Synchro .014"guide wire, 35 cm tip, Catalog: M00313010 recallZ-1909-2024 · Stryker Neurovascular · initiated Apr 23, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Stryker Neurovascular or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Neurovascular devices did not meet the minimum sample size required for endotoxin testing during manufacturing processes and are out of compliance with the AAMI ST72 standard, so these devices may exceed the acceptable limits of endotoxin levels; high levels of endotoxins may increase the risk of fever, shock and aseptic meningitis.
FDA root cause category: Process design
What the company is doing
On 4/23/24, recall notices were mailed to Supply Chain Management, Recall Coordinators, and Inventory Managers who were asked to do the following: 1) Cease further distribution or use of all affected devices. 2) Segregate the affected units in a secure location for return to the firm. 3) Circulate this Recall-Removal notice internally to all interested/affected parties. 4) Maintain awareness of this communication internally until all required actions have been completed within your facility. 5) Inform the firm if any of the subject devices have been distributed to other organizations. If yes, provide contact details so the firm can inform the recipients appropriately. 6) Inform the firm of any adverse events concerning the use of the subject devices. (US Customer Service: 1-855-91-NEURO (1-855-916-3876), Email: [email protected]) 7) Complete and return the response form via email to [email protected]
Recall details
- Recall number
- Z-1909-2024
- Status
- Open, Classified
- Initiated by the company
- Apr 23, 2024
- Posted by the FDA
- May 28, 2024
- Recalling firm
- Stryker Neurovascular
- Quantity
- 2859
- Product code
- DQX
- FDA submissions
- K023700, K002907
- Distribution
- Worldwide - US Nationwide distribution including in the states of MN, PA, IL, AK, NY, CA, MI, WI, TX, AZ, FL, KY, MA, IN, NE, MT, CT, NJ, TN, VA, AL, CO, DE, OH, UT, NC, ME, GA, SC, GU, WA, MO, OR, DC, OK, LA, MD, ID, NH, SD, NV, RI, ND, HI, MS, AR, IA, NM, VT, KS and the countries of Australia, Austria, Belgium, Brazil, Canada, Canary Islands, Chile, China, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, SAR China, Hungary, India, Ireland, Israel, Italy, Japan, Korea (South), Lithuania, Luxembourg, Netherlands, Northern Ireland, Norway, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, Spain, Sweden, Switzerland, Taiwan, Republic of China, Thailand, Turkey, United Arab Emirates, United Kingdom.
Product description
Synchro .014"guide wire, 35 cm tip, Catalog: M00313010 Synchro .014 guide wire, 300 cm length, Catalog: M00313310 SYNCHRO .010 WIRE SOFT STRAIGHT 200cm, Catalog: M00316310 SYNCHRO .010 WIRE SOFT STRAIGHT 300cm, Catalog: M00316330
Code information (models, lots, serial numbers)
Catalog/UDI-DI/Lot: M00313010/07613252186922/123279, 123364, 123404, 123985, 124214, 124216; M00313310/07613252186946/123494; M00316310/07613252187158/123851; M00316330/07613252187165/124156
Devices named in this recall
4 devices in the FDA device registry whose identifier appears in the recall's code information.
- SYNCHRO 14 M00313010Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls220 reports
- SYNCHRO 14 M00313310Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls64 reports
- SYNCHRO 10 M00316310Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls44 reports
- SYNCHRO 10 M00316330Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls4 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 94437).
- Z-1902-2024 · Open, ClassifiedTrevo XP ProVue Retriever 4 x 20, Catalog: 90182 Trevo XP ProVue Retriever 3 x 20, Catalog: 90183 Trevo XP ProVue Retriever 4 x 30, Catalog: 90185 Trevo XP ProVue Retriever 6 x 25,…
- Z-1903-2024 · Open, ClassifiedTrevo ProVue, Catalog: 90184
- Z-1904-2024 · Open, ClassifiedTrevo Pro 14 Microcatheter, Catalog: 90231; Trevo Pro 18 Microcatheter, Catalog: 90238
- Z-1905-2024 · Open, ClassifiedTrevo NXT ProVue 3 x 32, US, Catalog: 90312 Kits: TREVO NXT 3X32+TRAK21 2-PACK - CE, Catalog: 91412 TREVO NXT 4X28+TRAK21 2-PACK - CE, Catalog: 91413 TREVO NXT 4X41+TRAK21 2-PACK …
- Z-1906-2024 · Open, ClassifiedTrevo Trak 21 Microcatheter, Catalog: 90338
- Z-1907-2024 · Open, ClassifiedFlowGate2 8F x 85cm, Catalog: 90485 FlowGate2 8F x 95cm, Catalog: 90495
- Z-1908-2024 · Open, ClassifiedDAC 044 115cm, Catalog: 90760
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page