Skip to content

RefleXion Medical Radiotherapy System RXM1000 (UDI 00860003983812)Recalls & reported problems

Reflexion Medical, Inc. · The RefleXion Medical Radiotherapy System is indicated for treatment planning and precise delivery of image-guided radiation therapy, stereotactic radiotherapy, or stereotactic radiosurgery for tumors or other targeted tissues anywhere in the body when radiation treatment is indicated, while minimizing the delivery of radiation to vital healthy tissue. The megavoltage X-ray radiation is delivered in a rotational, modulated, image-guided format in accordance with the physician approved plan. The RefleXion Medical Radiotherapy System is also indicated for FDG-guided treatment which includes modeling, planning and precise delivery of FDG-guided radiation therapy, a type of Biology-guided Radiotherapy (BgRT), in five or fewer fractions for adults. It is indicated for tumor volumes in lung and bone subject to potential motion and positional uncertainty that have each been assessed with on-board PET/CT prior to delivery for adequate localization, sufficient FDG metabolic activity, local contrast and consistent biodistribution to meet the RMRS requirements, while minimizing the delivery of radiation to vital healthy tissue. BgRT involves the detection of signals from F18 during active beam delivery as a guide to deliver megavoltage X-ray radiotherapy in a rotational, modulated format in accordance with a physician approved treatment plan. For complete fludeoxyglucose F18 prescribing information, refer both to the current medical imaging agent labeling and to this device labeling under “FDG Medical Imaging Agent Information”. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This RefleXion radiotherapy system delivers image-guided radiation treatment for tumors. The 36 reports mostly describe software problems that can cause radiation to be delivered to the wrong location; they include 24 injuries and no deaths. There are 3 open recalls involving software that can lose positioning corrections, allow too many treatment fractions, or create dose errors. Reports rose to 32 in the latest 12 months from 3 in the previous 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
36
Deaths reported
0
Injuries reported
24
Recalls naming it
3

Summary

The FDA's MAUDE database contains 36 adverse event reports that list this exact device (DI 00860003983812), received between Nov 21, 2023 and Apr 3, 2026 — classified as 24 injuries, 12 malfunctions.

The most frequently reported specific device problem is “Program or Algorithm Execution Problem” (32 reports).

3 FDA recall records name this device's identifier, 3 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 11 reportsApr 2026: 21 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202621
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Feb 9, 2026Class IIOpen, Classified

    Due to a radiotherapy medical system software defect, when performing angular roll corrections followed by a repeat localization, the roll corrections applied in the initial localization scan are not carried through to treatment delivery leading to the treatment plan dose being delivered to the incorrect location, potentially a displacement of 5mm or higher, with a 10% to 20% underdose.

    Z-1656-2026 · Root cause: Software change control · Recall details · FDA record ↗

  2. Jul 22, 2025Class IIOpen, Classified

    Due to incorrect software configuration that potentially allows more than two (2) fractions within a 12-hour period (fractionLimits) and could potentially lead to adverse events (toxicity).

    Z-2486-2025 · Root cause: Software change control · Recall details · FDA record ↗

  3. Apr 16, 2024Class IIOpen, Classified

    A potential dose error exists for patients treated with an out of session SCINTIX partial fraction.

    Z-1933-2024 · Root cause: Software design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Program or Algorithm Execution Problem32
  • Use of Device Problem2
  • Use of Incorrect Control/Treatment Settings2
  • Alarm Not Visible1
  • Patient Device Interaction Problem1

Patient problems

  • Insufficient Information34
  • Encephalopathy1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 25 most recent of 36 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 3, 2026Program or Algorithm Execution Problem

    DURING ROUTINE TESTING OF THE REFLEXION MEDICAL RADIOTHERAPY SYSTEM (RMRS) ON SITE AT REFLEXION, THE FOLLOWING FAILURE MODE WAS IDENTIFIED. WHEN PERFORMING ANGULAR ROLL CORRECTIONS FOLLOWED BY A REPEAT LOCALIZATION, THE ROLL CORRECTIONS APPLIED IN THE INITIAL LOCALIZATION SCAN ARE NOT CARRIED THROUGH TO TREATMENT DELIVERY LEADING TO THE TREATMENT PLAN DOSE BEING DELIVERED TO THE INCORRECT LOCATION. THIS WAS DETERMINED TO BE THE RESULT OF A SOFTWARE DEFECT ASSOCIATED WITH THE CALCULATION AND COMMUNICATION OF CUMULATIVE VIRTUAL ROLL OFFSETS BETWEEN COMPONENTS IN THE SYSTEM. IT WAS DETERMINED THAT A VERY SMALL NUMBER OF AFFECTED TREATMENT SESSIONS MAY HAVE A DISPLACEMENT OF AS MUCH AS 5MM FROM THE INTENDED LOCATION FOR THE DELIVERED DOSE VOLUME. A DISPLACEMENT OF 5MM OR HIGHER MAY RESULT IN CLINICALLY SIGNIFICANT DOSIMETRIC ERRORS (POTENTIALLY AS HIGH AS 10% TO 20% UNDERDOSE). FOR CASES OF POTENTIAL CONCERN WHERE THE DISPLACEMENT WAS BELOW 5MM, WE FURTHER EVALUATED THE POTENTIAL RISK OF POINT DOSE MISMATCH WITHIN THE DISPLACEMENT REGION. FOLLOWING ADDITIONAL INVESTIGATION AND DISCUSSIONS WITH THE SITE AND USING INFORMATION AVAILABLE IN THE SYSTEM LOGS TO INVESTIGATE THE IMPACT OF THE

    Patient problems: Insufficient Information

    MAUDE report 24784116 ↗
  • InjuryApr 3, 2026Program or Algorithm Execution Problem

    DURING ROUTINE TESTING OF THE REFLEXION MEDICAL RADIOTHERAPY SYSTEM (RMRS) ON SITE AT REFLEXION, THE FOLLOWING FAILURE MODE WAS IDENTIFIED. WHEN PERFORMING ANGULAR ROLL CORRECTIONS FOLLOWED BY A REPEAT LOCALIZATION, THE ROLL CORRECTIONS APPLIED IN THE INITIAL LOCALIZATION SCAN ARE NOT CARRIED THROUGH TO TREATMENT DELIVERY LEADING TO THE TREATMENT PLAN DOSE BEING DELIVERED TO THE INCORRECT LOCATION. THIS WAS DETERMINED TO BE THE RESULT OF A SOFTWARE DEFECT ASSOCIATED WITH THE CALCULATION AND COMMUNICATION OF CUMULATIVE VIRTUAL ROLL OFFSETS BETWEEN COMPONENTS IN THE SYSTEM. IT WAS DETERMINED THAT A VERY SMALL NUMBER OF AFFECTED TREATMENT SESSIONS MAY HAVE A DISPLACEMENT OF AS MUCH AS 5MM FROM THE INTENDED LOCATION FOR THE DELIVERED DOSE VOLUME. A DISPLACEMENT OF 5MM OR HIGHER MAY RESULT IN CLINICALLY SIGNIFICANT DOSIMETRIC ERRORS (POTENTIALLY AS HIGH AS 10% TO 20% UNDERDOSE). FOR CASES OF POTENTIAL CONCERN WHERE THE DISPLACEMENT WAS BELOW 5MM, WE FURTHER EVALUATED THE POTENTIAL RISK OF POINT DOSE MISMATCH WITHIN THE DISPLACEMENT REGION. FOLLOWING ADDITIONAL INVESTIGATION AND DISCUSSIONS WITH THE SITE AND USING INFORMATION AVAILABLE IN THE SYSTEM LOGS TO INVESTIGATE THE IMPACT OF THE

