GoFlo Pump A127Recalls & reported problems
W.O.M. World of Medicine GmbH · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04056702003715), received between May 28, 2018 and Jun 17, 2018 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Calibration Problem” (1 report).
1 other recall cover devices cleared under the same FDA submission (K062380); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Calibration Problem1
- Device Operates Differently Than Expected1
- Pressure Problem1
Patient problems
- No Known Impact Or Consequence To Patient2
- Swelling1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 17, 2018Calibration Problem
IT WAS REPORTED CALIBRATION ISSUES HAVE CAUSED HIGH FLOW. A BACKUP DEVICE WAS AVAILABLE TO COMPLETE THE PROCEDURE WITH NO DELAY OR PATIENT INJURIES.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7608474 ↗MalfunctionMay 30, 2018Pressure Problem
IT WAS REPORTED THAT DURING SURGERY, THE WATER PRESSURE OF THE PUMP RISE RAPIDLY AND THE PUMP TURNED QUICKLY, SO THE WATER WAS LOWERED TO 5, BUT THE PUMP CONSTANTLY TURNED QUICKLY AND INTERRUPTED THE OP. OBSTRUCTION OF VISION. NO PATIENT INJURY REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7553443 ↗MalfunctionMay 28, 2018Device Operates Differently Than Expected
IT WAS REPORTED THAT DURING A HALFWAY INTO CASE THE SURGEON NOTICED THAT THE PATIENTS OPERATIVE SHOULDER WAS BEGINNING TO SWELL DUE TO EXTRAVASATION. HE REQUESTED TO HIS ASSISTANT TO REDUCE THE PRESSURE ( WAS RUNNING ON 40MMHG), ASSISTANT REDUCED THE PRESSURE TO 30MMHG. EXTRAVASATION CONTINUED RAPIDLY, SURGEON REMOVED THE TROCAR AND CANNULA FROM THE PATIENT WITH FLUID STILL RUNNING, FLUID FLOW WAS EXTREMELY FAST OUT OF TROCAR AT 300MMHG. THE SURGEON REQUESTED THE PUMP TO BE STOPPED AND RETURNED TO THE HOSPITALS TYPICAL FLUID PUMP SYSTEM ( LINVATEC 10K).
Patient problems: Swelling
MAUDE report 7547394 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K062380. One submission often covers many models, so check the product and lot details in each record.
- Z-2243-2024 · May 24, 2024 · Open, ClassifiedWOM Tube Set for lrrlgatlon, single-use- Intended to transport irrigation fluid to the patient in endoscopic surgeries. Distr. by Endoscopy Smith & Nephew Inc. REF Numbers: T0449-0
Other Arthroscope devices
- FMS 283512Medos International Sàrl · ARTHROSCOPE1,800 reports
- Arthrex® AR-6480Arthrex, Inc. · ARTHROSCOPE949 reports
- FMS VUE 284002Depuy Mitek, LLC · ARTHROSCOPE829 reports
- MITEK 242018Medos International Sàrl · Arthroscope731 reports
- METRx® System 9560100Medtronic Sofamor Danek, Inc. · Arthroscope667 reports
- FMS VUE II 284004Medos International Sàrl · Arthroscope562 reports
- VAPR COOLPULSE 228146Depuy Mitek, LLC · ARTHROSCOPE474 reports
- FMS DUO 284580Depuy Mitek, LLC · ARTHROSCOPE458 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page