RefleXion X1 Radiotherapy System Model RXM1000 recallZ-1933-2024 · Reflexion Medical, Inc. · initiated Apr 16, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Reflexion Medical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A potential dose error exists for patients treated with an out of session SCINTIX partial fraction.

FDA root cause category: Software design

What the company is doing

On 04/29/2024, the firm sent an "Urgent: Medical Device Recall Safety Notification" via email to customer informing them that during internal dosimetric testing of SCINTIX OOS partials in periodic tracking mode, it was observed that the delivered dose error of the full fraction can exceed 5% with respect to the upper bound of one or more static volumes within the DVH in the plan (non-tracked, non-target volumes in the patient frame of reference, i.e., the planning CT). This was determined to be the result of a software defect associated with completing interrupted. Customers are instructed to: 1. Notify all clinical staff of this recall 2. Utilized the Clinical Guidance Document that is include with this notification. For questions or concerns - contact Service and Support at 650-239-9070 or email [email protected]

Recall details

Recall number
Z-1933-2024
Status
Open, Classified
Initiated by the company
Apr 16, 2024
Posted by the FDA
May 30, 2024
Recalling firm
Reflexion Medical, Inc.
Quantity
7 systems
Product code
QVA
Distribution
U.S.: CA, CT, NJ, OR, PA and TX No O.U.S.

Product description

RefleXion X1 Radiotherapy System Model RXM1000 Software version 2.1.19-3 Radiation therapy system is a device that combines the functionality of an emission computed tomography detection system and a linear accelerator.

Code information (models, lots, serial numbers)

Model Number: RXM1000 UDI Codes that includes system serial numbers: (01)00860003983812(11)230504(21)X11001 (01)00860003983812(11)230518(21)X11002 (01)00860003983812(11)230505(21)X11003 (01)00860003983812(11)230530(21)X11004 (01)00860003983812(11)230811(21)X11006 (01)00860003983812(11)230410(21)X11007 (01)00860003983812(11)230803(21)X11008

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.