    Patient problems: Insufficient Information

    MAUDE report 24784115 ↗
  • InjuryApr 3, 2026Program or Algorithm Execution Problem

    DURING ROUTINE TESTING OF THE REFLEXION MEDICAL RADIOTHERAPY SYSTEM (RMRS) ON SITE AT REFLEXION, THE FOLLOWING FAILURE MODE WAS IDENTIFIED. WHEN PERFORMING ANGULAR ROLL CORRECTIONS FOLLOWED BY A REPEAT LOCALIZATION, THE ROLL CORRECTIONS APPLIED IN THE INITIAL LOCALIZATION SCAN ARE NOT CARRIED THROUGH TO TREATMENT DELIVERY LEADING TO THE TREATMENT PLAN DOSE BEING DELIVERED TO THE INCORRECT LOCATION. THIS WAS DETERMINED TO BE THE RESULT OF A SOFTWARE DEFECT ASSOCIATED WITH THE CALCULATION AND COMMUNICATION OF CUMULATIVE VIRTUAL ROLL OFFSETS BETWEEN COMPONENTS IN THE SYSTEM. IT WAS DETERMINED THAT A VERY SMALL NUMBER OF AFFECTED TREATMENT SESSIONS MAY HAVE A DISPLACEMENT OF AS MUCH AS 5MM FROM THE INTENDED LOCATION FOR THE DELIVERED DOSE VOLUME. A DISPLACEMENT OF 5MM OR HIGHER MAY RESULT IN CLINICALLY SIGNIFICANT DOSIMETRIC ERRORS (POTENTIALLY AS HIGH AS 10% TO 20% UNDERDOSE). FOR CASES OF POTENTIAL CONCERN WHERE THE DISPLACEMENT WAS BELOW 5MM, WE FURTHER EVALUATED THE POTENTIAL RISK OF POINT DOSE MISMATCH WITHIN THE DISPLACEMENT REGION. FOLLOWING ADDITIONAL INVESTIGATION AND DISCUSSIONS WITH THE SITE AND USING INFORMATION AVAILABLE IN THE SYSTEM LOGS TO INVESTIGATE THE IMPACT OF THE

    Patient problems: Insufficient Information

    MAUDE report 24784107 ↗
  • InjuryApr 3, 2026Program or Algorithm Execution Problem

    DURING ROUTINE TESTING OF THE REFLEXION MEDICAL RADIOTHERAPY SYSTEM (RMRS) ON SITE AT REFLEXION, THE FOLLOWING FAILURE MODE WAS IDENTIFIED. WHEN PERFORMING ANGULAR ROLL CORRECTIONS FOLLOWED BY A REPEAT LOCALIZATION, THE ROLL CORRECTIONS APPLIED IN THE INITIAL LOCALIZATION SCAN ARE NOT CARRIED THROUGH TO TREATMENT DELIVERY LEADING TO THE TREATMENT PLAN DOSE BEING DELIVERED TO THE INCORRECT LOCATION. THIS WAS DETERMINED TO BE THE RESULT OF A SOFTWARE DEFECT ASSOCIATED WITH THE CALCULATION AND COMMUNICATION OF CUMULATIVE VIRTUAL ROLL OFFSETS BETWEEN COMPONENTS IN THE SYSTEM. IT WAS DETERMINED THAT A VERY SMALL NUMBER OF AFFECTED TREATMENT SESSIONS MAY HAVE A DISPLACEMENT OF AS MUCH AS 5MM FROM THE INTENDED LOCATION FOR THE DELIVERED DOSE VOLUME. A DISPLACEMENT OF 5MM OR HIGHER MAY RESULT IN CLINICALLY SIGNIFICANT DOSIMETRIC ERRORS (POTENTIALLY AS HIGH AS 10% TO 20% UNDERDOSE). FOR CASES OF POTENTIAL CONCERN WHERE THE DISPLACEMENT WAS BELOW 5MM, WE FURTHER EVALUATED THE POTENTIAL RISK OF POINT DOSE MISMATCH WITHIN THE DISPLACEMENT REGION. FOLLOWING ADDITIONAL INVESTIGATION AND DISCUSSIONS WITH THE SITE AND USING INFORMATION AVAILABLE IN THE SYSTEM LOGS TO INVESTIGATE THE IMPACT OF THE

    Patient problems: Insufficient Information

    MAUDE report 24784106 ↗
  • InjuryApr 3, 2026Program or Algorithm Execution Problem

    DURING ROUTINE TESTING OF THE REFLEXION MEDICAL RADIOTHERAPY SYSTEM (RMRS) ON SITE AT REFLEXION, THE FOLLOWING FAILURE MODE WAS IDENTIFIED. WHEN PERFORMING ANGULAR ROLL CORRECTIONS FOLLOWED BY A REPEAT LOCALIZATION, THE ROLL CORRECTIONS APPLIED IN THE INITIAL LOCALIZATION SCAN ARE NOT CARRIED THROUGH TO TREATMENT DELIVERY LEADING TO THE TREATMENT PLAN DOSE BEING DELIVERED TO THE INCORRECT LOCATION. THIS WAS DETERMINED TO BE THE RESULT OF A SOFTWARE DEFECT ASSOCIATED WITH THE CALCULATION AND COMMUNICATION OF CUMULATIVE VIRTUAL ROLL OFFSETS BETWEEN COMPONENTS IN THE SYSTEM. IT WAS DETERMINED THAT A VERY SMALL NUMBER OF AFFECTED TREATMENT SESSIONS MAY HAVE A DISPLACEMENT OF AS MUCH AS 5MM FROM THE INTENDED LOCATION FOR THE DELIVERED DOSE VOLUME. A DISPLACEMENT OF 5MM OR HIGHER MAY RESULT IN CLINICALLY SIGNIFICANT DOSIMETRIC ERRORS (POTENTIALLY AS HIGH AS 10% TO 20% UNDERDOSE). FOR CASES OF POTENTIAL CONCERN WHERE THE DISPLACEMENT WAS BELOW 5MM, WE FURTHER EVALUATED THE POTENTIAL RISK OF POINT DOSE MISMATCH WITHIN THE DISPLACEMENT REGION. FOLLOWING ADDITIONAL INVESTIGATION AND DISCUSSIONS WITH THE SITE AND USING INFORMATION AVAILABLE IN THE SYSTEM LOGS TO INVESTIGATE THE IMPACT OF THE

    Patient problems: Insufficient Information

    MAUDE report 24783812 ↗
  • InjuryApr 3, 2026Program or Algorithm Execution Problem

