Skip to content

BD Insyte Autoguard recallZ-2458-2024 · Becton Dickinson Infusion Therapy Systems, Inc. · initiated Jun 26, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Becton Dickinson Infusion Therapy Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Catheters may contain a hole in the catheter tubing, which could result in leakage during the insertion process.

FDA root cause category: Process control

What the company is doing

On June 26, 2024, Becton Dickinson Infusion Therapy Systems Inc. issued a "Urgent: Medical Device Product Recall" Notification to affected consignees via: mail. BD ask consignees to take the following actions: 1. Ensure the contents of this Product Recall communication are read and understood by those within your organization. 2. Immediately review your inventory for Catalog Number 382533, Lot number 4127714. Destroy all unused product subject to the recall following your institution s process for destruction. 3. If affected device was previously used on a patient without incident, no further action is necessary. If the device is currently in use and issue has not been observed, continue to use as normal. 4. Notify any and all entities of this recall to whom the affected product may have been transferred. 5. Complete the attached Customer Response Form and return to your distributor, whether or not you have any of the impacted material.

Recall details

Recall number
Z-2458-2024
Status
Open, Classified
Initiated by the company
Jun 26, 2024
Posted by the FDA
Jul 31, 2024
Recalling firm
Becton Dickinson Infusion Therapy Systems, Inc.
Quantity
26,400 units
Product code
FOZ
FDA submissions
K201075
Distribution
US: IL OUS: N/A0

Product description

BD Insyte Autoguard BC, Shielded IV Catheter with Blood Control Technology, REF: 382533, 20 GA x 1.00 in (1.1x25 mm) 63 mL/min, Rx Only. BD Insyte Autoguard BC Shielded IV Catheter is intended to be inserted into a patient's peripheral vascular system for short term use to sample blood, monitor blood pressure, or administer fluids.

Code information (models, lots, serial numbers)

Lot # 4127714/ UDI-DI: 00382903825332

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

53 devices cleared under K201075 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page