Skip to content

Welch Allyn, Inc. 901049 (REF CP150-1ENB)Recalls & reported problems

Welch Allyn, Inc. · Welch Allyn CP 150 12-Lead Resting Electrocardiograph; Noninterpretive, AHA ECG Cable with Disposable Electrodes, CardioPerfect Workstation Connectivity via USB and Ethernet, Color Touchscreen Display, 100-Test Memory; 100–240 V, 50/60 Hz AC, IEC Plug Type-B · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Mar 24, 2025Class IIOpen, Classified

    There were customer reports of devices which experienced battery-related fires. The investigation found that the devices in question were utilizing third-party, after-market batteries that are not approved by Baxter.

    Z-1560-2025 · Root cause: Process design · Recall details · FDA record ↗

  2. Jun 3, 2024Class IIOpen, Classified

    During manufacturing, the step to install an Electromagnetic Interference (EMI) Absorber was omitted. The purpose of the EMI absorber is to prevent EMI leakage that might potentially affect other devices in the vicinity or the device itself may be vulnerable to EMI resulting in device malfunction or unavailability.

    Z-2323-2024 · Root cause: Employee error · Recall details · FDA record ↗

Other ELECTROCARDIOGRAPH devices

All ELECTROCARDIOGRAPH devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page