    DURING ROUTINE TESTING OF THE REFLEXION MEDICAL RADIOTHERAPY SYSTEM (RMRS) ON SITE AT REFLEXION, THE FOLLOWING FAILURE MODE WAS IDENTIFIED. WHEN PERFORMING ANGULAR ROLL CORRECTIONS FOLLOWED BY A REPEAT LOCALIZATION, THE ROLL CORRECTIONS APPLIED IN THE INITIAL LOCALIZATION SCAN ARE NOT CARRIED THROUGH TO TREATMENT DELIVERY LEADING TO THE TREATMENT PLAN DOSE BEING DELIVERED TO THE INCORRECT LOCATION. THIS WAS DETERMINED TO BE THE RESULT OF A SOFTWARE DEFECT ASSOCIATED WITH THE CALCULATION AND COMMUNICATION OF CUMULATIVE VIRTUAL ROLL OFFSETS BETWEEN COMPONENTS IN THE SYSTEM. IT WAS DETERMINED THAT A VERY SMALL NUMBER OF AFFECTED TREATMENT SESSIONS MAY HAVE A DISPLACEMENT OF AS MUCH AS 5MM FROM THE INTENDED LOCATION FOR THE DELIVERED DOSE VOLUME. A DISPLACEMENT OF 5MM OR HIGHER MAY RESULT IN CLINICALLY SIGNIFICANT DOSIMETRIC ERRORS (POTENTIALLY AS HIGH AS 10% TO 20% UNDERDOSE). USING INFORMATION AVAILABLE IN THE SYSTEM LOGS TO INVESTIGATE THE IMPACT OF THE ABOVE PROBLEM, THE FOLLOWING CASE WAS IDENTIFIED AS MEETING A GREATER THAN 5MM DISPLACEMENT REPRESENTING A POTENTIALLY SIGNIFICANT DOSIMETRIC ERROR. THE PATIENT PLAN CONSISTED OF 42.56 GY TO BE DELIVERED IN 16 FRACTIONS, WITH A 2.66

    Patient problems: Insufficient Information

    MAUDE report 24775477 ↗
  • InjuryApr 1, 2026Program or Algorithm Execution Problem

    DURING ROUTINE TESTING OF THE REFLEXION MEDICAL RADIOTHERAPY SYSTEM (RMRS) ON SITE AT REFLEXION, THE FOLLOWING FAILURE MODE WAS IDENTIFIED. WHEN PERFORMING ANGULAR ROLL CORRECTIONS FOLLOWED BY A REPEAT LOCALIZATION, THE ROLL CORRECTIONS APPLIED IN THE INITIAL LOCALIZATION SCAN ARE NOT CARRIED THROUGH TO TREATMENT DELIVERY LEADING TO THE TREATMENT PLAN DOSE BEING DELIVERED TO THE INCORRECT LOCATION. THIS WAS DETERMINED TO BE THE RESULT OF A SOFTWARE DEFECT ASSOCIATED WITH THE CALCULATION AND COMMUNICATION OF CUMULATIVE VIRTUAL ROLL OFFSETS BETWEEN COMPONENTS IN THE SYSTEM. IT WAS DETERMINED THAT A VERY SMALL NUMBER OF AFFECTED TREATMENT SESSIONS MAY HAVE A DISPLACEMENT OF AS MUCH AS 5MM FROM THE INTENDED LOCATION FOR THE DELIVERED DOSE VOLUME. A DISPLACEMENT OF 5MM OR HIGHER MAY RESULT IN CLINICALLY SIGNIFICANT DOSIMETRIC ERRORS (POTENTIALLY AS HIGH AS 10% TO 20% UNDERDOSE). USING INFORMATION AVAILABLE IN THE SYSTEM LOGS TO INVESTIGATE THE IMPACT OF THE ABOVE PROBLEM, THE FOLLOWING CASE WAS IDENTIFIED AS MEETING A GREATER THAN 5MM DISPLACEMENT REPRESENTING A POTENTIALLY SIGNIFICANT DOSIMETRIC ERROR. THE PATIENT PLAN CONSISTED OF 24 GY TO BE DELIVERED IN 12 FRACTIONS, WITH A 2 GY PLA

    Patient problems: Insufficient Information

    MAUDE report 24760525 ↗
  • InjuryApr 1, 2026Program or Algorithm Execution Problem

    DURING ROUTINE TESTING OF THE REFLEXION MEDICAL RADIOTHERAPY SYSTEM (RMRS) ON SITE AT REFLEXION, THE FOLLOWING FAILURE MODE WAS IDENTIFIED. WHEN PERFORMING ANGULAR ROLL CORRECTIONS FOLLOWED BY A REPEAT LOCALIZATION, THE ROLL CORRECTIONS APPLIED IN THE INITIAL LOCALIZATION SCAN ARE NOT CARRIED THROUGH TO TREATMENT DELIVERY LEADING TO THE TREATMENT PLAN DOSE BEING DELIVERED TO THE INCORRECT LOCATION. THIS WAS DETERMINED TO BE THE RESULT OF A SOFTWARE DEFECT ASSOCIATED WITH THE CALCULATION AND COMMUNICATION OF CUMULATIVE VIRTUAL ROLL OFFSETS BETWEEN COMPONENTS IN THE SYSTEM. IT WAS DETERMINED THAT A VERY SMALL NUMBER OF AFFECTED TREATMENT SESSIONS MAY HAVE A DISPLACEMENT OF AS MUCH AS 5MM FROM THE INTENDED LOCATION FOR THE DELIVERED DOSE VOLUME. A DISPLACEMENT OF 5MM OR HIGHER MAY RESULT IN CLINICALLY SIGNIFICANT DOSIMETRIC ERRORS (POTENTIALLY AS HIGH AS 10% TO 20% UNDERDOSE). USING INFORMATION AVAILABLE IN THE SYSTEM LOGS TO INVESTIGATE THE IMPACT OF THE ABOVE PROBLEM, THE FOLLOWING CASE WAS IDENTIFIED AS MEETING A GREATER THAN 5MM DISPLACEMENT REPRESENTING A POTENTIALLY SIGNIFICANT DOSIMETRIC ERROR. THE PATIENT PLAN CONSISTED OF 36 GY TO BE DELIVERED IN 20 FRACTIONS, WITH A 1.8 GY P

    Patient problems: Insufficient Information

    MAUDE report 24760516 ↗
  • InjuryApr 1, 2026Program or Algorithm Execution Problem

