Ventilator HAMILTON-C6, PN: 160021 recallZ-2020-2024 · Hamilton Medical AG · initiated May 15, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Hamilton Medical AG or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Ventilator may enter sensor fail mode, ventilation may not re-initiate, after patient is reconnected, which may cause hypoxia, if this sequence occurs: 1)User presses O2 enrichment, 2)User disconnects endotracheal tube for open suctioning, 3)Sensor error initiated, sensor fail mode occurs, 4)Patient is re-connected with sensor fail mode active, 5) Ventilation is not re-initiated by the ventilator.

FDA root cause category: Packaging change control

What the company is doing

On 5/24/24, correction notices were emailed to customers and distributors and they were asked to do the following: If ventilation is not re-initiated due to this issue, there are 4 ways to reinitiate ventilation: 1) Select and confirm a control setting (even without a change of the value), or 2) Select and confirm a new ventilation mode, or 3)Switch the ventilator to standby mode and restart ventilation from standby mode, or 4) Switch the ventilator off and switch it on again If the user is not able to manually re-initiate ventilation in due time as per one of the 4 ways described above, they might decide to ventilate the patient by alternative means. - Read and sign the confirmation sheet, which can be returned to [email protected] - Ensure that your medical staff is aware of the notice content - Attach this letter to each affected device's operator's manual in order to ensure quick access. - The newest available software version 1.2.3 contains the required fix and shall be installed without undue delay. Your local distribution partner will support you with updating your affected devices with high priority. Questions, contact [email protected]

Recall details

Recall number
Z-2020-2024
Status
Open, Classified
Initiated by the company
May 15, 2024
Posted by the FDA
Jun 18, 2024
Recalling firm
Hamilton Medical AG
Quantity
68
Product code
CBK
FDA submissions
K201658
Distribution
US Nationwide distribution in the states of GA, KY, FL, IN, NV, MI, VA, OR, TX, SC, CA, PA.

Product description

Ventilator HAMILTON-C6, PN: 160021

Code information (models, lots, serial numbers)

UDI: 07630002808590, Software Versions: SW v1.1.4, SW v1.1.5, SW v1.1.6

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.