Skip to content

Welch Allyn CP150 Electrocardiograph with Spirometry… recallZ-2323-2024 · Baxter Healthcare Corporation · initiated Jun 3, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Baxter Healthcare Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

During manufacturing, the step to install an Electromagnetic Interference (EMI) Absorber was omitted. The purpose of the EMI absorber is to prevent EMI leakage that might potentially affect other devices in the vicinity or the device itself may be vulnerable to EMI resulting in device malfunction or unavailability.

FDA root cause category: Employee error

What the company is doing

On 06/03/2024, Baxter Healthcare's Customer Service team contacted the consignee, via telephone call, to discuss the recall action and request the affected unit be returned to Baxter.

Recall details

Recall number
Z-2323-2024
Status
Open, Classified
Initiated by the company
Jun 3, 2024
Posted by the FDA
Jul 10, 2024
Recalling firm
Baxter Healthcare Corporation
Quantity
1 unit
Product code
DPS
FDA submissions
K141582
Distribution
Distribution to Illinois.

Product description

Welch Allyn CP150 Electrocardiograph with Spirometry Option, Product Code: CP150A-1ENB. The Electrocardiograph is one of the tools that clinicians use to evaluate, diagnose, and monitor patient cardiac function.

Code information (models, lots, serial numbers)

UDI/DI: 00732094177640. Product Code: CP150A-1ENB. Serial Number: 100013892024

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

3 devices cleared under K141582 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page