    DURING ROUTINE TESTING OF THE REFLEXION MEDICAL RADIOTHERAPY SYSTEM (RMRS) ON SITE AT REFLEXION, THE FOLLOWING FAILURE MODE WAS IDENTIFIED. WHEN PERFORMING ANGULAR ROLL CORRECTIONS FOLLOWED BY A REPEAT LOCALIZATION, THE ROLL CORRECTIONS APPLIED IN THE INITIAL LOCALIZATION SCAN ARE NOT CARRIED THROUGH TO TREATMENT DELIVERY LEADING TO THE TREATMENT PLAN DOSE BEING DELIVERED TO THE INCORRECT LOCATION. THIS WAS DETERMINED TO BE THE RESULT OF A SOFTWARE DEFECT ASSOCIATED WITH THE CALCULATION AND COMMUNICATION OF CUMULATIVE VIRTUAL ROLL OFFSETS BETWEEN COMPONENTS IN THE SYSTEM. IT WAS DETERMINED THAT A VERY SMALL NUMBER OF AFFECTED TREATMENT SESSIONS MAY HAVE A DISPLACEMENT OF AS MUCH AS 5MM FROM THE INTENDED LOCATION FOR THE DELIVERED DOSE VOLUME. A DISPLACEMENT OF 5MM OR HIGHER MAY RESULT IN CLINICALLY SIGNIFICANT DOSIMETRIC ERRORS (POTENTIALLY AS HIGH AS 10% TO 20% UNDERDOSE). USING INFORMATION AVAILABLE IN THE SYSTEM LOGS TO INVESTIGATE THE IMPACT OF THE ABOVE PROBLEM, THE FOLLOWING CASE WAS IDENTIFIED AS MEETING A GREATER THAN 5MM DISPLACEMENT REPRESENTING A POTENTIALLY SIGNIFICANT DOSIMETRIC ERROR. THE PATIENT PLAN CONSISTED OF 60 GY TO BE DELIVERED IN 30 FRACTIONS, WITH A 2 GY PLA

    Patient problems: Insufficient Information

    MAUDE report 24760513 ↗
  • InjuryApr 1, 2026Program or Algorithm Execution Problem

    DURING ROUTINE TESTING OF THE REFLEXION MEDICAL RADIOTHERAPY SYSTEM (RMRS) ON SITE AT REFLEXION, THE FOLLOWING FAILURE MODE WAS IDENTIFIED. WHEN PERFORMING ANGULAR ROLL CORRECTIONS FOLLOWED BY A REPEAT LOCALIZATION, THE ROLL CORRECTIONS APPLIED IN THE INITIAL LOCALIZATION SCAN ARE NOT CARRIED THROUGH TO TREATMENT DELIVERY LEADING TO THE TREATMENT PLAN DOSE BEING DELIVERED TO THE INCORRECT LOCATION. THIS WAS DETERMINED TO BE THE RESULT OF A SOFTWARE DEFECT ASSOCIATED WITH THE CALCULATION AND COMMUNICATION OF CUMULATIVE VIRTUAL ROLL OFFSETS BETWEEN COMPONENTS IN THE SYSTEM. IT WAS DETERMINED THAT A VERY SMALL NUMBER OF AFFECTED TREATMENT SESSIONS MAY HAVE A DISPLACEMENT OF AS MUCH AS 5MM FROM THE INTENDED LOCATION FOR THE DELIVERED DOSE VOLUME. A DISPLACEMENT OF 5MM OR HIGHER MAY RESULT IN CLINICALLY SIGNIFICANT DOSIMETRIC ERRORS (POTENTIALLY AS HIGH AS 10% TO 20% UNDERDOSE). USING INFORMATION AVAILABLE IN THE SYSTEM LOGS TO INVESTIGATE THE IMPACT OF THE ABOVE PROBLEM, THE FOLLOWING CASE WAS IDENTIFIED AS MEETING A GREATER THAN 5MM DISPLACEMENT REPRESENTING A POTENTIALLY SIGNIFICANT DOSIMETRIC ERROR. THE PATIENT PLAN CONSISTED OF 30 GY TO BE DELIVERED IN 3 FRACTIONS, WITH A 10 GY PLA

    Patient problems: Insufficient Information

    MAUDE report 24760505 ↗
  • InjuryApr 1, 2026Program or Algorithm Execution Problem

    DURING ROUTINE TESTING OF THE REFLEXION MEDICAL RADIOTHERAPY SYSTEM (RMRS) ON SITE AT REFLEXION, THE FOLLOWING FAILURE MODE WAS IDENTIFIED. WHEN PERFORMING ANGULAR ROLL CORRECTIONS FOLLOWED BY A REPEAT LOCALIZATION, THE ROLL CORRECTIONS APPLIED IN THE INITIAL LOCALIZATION SCAN ARE NOT CARRIED THROUGH TO TREATMENT DELIVERY LEADING TO THE TREATMENT PLAN DOSE BEING DELIVERED TO THE INCORRECT LOCATION. THIS WAS DETERMINED TO BE THE RESULT OF A SOFTWARE DEFECT ASSOCIATED WITH THE CALCULATION AND COMMUNICATION OF CUMULATIVE VIRTUAL ROLL OFFSETS BETWEEN COMPONENTS IN THE SYSTEM. IT WAS DETERMINED THAT A VERY SMALL NUMBER OF AFFECTED TREATMENT SESSIONS MAY HAVE A DISPLACEMENT OF AS MUCH AS 5MM FROM THE INTENDED LOCATION FOR THE DELIVERED DOSE VOLUME. A DISPLACEMENT OF 5MM OR HIGHER MAY RESULT IN CLINICALLY SIGNIFICANT DOSIMETRIC ERRORS (POTENTIALLY AS HIGH AS 10% TO 20% UNDERDOSE). USING INFORMATION AVAILABLE IN THE SYSTEM LOGS TO INVESTIGATE THE IMPACT OF THE ABOVE PROBLEM, THE FOLLOWING CASE WAS IDENTIFIED AS MEETING A GREATER THAN 5MM DISPLACEMENT REPRESENTING A POTENTIALLY SIGNIFICANT DOSIMETRIC ERROR. THE PATIENT PLAN CONSISTED OF 14 GY TO BE DELIVERED IN 4 FRACTIONS, WITH A 3.5 GY PL

    Patient problems: Insufficient Information

    MAUDE report 24760501 ↗
  • InjuryApr 1, 2026Program or Algorithm Execution Problem

    DURING ROUTINE TESTING OF THE REFLEXION MEDICAL RADIOTHERAPY SYSTEM (RMRS) ON SITE AT REFLEXION, THE FOLLOWING FAILURE MODE WAS IDENTIFIED. WHEN PERFORMING ANGULAR ROLL CORRECTIONS FOLLOWED BY A REPEAT LOCALIZATION, THE ROLL CORRECTIONS APPLIED IN THE INITIAL LOCALIZATION SCAN ARE NOT CARRIED THROUGH TO TREATMENT DELIVERY LEADING TO THE TREATMENT PLAN DOSE BEING DELIVERED TO THE INCORRECT LOCATION. THIS WAS DETERMINED TO BE THE RESULT OF A SOFTWARE DEFECT ASSOCIATED WITH THE CALCULATION AND COMMUNICATION OF CUMULATIVE VIRTUAL ROLL OFFSETS BETWEEN COMPONENTS IN THE SYSTEM. IT WAS DETERMINED THAT A VERY SMALL NUMBER OF AFFECTED TREATMENT SESSIONS MAY HAVE A DISPLACEMENT OF AS MUCH AS 5MM FROM THE INTENDED LOCATION FOR THE DELIVERED DOSE VOLUME. A DISPLACEMENT OF 5MM OR HIGHER MAY RESULT IN CLINICALLY SIGNIFICANT DOSIMETRIC ERRORS (POTENTIALLY AS HIGH AS 10% TO 20% UNDERDOSE). USING INFORMATION AVAILABLE IN THE SYSTEM LOGS TO INVESTIGATE THE IMPACT OF THE ABOVE PROBLEM, THE FOLLOWING CASE WAS IDENTIFIED AS MEETING A GREATER THAN 5MM DISPLACEMENT REPRESENTING A POTENTIALLY SIGNIFICANT DOSIMETRIC ERROR. THE PATIENT PLAN CONSISTED OF 20.5 GY TO BE DELIVERED IN 5 FRACTIONS, WITH A 4.1 GY

