOTE'.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8802338 ↗MalfunctionJul 18, 2019Material Twisted/Bent; Material Twisted/Bent
THEY CHECKED THAT MASS LOCATION WAS BODY OF PANCREAS BEFORE THE PROCEDURE IN EUS. THEY ADVANCED ECHO-19 TO MASS IN THROUGH THE SCOPE. HOWEVER, THE NEEDLE DIDN'T PUNCTURE TO MASS IN EUS. CONSEQUENTLY, THE NEEDLE REMOVED OUT OF THE SCOPE. SO THEY USED ANOTHER ECHO-19. DEVICE RELATED TO THIS COMPLAINT WAS EVALUATED ON 20-JUN-2019. THE TIP OF THE NEEDLE WAS NOTED TO BE DEFORMED - OVERALL RISK ASSESSED AS CATEGORY IIB (MODERATE). NO ADVERSE EFFECTS TO THE PATIENT WAS REPORTED AS OCCURRING. FDA MDR REPORTING REQUIRED AS THE RISK IS NOT CONSIDERED TO BE 'REMOTE'.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8802281 ↗InjuryJul 12, 2019Break; Break
ABOUT 3 INCHES BROKE OFF OF THE NEEDLE INSIDE A PATIENT. THE NEEDLE WAS USED TO ACCESS THE CYST-GASTRIC CAVITY. THE PHYSICIAN WAS ABLE TO RETRIEVE THE PIECE THAT WAS LEFT IN THE PATIENT. STAFF NOR PHYSICIAN NOTED ANYTHING UNTOWARD WITH THE NEEDLE DURING SET-UP OR INITIAL USE.
Patient problems: Foreign Body In Patient; No Consequences Or Impact To Patient
MAUDE report 8783927 ↗MalfunctionJun 13, 2019Retraction Problem; Retraction Problem
DURING THE HEPATOGASTROSTOMY PROCEDURE, THE NEEDLE ECHO-19 WAS USED. THE NEEDLE WAS CHECKED BEFORE THE PROCEDURE. BOTH THE PACKAGING AND THE SHEATH WERE INTACT. AFTER USE, THE NEEDLE WAS WITHDRAWN AND THE NEEDLE-OUT MARKER PLACED IN AND LOCKED INTO THE "ZERO" POSITION. AFTER REMOVING THE NEEDLE FROM THE ENDOSCOPE CHANNEL, IT TURNED OUT THAT THE TIP OF THE NEEDLE DID NOT HIDE INTO THE SHEATH, SO THAT WHEN IT WAS REMOVED, THE ENDOSCOPE'S BIOPSY CANAL WAS DAMAGED. THE ENDOSCOPE HAS BEEN SENT BACK FOR REPAIR.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8695632 ↗MalfunctionMay 23, 2019Material Twisted/Bent; Material Twisted/Bent
WHEN DOCTOR TRIED TO PUNCTURE THE PREUDOCYST, IT DID NOT PUNCTURE. THE DOCTOR REMOVED NEEDLE FROM THE SCOPE AND FOUND THAT NEEDLE WAS BEND FROM THE TIP. DOCTOR USED ANOTHER ECHO-19 TO COMPLETE THE PROCEDURE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8636621 ↗MalfunctionMay 16, 2019Material Twisted/Bent; Material Twisted/Bent
AS PER SOURCE DOCUMENT: IMPOSSIBLE TO ADVANCE THE NEEDLE. AS PER COMPLAINT FORM: "IT IS IMPOSSIBLE TO GET THE NEEDLE INTO THE LESION."
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8615764 ↗
FDA clearance and classification
The FDA regulates product code FCG (biopsy needle) under 21 CFR 876.1075 as a Class II device. 21 CFR 876.1075 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the EchoTip G31520
Has the EchoTip G31520 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-1833-2024 (Class II), began on Apr 1, 2024: Devices may contain elevated levels of bacterial endotoxin. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the EchoTip G31520?
The FDA's MAUDE database lists 84 adverse event reports for this device, received between Feb 16, 2016 and Aug 11, 2025. In those reports, the most common specific device problems are “Break” (38), “Retraction Problem” (15) and “Material Twisted/Bent” (13); the most common patient problem is “Foreign Body In Patient” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the EchoTip G31520?
None of the reports are classified as deaths, 9 as injuries and 75 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the EchoTip G31520, and is it still sold?
The FDA's device database lists Cook Ireland Ltd as the company responsible for it (brand EchoTip, model G31520). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Biopsy needle devices
- MAGNUM MN1410Bard Peripheral Vascular, Inc. · Biopsy needle1 recall119 reports
- Single Use Aspiration Needle NA-U401SX NA-U401SX-4021Olympus Medical Systems Corp. · Biopsy needle1 recall94 reports
- Single Use Aspiration Needle NA-201SX-4021Olympus Medical Systems Corp. · Biopsy needle1 death126 reports
- EchoTip ProCore G34785Cook Ireland Ltd · Biopsy needle97 reports
- MAGNUM MN1820Bard Peripheral Vascular, Inc. · Biopsy needle1 recall37 reports
- Single Use Aspiration Needle NA-U401SX NA-U401SX-4022Olympus Medical Systems Corp. · Biopsy needle1 recall28 reports
- Interject M00518250Boston Scientific Corporation · Biopsy needle70 reports
- EchoTip G52012Cook Ireland Ltd · Biopsy needle1 recall18 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page