Skip to content

WOM Tube Set for lrrlgatlon, single-use- Intended to… recallZ-2243-2024 · WOM World Of Medicine AG · initiated May 24, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from WOM World Of Medicine AG or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Incorrect transponder data was written on the RFID tags triggers the error message E03, Tube Set lnvalid, may result in a delay of surgery

FDA root cause category: Process control

What the company is doing

Smith & Nephew Issued Urgent Field Safety Notice (FSN) - MEDICAL DEVICE RECALL WOM reference no.: 2024-0001 via email on 5/24/23. Letter states reason fo recall, health risk and action to take; lnform individuals within your organization who need to be aware of this recall. - Check all stock areas and/or operating room storage to determine if any tube sets with lol numbers from Annex 'l are at your facility, - Dispose of all affected tube sets properly. Confirm this on the customer response form. Return the form to [email protected]. Please send any questions to [email protected]

Recall details

Recall number
Z-2243-2024
Status
Open, Classified
Initiated by the company
May 24, 2024
Posted by the FDA
Jul 1, 2024
Recalling firm
WOM World Of Medicine AG
Quantity
1700 units
Product code
HRX
FDA submissions
K062380
Distribution
US Nationwide distribution.

Product description

WOM Tube Set for lrrlgatlon, single-use- Intended to transport irrigation fluid to the patient in endoscopic surgeries. Distr. by Endoscopy Smith & Nephew Inc. REF Numbers: T0449-01

Code information (models, lots, serial numbers)

UDI: 04056702003678 Lot Numbers: 4030902, 4030903

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

1 device cleared under K062380 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page