Proclaim recallZ-2369-2024 · Abbott Medical · initiated May 16, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Abbott Medical or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The duration between the implantable pulse generator reaching the elective replacement indicator threshold and end of service may be shorter than indicated in product labeling.
FDA root cause category: Vendor change control
What the company is doing
On June 6, 2024, Abbott issued a Urgent Medical Device Correction notification to affected consignees. Abbott asked consignees to take the following actions: 1. This issue will continue to affect your patients for the foreseeable future. Abbott is working on an update to align the product labeling with actual device performance. 2. When an ERI notification appears, see the following considerations to make a determination with your patients on timing of IPG replacement within the customer notification. 3. See the customer notification for recommended strategies for preventing lapse in therapy. 4. Please return a completed Acknowledgement Form and maintain a record of this notice along with a copy of the completed Acknowledgement Form. 5. For questions about this issue or reported complaints, please contact your Abbott representative, or Abbott Technical Support at 1-800-727-7846. 6. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail or by fax.
Recall details
- Recall number
- Z-2369-2024
- Status
- Open, Classified
- Initiated by the company
- May 16, 2024
- Posted by the FDA
- Jul 16, 2024
- Recalling firm
- Abbott Medical
- Quantity
- 38,530 units
- Product code
- LGW
- FDA submissions
- P010032S151
- Distribution
- Worldwide - US Nationwide distribution including in the states of ALABAMA, ALASKA, ARIZONA, ARKANSAS, CALIFORNIA, COLORADO, CONNECTICUT, DELAWARE, DISTRICT OF COLUMBIA, FLORIDA, GEORGIA, GUAM, HAWAII, IDAHO, ILLINOIS, INDIANA, IOWA, KANSAS, KENTUCKY, LOUISIANA,MAINE, MARYLAND, MASSACHUSETTS, MICHIGAN, MINNESOTA, MISSISSIPPI, MISSOURI, MONTANA, NEBRASKA, NEVADA, NEW HAMPSHIRE, NEW JERSEY, NEW MEXICO, NEW YORK, NORTH CAROLINA, NORTH DAKOTA, OHIO, OKLAHOMA, OREGON, PENNSYLVANIA, PUERTO RICO, RHODE ISLAND, SOUTH CAROLINA, SOUTH DAKOTA, TENNESSEE, TEXAS, UTAH, VERMONT, VIRGINIA, WASHINGTON, WEST VIRGINIA, WISCONSIN, WYOMING and the countries of Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Georgia, Germany, Greece, Hungary, Ireland, Israel, Italy , Japan, Korea, Kuwait, Lithuania, Luxembourg, Mexico, Netherlands, New Zealand, Norway, Poland, Portugal, Reunion Russian Federation, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom.
Product description
Proclaim 5 Elite Implantable Pulse Generator, REF 3660, Sterile EO, Rx Only
Code information (models, lots, serial numbers)
05415067020192/ All Serial Numbers
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Proclaim™ 5 EliteSt. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1 recall10 deaths9,661 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 94798).
- Z-2358-2024 · Open, ClassifiedAbbott Proclaim XR 5 Implantable Pulse Generator, REF 3660, Sterile EO, Rx Only
- Z-2359-2024 · Open, ClassifiedAbbott Proclaim 5 Implantable Pulse Generator, REF 3661, Sterile EO, Rx Only
- Z-2360-2024 · Open, ClassifiedAbbott Proclaim XR 7 Implantable Pulse Generator, REF 3662, Sterile EO, Rx Only
- Z-2361-2024 · Open, ClassifiedAbbott Proclaim 7 Implantable Pulse Generator, REF 3663, Sterile EO, Rx Only
- Z-2362-2024 · Open, ClassifiedAbbott Proclaim 5 Implantable Pulse Generator, REF 3665, Sterile EO, Rx Only
- Z-2363-2024 · Open, ClassifiedAbbott Proclaim 7 Implantable Pulse Generator, REF 3667, Sterile EO, Rx Only
- Z-2364-2024 · Open, ClassifiedAbbott Proclaim Plus 5 Implantable Pulse Generator, REF 3670, Sterile EO, Rx Only
- Z-2365-2024 · Open, ClassifiedAbbott Proclaim Plus 5 Implantable Pulse Generator, REF 3671, Sterile EO, Rx Only
- Z-2366-2024 · Open, ClassifiedAbbott Proclaim Plus 7 Implantable Pulse Generator, REF 3672, Sterile EO, Rx Only
- Z-2367-2024 · Open, ClassifiedAbbott Proclaim DRG Implantable Pulse Generator, REF 3664, Sterile EO, Rx Only
- Z-2368-2024 · Open, ClassifiedProclaim 7 Elite Implantable Pulse Generator, REF 3662, Sterile EO, Rx Only
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page