Vaporizer Sevoflurane Maquet Filling, Product code/REF… recallZ-1877-2024 · Getinge Usa Sales Inc · initiated May 1, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Getinge Usa Sales Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Sevoflurane used with the vaporizer may degrade to hydrogen fluoride, which could present a risk to patients or health care providers through inhalation and/or skin exposure. Potential harms include irritation of respiratory tract, which in worst case may lead to lung edema and/or severe hypocalcemia, blistering, superficial ulceration, and/or hypomagnesemia.
FDA root cause category: Under Investigation by firm
What the company is doing
On May 1, 2024 URGENT MEDICAL DEVICE REMOVAL letters were sent to customers. Actions to be taken by customer: Until further notice, do NOT use the Vaproizer with Sevoflurane provided by Poramal and Baxter Sevoflurane. Do NOT use the Vaporizer if it was ever used with Sevoflurane Piramal and/or Baxter Sevoflurane, even if it currently is used with AbbVie Ultane. Our records indicate that you have received the Vaporizer Sevoflurane, Maquet Filling having one or more of the serial numbers that are affected by this recall. 1. Please examine your inventory immediately to determine if you have any of the Vaporizer Sevoflurane, Maquet Filling with the product codes/serial numbers listed in this notice. The Product Code/REF Number and serial number (SN) are found on the label located underneath each vaporizer. Should you have any affected product, please remove from areas of use. 2. If you have ONLY used Vaporizer Product Code/REF Number 6682282, serial number 17003-23454 with AbbVie Ultane, you can continue to use the vaporizer, by certifying that you only use AbbVie Ultane on the Medical Device Recall Response Form. If you are uncertain whether you have previously used the vaporizer with Piramal Sevoflurane and/or Baxter Sevoflurane, discontinue use of the vaporizer. 3. Persuant to the previous recall for the Vaporizer Product Code/REF Number 6886601, you can continue to use this vaporizer so long as it has ONLY been used with Abbvie Ultane. If you are uncertain whether you have previously used the vaporizer with Piramal Sevoflurane and/or Baxter Sevoflurane, discontinue use of the vaporizer. 4. If at any time, the agent in the vaporizer looks cloudy and yellowish, you should immediately stop use of the affected vaporizer. 5. If you have affected product you are entitled to a credit or replacement. Please contact Getinge Customer Service at 888-943-8872 (press option 2) to request a return authorization (RMA) and shipping instructions to return any affected product. Pack
Recall details
- Recall number
- Z-1877-2024
- Status
- Open, Classified
- Initiated by the company
- May 1, 2024
- Posted by the FDA
- May 31, 2024
- Recalling firm
- Getinge Usa Sales Inc
- Quantity
- 120 units (US)
- Product code
- BSZ
- FDA submissions
- K191027
- Distribution
- Domestic: CA, CO, IN, MO, OR, VA, WA. International: Aruba, Australia, Austria, Belgium, Bolivia, Bosnia-Herzegovina, Brazil, Bulgaria, Chile, China, Costa Rica, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iraq, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Lebanon, Libya, Lithuania, Macau, Macedonia, Malaysia, Mexico, Moldavia, Morocco, Mozambique, Namibia, Netherlands, Nicaragua, Norway, Oman, Pakistan, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Vietnam Please note that a list of foreign consignees was not provided by the manufacturer and it is not available at this time.
Product description
Vaporizer Sevoflurane Maquet Filling, Product code/REF number (6682282), Component of: Flow-i C20, Flow-i C30,Flow-i C40, Flow-c, Flow-e.
Code information (models, lots, serial numbers)
Part Number 6682282; UDI/DI: 07325710000212: Serial Numbers: 17003 - 23454.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Vaporizer Sevoflur. Maquet filling 6682282Maquet Critical Care AB · Gas-Machine, Anesthesia1 recall0 reports
Other devices under the same FDA submission
6 devices cleared under K191027 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Flow-c Anesthesia System 6887700Maquet Critical Care AB · Gas-Machine, Anesthesia6 recalls72 reports
- Flow-e Anesthesia System 6887900Maquet Critical Care AB · Gas-Machine, Anesthesia5 recalls19 reports
- Flow-i C20 Anesthesia System 6888520Maquet Critical Care AB · Gas-Machine, Anesthesia1 recall2 deaths88 reports
- Flow-i C30 Anesthesia System 6888530Maquet Critical Care AB · Gas-Machine, Anesthesia1 recall13 reports
- Flow-i C40 Anesthesia System 6888540Maquet Critical Care AB · Gas-Machine, Anesthesia1 recall9 reports
- Vaporizer Sevoflurane, MAQUET filling 6886601Maquet Critical Care AB · Gas-Machine, Anesthesia1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page