    Patient problems: Insufficient Information

    MAUDE report 24760493 ↗
  • InjuryApr 1, 2026Program or Algorithm Execution Problem

    DURING ROUTINE TESTING OF THE REFLEXION MEDICAL RADIOTHERAPY SYSTEM (RMRS) ON SITE AT REFLEXION, THE FOLLOWING FAILURE MODE WAS IDENTIFIED. WHEN PERFORMING ANGULAR ROLL CORRECTIONS FOLLOWED BY A REPEAT LOCALIZATION, THE ROLL CORRECTIONS APPLIED IN THE INITIAL LOCALIZATION SCAN ARE NOT CARRIED THROUGH TO TREATMENT DELIVERY LEADING TO THE TREATMENT PLAN DOSE BEING DELIVERED TO THE INCORRECT LOCATION. THIS WAS DETERMINED TO BE THE RESULT OF A SOFTWARE DEFECT ASSOCIATED WITH THE CALCULATION AND COMMUNICATION OF CUMULATIVE VIRTUAL ROLL OFFSETS BETWEEN COMPONENTS IN THE SYSTEM. IT WAS DETERMINED THAT A VERY SMALL NUMBER OF AFFECTED TREATMENT SESSIONS MAY HAVE A DISPLACEMENT OF AS MUCH AS 5MM FROM THE INTENDED LOCATION FOR THE DELIVERED DOSE VOLUME. A DISPLACEMENT OF 5MM OR HIGHER MAY RESULT IN CLINICALLY SIGNIFICANT DOSIMETRIC ERRORS (POTENTIALLY AS HIGH AS 10% TO 20% UNDERDOSE). USING INFORMATION AVAILABLE IN THE SYSTEM LOGS TO INVESTIGATE THE IMPACT OF THE ABOVE PROBLEM, THE FOLLOWING CASE WAS IDENTIFIED AS MEETING A GREATER THAN 5MM DISPLACEMENT REPRESENTING A POTENTIALLY SIGNIFICANT DOSIMETRIC ERROR. THE PATIENT PLAN CONSISTED OF 30 GY TO BE DELIVERED IN 15 FRACTIONS, WITH A 2 GY PLA

    Patient problems: Insufficient Information

    MAUDE report 24760488 ↗
  • InjuryApr 1, 2026Program or Algorithm Execution Problem

    DURING ROUTINE TESTING OF THE REFLEXION MEDICAL RADIOTHERAPY SYSTEM (RMRS) ON SITE AT REFLEXION, THE FOLLOWING FAILURE MODE WAS IDENTIFIED. WHEN PERFORMING ANGULAR ROLL CORRECTIONS FOLLOWED BY A REPEAT LOCALIZATION, THE ROLL CORRECTIONS APPLIED IN THE INITIAL LOCALIZATION SCAN ARE NOT CARRIED THROUGH TO TREATMENT DELIVERY LEADING TO THE TREATMENT PLAN DOSE BEING DELIVERED TO THE INCORRECT LOCATION. THIS WAS DETERMINED TO BE THE RESULT OF A SOFTWARE DEFECT ASSOCIATED WITH THE CALCULATION AND COMMUNICATION OF CUMULATIVE VIRTUAL ROLL OFFSETS BETWEEN COMPONENTS IN THE SYSTEM. IT WAS DETERMINED THAT A VERY SMALL NUMBER OF AFFECTED TREATMENT SESSIONS MAY HAVE A DISPLACEMENT OF AS MUCH AS 5MM FROM THE INTENDED LOCATION FOR THE DELIVERED DOSE VOLUME. A DISPLACEMENT OF 5MM OR HIGHER MAY RESULT IN CLINICALLY SIGNIFICANT DOSIMETRIC ERRORS (POTENTIALLY AS HIGH AS 10% TO 20% UNDERDOSE). USING INFORMATION AVAILABLE IN THE SYSTEM LOGS TO INVESTIGATE THE IMPACT OF THE ABOVE PROBLEM, THE FOLLOWING CASE WAS IDENTIFIED AS MEETING A GREATER THAN 5MM DISPLACEMENT REPRESENTING A POTENTIALLY SIGNIFICANT DOSIMETRIC ERROR. THE PATIENT PLAN CONSISTED OF 30 GY TO BE DELIVERED IN 5 FRACTIONS, WITH A 6 GY PLAN

    Patient problems: Insufficient Information

    MAUDE report 24760478 ↗
  • InjuryApr 1, 2026Program or Algorithm Execution Problem

    DURING ROUTINE TESTING OF THE REFLEXION MEDICAL RADIOTHERAPY SYSTEM (RMRS) ON SITE AT REFLEXION, THE FOLLOWING FAILURE MODE WAS IDENTIFIED. WHEN PERFORMING ANGULAR ROLL CORRECTIONS FOLLOWED BY A REPEAT LOCALIZATION, THE ROLL CORRECTIONS APPLIED IN THE INITIAL LOCALIZATION SCAN ARE NOT CARRIED THROUGH TO TREATMENT DELIVERY LEADING TO THE TREATMENT PLAN DOSE BEING DELIVERED TO THE INCORRECT LOCATION. THIS WAS DETERMINED TO BE THE RESULT OF A SOFTWARE DEFECT ASSOCIATED WITH THE CALCULATION AND COMMUNICATION OF CUMULATIVE VIRTUAL ROLL OFFSETS BETWEEN COMPONENTS IN THE SYSTEM. IT WAS DETERMINED THAT A VERY SMALL NUMBER OF AFFECTED TREATMENT SESSIONS MAY HAVE A DISPLACEMENT OF AS MUCH AS 5MM FROM THE INTENDED LOCATION FOR THE DELIVERED DOSE VOLUME. A DISPLACEMENT OF 5MM OR HIGHER MAY RESULT IN CLINICALLY SIGNIFICANT DOSIMETRIC ERRORS (POTENTIALLY AS HIGH AS 10% TO 20% UNDERDOSE). USING INFORMATION AVAILABLE IN THE SYSTEM LOGS TO INVESTIGATE THE IMPACT OF THE ABOVE PROBLEM, THE FOLLOWING CASE WAS IDENTIFIED AS MEETING A GREATER THAN 5MM DISPLACEMENT REPRESENTING A POTENTIALLY SIGNIFICANT DOSIMETRIC ERROR. THE PATIENT PLAN CONSISTED OF 20 GY TO BE DELIVERED IN 5 FRACTIONS, WITH A 4 GY PLAN

