MAQUET CARDIOHELP recallZ-1831-2024 · Maquet Medical Systems USA · initiated Apr 15, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Maquet Medical Systems USA or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The firm identified that the measured patient leakage current in some devices significantly exceeded expected levels.
FDA root cause category: Under Investigation by firm
What the company is doing
Consignees received an "URGENT MEDICAL DEVICE - Removal" email, dated 4/15/2024, notifying them of this recall event. Consignees are asked to examine their inventory to identify if they possess any affected devices and all consignees are asked to return the provided Response Form via email to identify where affected devices are located. Consignees that report having any affected devices will be contacted by Getinge to arrange for return of the affected device to upgrade the system's sensor panel and perform testing to ensure that leakage current is within specifications. Technical Support can also be contacted at 888-943-8872 (option 4, option 2, option 1) Monday through Friday between 8:00 AM and 6:00 PM EST to make service arrangements. Consignees are asked to not use affected devices for any additional patients until it has been cleared by the MCP/Getinge system check, repaired, or replaced. The provided recall notice is to be forwarded to users of the CARDIOHELP-i system and if the device was further distributed.
Recall details
- Recall number
- Z-1831-2024
- Status
- Open, Classified
- Initiated by the company
- Apr 15, 2024
- Posted by the FDA
- May 16, 2024
- Recalling firm
- Maquet Medical Systems USA
- Quantity
- 148 units (US: 5; OUS: 143)
- Product code
- DTQ
- FDA submissions
- K133598
- Distribution
- US Nationwide distribution in the states of NJ & TX.
Product description
CARDIOHELP-i System. Blood oxygenation and carbon dioxide removal system.
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
Part No. 70104.8012 and 70104.8012R; UDI-DI: 04037691658384; Serial no. 90410021, 90410029, 90410165, 90410254, 90410255, 90410024, 90410025, 90410026, 90410028, 90410030, 90410031, 90410034, 90410035, 90410036, 90410043, 90410044, 90410045, 90410046, 90410047, 90410048, 90410049, 90410050, 90410051, 90410052, 90410053, 90410054, 90410055, 90410056, 90410057, 90410058, 90410059, 90410060, 90410061, 90410062, 90410063, 90410064, 90410065, 90410067, 90410068, 90410069, 90410070, 90410071, 90410072, 90410073, 90410074, 90410075, 90410076, 90410077, 90410078, 90410079, 90410080, 90410081, 90410082, 90410084, 90410085, 90410086, 90410087, 90410088, 90410089, 90410091, 90410093, 90410094, 90410095, 90410096, 90410097, 90410098, 90410099, 90410100, 90410101, 90410102, 90410103, 90410104, 90410105, 90410106, 90410107, 90410108, 90410109, 90410110, 90410111, 90410112, 90410113, 90410114, 90410115, 90410116, 90410117, 90410118, 90410119, 90410120, 90410121, 90410122, 90410123, 90410124, 90410125, 90410126, 90410127, 90410130, 90410131, 90410132, 90410133, 90410134, 90410135, 90410136, 90410137, 90410138, 90410139, 90410140, 90410141, 90410142, 90410143, 90410144, 90410146, 90410147, 90410148, 90410149, 90410151, 90410152, 90410153, 90410154, 90410155, 90410156, 90410157, 90410158, 90410159, 90410160, 90410161, 90410162, 90410163, 90410166, 90410167, 90410168, 90410169, 90410170, 90410172, 90410173, 90410174, 90410175, 90410176, 90410177, 90410178, 90410179, 90410180, 90410181, 90410182, 90410183, 90410184, 90410185, 90410186, 90410187.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- MAQUET CARDIOHELP CARDIOHELP SystemMaquet Cardiopulmonary AG · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS4 recalls18 deaths107 reports
Product lines
Other devices under the same FDA submission
4 devices cleared under K133598 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- MAQUET CARDIOHELP CARDIOHELP-iMaquet Cardiopulmonary AG · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS5 recalls2 deaths64 reports
- MAQUET CARDIOHELP BS 3/8x3/32 L1.7Maquet Cardiopulmonary AG · Console, Heart-Lung Machine, Cardiopulmonary Bypass2 recalls0 reports
- MAQUET CARDIOHELP CARDIOHELP Emergency Drive (w/o holder)Maquet Cardiopulmonary AG · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS1 recall0 reports
- MAQUET CARDIOHELP CARDIOHELP Base UnitMaquet Cardiopulmonary AG · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS2 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page