    Patient problems: Insufficient Information

    MAUDE report 24760461 ↗
  • InjuryApr 1, 2026Program or Algorithm Execution Problem

    DURING ROUTINE TESTING OF THE REFLEXION MEDICAL RADIOTHERAPY SYSTEM (RMRS) ON SITE AT REFLEXION, THE FOLLOWING FAILURE MODE WAS IDENTIFIED. WHEN PERFORMING ANGULAR ROLL CORRECTIONS FOLLOWED BY A REPEAT LOCALIZATION, THE ROLL CORRECTIONS APPLIED IN THE INITIAL LOCALIZATION SCAN ARE NOT CARRIED THROUGH TO TREATMENT DELIVERY LEADING TO THE TREATMENT PLAN DOSE BEING DELIVERED TO THE INCORRECT LOCATION. THIS WAS DETERMINED TO BE THE RESULT OF A SOFTWARE DEFECT ASSOCIATED WITH THE CALCULATION AND COMMUNICATION OF CUMULATIVE VIRTUAL ROLL OFFSETS BETWEEN COMPONENTS IN THE SYSTEM. IT WAS DETERMINED THAT A VERY SMALL NUMBER OF AFFECTED TREATMENT SESSIONS MAY HAVE A DISPLACEMENT OF AS MUCH AS 5MM FROM THE INTENDED LOCATION FOR THE DELIVERED DOSE VOLUME. A DISPLACEMENT OF 5MM OR HIGHER MAY RESULT IN CLINICALLY SIGNIFICANT DOSIMETRIC ERRORS (POTENTIALLY AS HIGH AS 10% TO 20% UNDERDOSE). USING INFORMATION AVAILABLE IN THE SYSTEM LOGS TO INVESTIGATE THE IMPACT OF THE ABOVE PROBLEM, THE FOLLOWING CASE WAS IDENTIFIED AS MEETING A GREATER THAN 5MM DISPLACEMENT REPRESENTING A POTENTIALLY SIGNIFICANT DOSIMETRIC ERROR. THE PATIENT PLAN CONSISTED OF 40 GY TO BE DELIVERED IN 10 FRACTIONS, WITH A 4 GY PLA

    Patient problems: Insufficient Information

    MAUDE report 24760450 ↗
  • InjuryApr 1, 2026Program or Algorithm Execution Problem

    DURING ROUTINE TESTING OF THE REFLEXION MEDICAL RADIOTHERAPY SYSTEM (RMRS) ON SITE AT REFLEXION, THE FOLLOWING FAILURE MODE WAS IDENTIFIED. WHEN PERFORMING ANGULAR ROLL CORRECTIONS FOLLOWED BY A REPEAT LOCALIZATION, THE ROLL CORRECTIONS APPLIED IN THE INITIAL LOCALIZATION SCAN ARE NOT CARRIED THROUGH TO TREATMENT DELIVERY LEADING TO THE TREATMENT PLAN DOSE BEING DELIVERED TO THE INCORRECT LOCATION. THIS WAS DETERMINED TO BE THE RESULT OF A SOFTWARE DEFECT ASSOCIATED WITH THE CALCULATION AND COMMUNICATION OF CUMULATIVE VIRTUAL ROLL OFFSETS BETWEEN COMPONENTS IN THE SYSTEM. IT WAS DETERMINED THAT A VERY SMALL NUMBER OF AFFECTED TREATMENT SESSIONS MAY HAVE A DISPLACEMENT OF AS MUCH AS 5MM FROM THE INTENDED LOCATION FOR THE DELIVERED DOSE VOLUME. A DISPLACEMENT OF 5MM OR HIGHER MAY RESULT IN CLINICALLY SIGNIFICANT DOSIMETRIC ERRORS (POTENTIALLY AS HIGH AS 10% TO 20% UNDERDOSE). USING INFORMATION AVAILABLE IN THE SYSTEM LOGS TO INVESTIGATE THE IMPACT OF THE ABOVE PROBLEM, THE FOLLOWING CASE WAS IDENTIFIED AS MEETING A GREATER THAN 5MM DISPLACEMENT REPRESENTING A POTENTIALLY SIGNIFICANT DOSIMETRIC ERROR. THE PATIENT PLAN CONSISTED OF 25 GY TO BE DELIVERED IN 5 FRACTIONS, WITH A 5 GY PLAN

    Patient problems: Insufficient Information

    MAUDE report 24760445 ↗
  • InjuryApr 1, 2026Program or Algorithm Execution Problem

    DURING ROUTINE TESTING OF THE REFLEXION MEDICAL RADIOTHERAPY SYSTEM (RMRS) ON SITE AT REFLEXION, THE FOLLOWING FAILURE MODE WAS IDENTIFIED. WHEN PERFORMING ANGULAR ROLL CORRECTIONS FOLLOWED BY A REPEAT LOCALIZATION, THE ROLL CORRECTIONS APPLIED IN THE INITIAL LOCALIZATION SCAN ARE NOT CARRIED THROUGH TO TREATMENT DELIVERY LEADING TO THE TREATMENT PLAN DOSE BEING DELIVERED TO THE INCORRECT LOCATION. THIS WAS DETERMINED TO BE THE RESULT OF A SOFTWARE DEFECT ASSOCIATED WITH THE CALCULATION AND COMMUNICATION OF CUMULATIVE VIRTUAL ROLL OFFSETS BETWEEN COMPONENTS IN THE SYSTEM. IT WAS DETERMINED THAT A VERY SMALL NUMBER OF AFFECTED TREATMENT SESSIONS MAY HAVE A DISPLACEMENT OF AS MUCH AS 5MM FROM THE INTENDED LOCATION FOR THE DELIVERED DOSE VOLUME. A DISPLACEMENT OF 5MM OR HIGHER MAY RESULT IN CLINICALLY SIGNIFICANT DOSIMETRIC ERRORS (POTENTIALLY AS HIGH AS 10% TO 20% UNDERDOSE). USING INFORMATION AVAILABLE IN THE SYSTEM LOGS TO INVESTIGATE THE IMPACT OF THE ABOVE PROBLEM, THE FOLLOWING CASE WAS IDENTIFIED AS MEETING A GREATER THAN 5MM DISPLACEMENT REPRESENTING A POTENTIALLY SIGNIFICANT DOSIMETRIC ERROR. THE PATIENT PLAN CONSISTED OF 66 GY TO BE DELIVERED IN 33 FRACTIONS, WITH A 2 GY PLA

    Patient problems: Insufficient Information

    MAUDE report 24760436 ↗
  • InjuryApr 1, 2026Program or Algorithm Execution Problem

    DURING ROUTINE TESTING OF THE REFLEXION MEDICAL RADIOTHERAPY SYSTEM (RMRS) ON SITE AT REFLEXION, THE FOLLOWING FAILURE MODE WAS IDENTIFIED. WHEN PERFORMING ANGULAR ROLL CORRECTIONS FOLLOWED BY A REPEAT LOCALIZATION, THE ROLL CORRECTIONS APPLIED IN THE INITIAL LOCALIZATION SCAN ARE NOT CARRIED THROUGH TO TREATMENT DELIVERY LEADING TO THE TREATMENT PLAN DOSE BEING DELIVERED TO THE INCORRECT LOCATION. THIS WAS DETERMINED TO BE THE RESULT OF A SOFTWARE DEFECT ASSOCIATED WITH THE CALCULATION AND COMMUNICATION OF CUMULATIVE VIRTUAL ROLL OFFSETS BETWEEN COMPONENTS IN THE SYSTEM. IT WAS DETERMINED THAT A VERY SMALL NUMBER OF AFFECTED TREATMENT SESSIONS MAY HAVE A DISPLACEMENT OF AS MUCH AS 5MM FROM THE INTENDED LOCATION FOR THE DELIVERED DOSE VOLUME. A DISPLACEMENT OF 5MM OR HIGHER MAY RESULT IN CLINICALLY SIGNIFICANT DOSIMETRIC ERRORS (POTENTIALLY AS HIGH AS 10% TO 20% UNDERDOSE). USING INFORMATION AVAILABLE IN THE SYSTEM LOGS TO INVESTIGATE THE IMPACT OF THE ABOVE PROBLEM, THE FOLLOWING CASE WAS IDENTIFIED AS MEETING A GREATER THAN 5MM DISPLACEMENT REPRESENTING A POTENTIALLY SIGNIFICANT DOSIMETRIC ERROR. THE PATIENT PLAN CONSISTED OF 30 GY TO BE DELIVERED IN 5 FRACTIONS, WITH A 6 GY PLAN

    Patient problems: Insufficient Information

    MAUDE report 24760215 ↗
  • InjuryApr 1, 2026Program or Algorithm Execution Problem

    DURING ROUTINE TESTING OF THE REFLEXION MEDICAL RADIOTHERAPY SYSTEM (RMRS) ON SITE AT REFLEXION, THE FOLLOWING FAILURE MODE WAS IDENTIFIED. WHEN PERFORMING ANGULAR ROLL CORRECTIONS FOLLOWED BY A REPEAT LOCALIZATION, THE ROLL CORRECTIONS APPLIED IN THE INITIAL LOCALIZATION SCAN ARE NOT CARRIED THROUGH TO TREATMENT DELIVERY LEADING TO THE TREATMENT PLAN DOSE BEING DELIVERED TO THE INCORRECT LOCATION. THIS WAS DETERMINED TO BE THE RESULT OF A SOFTWARE DEFECT ASSOCIATED WITH THE CALCULATION AND COMMUNICATION OF CUMULATIVE VIRTUAL ROLL OFFSETS BETWEEN COMPONENTS IN THE SYSTEM. IT WAS DETERMINED THAT A VERY SMALL NUMBER OF AFFECTED TREATMENT SESSIONS MAY HAVE A DISPLACEMENT OF AS MUCH AS 5MM FROM THE INTENDED LOCATION FOR THE DELIVERED DOSE VOLUME. A DISPLACEMENT OF 5MM OR HIGHER MAY RESULT IN CLINICALLY SIGNIFICANT DOSIMETRIC ERRORS (POTENTIALLY AS HIGH AS 10% TO 20% UNDERDOSE). USING INFORMATION AVAILABLE IN THE SYSTEM LOGS TO INVESTIGATE THE IMPACT OF THE ABOVE PROBLEM, THE FOLLOWING CASE WAS IDENTIFIED AS MEETING A GREATER THAN 5MM DISPLACEMENT REPRESENTING A POTENTIALLY SIGNIFICANT DOSIMETRIC ERROR. THE PATIENT PLAN CONSISTED OF 50 GY TO BE DELIVERED IN 25 FRACTIONS, WITH A 2 GY PLA

    Patient problems: Insufficient Information

    MAUDE report 24759959 ↗
  • InjuryApr 1, 2026Program or Algorithm Execution Problem

    DURING ROUTINE TESTING OF THE REFLEXION MEDICAL RADIOTHERAPY SYSTEM (RMRS) ON SITE AT REFLEXION, THE FOLLOWING FAILURE MODE WAS IDENTIFIED. WHEN PERFORMING ANGULAR ROLL CORRECTIONS FOLLOWED BY A REPEAT LOCALIZATION, THE ROLL CORRECTIONS APPLIED IN THE INITIAL LOCALIZATION SCAN ARE NOT CARRIED THROUGH TO TREATMENT DELIVERY LEADING TO THE TREATMENT PLAN DOSE BEING DELIVERED TO THE INCORRECT LOCATION. THIS WAS DETERMINED TO BE THE RESULT OF A SOFTWARE DEFECT ASSOCIATED WITH THE CALCULATION AND COMMUNICATION OF CUMULATIVE VIRTUAL ROLL OFFSETS BETWEEN COMPONENTS IN THE SYSTEM. IT WAS DETERMINED THAT A VERY SMALL NUMBER OF AFFECTED TREATMENT SESSIONS MAY HAVE A DISPLACEMENT OF AS MUCH AS 5MM FROM THE INTENDED LOCATION FOR THE DELIVERED DOSE VOLUME. A DISPLACEMENT OF 5MM OR HIGHER MAY RESULT IN CLINICALLY SIGNIFICANT DOSIMETRIC ERRORS (POTENTIALLY AS HIGH AS 10% TO 20% UNDERDOSE). USING INFORMATION AVAILABLE IN THE SYSTEM LOGS TO INVESTIGATE THE IMPACT OF THE ABOVE PROBLEM, THE FOLLOWING CASE WAS IDENTIFIED AS MEETING A GREATER THAN 5MM DISPLACEMENT REPRESENTING A POTENTIALLY SIGNIFICANT DOSIMETRIC ERROR. THE PATIENT PLAN CONSISTED OF 30 GY TO BE DELIVERED IN 5 FRACTIONS, WITH A 6 GY PLAN

    Patient problems: Insufficient Information

    MAUDE report 24758959 ↗
  • MalfunctionMar 5, 2026Program or Algorithm Execution Problem

    DURING ROUTINE TESTING OF THE REFLEXION MEDICAL RADIOTHERAPY SYSTEM (RMRS) ON SITE AT REFLEXION, THE FOLLOWING FAILURE MODE WAS IDENTIFIED. WHEN PERFORMING ANGULAR ROLL CORRECTIONS FOLLOWED BY A REPEAT LOCALIZATION, THE ROLL CORRECTIONS APPLIED IN THE INITIAL LOCALIZATION SCAN ARE NOT CARRIED THROUGH TO TREATMENT DELIVERY LEADING TO THE TREATMENT PLAN DOSE BEING DELIVERED TO THE INCORRECT LOCATION. THIS WAS DETERMINED TO BE THE RESULT OF A SOFTWARE DEFECT ASSOCIATED WITH THE CALCULATION AND COMMUNICATION OF CUMULATIVE VIRTUAL ROLL OFFSETS BETWEEN COMPONENTS IN THE SYSTEM. IT WAS DETERMINED THAT MOST AFFECTED TREATMENT SESSIONS WILL HAVE A DISPLACEMENT OF LESS THAN 1MM FROM THE INTENDED LOCATION FOR THE DELIVERED DOSE VOLUME; HOWEVER, A VERY SMALL NUMBER OF AFFECTED TREATMENT SESSIONS MAY HAVE A DISPLACEMENT OF AS MUCH AS 5MM FROM THE INTENDED LOCATION FOR THE DELIVERED DOSE VOLUME. A DISPLACEMENT OF 5MM OR HIGHER MAY RESULT IN CLINICALLY SIGNIFICANT DOSIMETRIC ERRORS (POTENTIALLY AS HIGH AS 10% TO 20% UNDERDOSE). A RECALL HAS BEEN FILED FOR THIS ISSUE ON 12-FEB-2026 (RES # 98518, REFLEXION REPORT # 3011716550-02122026-C-00001). SEE H11 FOR ADDITIONAL INVESTIGATION INFORMATION.

    Patient problems: Insufficient Information

    MAUDE report 24527133 ↗
  • MalfunctionMar 5, 2026Program or Algorithm Execution Problem

    DURING ROUTINE TESTING OF THE REFLEXION MEDICAL RADIOTHERAPY SYSTEM (RMRS) ON SITE AT REFLEXION, THE FOLLOWING FAILURE MODE WAS IDENTIFIED. WHEN PERFORMING ANGULAR ROLL CORRECTIONS FOLLOWED BY A REPEAT LOCALIZATION, THE ROLL CORRECTIONS APPLIED IN THE INITIAL LOCALIZATION SCAN ARE NOT CARRIED THROUGH TO TREATMENT DELIVERY LEADING TO THE TREATMENT PLAN DOSE BEING DELIVERED TO THE INCORRECT LOCATION. THIS WAS DETERMINED TO BE THE RESULT OF A SOFTWARE DEFECT ASSOCIATED WITH THE CALCULATION AND COMMUNICATION OF CUMULATIVE VIRTUAL ROLL OFFSETS BETWEEN COMPONENTS IN THE SYSTEM. IT WAS DETERMINED THAT MOST AFFECTED TREATMENT SESSIONS WILL HAVE A DISPLACEMENT OF LESS THAN 1MM FROM THE INTENDED LOCATION FOR THE DELIVERED DOSE VOLUME; HOWEVER, A VERY SMALL NUMBER OF AFFECTED TREATMENT SESSIONS MAY HAVE A DISPLACEMENT OF AS MUCH AS 5MM FROM THE INTENDED LOCATION FOR THE DELIVERED DOSE VOLUME. A DISPLACEMENT OF 5MM OR HIGHER MAY RESULT IN CLINICALLY SIGNIFICANT DOSIMETRIC ERRORS (POTENTIALLY AS HIGH AS 10% TO 20% UNDERDOSE). A RECALL HAS BEEN FILED FOR THIS ISSUE ON 12-FEB-2026 (RES# (B)(4), REFLEXION REPORT # (B)(4)). SEE H11 FOR ADDITIONAL INVESTIGATION INFORMATION.

    Patient problems: Insufficient Information

    MAUDE report 24527130 ↗
  • MalfunctionMar 5, 2026Program or Algorithm Execution Problem

    DURING ROUTINE TESTING OF THE REFLEXION MEDICAL RADIOTHERAPY SYSTEM (RMRS) ON SITE AT REFLEXION, THE FOLLOWING FAILURE MODE WAS IDENTIFIED. WHEN PERFORMING ANGULAR ROLL CORRECTIONS FOLLOWED BY A REPEAT LOCALIZATION, THE ROLL CORRECTIONS APPLIED IN THE INITIAL LOCALIZATION SCAN ARE NOT CARRIED THROUGH TO TREATMENT DELIVERY LEADING TO THE TREATMENT PLAN DOSE BEING DELIVERED TO THE INCORRECT LOCATION. THIS WAS DETERMINED TO BE THE RESULT OF A SOFTWARE DEFECT ASSOCIATED WITH THE CALCULATION AND COMMUNICATION OF CUMULATIVE VIRTUAL ROLL OFFSETS BETWEEN COMPONENTS IN THE SYSTEM. IT WAS DETERMINED THAT MOST AFFECTED TREATMENT SESSIONS WILL HAVE A DISPLACEMENT OF LESS THAN 1MM FROM THE INTENDED LOCATION FOR THE DELIVERED DOSE VOLUME; HOWEVER, A VERY SMALL NUMBER OF AFFECTED TREATMENT SESSIONS MAY HAVE A DISPLACEMENT OF AS MUCH AS 5MM FROM THE INTENDED LOCATION FOR THE DELIVERED DOSE VOLUME. A DISPLACEMENT OF 5MM OR HIGHER MAY RESULT IN CLINICALLY SIGNIFICANT DOSIMETRIC ERRORS (POTENTIALLY AS HIGH AS 10% TO 20% UNDERDOSE). A RECALL HAS BEEN FILED FOR THIS ISSUE ON 12-FEB-2026 (RES# (B)(4), REFLEXION REPORT # (B)(4)). SEE H11 FOR ADDITIONAL INVESTIGATION INFORMATION.

    Patient problems: Insufficient Information

    MAUDE report 24527129 ↗
  • MalfunctionMar 5, 2026Program or Algorithm Execution Problem

    DURING ROUTINE TESTING OF THE REFLEXION MEDICAL RADIOTHERAPY SYSTEM (RMRS) ON SITE AT REFLEXION, THE FOLLOWING FAILURE MODE WAS IDENTIFIED. WHEN PERFORMING ANGULAR ROLL CORRECTIONS FOLLOWED BY A REPEAT LOCALIZATION, THE ROLL CORRECTIONS APPLIED IN THE INITIAL LOCALIZATION SCAN ARE NOT CARRIED THROUGH TO TREATMENT DELIVERY LEADING TO THE TREATMENT PLAN DOSE BEING DELIVERED TO THE INCORRECT LOCATION. THIS WAS DETERMINED TO BE THE RESULT OF A SOFTWARE DEFECT ASSOCIATED WITH THE CALCULATION AND COMMUNICATION OF CUMULATIVE VIRTUAL ROLL OFFSETS BETWEEN COMPONENTS IN THE SYSTEM. IT WAS DETERMINED THAT MOST AFFECTED TREATMENT SESSIONS WILL HAVE A DISPLACEMENT OF LESS THAN 1MM FROM THE INTENDED LOCATION FOR THE DELIVERED DOSE VOLUME; HOWEVER, A VERY SMALL NUMBER OF AFFECTED TREATMENT SESSIONS MAY HAVE A DISPLACEMENT OF AS MUCH AS 5MM FROM THE INTENDED LOCATION FOR THE DELIVERED DOSE VOLUME. A DISPLACEMENT OF 5MM OR HIGHER MAY RESULT IN CLINICALLY SIGNIFICANT DOSIMETRIC ERRORS (POTENTIALLY AS HIGH AS 10% TO 20% UNDERDOSE). A RECALL HAS BEEN FILED FOR THIS ISSUE ON 12-FEB-2026 (RES# (B)(4), REFLEXION REPORT # (B)(4)). SEE H11 FOR ADDITIONAL INVESTIGATION INFORMATION.

    Patient problems: Insufficient Information

    MAUDE report 24527107 ↗

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page